NORPORT CT-PC, LUCENT NON-CORING NEEDEL INFUSION SET
K111101 · Norfolk Medical · LJT · Aug 25, 2011 · General Hospital
Device Facts
Record ID
K111101
Device Name
NORPORT CT-PC, LUCENT NON-CORING NEEDEL INFUSION SET
Applicant
Norfolk Medical
Product Code
LJT · General Hospital
Decision Date
Aug 25, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5965
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NorPort CT.PC Family of Implanted Vascular Access Devices is indicated for use when the patient requires the following: repeated access to the vascular system for injections, infusion of drugs, administration of blood or blood products, and/or withdrawal of blood as part of the therapy regimen. When used with a appropriately labeled power injectable needle infusion set, the NorPort CT.PC Port is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/s with a 19 and 20 gauge non-coring power injectable needle and 2 ml/s with a 22-gauge noncoring power injectable needle.
Device Story
NorPort CT.PC is an implantable vascular access port and catheter system; Lucent Non-Coring Needle is an infusion set for port access. Port reservoir consists of titanium or polysulfone chamber with silicone septum; catheter is radiopaque. System implanted subcutaneously by physician; catheter tunneled to superior vena cava/right atrium junction. Physician accesses port via percutaneous needle puncture of septum. Device enables repeated delivery of fluids/medications or blood withdrawal. Power injection capability allows delivery of contrast media at specified flow rates (up to 5 ml/s for 19/20G needles; 2 ml/s for 22G needles). Port features machine-etched 'CT' symbol for radiographic identification. Benefits include reliable long-term vascular access and reduced need for repeated venipuncture.
Clinical Evidence
Bench testing only. No clinical data presented. Performance verified via catheter-to-port connection testing, septum puncture/leak testing, patency verification, and high-pressure static testing to confirm suitability for power injection.
Technological Characteristics
Materials: Titanium or polysulfone reservoir, silicone septum, radiopaque catheter. Design: Triangular port with recessed septum ring and suture holes. Connectivity: Standalone implantable device. Sterilization: Ethylene oxide (ISO 11135-1). Standards: BS EN ISO 10555-1 (catheters), ISO 10993-7 (EO residuals), ISO 14971 (risk management).
Indications for Use
Indicated for patients requiring repeated vascular access for drug infusion, blood/blood product administration, blood sampling, and power injection of contrast media. Contraindicated for patients where catheter placement may result in 'pinch-off' between the clavicle and first rib.
Regulatory Classification
Identification
A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.
K070003 — POWER INJECTABLE IMPLANTABLE INFUSION PORT WITH SILICONE LUMEN (ATTACHABLE AND PRE-ATTACHED) · Medcomp · May 15, 2007
K072481 — JET PORT PLUS HP CATHETER SYSTEM · Pfm Medical, Inc. · Mar 7, 2008
K112713 — NORFOLK MEDICAL-SPORTPORT-FAMILY OF PORTS · Norfolk Medical Products, Inc. · Oct 27, 2011
K122767 — NMI PORT · Navilyst Medical, Inc. · Apr 30, 2013
K132177 — MEDCOMP GEN III POWER INJECTABLE PORT · Medcomp ( Medical Components) · Dec 12, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
## Norfolk & Me
7350 N. Ridgeway, Skokie, IL 60076 • 847-674-7075
## Section 5
AUG 2 5 2011
KIIIIIII
### 510(k) Summary
### NorPort CT.PC Port Family of Implanted Vascular Access Devices, and Lucent Non-Coring Needle Infusion Set
510 (k) Summary of Safety and Effectiveness Information 21 CFR 807.92
### Part I General Information
#### 5.1 Submitter/Sponsor Information
Submitter Name: FDA Establishment Registration Number: Address:
Telephone Number: Fax Number: Contact Person: Date of Preparation:
#### 5.2 Device Name
Device Name:
Trade Name:
Device Common or Usual Name:
Norfolk Medical Products, Inc.
