K122265 · Yukon Medical, LLC · LHI · Aug 14, 2012 · General Hospital
Device Facts
Record ID
K122265
Device Name
VIALOK SINGLE VIAL ACCESS DEVICE VENTED
Applicant
Yukon Medical, LLC
Product Code
LHI · General Hospital
Decision Date
Aug 14, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The Vented Single Vial Access Device is a stand-alone, single-use, disposable device which permits access to a medication vial without the use of a needle. The device is intended for use by healthcare professionals in a wide variety of healthcare environments, including hospitals, healthcare facilities, and pharmacies. The Vented Single Vial Access Device is indicated for use with standard medication vials and mating luer access devices for withdrawal and/or injection of fluid.
Device Story
Sterile, single-use, disposable needle-free vial access device; enables fluid transfer/mixing between standard medication vials and luer-compatible delivery systems. Device features dual-lumen spike for fluid path and pressure-equalizing filtered air path; includes shroud (20mm/13mm models) or universal design for vial attachment. Operated by healthcare professionals in clinical/pharmacy settings. Facilitates safe drug reconstitution/withdrawal; eliminates needle-stick risk during vial access. Performance verified via bench testing including attachment/detachment force, flow rate, leak testing, and filter integrity.
Clinical Evidence
Bench testing only. No clinical data. Testing included attachment/detachment force, flow rate, pressurization/vacuum leak tests, priming volume, filter integrity, septum coring, multiple access, ISO 594-2 luer performance, and ISO 10993 biocompatibility (cytotoxicity, irritation, sensitization, systemic toxicity, hemocompatibility).
Technological Characteristics
MABS body; filter sub-assembly (nylon substrate, polyethylene housing); polyethylene cap/spike cover. Fluorosilicone polymer spike lubricant. Dual-lumen spike for fluid/vented air. Dimensions: 20mm, 13mm, or universal. Connectivity: mechanical luer interface. Sterilization: not specified. Biocompatibility: ISO 10993 compliant.
Indications for Use
Indicated for healthcare professionals in hospitals, clinics, and pharmacies for needle-free access to standard medication vials for fluid withdrawal or injection using mating luer access devices.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
K072511 — SWABABLE VIAL ADAPTER 13MM, SWABABLE VIAL ADAPTER 20MM · Medimop Medical Projects, Ltd. · Dec 13, 2007
K080989 — UNIVERSAL VIAL ACCESS SPIKE · Icu Medical, Inc. · May 2, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
K 12 2265
510(k) Notification
Yukon Medical, LLC
Vented Single Vial Access Device
### 5. 510(k) Summary
5.2.
### 5.1. Submitter Information
# AUG 1 4 2012
| Company Name: | Yukon Medical, LLC |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Address: | 2200 Gateway Centre Blvd<br>Suite 208<br>Morrisville, NC 27560 |
| Company Phone: | (919) 595-8250 |
| Company Fax: | (919) 595-8251 |
| Contact Person: | Carl Dupper, Director of Engineering and Quality<br>Yukon Medical<br>Phone: 919-595-8257<br>Fax: 919-595-8251<br>Email: cdupper@yukonmedical.com |
| Prepared By: | Kristin Benokraitis, Director of Programs and Quality<br>Management<br>Gilero, LLC<br>4022 Stirrup Creek Drive Suite 300<br>Durham, NC 27703<br>Phone: 919-595-8223<br>Fax: 919-595-8221<br>Email: kristinb@gilero.com |
| Date Summary Prepared: | June 29, 2012 |
| Device Identification | |
| Trade/Proprietary<br>Name: | ViaLokTM 20 mm Single Vial Access Device,<br>Vented, Female Luer |
| | ViaLokTM 13 mm Single Vial Access Device,<br>Vented, Female Luer |
| | Universal Single Vial Access Device, Vented,<br>Female Luer |
| Common Name: | Vial Access Device |
{1}------------------------------------------------
| Classification | Intravascular Administration Set |
|----------------------|----------------------------------|
| Name: | 21 CFR 880.5440, Class II |
| Classification Panel | General Hospital |
| Product Code: | LHI, I.V. Fluid Transfer Set |
#### 5.3. Predicate Device
The Vented Single Vial Access Device is substantially equivalent to the following predicate device:
| Device | Manufacturer | 510(k) | Date Cleared |
|-------------------------------------------|-----------------------------------|---------|------------------|
| ViaLokTM<br>Non-vented | Yukon Medical | K121182 | May 3, 2012 |
| Vented Vial<br>Adapter Transfer<br>Device | Medimop Medical<br>Projects, LTD. | K062482 | November 3, 2006 |
