K162804 · Quest Medical, Inc. · FPA · Mar 8, 2017 · General Hospital
Device Facts
Record ID
K162804
Device Name
Q2 Low Pressure Power Injection Extension Set
Applicant
Quest Medical, Inc.
Product Code
FPA · General Hospital
Decision Date
Mar 8, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
For administration of intravenous fluids to a patient's vascular system utilizing needle-free components and provide short-term intravascular or subcutaneous access using a variety of infusates (i.e. anesthesia drugs, chemotherapeutics, drugs, antibiotics, blood products, commonly used prep solutions). The device may also be used with low pressure power injectors rated for a maximum setting of 325 psi.
Device Story
Sterile, single-use IV extension set; 7-inch tubing with male/female luer connectors and clamp; includes swabable needle-free port. Connects to vascular access devices for fluid/drug administration or aspiration. Used in clinical settings by healthcare providers. Supports low-pressure power injection (max 325 psi, 10 mL/s). Facilitates safe, needle-free access; reduces needlestick risk; DEHP-free construction.
Clinical Evidence
Bench testing only. Performance validated via particulate testing (ISO 8536-4:2010), simulated shipping, package burst, priming volume, backpressure leak, flow rate, and simulated maximum use for power injection. Biocompatibility verified per ISO 10993-1:2009. Sterilization validated per ISO 11135; EtO residuals tested per ISO 10993-7:2008. Pyrogen testing performed via kinetic chromogenic LAL method.
Technological Characteristics
7-inch PVC (non-DEHP) tubing; co-polyester luers; polycarbonate/silicone needleless connector; ABS spin lock. Manual operation. Sterilized via 100% Ethylene Oxide (EtO) to 10^-6 SAL. Single-use, disposable. No software or electronic components.
Indications for Use
Indicated for single use in patients requiring direct injection, intermittent infusion, continuous infusion, or aspiration of fluids/drugs. Compatible with power injector procedures up to 325 psi at 10ml/sec flow rate.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
Carefusion MaxPlus Pressure Rated Extension Set with removable clear needleless connector (K140831)
Reference Devices
Quest Medical Inc. Q2 Multiport IV Administration Sets and Extension Sets (K162304)
CheckMate IV Administration Sets and Extension Sets (K162304)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 8, 2017
Quest Medical, Inc. Amy Clendening-Wheeler Senior Regulatory Affairs Specialist One Allentown Parkway Allen, TX 75002
Re: K162804
Trade/Device Name: Q2® Low Pressure Power Injection Extension Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Sets Regulatory Class: Class II Product Code: FPA Dated: February 3, 2017 Received: February 6, 2017
Dear Amy Clendening-Wheeler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162804
Device Name
Q2 Low Pressure Power Injection Extension Set
#### Indications for Use (Describe)
Pressure Rated: The Q2 Low Pressure Power Injector Extension Set with needleless connector is for single use only. The extension set may be used for direct injection, infusion, continuous infusion or aspiration. This set may be used with power injector procedures to a maximum pressure of 325 psi at a flow rate of 10ml per second. Non Pressure Rated: The Q2 Low Pressure Power Injector Extension Set with needleless connector is for single use only. The extension set may be used for direct injection, intermittent infusion, continuous infusion or aspiration.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY K162804
| DATE PREPARED: | March 8, 2017 | |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| SUBMITTER: | Quest Medical, Inc.<br>One Allentown Parkway<br>Allen, TX 75002 USA<br>972-390-9800/800-627-0226 | |
| | Contact: Amy Clendening-Wheeler<br>Phone: (972) 332-6228<br>Fax: (469) 795-2338<br>Email: awheeler@questmedical.com | |
| DEVICE NAME: | Q2® Low Pressure Power Injection Extension Set | |
| COMMON NAME: | IV Extension Sets | |
| CLASSIFICATION NAME: | Intravascular Administration Sets | |
| PRODUCT CODE: | FPA | |
| REGULATION: | 880.5440 | |
| CLASS: | II | |
| PREDICATE DEVICE: | Carefusion MaxPlus® Pressure Rated Extension Set with removable clear<br>needleless connector (K140831) | |
| REFERENCE DEVICE: | Quest Medical Inc. Q2® Multiport IV Administration Sets and Extension<br>Sets and CheckMate® IV Administration Sets and Extension Sets<br>(K162304) | |
| DESCRIPTION: | Sterile, single use non-pyrogenic intravenous administration extension<br>set comprised of 7 inch tubing bonded to a male luer on one end and a<br>female luer on the other end with a clamp in between. A swabable,<br>needle-free port is connected to the female luer. This device is not<br>made with the plasticizer Diethylhexylphthalate (DEHP). | |
| INTENDED USE: | For administration of intravenous fluids to a patient's vascular system<br>utilizing needle-free components and provide short-term intravascular<br>or subcutaneous access using a variety of infusates (i.e. anesthesia | |
{4}------------------------------------------------
drugs, chemotherapeutics, drugs, antibiotics, blood products, commonly used prep solutions). The device may also be used with low pressure power injectors rated for a maximum setting of 325 psi.