1450392 7350 N. Ridgeway Skokie, IL 60076 (847) 674-7075 (847) 674-7066 Michael J. Dalton/President April 15, 2011
Implanted Drug Delivery Device (Port) and Huber Needle Intravascular Administration Set (Needle)
NorPort CT.PC Family of Ports and Lucent Non-Coring Needle Infusion Set
Implantable Infusion Ports and Intravascular Administration Set
{1}------------------------------------------------
## Norfolk C Medical KIIIIol
## 7350 N. Ridgeway, Skokie, IL 60076 • 847-674-7075
AUG 25 2011
#### Classification Name:
Catheter/Port:
LJT- Port & Catheter, Implanted, Subcutaneous, Intravascular 21 CFR 880.5965- Subcutaneous, Implanted, Intravascular Infusion Port and Catheter, Class II
Needle:
FPA- Huber Needle Intravascular Administration Set 21 CFR 880.5440-Intravascular Administration Set
#### 5.3 Predicate Device Name
Catheter/Port:
Device Name: Premarket Notification:
Device Name: Premarket Notification:
Device Name:
Premarket Notification:
Device Name:
Premarket Notification:
Device Name: Premarket Notification:
Device Name: Premarket Notification: Norfolk Medical NorPort Port 510 (k)# 102385, 2010
Bard Power Port Titanium Port w/8Fr 510 (k)# 060812, 2006
Smiths Medical PORT-A-CATH Titanium Venous Access System 510 (k)# 070116, 2007
Angiodynamics Smartport CT Series Port Access System 510 (k)# 081472, 2008
PHS C-PortHP "Power Injectable" Port 510 (k)# 091099, 2009
MEDCOMP ProFUSECT 510 (k)# 070003, 2007
{2}------------------------------------------------
# Norfolk ( Medical
7350 N. Ridgeway, Skokie, IL 60076 . 847-674-7075
Needle Infusion Set:
| Device Name: | Smiths Medical Gripper Plus Power P.A.C Needle |
|-------------------------|------------------------------------------------|
| Premarket Notification: | 510 (k)# 070116, 2007 |
| Device Name: | Angiodynamics Lifeguard CT Safety Infusion Set |
| Premarket Notification: | 510 (k)# 072375, 2007 |
| Device Name: | Bard PowerLoc Clear Safety Infusion Set |
| Premarket Notification: | 510 (k)# 082306, 2008 |
| Device Name: | MEDCOMP CT Injectable Safety Huber Needle |
| Premarket Notification: | 510 (k)# 080544, 2008 |
#### 5.4 Device Description
## NorPort CT.PC Port Family of Implanted Vascular Access Devices:
The Norfolk Medical NorPort CT-PC Port is composed of a port reservoir with an attachable catheter system. The Port reservoir consists of a titanium or polysulfone chamber with a silicone septum and an outlet pin (510 (k)#102385). The silicone septum is designed to allow multiple needle puncturing while maintaining the leak-tight integrity.
The NorPort CT·PC comes in standard design to enable regular use as an infusion/withdrawal port and has high-pressure injection capability to aid in the delivery of large volumes of specialty fluids such as contrast media. The NorPort CT·PC is triangular shaped with a recessed septum ring. The body port has a "CT" symbol machine etched on the bottom end to enable easy recognition of the port with most imaging systems. The port has 3 holes for suture fixation to the tissue. A catheter comes with the port and is inserted in the vascular system. The catheter is usually inserted into a venous vessel and fed down into the superior vena cava. The placement of the catheter is checked by radiographic technique. The catheter is radiopaque to enable visualization for placement. The kit provided to aid in insertion of the catheter and placement of the port may include introducer items like needles, sheathes, vein picks, guidewires, straighteners, and dilators. Likewise the, "cut-down", kit for port placement may contain tunnelers, infusion sets with 90 degree Huber point needles, and syringes along with the port and catheter.
{3}------------------------------------------------
# - Norfolk C Medical
7350 N. Ridgeway, Skokie, IL 60076 • 847-674-7075
## Lucent Non-Coring Needle Infusion Set:
The Lucent Non-Coring Needle Infusion Set is similar to the following predicate devices: Smiths Medical Gripper Plus Power P.A.C Needle (510 (k)# 070116), the Angiodynamics Lifeguard CT Safety Infusion Set (510 (k)#072375), the Bard the rater of the Safety Infusion Set (510 (k)# 082306), and the MEDCOMP CT Injectable Safety Huber Needle (510 (k)# 080544). The Lucent Non-coring and the predicated devices are supplied sterile and non-pyrogenic and are intended for the administration into or withdrawal of fluids from the Norfolk Medical line of implanted ports. When a "Power Injectable" NorPort CT-PC Port is used, the Lucent Non-coring and the predicate devices are indicated for power injection of contrast media.
