K053554 · Milestone Scientific, Inc. · FRN · Jul 10, 2006 · General Hospital
Device Facts
Record ID
K053554
Device Name
COMPUFLO
Applicant
Milestone Scientific, Inc.
Product Code
FRN · General Hospital
Decision Date
Jul 10, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CompuFlo is intended for use in delivering medication and other fluids in a controlled manner. The CompuFlo is indicated for use in adults for the continuous or intermittent delivery of medications and other fluids through intravenous, intra-arterial, subcutaneous, epidural and enteral routes.
Device Story
CompuFlo is a motor-driven piston syringe pump for controlled delivery of medications and fluids. System inputs include user-provided syringes and an external pressure transducer (Meritrans); internal pressure transducer monitors system performance. Device features a computer processor, LCD, and foot pedal control. It calculates pressure at the needle tip using a specific algorithm. Used in clinical settings to deliver fluids via intravenous, intra-arterial, subcutaneous, epidural, or enteral routes. Output allows healthcare providers to monitor and control infusion rates and pressure, facilitating accurate medication administration and potentially improving patient safety through controlled delivery.
Clinical Evidence
Bench testing only. Evaluation included syringe plunger travel distance, volume dispensed, flow rate accuracy, force measurements on the syringe platform, and accuracy of internal and external pressure sensors. Static pressure testing was performed. Results demonstrated the device functions as intended and accurately calculates needle-tip pressure.
Technological Characteristics
Motor-driven piston syringe pump; includes internal pressure transducer, external pressure transducer (Meritrans), syringe retainer, computer processor, LCD, and power supply. Housed in a plastic cabinet. Operates with Luer Lock plastic syringes, tubing, and needles. Connectivity includes foot pedal interface. Software performs pressure calculations.
Indications for Use
Indicated for adults requiring continuous or intermittent delivery of medications and fluids via intravenous, intra-arterial, subcutaneous, epidural, or enteral routes.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
K982640 — MEDEX 3000 SERIES SYRINGE INFUSION PUMP · Medex, Inc. · Apr 6, 1999
K982170 — FLUIDSENSE FLUID DELIVERY SYSTEM · Fluidsense Corp. · Nov 30, 1998
K093332 — DELPHI INFUSION PUMP/ DELPHI ADMINISTRATION SET MODEL IV-01000/ MODEL IV-02XXXX (WHERE X IS ANY NUMBER BETWEEN 0-9) · Delphi Medical Systems · Dec 9, 2009
K042696 — BODYGUARD INFUSION PUMP SYSTEM · Caesarea Medical Electronics , Ltd. · Mar 11, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
CompuFlo 510(k) Page 16
Kos 355 4
#### 510(k) SUMMARY
#### JUL 1 0 2006 Milestone's CompuFlo Syringe Pump
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Howard Holstein Hogan & Hartson LLP Columbia Square 555 Thirteenth Street, NW Washington DC 20004-1109
Phone: (202) 637-5600
Date Prepared: December 20, 2005
### Name of Device and Name/Address of Sponsor
CompuFlo™ Infusion Pump
Milestone Scientific, Inc. 220 S. Orange Avenue Livingston, NJ 07039
#### Common or Usual Name
Syringe Infusion Pump
#### Classification Name
Infusion Pump, Class II, 21 C.F.R. § 880.5725
#### Predicate Devices
- · Harvard Clinical Technology, Inc.'s Harvard 2 Syringe Pump (K050107)
- · Alaris Medical Systems, Inc.'s Medley Syringe Pump Module (K023264)
- · Medex, Inc.'s Medex 3000 Series Infusion Pump (K040899)
{1}------------------------------------------------
# Intended Use / Indications for Use
The CompuFlo is intended for use in delivering medication and other fluids in a controlled manner. The CompuFlo is indicated for use in adults for the continuous or intermittent delivery of medications and other fluids through intravenous, intra-arterial, subcutaneous, epidural and enteral routes.
## Technological Characteristics
The CompuFlo unit consists of a motor-driven piston syringe pump with an internal pressure transducer, an external pressure transducer (Meritrans™ Disposable Pressure Transducer (K920977)), a syringe retainer, a computer processor, a liquid crystal display (LCD), and a power supply. All but the external pressure transducer are contained in a plastic cabinet. A foot pedal and power cord are also included. The unit operates with a variety of user-provided disposable supplies, including Luer Lock plastic piston syringes, plastic syringe tubing, and needles.
## Performance Data
Machine performance of the CompuFlo was evaluated, in addition to the software items. The evaluation of machine performance included: travel distance of the syringe plunger and the volume dispensed; the volumes dispensed; the flow rates; the force measurements on the syringe platform; and, assessing the accuracy of the internal and external pressure sensors. Pressure testing was conducted in a static fashion. With the exception of several minor tests not directly related to device performance and explained in detail in Section XVI, the CompuFlo functioned as intended. These test results demonstrate that the CompuFlo can accurately dispense a desired quantity of medication or fluid, and is capable of calculating the pressure at the needle tip according to the equation described in this 510(k) notice.
## Substantial Equivalence
The CompuFlo raises no new questions of safety or effectiveness as compared to Harvard Clinical Technology, Inc.'s Harvard 2 Syringe Pump (K050107), Alaris Medical Systems, Inc.'s Medley Syringe Pump Module (K023264), and Medex, Inc.'s Medex 3000 Series Infusion Pump (K040899). The CompuFlo has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate devices. The minor technological differences between the CompuFlo and its predicate devices raise no new issues of safety or effectiveness. Thus, the CompuFlo is substantially equivalent.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features the department's name in a circular arrangement around the perimeter. In the center of the seal is a stylized representation of a human figure, composed of three overlapping profiles, symbolizing health and human services.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 1 0 2006
Milestone Scientific, Incorporated C/O Mr. Howard Holstein Hogan & Hartson L.L.P. 555 Thirteenth Street, NW Washington, D.C. 20004
Re: K053554
Trade/Device Name: CompuFlo Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN Dated: June 27, 2006 Received: June 27, 2006
Dear Mr. Holstein:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2 - Mr. Holstein
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Suyite Y. Michino m.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### Indications for Use Statement
Kes 355 y 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: CompuFlo
Indications for Use:
The CompuFlo is intended for use in delivering medication and other fluids in a controlled manner. The CompuFlo is indicated for use in adults for the continuous or intermittent delivery of medications and other fluids through intravenous, intra-arterial, subcutaneous, epidural and enteral routes.
Prescription Use _ X (Part 21 C.F.R. 801 Subpart D) AND/OR
Over-The-Counter Use (21 C.F.R. 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Cinta V.V.
i Chan-Off) icn of Anesthesiology, General Hospital, Jon Control. Dental Devices
Number: K45 5554
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.