INDICATIONS FOR USE: Pressure Rated: The Q2 Low Pressure Power Injector Extension Set with needleless connector is for single use only. The extension set may be used for direct injection, intermittent infusion, continuous infusion or aspiration. This set may be used with power injector procedures to a maximum pressure of 325 psi at a flow rate of 10ml per second.
> Non Pressure Rated: The Q2 Low Pressure Power Injector Extension Set with needleless connector is for single use only. The extension set may be used for direct injection, intermittent infusion, continuous infusion or aspiration.
### SUBSTANTIAL EQUIVALENCE
### Bench Testinq
Functional performance testing including particulate testing per ISO 8536-4:2010, simulated shipping, package burst, priming volume, backpressure leak and flow rate for both normal and low pressure power injection use, and simulated maximum use for both normal and low pressure power injection use was completed for the proposed Q2 Low Pressure Power Injector Extension Set to demonstrate the device performs as intended and to manufacturer specification. Results of testing successfully demonstrate that the Q2 Low Pressure Power Injector Extension Set performs similarly to the predicate device to meet the functional requirements for both non-pressure use and low pressure power injection.
## Sterilization
There is no change to the sterilization process for the proposed new Q2 Low Pressure Power Injection Extension Set from the other Quest Medical, Inc. IV Administration and Extension Set product family. The sterilization process is validated per ISO 11135 and Ethylene Oxide residuals testing was performed for the product line per ISO 10993-7:2008 "Biological Evaluation of Medical Devices – Part 7: Ethylene Oxide Sterilization Residuals." Although the method of sterilization is different from the predicate device, both devices are sterile, non-pyrogenic, single-use devices with a minimum 10 ° SAL.
# Shelf Life
Shelf life for the Q2 Low Pressure Power Injection Set is qualified to 1 year. The predicate device is qualified to 5 years shelf life. The difference in shelf life is not a significant
{5}------------------------------------------------
difference as this attribute has no impact to the function of the device. The shelf life of each device is validated to meet each manufacturer's specification.
# Biocompatibility
The materials of construction for the Model 95714 are identical to that of the Quest Medical Inc. Reference Devices of K162304, where the fully assembled representative IV Administration Set was tested according to ISO 10993-1:2009. Hemocompatibility, Cytotoxicity, Sensitization, Irritation/Intracutaneous Reactivity, Systemic Toxicity, Subchronic Toxicity, and Materialmediated Pyrogenicity tests were performed. Test results successfully verified that the IV Administration Set and Extension Set materials of construction are biocompatible for their clinical application.
# Pyrogen
Pyrogen testing for bacterial endotoxins was performed via the kinetic chromogenic LAL (Limulous Amebocyte Lysate) method. The results found that the proposed new materials do not introduce a level of endotoxin that exceed established guidelines.
# Comparison to Predicate:
The following table shows a comparison between the Q2® Low Pressure Power Injection Extension Set subject device to the Carefusion MaxPlus Pressure Rated Extension Set with removable clear needleless connector predicate device. The subject and predicate devices are similar in physical properties, materials, and configuration.