#### Indications for Use Statement and Product Function 5.5
#### Catheter/ Port:
The NorPort CT PC Family of Implanted Vascular Access Devices is indicated for use when the patient requires repeated access to the vascular system for injections, infusion drugs, administration of blood or blood products, and/or withdrawal of blood as part of the therapy regimen.
The NorPort CT PC Port is indicated for power injection of contrast media. For rne ron ection of contrast media, the maximum recommended infusion rate is 5 power injection of contained on coring power injectable needle and 2 ml/s with a 22-gauge non-coring power injectable needle.
### Needle:
The Lucent Non-Coring needle Infusion Set is intended for the administration of fluids and drugs as well as blood sampling through implanted vascular ports.
When used with the NorPort CT·PC Family of Implanted Vascular Access Devices, the Lucent Non-Coring Needle Infusion Set is also indicated for power injection of contrast media into the central venous system.
#### Contraindications or Cautions for Use 5.6
A detailed list of possible implantation complications and contraindications of the NorPort CT PC family of ports are listed in the "User's Manual" which is supplied with each NorPort CT Kit. Some of the possible complications due to supplied with each the Norport CT-PC include, but are not limited to, infection, occlusion, embolization, catheter fragmentation, erosion, extrusion of the device, hematomas, clot formation and thrombosis.
{4}------------------------------------------------
# Norfolk @ Medical
7350 N. Ridgeway, Skokie, IL 60076 • 847-674-7075
Improper placement of the catheter in the body has been shown to cause the catheter to be cut off from a, "pinching" effect by the clavicle and the first rib. Exercise caution when placing the catheter into the vein and the superior vena cava to ensure that the catheter does not pass between the juncture of the clavicle and the first rib. This "Pinch-off" complication is well documented and a well understood potential complication of the surgical implantation of the Port system.
#### 5.7 Methods of Application
The method of application is to prepare and insert the catheter into the vein (using "cut down" method or "percutaneous" method) and on into the junction of the superior vena cava and the right atrium. Then the proximal end of the catheter is tunneled subcutaneously to an area of cut down where the port is to be placed beneath the skin and secured to the muscle tissue. The catheter is joined to the port and the port is sutured to the muscle. All incisions are sutured normally by the physician.
Intravenous fluids, medications, blood products, or nutritional fluids may then be administered by needle puncture of the septum in the port or periodic blood samples may be acquired if appropriate flushing techniques are followed. A complete description of the uses of the port is contained in the Instructions for Use/User's Manual in Appendix B.
When used with a power injectable needle infusion set, the NorPort CT-PC Port is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/s with a 19 and 20 gauge non-coring power injectable needle and 2 ml/s with a 22-gauge non-coring power injectable needle.
#### 5.8 Special Precautions for Disposal
Any sharp instrument used in the procedure should be properly disposed of according to the institution policy. Any remaining parts or empty packages do not require special handling when disposing.
#### રું. કે જેવી સાંતર તાલુકામાં આવેલું એક ગામનાં મુખ્યત્વે ખેતી, ખાતે કરવામાં આવેલું એક ગામનાં મુખ્યત્વે ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં મુખ્યત્વે ખેતી, ખેતમજૂરી તેમ Sterilization Information
The NorPort CT-PC Port and the Lucent Non-coring needle infusion set are provided sterile and are recommended for single use only. In the event that the product becomes contaminated (not by bodily fluids or tissue) prior to use, replace it with sterile product and return the device to Norfolk Medical for repackaging and re-sterilization.
। ર
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the words "Norfolk Medical" in a bold, serif font. There is a logo between the two words that is a stylized letter "C". There is a trademark symbol below the logo.
7350 N. Ridgeway, Skokie, IL 60076 • 847-674-7075
#### 5.10 Validity Period
.
As long as the package remains unopened and undamaged the product is valid for use. The materials included with the NorPort kit, the Lucent Non-coring needle and the NorPort CT·PC Port do not degrade over time.
