ACCU-CHEK FLEXLINK PLUS INFUSION SET, ACCU-CHECK LINKASSIST PLUS INSERTION DEVICE
K100704 · Disetronic Medical Systems AG · FRN · May 14, 2010 · General Hospital
Device Facts
Record ID
K100704
Device Name
ACCU-CHEK FLEXLINK PLUS INFUSION SET, ACCU-CHECK LINKASSIST PLUS INSERTION DEVICE
Applicant
Disetronic Medical Systems AG
Product Code
FRN · General Hospital
Decision Date
May 14, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
ACCU-CHEK® FlexLink Plus is an infusion set for the subcutaneous infusion of insulin administered with micro dosage insulin pumps. ACCU-CHEK® LinkAssist Plus is an insertion device, which is intended specifically for placement of compatible ACCU-CHEK® FlexLink Plus infusion sets.
Device Story
ACCU-CHEK FlexLink Plus is a disconnectable infusion set with a soft cannula perpendicular to an adhesive patch; used for subcutaneous insulin delivery from external micro-dosage pumps. Includes a needle protection feature that automatically retracts the introducer needle post-insertion. ACCU-CHEK LinkAssist Plus is an optional, non-invasive, non-sterile, reusable mechanical insertion aid designed to facilitate the placement of the FlexLink Plus infusion set. Both devices are used by patients with insulin-dependent diabetes mellitus. The infusion set interfaces with standard insulin pumps; the insertion device is operated by the patient to ensure proper cannula placement. The system benefits patients by providing a reliable, disconnectable method for insulin delivery with integrated needle safety features to prevent accidental sticks.
Clinical Evidence
No clinical studies were performed. Evidence consists of in vitro functional testing, biocompatibility testing, and human factors usability testing. Usability evaluations involved patients who routinely use insulin pumps and infusion sets; results confirmed that customer requirements were met, primary operating functions performed as intended, and no needle stick injuries occurred during use.
Technological Characteristics
Infusion set: soft cannula, flexible catheter, adhesive patch, luer connector, automatic needle retraction mechanism. Insertion device: mechanical, non-sterile, reusable, includes safety catch. Materials: biocompatible. Connectivity: interfaces with commercially available insulin infusion pumps. No software or electronic components.
Indications for Use
Indicated for patients with insulin dependent diabetes mellitus requiring subcutaneous insulin infusion via micro dosage insulin pumps.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
ACCU-CHEK Ultraflex infusion set
ACCU-CHEK LinkAssist insertion device
Related Devices
K122686 — ACCU-CHEK ULTRAFLEX INFUSION SET · Roche Diabetes Care AG · Nov 16, 2012
K063146 — ACCU-CHEK LINKASSIST · Disetronic Medical Systems AG · Nov 27, 2006
K070189 — ACCU-CHEK ULTRAFLEX INFUSION SET · Disetronic Medical Systems AG · Feb 21, 2007
K101196 — ACCU-CHEK ULTRFLEX INFUSION SET · Roche Diagnostics Corp. · Aug 27, 2010
K143446 — Accu-Chek FlexLink Plus infusion set · Roche Diagnostics GmbH · Aug 13, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
KUUZOY
····
; .
ﺑ
MAY 1 4 2010
| 510(k) Summary | |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Introduction | According to the requirements of 21 CFR 807.92, the following information<br>provides sufficient detail to understand the basis for a determination of<br>substantial equivalence. |
| Submitter | Submitted by:<br>Disetronic Medical Systems AG<br>Kirchbergstrasse 190, Postfach<br>CH-3401 Burgdorf, Switzerland |
| | United States Contact Person:<br>Scott Thiel<br>Roche Diagnostics<br>9115 Hague Road<br>Indianapolis, Indiana 46250<br>317-521-3362<br>scott.thiel@roche.com<br>Date Prepared: April 12, 2010 |
| Device name | Proprietary name: ACCU-CHEK® FlexLink Plus infusion set<br>Common name: subcutaneous infusion set<br>Classification name: intravascular administration set<br>Product Code: FPA<br><br>Proprietary name: ACCU-CHEK® LinkAssist Plus insertion device<br>Common name: infusion set insertion device<br>Classification name: syringe needle introducer<br>Product Code: KZH |
| Predicate<br>device | We claim substantial equivalence to the current legally marketed ACCU-<br>CHEK® Ultraflex infusion set and ACCU-CHEK® LinkAssist insertion device. |
Continued on next page
- --
{1}------------------------------------------------
# 510(k) Summary, Continued
| Device description | The ACCU-CHEK® FlexLink Plus is a disconnectable infusion set with soft cannula perpendicular to the adhesive, for transfusion of insulin into the subcutaneous tissue. The unit is designed to interface with commercially available insulin infusion pumps with suitable connections. The insulin infusion pump systems are designed to control the delivery of insulin as prescribed by a health care professional. The system (infusion set, insulin infusion pump, and insulin) is indicated for patients with insulin dependent diabetes mellitus. |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The ACCU-CHEK® LinkAssist Plus is an insertion aid for automatic application of ACCU-CHEK® infusion sets equipped with a compatible adapter. The ACCU-CHEK® LinkAssist device is non-invasive, non-sterile and intended for multiple uses by the same patient. |
