CADD-PCA, -MICRO, -PRIZM AMBULATORY INFUSION PUMP

K955180 · Sims Deltec, Inc. · FRN · May 23, 1996 · General Hospital

Device Facts

Record IDK955180
Device NameCADD-PCA, -MICRO, -PRIZM AMBULATORY INFUSION PUMP
ApplicantSims Deltec, Inc.
Product CodeFRN · General Hospital
Decision DateMay 23, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5725
Device ClassClass 2
AttributesTherapeutic

Indications for Use

CADD-PCA® Model 5800 or 5800R Pump: The CADD-PCA pump is indicated for intravenous, subcutaneous, epidural space or subarachnoid space infusions. CADD-Micro® Model 5900 Pump: The CADD-Micro pump is indicated for intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural space or subarachnoid space microinfusions. CADD-Prizm™ Model 6100 Pump: The CADD-Prizm pump is indicated for intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural space or subarachnoid space infusions.

Device Story

Microcomputer-controlled ambulatory infusion pumps deliver medication via parenteral routes (intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural, subarachnoid). Devices operate via programmed infusion parameters to deliver medication through compatible catheters. Used in clinical or ambulatory settings by healthcare providers or patients. Output consists of controlled fluid delivery; informs clinical pain management or therapeutic infusion regimens. Benefit includes expanded site-specific delivery options for existing pump platforms.

Clinical Evidence

No clinical studies or functional testing were performed. The submission relies on the existing clearance of the pump hardware and the established clinical indication for subarachnoid use associated with compatible catheter systems.

Technological Characteristics

Microcomputer-controlled ambulatory infusion pumps. No hardware or software modifications were made for this labeling expansion. Connectivity and materials are consistent with previously cleared versions of the CADD-PCA, CADD-Micro, and CADD-Prizm pump lines.

Indications for Use

Indicated for patients requiring medication delivery via intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural, or subarachnoid routes using microcomputer-controlled ambulatory infusion pumps.

Regulatory Classification

Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} K955180 510(k) Summary Date Prepared: November 8, 1995 ## I. GENERAL INFORMATION Applicant's name and address: SIMS Deltec, Inc. 1265 Grey Fox Road St. Paul, MN 55112 Contact: David H. Short Manager, Regulatory Affairs (612)628-7214 ## II. PRODUCT INFORMATION Common/Usual Name: Ambulatory Infusion Pumps Classification Name: Infusion Pump Proprietary Name: CADD-PCA® Model 5800 Pump CADD-PCA® Model 5800R Pump CADD-Micro® Model 5900 Pump CADD-Prizm™ Model 6100 Pump ## III. EQUIVALENCE DEVICE COMPARISON IPI AlgoLine™ Catheter Kits manufactured by PS Medical Coporation, Goleta, CA. ## IV. DEVICE DESCRIPTION The subject devices are microcomputer controlled ambulatory infusion pumps intended for the delivery of medication via various parenteral routes, i.e., intravenous, intra-arterial, subcutaneous, intraperitoneal, or epidural space infusion. The purpose of this 510(k) Notification is to add subarachnoid space infusion to the labeling and thereby expanding the intended use of each pump. The subject devices, as a matter of course, will be used for the delivery of medication through catheters that have this indication. Page 1 of 3 241 {1} 510(k) Summary Date Prepared: November 8, 1995 ## IV. DEVICE COMPARISON Provided in Table 1 is a comparison of devices and their respective indications for use. Table 1. Comparison of indications for use of various devices. | | CADD-Prizm™ Pump | CADD-PCA® Pump (Model 5800 & 5800R) | CADD-Micro® Pump | AngoLine™ Catheter Kits | | --- | --- | --- | --- | --- | | Manufacturer | SIMS Deltec, Inc. | SIMS Deltec, Inc. | SIMS Deltec, Inc. | IPI PS Medical Corporation | | SYSTEM | | | | | | Microcomputer controlled pump | YES | YES | YES | n/a | | INDICATIONS FOR USE | | | | | | Intravenous | YES | YES | YES | n/a | | Intra-arterial | YES | NO | YES | n/a | | Subcutaneous | YES | YES | YES | n/a | | Intraperitoneal | YES | NO | YES | n/a | | Epidural space | YES | YES | YES | YES | | Subarachnoid space | NO | NO | NO | YES | ## V. INTENDED USE CADD-PCA® Model 5800 or 5800R Pump: The CADD-PCA pump is indicated for intravenous, subcutaneous, epidural space or subarachnoid space infusions. CADD-Micro® Model 5900 Pump: The CADD-Micro pump is indicated for intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural space or subarachnoid space microinfusions. CADD-Prizm™ Model 6100 Pump: The CADD-Prizm pump is indicated for intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural space or subarachnoid space infusions. Page 2 of 3 {2} 510(k) Summary Date Prepared: November 8, 1995 ## VI. SUMMARY OF STUDIES ### A. Functional Testing No software and/or functional modifications were required as it relates to the addition of the subarachnoid space infusion indication to the labeling of each subject device, therefore, no functional testing was deemed necessary. ### B. Clinical Studies Clinical studies were not deemed necessary. ## VII. CONCLUSION Based upon the information provided above, the addition of subarachnoid space infusion to the labeling of each subject device is substantially equivalent to the labeling of commercially available catheters, such as the IPI AlgoLine™ Catheter Kits manufactured by PS Medical Coporation, with this indication. Page 3 of 3
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...