#### Special Precautions for Handling and Storage 5.11
Products should be stored in a cool and dry place. It should only be open prior to use.
#### Description of Package (NorPort CT PC Port Kit) · 5.12
The NorPort CT-PC Port Kit includes a port and catheter and the following inside the sterile package:
- 1 Non-coring Point Needle (straight needle)
- 1-Vein Pick (retraction/introduction device)
- 1- Tunneling trocar (atraumatic tip)
- 1- Blunt needle
- 1-10 mL syringe
- 1- Locking Mechanism
- 1- Patient chart Sticker
- 1-. User's Manual
The plastic tray is contained within a heat-sealed polyethylene/nylon Tyvek® header bag (pouch) and sterilized. If requested a percutaneous introducer kit is included in the NorPort CT PC Port kit package. This Port/Introducer Kit is listed with its own catalog number. This kit might be ordered separate or as part of the kit as requested by the physician.
The percutaneous introducer kit, or "Full-Kit" as described in the brochure and catalog, contains the following:
- 1- Basic set
- 1- "Split Sheath" percutaneous introducer
- 1- J-Flex guide wire with thumb advancer
- 1- Introducer needle
In addition, the NorPort CT·PC Port "Full-Kit" is packaged with a "User's Manual" and "Patient Implant" stickers.
Each Kit.is packaged in an external fiberboard box. Each box is labeled accordingly to match the item inside the box.
{6}------------------------------------------------
# Norfolk C Medical
7350 N. Ridgeway, Skokie, IL 60076 • 847-674-7075
#### 5.13 Labeling
All products are properly labeled on its plastic container with a stick on label . containing the following information. Refer to appendix C for sample labels.
- Company Name, Address and Contact Number
- Product Name (Brand Name and Common Name
- Catalog Number
- Catheter Size
- Lot Number
- Units included
- Restricted Device Note: Federal Law restricts this device to sale by or on the order of a physician
- Sterile unopened undamaged package
- For single use only statement
#### Instruction for Use 5.14
Note: Detailed instructions for use and care of the NorPort CT·PC are in the "USER'S MANUAL. A simple listing of implant instruction is listed here:
1) Before implanting inspect the port thoroughly. Do not use if holes, cracks, or surface contaminations are present.
2) Flush all air from the port prior to placement using the 20ga Non-coring point needle and syringe with heparinized saline, which is provided with the kit.
3) The selected site for the reservoir body should be over a bony structure and in a location that is convenient and comfortable for the patient.
4) Place the catheter into the vein using the "cut-down" technique or by using a percutaneous introducer.
5) Place the tip of the catheter in an area of high blood flow when placing it in the venous system. Fluoroscopy is highly recommended to verify proper placement of the catheter tip in the junction of the superior vena cava/high right atrium.
6) Take special care not to serrate the catheter tip or occlude it during the catheter placement process. Leave sufficient slack upon placement so the patient movement does no stress the catheter.
7) Position the pocket for the reservoir so that the suture line is not directly over the port. Do not place the port too deep or to shallow. A depth of approximately 5mm under the skin surface is recommended as the optimal placement of depth.
{7}------------------------------------------------
## Norfolk ( Medical
7350 N. Ridgeway, Skokie, IL 60076 • 847-674-7075
8) Cut the catheter to the proper length and moisten all components with saline.
9) Slide the catheter lock over the catheter.
10) Slide the catheter over the barbed outlet tube (pin connector) of the reservoir.
11) Slide the catheter lock and catheter forward until the catheter and the outlet tube are completely covered.
12) Test the connection by gently tugging on the catheter.
13) Secure the port to the underlying fascia with at least three non-absorbable sutures.
14) After suturing has been satisfactorily completed, flush the incision with an appropriate antibiotic to ensure a sterile pocket.
I 5) Before closure, check patency and flow through the NorPort CT PC Port by x-ray, fluoroscopy, or by imaging technique of choice.
16) After each use, always leave the NorPort CT·PC Port filled with a heparinized saline solution in a concentration recommended by your institution.
17) When the NorPort CT·PC Port is to be used for "power injection" of special fluids, a power injectable needle infusion set must be used. For power injection of contrast media, the maximum recommended infusion rate is 5ml/s with a 19 or 20 gauge non-coring power injectable needle and 2 ml/s with a 22 gauge needle.
