K982123 · Medex, Inc. · FPA · Jul 21, 1998 · General Hospital
Device Facts
Record ID
K982123
Device Name
CONTRAST CONTROL DEVICE
Applicant
Medex, Inc.
Product Code
FPA · General Hospital
Decision Date
Jul 21, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The Control Device (CCD) is intended for use in the IV delivery of contrast media during imaging procedures with a secondary intended use of conservation of contrast media.
Device Story
Contrast Control Device (CCD) facilitates IV delivery of contrast media during imaging procedures; enables conservation of contrast media. Device consists of two parts: Part A (reusable spike assembly for contrast container) and Part B (disposable valve and tubing set for patient). Two in-line backcheck valves in Part B prevent retrograde flow, protecting Part A from contamination. Clinician spikes contrast bottle with Part A; attaches Part B for patient procedure. After case, Part B is detached and Part A is capped with sterile dead-ender. Part A can be reused for up to three patients within a six-hour window by attaching new sterile Part B sets. Used in clinical imaging settings by clinicians.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Device comprises a reusable spike assembly (Part A) and disposable valve/tubing set (Part B). Features two in-line backcheck valves to prevent retrograde flow. Designed for multi-patient use (up to 3 patients) within a 6-hour timeframe. Non-electronic, mechanical fluid delivery system.
Indications for Use
Indicated for IV delivery of contrast media during imaging procedures. Patient population not specified.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
K984231 — ANGIOGRAPHIC CONTRAST MANAGEMENT SYSTEM, MODEL PART A AND PART B · Invasatec · May 31, 2000
K971624 — DYE STOP AND SAVE - QUICKFILL · Shepherd Laboratories, Inc. · Sep 8, 1997
K955179 — CONTRAST MEDIA SET · B.Braun Medical, Inc. · Mar 12, 1997
K983314 — PREMICA CT CONTRAST MEDIA DELIVERY SYSTEM · Nycomed Amersham Imaging · Feb 19, 1999
K173913 — MEDRAD® Imaging Bulk Package Transfer Set · Bayer U.S., LLC · May 4, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
K982123
### 510 (k) Summary
This summary regarding 510(k) safety and effectiveness and being submitted in accordance with the requirements of SMDA 1990 and 21 CFR part 807.92.
## 807.92(a)(1) Submitter's (and Contact) Name, Address, Telephone No., Summary Date
Cathy Chenetski Director, Regulatory Affairs Medex 3637 Lacon Road Hilliard, OH 43026 (614) 529-3932
June 11, 1998
## 807.92(a)(2) Device Name (Including Trade Name), Common Name, Classification Name
| Classification Name: | Intravascular Administration Set (80 FPA) |
|-----------------------|-------------------------------------------|
| Common/Usual Name: | Intravascular Fluid Delivery Set |
| Trade/Propriety Name: | Not yet determined. |
| Part Number: | Not yet determined |
#### 807.92(a)(3) Legally Marketed Predicate Device to Which Equivalence is Claimed
The modified Contrast Control Device, for purposes as defined under Section 510(k) of the Federal Food, Drug and Cosmetic Act, are substantially equivalent to the Medex IV Administration Set, K883318 and pre-amendment (primary predicate). Additionally, the Contrast Control Device is substantially equivalent to the Namic Contrast Savings System. K903493 (secondary predicate).
#### 807.92(a)(4) Description of the Premarket Notification Device and 807.92 (a) (5) Intended Use
The CCD is made up of two parts: Part A, which is the reusable spike assembly (intended for use on only one container of contrast media/dye and considered an extension of the contrast container) and; Part B, which is the disposable valve and tubing set (intended for use on only one patient). The two in-line backcheck valves in Part B prevent inadvertent retrograde flow thereby protecting Part A from contamination.
The clinician spikes the bottle of contrast media with Part A and then dead-ends the spike assembly until the start of the case. The clinician then attaches Part B and proceeds with the first case. When that case is complete the clinician detaches Part B and caps Part A with a new sterile dead-ender. When the clinician is ready to start the next case he attaches a new sterile Part B and proceeds. This can be repeated up to two times (for a total of three patients) within a maximum six-hour time frame.
The Control Device (CCD) is intended for use in the IV delivery of contrast media during imaging procedures with a secondary intended use of conservation of contrast media.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 21 1998
Ms. Cathy Chenetski Director, Regulatory Affairs Medex, Incorporated 3637 Lacon Road Hilliard, Ohio 43026
K982123 Re : Contrast Control Device Trade Name: Requlatory Class: II Product Code: FPA June 11, 1997 Dated: Received: June 16, 1997
Dear Ms. Chenetski:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in Please note: this response to your the Federal Register. premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
Page 2 - Ms. Chenetski
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA described in your sitem privalence of your device to a legally marketed predicate device results in a classification for your markets and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A. Ditrain gr
Timothy A. Ulatowski Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
K98212コ
Page 1 of 1
510(k) Number (if known) ____________________________________________________________________________________________________________________________________________________
Device Name: Contrast Control Device
Indications for Use:
The Control Device (CCD) is intended for use in the IV delivery of contrast media during The Ountrast Oontrol Donot (OOD) is intented use of conservation of contrast media.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use_ (Per 21 CFR 801.109) OR
Over-The-Counter Use_
Patricia Cucerdl
(Division Sign-Off)
(Optional Format 1-2-96)
Division of Dental, Infection Control, and General Hospital Devices
510(k) Number K982123
000013
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.