| Intended use | ACCU-CHEK® FlexLink Plus is an infusion set for the subcutaneous infusion of insulin administered with micro dosage insulin pumps. |
| | ACCU-CHEK® LinkAssist Plus is an insertion device, which is intended specifically for placement of compatible ACCU-CHEK® FlexLink Plus infusion sets. |
| | Continued on next page |
{2}------------------------------------------------
### 510(k) Summary, Continued
The ACCU-CHEK® FlexLink Plus infusion set was compared to the ACCU-Device CHEK® Ultraflex infusion set. The ACCU-CHEK® FlexLink Plus is Comparisons substantially equivalent to this product by having the same intended use, same storage conditions, same luer connector, same operating conditions, a flexible catheter and needle for insertion into the subcutaneous tissue and separate extension tubing with detachable connector. Both sets have an adhesive patch that secures the headset to the skin. Prior to infusion, both sets require removal of the introducer needle. The ACCU-CHEK® FlexLink Plus incorporates a needle protection feature that retracts the needle automatically following insertion of the needle and cannula. The optional ACCU-CHEK® LinkAssist Plus insertion device was compared to the optional ACCU-CHEK® LinkAssist insertion device. The ACCU-CHEK® FlexLink Plus is substantially equivalent to this product by having the same intended use, same storage conditions, same operating conditions, and same general functionality. The overall dimensions of the ACCU-CHEK® FlexLink Plus are slightly wider and longer than the ACCU-CHEK® FlexLink device. Both devices include a safety catch; the ACCU-CHEK® FlexLink Plus incorporates the safety catch in the base of the device
such that it is depressed when placing the device into position on the body.
Continued on next page
{3}------------------------------------------------
### 510(k) Summary, Continued
#### Summary of Functional Testing:
In vitro functional testing of the ACCU-CHEK® FlexLink Plus infusion set and the optional ACCU-CHEK® LinkAssist Plus insertion device was conducted. Biocompatibility testing was performed on the materials used in both devices.
#### Clinical Studies:
Human clinical studies were not deemed necessary to evaluate the safety or effectiveness of the ACCU-CHEK® FlexLink Plus infusion set or the optional ACCU-CHEK® LinkAssist Plus insertion device.
System validation testing included human factors usability testing of the customer requirements and primary operating function of the ACCU-CHEK® FlexLink Plus infusion set and the optional ACCU-CHEK® LinkAssist Plus insertion device. Individuals who participated in the evaluation included patients who routinely use an insulin pump and infusion set. The usability evaluations determined the customer requirements were met and that the primary operating function operated as intended. No needle stick injuries were reported associated with unintentional ejection.
Study Conclusions
Studies
#### Functional Testing:
The results of the testing conducted indicate the ACCU-CHEK® FlexLink Plus infusion set and the optional ACCU-CHEK® LinkAssist Plus insertion device functioned according to specifications and the materials used in the devices are biocompatible.
### Human factors usability testing:
The design validation of ACCU-CHEK® FlexLink Plus & ACCU-CHEK® LinkAssist Plus confirmed that the devices fulfill its intended use, customer requirements and primary operating functions.
Based upon these results, the product is considered acceptable for human use.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Disetronic Medical System AG C/O Mr. Scott Thiel Regulatory Affairs Program Manager Roshe Diagnostics 9115 Hague Road Indianapolis, Indiana 46250-0457
MAY 1 4 2010
Re: K100704
Trade/Device Name: ACCU-CHEK® FlexLink Plus infusion set with optional ACCU- CHEK® LinkAssist Plus insertion device Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FRN Dated: April 14, 2010 Received: April 16, 2010
Dear Mr. Thiel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2- Mr. Thiel
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Nh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital,
Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use Statement
510(k) Number (if known): K100704
Device Name: ACCU-CHEK® FlexLink Plus infusion set with optional ACCU-CHEK® LinkAssist Plus insertion device
Indications for Use:
ACCU-CHEK® FlexLink Plus is an infusion set for the subcutaneous infusion of insulin administered with micro dosage insulin pumps.
ACCU-CHEK® LinkAssist Plus is an insertion device, which is intended specifically for placement of compatible ACCU-CHEK® FlexLink Plus infusion sets.
Prescription Use_ × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
---
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K100704
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.