A copy of the Instructions for User's Manual is attached in Appendix B, which contains detailed instructions.
#### Technological Characteristics Summary ર્ . 15
The following Summary Information refers to the Norfolk Medical NorPort CT.PC Port and catheter, and the Lucent Non-coring needle
## Is the new device compared to Marketed Device?
- Yes, the Norfolk Medical NorPort CT·PC Port and catheter, and the Lucent Non-Coring needle Infusion Set are compared to legally marketed predicate devices.
{8}------------------------------------------------
## Norfolk (
7350 N. Ridgeway, Skokie, IL 60076 • 847-674-7075
### Does the new device have the same indications for use statement?
- Yes, the Norfolk Medical NorPort CT PC Port and catheter, and the Lucent Non-coring needle have the same Indications For Use statement to legally marketed predicate devices.
## Do the differences alter the intended therapeutic/diagnostic/etc. effect (i.e. in deciding, impact on safety and effectiveness may be considered)?
- No, the differences do not alter the intended use of the port.
### Does the new device have the same technological characteristics, e.g. design, materials, etc?
- No. not in all regards. The NorPort CT·PC Port implanted Port has some minor differences from the predicate devices, however, the fundamental scientific technology of the port/catheter has not changed.
The Lucent Non-coring needle infusion set has some minor differences from the predicate devices, however, the fundamental scientific technology has not changed.
#### Could the new characteristics affect safety and effectiveness?
No, there are no new significant characteristics that would affect safety and effectiveness in the new device.
### Do accepted scientific methods exist for assessing effects of the new characteristics?
Yes. FDA. Guidance on 510(k) Submissions for Implanted Infusion Ports, dated October 1990 was used to as a reference for the performance testing.
BS EN ISO 10555-1:2009, Sterile, single-use intravascular catheters Part I : General requirements, was used as a testing reference.
ISO 11135-1:2007, Sterilization of health care products -Ethylene oxide- Part 1, Requirements for development, validation and routine control of a sterilization process for medical devices. Specifies requirements for the development, validation and routine control of an ethylene oxide sterilization process for medical devices.
{9}------------------------------------------------
# -Norfolk @ Medical
## 7350 N. Ridgeway, Skokie, IL 60076 • 847-674-7075
ISO 10993-7: 2008, Biological Evaluation of Medical Devices- Part 7: Ethylene Oxide Sterilization Residuals.
ISO 14971: 2007, Medical Devices- Application of Risk Management to Medical Devices.
## Are performance data available to assess effects of new characteristics?
Yes, verification testing was performed according to protocols based on the above referenced guidance document recommendation and additional standards, with positive performance examination results available as objective evidence in each category.
## Do performance data demonstrate equivalence?
Yes, performance data gathered in the design verification testing demonstrated that the NorPort family of products is substantially equivalent to the predicate devices mentioned in this submission.
#### Non-Clinical Data 5.16
The data collected from the non-clinical tests demonstrated that the functionality and performance characteristics of the NorPort CT-PC implanted vascular port and the Lucent Non-coring needle infusion set are comparable to the currently marketed vascular ports and power capable needles. The tests performed include: catheter to port connection, septum puncture/port leak, patency verification, and high pressure static testing.
#### 5.17 Conclusion
The NorPort CT·PC Port and the Lucent Non-coring Needle Infusion Set have met all predetermined acceptance criteria of design verification evaluations through testing examination. Based on the FDA's decision tree, it is logically concluded through evidence that the above mentioned medical devices are substantially equivalent to the predicate devices: Norfolk Medical NorPort 510 (k)#102385, 2010; the Angiodynamics SmartPort CT Port Access System 510 (k)#081472, 2008, the Smiths Medical PORT-A-CATH Titanium Venous Access System 510 (k)# 070116, 2007, the Bard PowerPort Implanted Titanium Port Chronoflex with 8 Fr. Catheter 510 (k)# 060812, 2006, the MEDCOMP ProFUSE Port 510 (k)# 070003, 2007, the PHS Medical C-Port 510 (k)# 091099, 2009 and the Norfolk Medical NorPort 510 (k)# 102385, 2010; Smiths Medical Gripper Plus Power P.A.C Needle (510 (k)# 070116), the Angiodynamics Lifeguard CT Safety Infusion Set (510 (k)#072375), the Bard PowerLoc Clear Safety Infusion Set (510 (k)# 082306), and the MEDCOMP CT Injectable Safety Non- coring Needle (510 (k)# 080544).