#### Device Description 5.4.
The Vented Single Vial Access Device (ViaLokTM / Universal) is a sterile, stand-alone, single-use, disposable device which permits access to a medication vial without the use of a needle. These devices are inserted into a stopper of a medication vial. The healthcare provider uses the Vented Single Vial Access Device to transfer and mix drugs contained in standard medication vials. The Vented Single Vial Access Device includes three product configurations in this submission:
- ViaLok™ 20 mm Single Vial Access Device, Vented, Female Luer .
- . ViaLok™ 13 mm Single Vial Access Device, Vented, Female Luer
- . Universal Single Vial Access Device, Vented, Female Luer
The 20mm and 13mm product configurations are intended to mate with standard medication vial enclosure sizes (20mm and 13mm, respectively). The Universal device does not contain a shroud and can be used independent of enclosure size. Each Vented Single Vial Access Device configuration is offered with a female luer.
#### 5.5. Intended Use
The Vented Single Vial Access Device is a stand-alone, single-use, disposable device which permits access to a medication vial without the use of a needle. The device is intended for use by healthcare professionals in a wide variety of healthcare environments, including hospitals, healthcare facilities, and pharmacies.
{2}------------------------------------------------
The Vented Single Vial Access Device is indicated for use with standard " medication vials and mating luer access devices for withdrawal and/or injection of fluid.
The indications statement for the Vented Single Vial Access Device is the same as its predicate device, Yukon Medical's ViaLok™ Non-vented (K121182).
### 5.6. Predicate Device Comparison - Technical Characteristics Equivalency of technical characteristics is demonstrated through a direct comparison of the Vented Single Vial Access Device and the predicate devices listed in the table below.
| | Device | | |
|-----------------------------|--------------------------------------------------------|-----------------------------------------------|-------------------------------------------------------------------|
| Technical<br>Characteristic | Subject Device:<br>Vented Single Vial<br>Access Device | Predicate: ViaLok™<br>Non-vented<br>(K121182) | Predicate:<br>Vented Vial Adapter<br>Transfer Device<br>(K062482) |
| Spike | Yes | Yes | Yes |
| Shroud | 20mm and 13mm - Yes<br>Universal - No | Yes | Yes |
| Luer Access | Yes | Yes | Yes |
| Filter | Yes | No | Yes |
### Spike
The spike is used to penetrate a standard medication vial stopper and provide fluid and filtered air paths.
The subject device and the predicate (Vented Vial Adapter Transfer Device) both have equivalent dual lumen spikes; one lumen for fluid transfer and the other which allows for pressure equalization with vented air.
### Shroud
The purpose of the shroud is to secure the device to a standard medication vial after the stopper is penetrated. The 20mm and 13mm product configurations contain this feature which is equivalent to that of the ViaLokTM Non-vented design.
### Luer Access
All configurations of the subject device facilitate connection of mating luer access devices. This is equivalent to referenced predicate devices.
### Filter
Air flows into and out of the vial through a filter during drug transfer/withdrawal.
{3}------------------------------------------------
The subject device and the predicate (Vented Vial Adapter Transfer Device) both provide equivalent means of air filtration.
### Materials
The Vented Single Vial Access Device is an assembly of a molded methyl methacrylate acrylonitrile butadiene styrene copolymer (MABS) body and filter sub-assembly. The filter sub-assembly is comprised of a non-woven nylon substrate secured to a molded polyethylene housing.
The Vented Single Vial Access Device may be provided with a molded polyethylene cap and/or a molded polyethylene spike cover.
The subject device's spike may also be lubricated with a fluorosilicone polymer.
The Vented Single Vial Access Device materials do not contain natural rubber latex.
