K060986 · Steven H. Warnock, M.D. · FRP · Jul 11, 2006 · General Hospital
Device Facts
Record ID
K060986
Device Name
HEAD BED
Applicant
Steven H. Warnock, M.D.
Product Code
FRP · General Hospital
Decision Date
Jul 11, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5680
Device Class
Class 1
Attributes
Therapeutic, Pediatric
Indications for Use
Deformational plagiocephaly is an abnormal flattening of the infant's head resulting from lying in one position. Some medical conditions predispose infants to this condition. Frequent changes in head position can help prevent or correct deformational plagiocephly. The Head Bed device is intended for use as an infant sleep positioner. The Head Bed device is used to position the head so as to aid in the prevention of deformational plagiocephaly. It is intended for use in infants less than 6 months of age or in those infants who cannot roll from back to front. The device is placed on the mattress, around the head to control head position. This selective positioning of the head allows some control over the forces that create flattening of the head.
Device Story
Head Bed is a U-shaped soft polyurethane foam cushion with a flame-retardant cloth cover. Device placed on mattress around infant's head to maintain position; prevents deformational plagiocephaly by controlling forces causing head flattening. Used for infants <6 months or those unable to roll from back to front. Operated by caregivers in home environment. Provides passive support to maintain head orientation.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
U-shaped polyurethane foam cushion; flame-retardant fabric cover; dimensions 17.5cm x 18.5cm x 4cm; Velcro closure. Passive mechanical support device; no energy source or software.
Indications for Use
Indicated for infants aged 0-6 months to aid in the prevention of deformational plagiocephaly. Intended for infants who cannot roll from back to front.
Regulatory Classification
Identification
A pediatric position holder is a device used to hold an infant or a child in a desired position for therapeutic or diagnostic purposes, e.g., in a crib under a radiant warmer, or to restrain a child while an intravascular injection is administered.
Predicate Devices
NightForm Infant Sleep Positioner
Related Devices
K062143 — KOZY COMFORT INFANT POSITIONER · Alfred E Mann Inst. For Biomedical Engineering · Oct 24, 2006
K082367 — INFANT SLEEP BEANIE · Life Innovations, LLC · Jun 12, 2009
K041996 — NIGHTFORM · Plastic Surgery Solutions, LLC · Aug 3, 2004
K051300 — ROBIN HOOD VEST · Kamber Corporation · Dec 8, 2005
K013452 — DANMAR PRODUCTS CRANIAL ADJUSTIVE PROSTHESIS · Danmar Products, Inc. · Oct 31, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
Kα60986
page 1 of 2
### 510(k) Summary
JUL 1 1 2006
#### Applicant Information
Submitter
Contact information
Steven H. Warnock, M.D., P.C. 9829 S. 1300 E. Suite 200 Sandy, UT 84094
Steven Warnock, M.D. Phone: 801-571-2020 Fax: 801-571-6899 email: swarnockmd(a)yahoo.com
#### Submitter Registration Number: K060986
### Device Name and Classification:
| Common Name: | Infant Sleep Positioner |
|----------------------|-------------------------------------------------|
| Classification Name: | Holder, Infant Position, Code FRP, CFR 880.5680 |
| Trade Name: | Head Bed™ |
| Class: | Class 1 |
#### Device Description:
The Head Bed device is a U-shaped soft polyurethane foam cushion covered in flame retardant cloth material. Outside dimensions are 17.5cm (w) x 18.5cm (1) x 4cm (h). Inside dimensions are 13.5cm (h) x 6.5cm (w). The device is adequately soft so as to provide support for the infants head without being abrasive. The removable cloth cover is flame retardant and machine washable.
#### Intended Use:
Deformational plagiocephaly is an abnormal flattening of the infant's head resulting from lying in one position. Some medical conditions predispose infants to this condition. Frequent changes in head position can help prevent or correct deformational plagiocephly. The Head Bed device is intended for use as an infant sleep positioner. The Head Bed device is used to position the head so as to aid in the prevention of deformational plagiocephaly. It is intended for use in infants less than 6 months of age or in those infants who cannot roll from back to front. The device is placed on the mattress, around the head to control head position. This selective positioning of the head allows some control over the forces that create flattening of the head.
#### Predicate Device:
The subject device is substantially equivalent to a predicate device NightForm Infant Sleep Positioner.
{1}------------------------------------------------
KUCO980
#### ﻴﺮ ﺍﻟﻤﺮﺍﺟﻊ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ 2010 C
.
## Predicate Comparison:
.
| Comparison | Head Bed™ Positioner | NightForm™ Infant Sleep Positioner |
|-------------------|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| Application | Sleep positioning | Sleep positioning |
| Target population | Infants 0-6 months | Infants 0-9 months |
| Purpose | To maintain infant sleep position for prevention of deformational plagiocephaly. | To maintain infant sleep position for prevention of deformational plagiocephaly. |
| Materials | Fire-resistant foam and a fire retardant fabric cover with Velcro closure. | Fire-retardant foam and fabric (cotton or poly/cotton blend), mechanical hook & loop fasteners, nylon zipper. |
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three heads facing to the right. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" are arranged in a circular fashion around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
### JUL 1 1 2006
Dr. Steven H. Warnock President Steven H. Warnock, M.D. 9829 South 1300 East, Suite 200 Sandy, Utah 84094
Re: K060986
Trade/Device Name: Head Bed™ Infant Positioner Regulation Number: 880.5680 Regulation Name: Pediatric Position Holder Regulatory Class: I Product Code: FRP Dated: March 6, 2006 Received: March 18, 2006
Dear Dr. Warnock:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (first the indications for use stated in the enclosure) to legally marketed predicate devices may keted in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the gremeral controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Codc of Federal Regulations, Title 21, Parts 8010 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2 -Dr. Warnock
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours.
Cure Salion for
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K06 0986
# Indications For Use
510(k) Number (if known): K060986
Head Bed™ infant positioner Device Name:
Indications For Use:
The Head Bed™ infant positioner is indicated for infants aged 0-6 months to aid in the prevention of deformational plagiocephaly.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use X (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for Hach
Sign-Off)
n of Anesthesiology, General Hospital,
won Control, Dental Devices
0906
i, humbat: K. D.C.R.C.
Page 1 of
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.