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle, there is a stylized symbol that resembles an abstract caduceus, with three curved lines representing the serpent and a single vertical line representing the staff.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Michael J. Dalton President Norfolk Medical 7350 North Ridgeway Skokie, Illinois 60076
## AUG 25 2011
Re: K111101
Trade/Device Name: Lucent Non-coring Needle, Huber Needle Intravascular Administration Set-FPA, Nor-Port CT-PC (Family of Implanted Vascular Access Devices) Port and Catheter, Implanted, Subcutaneous, Intravascular-LJT Regulation Number: 21 CFR 880.5965 Regulation Name: Subcutaneous, Implanted, Intravascular Infusion Port and Catheter Regulatory Class: II Product Code: LJT Dated: August 9, 2011 Received: August 10, 2011
Dear Mr. Dalton:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave rottowed your sociation mined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device mensiale conmisered proct to may 20, 10, 10 reclassified in accordance with the provisions of Anchuments, or to actives and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing procities requirements for annual regionant misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your (I MA), It may be subject to additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{11}------------------------------------------------
#### Page 2 - Mr. Dalton
Please be advised that FDA's issuance of a substantial equivalence determination does not Flease be advised that I Dr. o lesualites on that your device complies with other requirements mean that I DA nas made a determinations administered by other Federal agencies. of the Act of ally I edelar statues and regisments, including, but not limited to: registration You must comply with an the reet 8 recess and 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803), good manufacturing (reporting of medical device teated the quality systems (QS) regulation (21 CFR Part 820); practice requirements as set return in the qualisy of provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), II you desire specific actice to: JoboutFDA/CentersOffices/CDRH/CDRHOffices please go to mtp.//www.lenter for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to Compilatic. Also, produce note the regar. 97). For questions regarding the reporting of premarket netification of the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's http://www.lad.goveree and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the You may onlain other general miorniational and Consumer Assistance at its toll-free Division of Siman Manatalations (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
ph for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
## Enclosure
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the words "Norfolk Medical" in a bold, serif font. There is a logo between the two words that is a black, stylized image of a heart with a curved line around it. The letters are all black and the logo is also black.
7350 N. Ridgeway, Skokie, IL 60076 • 847-674-7075
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Section 4
INDICATIONS FOR USE STATEMENT 510(k) Number:
510(k) Number (if known): K111101
## Device Name:
Lucent Non-coring Needle, Huber Needle Intravascular Administration Set-FPA
Needle Set:
The Lucent Non-coring needle set is intended for the administration of fluids and drugs as well as blood sampling through implanted vascular ports.
When used with the NorPort CT·PC Family of Implanted Vascular Access Device is also indicated for power injection of contrast media into the central venous system. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/s with a 19 and 20 gauge non-coring power injectable needle and 2 ml/s with a 22-gauge non-coring power injectable needle.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ali Chy 8/25/11
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: _______________________________________________________________________________________________________________________________________________________________
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the words "Norfolk Medical" in a serif font. The word "Norfolk" is on the left, followed by a logo in the shape of a stylized ear. The word "Medical" is on the right. The logo is black, and the words are black.
(Division Sign-Off) 7350 N. Ridgeway, Skokie, IL 60076 . 847-674-7075 Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
Section 4
## INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K111101
## Device Name:
NorPort CT-PC (Family of Implanted Vascular Access Devices) Port and Catheter, Implanted, Subcutaneous, Intravascular-LJT
#### Catheter /Port:
The NorPort CT.PC Family of Implanted Vascular Access Devices is indicated for use when the patient requires the following: repeated access to the vascular system for injections, infusion of drugs, administration of blood or blood products, and/or withdrawal of blood as part of the therapy regimen.
When used with a appropriately labeled power injectable needle infusion set, the NorPort CT.PC Port is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/s with a 19 and 20 gauge non-coring power injectable needle and 2 ml/s with a 22-gauge noncoring power injectable needle.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: 仅11///////
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.