The Vented Single Vial Access Device has been tested and meets the biological requirements outlined in ISO 10993-1, ISO 10993-4, ISO 10993-5, ISO 10993-9, ISO 10993-10, and ISO 10993-11. A summary of these test results is provided in Section 15 - Biocompatibility.
#### 5.7. Predicate Device Comparison - Performance Characteristics
The performance data supplied with this submission demonstrates that the Vented Single Vial Access Device meets all specified requirements and is substantially equivalent to the predicate device(s).
Performance data for the Vented Vial Adapter Transfer Device was not available. Therefore, samples of the predicate device were tested along with the Vented Single Vial Access Device.
The following tests were conducted on the Vented Single Vial Access Device to demonstrate equivalency of the performance characteristics to the predicate device(s):
- Attachment Force .
- Flow Rate .
- Pressurization Leak Test .
- Vacuum Leak Test .
- Priming Volume .
- Detachment Force ●
- Filter Integrity .
- Septum Coring .
- Multiple Access .
{4}------------------------------------------------
510(k) Notification Yukon Medical, LLC Vented Single Vial Access Device
- ISO 594-2 Test Methods (Vented Single Vial Access Device tested . only)
- Luer Leakage (Air Ingress) o
- Luer Leakage (Fluid Ingress) ্ত
- Luer Attachment o
- Luer Separation Force o
- o Unscrewing Torque
- Resistance to Overriding ্
- O Stress Cracking
- Biocompatibility ISO 10993 (Vented Single Vial Access Device . tested only)
- o Cytotoxicity by Elution Test (Cytotoxicity)
- Intracutaneous Reactivity (Irritation or Intracutaneous ০ Reactivity)
- Maximization Test for Delayed Hypersensivity (Sensitization) ୍
- Acute Systemic Toxicity (Systemic Toxicity (Acute)) ୍
- Evaluation of Hemocompatibility: Interaction with Blood o (Hemocompatibility)
Test results demonstrate that the Vented Single Vial Access Device is as safe and effective as the legally marketed devices designated as predicate device(s).
#### 5.8. Conclusion
Test results demonstrate that the Vented Single Vial Access Device is as effective, and performs at least as safely and effectively as the legally marketed devices designated as predicate devices.
Based on comparisons of the device's intended use, technology and performance characteristics, the Vented Single Vial Access Device is substantially equivalent to the indicated predicate devices.
Any differences between the Vented Single Vial Access Device and the equivalent predicate devices have no significant influence on safety or effectiveness.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans. The eagle is a stylized design, with simple lines and curves.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Yukon Medical, Limited Liability Company AUG 1 4 2012 C/O Mark Job Responsible Third Party Official Regulatory Technology Services, Limited Liability Company 1394 250 Street, North West Buffalo, Minnesota 55313 Re: K12265 Trade/Device Name: ViaLok13 20 mm Single Vial Access Device, Vented, Female Luer ViaLok1M 13 mm Single Vial Access Device, Vented, Female Luer Universal Single Vial Access Device, Vented, Female Luer Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: LHI Dated: July 27, 2012 Received: July 30, 2012
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In
{6}------------------------------------------------
Page 2- Mr. Job
addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
th for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### Indications for Use Statement 4.
510(k) Number (if known):
Device Name:
- ViaLok™ 20 mm Single Vial Access Device, Vented, Female Luer ●
- ViaLok™ 13 mm Single Vial Access Device, Vented, Female Luer .
- . Universal Single Vial Access Device, Vented, Female Luer
Indications for Use: The Vented Single Vial Access Device is a stand-alone, singleuse, disposable device which permits access to a medication vial without the use of a needle. The device is intended for use by healthcare professionals in a wide variety of healthcare environments, including hospitals, healthcare facilities, and pharmacies.
The Vented Single Vial Access Device is indicated for use with standard medication vials and mating luer access devices for withdrawal and/or injection of fluid.
Prescription Use ਮ (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
hild C. Chapman 8/13/12
(Division Sign-Off) Division of Anesthesiology, General Hospital infection Control, Dental Devices
510(k) Number: k122265
4-1
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.