| | Predicate Device | Subject Device |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) | K140831 | K162804 |
| Model # | MP5303-C | 95714 |
| Brand Name | MaxPlus Pressure Rated Extension Set with<br>removable clear needleless connector | Q2 Low Pressure Power Injector Extension Set |
| Manufacturer | CareFusion | Quest Medical, Inc. |
| FDA Regulation<br>Number | 21CFR 880.5440 | SAME |
| FDA Class | II | SAME |
| FDA Product<br>Code | FPA | SAME |
| Device<br>Description | Intravascular extension sets intended for<br>single patient use, including pediatrics and<br>immunocompromised patients, for direct<br>injection, intermittent infusion, continuous<br>infusion or aspiration of drugs, blood, and<br>fluids. The extension sets include a<br>needleless connector to allow thorough and | The device is an intravascular extension set<br>available in one configuration that includes a<br>needleless connector. The device is not made<br>with the plasticizer Diethylhexylphthalate<br>(DEHP). |
| | Predicate Device | Subject Device |
| | easy disinfection due to a solid, flat, smooth<br>surface and eliminates the risk of<br>needlestick injuries. The extension sets are<br>not made from natural rubber latex or<br>DEHP. Configurations are available with<br>needleless connectors, split septum ports,<br>filters, stopcocks, manifolds, and/or T-<br>connectors. | |
| Intended Use | For direct injection, intermittent infusion,<br>continuous infusion or aspiration. | For administration of intravenous fluids to a<br>patient's vascular system utilizing needle-free<br>components and provide short-term<br>intravascular or subcutaneous access using a<br>variety of infusates (i.e. anesthesia drugs,<br>chemotherapeutics, drugs, antibiotics, blood<br>products, commonly used prep solutions). The<br>device may also be used with low pressure<br>power injectors rated for a maximum setting<br>of 325 psi. |
| Indications for<br>Use | Pressure Rated: The MaxZero multi fuse<br>extension set with needleless connector(s) is<br>for single use only. The extension set may be<br>used for direct injection, intermittent<br>infusion, continuous infusion or aspiration.<br>This set may be used with power injector<br>procedures to a maximum pressure of 325<br>psi at a flow rate of 10ml per second.<br>Non Pressure Rated: The MaxZero multi<br>fuse extension set with needleless<br>connector(s) is for single use only. The<br>extension set may be used for direct<br>injection, intermittent infusion, continuous<br>infusion or aspiration. | Pressure Rated: The Q2 Low Pressure Power<br>Injector Extension Set with needleless<br>connector is for single use only. The extension<br>set may be used for direct injection,<br>intermittent infusion, continuous infusion or<br>aspiration. This set may be used with power<br>injector procedures to a maximum pressure of<br>325 psi at a flow rate of 10ml per second.<br>Non Pressure Rated: The Q2 Low Pressure<br>Power Injector Extension Set with needleless<br>connector is for single use only. The extension<br>set may be used for direct injection,<br>intermittent infusion, continuous infusion or<br>aspiration.<br>(SAME) |
| Device Design | | |
| Components | Tubing, Luer, Check valve, Needleless<br>Connector, Male spin lock connector<br>Non-fluid contacting: Pinch clamp, vented<br>female luer lock cap | SAME |
| Dimensions | Length: 7 inch (18cm) | SAME |
| Device<br>Materials | | |
| Tubing | PVC- Purple striped, non-DEHP, standard<br>bore | Identical to Quest Medical, Inc. Reference<br>Device K162304: PVC - clear, non-DEHP Alpha<br>Gary 2235L-80, standard bore |
| Luer | Co-Polyester | Identical to Quest Medical, Inc. Reference |
| | Predicate Device | Subject Device |
| Needleless<br>Connector | MaxZero Clear- Polycarbonate/silicone | Device K162304: Co-polyester<br>Identical to Quest Medical, Inc. Reference<br>Device K162304: SwabSite Valve –<br>Polycarbonate/silicone |
| Male spin luer<br>lock/hub | ABS thermoplastic | Identical to Quest Medical, Inc. Reference<br>Device K162304: CYRO XT 250-301 |
| Technology | | |
| Use | Use with low pressure power injectors up to<br>325 psi and maximum flow rate of 10<br>mL/second | SAME |
| Priming volume | < 1 mL | 1.0 mL |
| Energy Source | User Operated | SAME |
| Principle of<br>Operation | Luer activation; direct injection, intermittent<br>infusion, continuous infusion, aspiration | SAME |
| Sterilization and<br>Package | | |
| Method | Irradiation; E-Beam | 100% EtO |
| Minimum SAL | 1 x 10-6 | SAME |
| Packaging | Tyvek pouch | SAME |
| Shelf Life | 5 year | 1 year |
| Disposable or<br>Reusable | Disposable | SAME |
{6}------------------------------------------------
{7}------------------------------------------------
# CONCLUSION:
Results of all functional performance testing conducted successfully demonstrate that the proposed new Q2 Low Pressure Power Injection Extension Set is substantially equivalent to the legally marketed predicate device.