K030083 · Boston Scientific Corp · LJT · Jan 28, 2003 · General Hospital
Device Facts
Record ID
K030083
Device Name
VAXCEL WITH PASV PORT
Applicant
Boston Scientific Corp
Product Code
LJT · General Hospital
Decision Date
Jan 28, 2003
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5965
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Vaxcel™ w/ PASV® Port device is intended for use in patients who require long-term access to the central venous system for administration of fluids including but not limited to hydration fluids, antibiotics, chemotherapy, analgesics, nutritional therapy, and blood products. The device is also indicated for blood specimen withdrawal.
Device Story
Vaxcel™ with PASV® implantable port system provides long-term central venous access; enables administration of fluids/medications and blood specimen withdrawal. System comprises subcutaneous port and catheter; available in standard or mini configurations; constructed from titanium or polysulfone. Device implanted by physician; used in clinical settings. Port allows percutaneous access to venous system via needle; PASV® technology facilitates fluid flow and blood sampling. Benefits include reliable, repeated access for chronic therapies; reduces need for frequent venipuncture.
Clinical Evidence
Bench testing only; biocompatibility testing performed per ISO 10993.
Technological Characteristics
Implantable subcutaneous port and catheter system. Materials: Titanium or Polysulfone. Biocompatibility: ISO 10993 compliant. Configuration: Standard or mini port sizes. Non-electronic, mechanical device.
Indications for Use
Indicated for patients requiring long-term central venous access for fluid administration (hydration, antibiotics, chemotherapy, analgesics, nutritional therapy, blood products) and blood specimen withdrawal.
Regulatory Classification
Identification
A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.
Predicate Devices
Vaxcel™ Implantable port system
PASV® implantable port system
Vaxcel™ with PASV® PICC
Related Devices
K032008 — VAXCEL WITH PASV PORT, MODEL 45-213 · Boston Scientific Corp · Jul 14, 2003
K031844 — VAXCEL WITH PASV PORT (LOW PROFILE), MODELS 45-236 AND 45-238 · Boston Scientific Corp · Jul 15, 2003
K032375 — VAXCEL PORT, MODELS 45-336 AND 45-338 · Boston Scientific Corp · Aug 19, 2003
K032600 — VAXCEL PORT, MODEL 45-333 AND VAXCEL W/PASV PORT, MODEL 45-233 · Boston Scientific Corp · Sep 9, 2003
K982888 — VAXCEL IMPANTABLE VASCULAR ACCESS SYSTEM · Boston Scientific Corp · Oct 20, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
JAN 2 8 2003
Image /page/0/Picture/1 description: The image shows the text "KO30083" in a handwritten style. The text is written in black ink on a white background. To the right of the text is the word "VaxcelTM".
## Summary of Safety and Effectiveness
| General Provisions | Trade Name: Vaxcel™ with PASV® implantable port system |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Classification Name: Implanted Subcutaneous Port and Catheter, 80 LJT |
| Name of Predicate Devices | Vaxcel™ Implantable port system<br>PASV® implantable port system<br>Vaxcel™ with PASV® PICC |
| Classification | Class II |
| Performance Standards | Performance Standards have not been established by FDA under Section 514 of the Food, Drug and Cosmetic Act |
| Intended Use and Device Description | The Vaxcel™ w/ PASV® Port device is intended for use in patients who require long-term access to the central venous system for administration of fluids including but not limited to hydration fluids, antibiotics, chemotherapy, analgesics, nutritional therapy, and blood products. The device is also indicated for blood specimen withdrawal. The device is available in standard and mini port configurations made of either Titanium or Polysulfone materials. |
| Biocompatibility | The Vaxcel™ with PASV® implantable port system have been tested for biocompatibility per ISO 10993. All data demonstrate this device is biocompatible for its intended use. |
| Summary of Substantial Equivalence | The Vaxcel™ with PASV® implantable port system have been tested and compared to the predicate device. All data gathered demonstrate this device as substantially equivalent. No new issues of safety or efficacy have been raised. |
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three lines representing its head and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 8 2003
·Ms. Jodi L. Greenizen Regulatory Affairs Project Manager Boston Scientific Corporation NAMIC Technology Center 10 Glens Falls Technical Park, Dix Avenue Glens Falls, New York 12801
Re: K030083
Trade/Device Name: Vaxcel™ with PASV® Implantable Port System Regulation Number: 880. 5965 Regulation Name: Subcutaneous, Implanted, Intravascular Infusion Port Catheter Regulatory Class: II Product Code: LJT Dated: January 8, 2003 Received: January 9, 2003
Dear Ms. Greenizen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
## Page 2 - Greenizer
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if annlicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours,
Susan Runpe
Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{3}------------------------------------------------
·
## Indications For Use
| 510(k) Number (if known) | Unknown K 030083 |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Vaxcel™ with PASV® implantable port system |
| Indications for Use | The Vaxcel™ w/ PASV® Port device is intended for use in patients who require long-term access to the central venous system for administration of fluids including but not limited to hydration fluids, antibiotics, chemotherapy, analgesics, nutritional therapy, and blood products. The device is also indicated for blood specimen withdrawal. |
## (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
| | Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--|--------------------------------------------------------|
|--|--------------------------------------------------------|
| Prescription Use | <span></span> | OR | Over-The Counter Use |
|----------------------|---------------|----|--------------------------|
| (Per 21 CFR 801.109) | | | (Optional Format 1-2-96) |
| | <i>Patricia Cucente</i> |
|-----------------------------------------------|------------------------------------------------------------------------------------|
| | (Division Sign-Off) |
| Confidential<br>Boston Scientific Corporation | Division of Anesthesiology, General Hospital,<br>Infection Control, Dental Devices |
| 510(k) Number: | K030083 |
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.