INTRAVSCULAR ADMINISTRATION SET

K100946 · Terumo Europe N.V. · FPA · Aug 2, 2010 · General Hospital

Device Facts

Record IDK100946
Device NameINTRAVSCULAR ADMINISTRATION SET
ApplicantTerumo Europe N.V.
Product CodeFPA · General Hospital
Decision DateAug 2, 2010
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.5440
Device ClassClass 2

Indications for Use

The Surflo Winged Infusion Sets with Needle Protection (Surshield) are intended to access the peripheral vascular system, for intravenous administration of fluids and/or withdrawal of blood specimens using a syringe, luer adapter, or other compatible/appropriate devices, Additionally, after withdrawal of the needle from the patient's vein. the shield cover can be manually activated to cover the needle to minimize risk of accidental needle stick.

Device Story

Sterile, single-use winged infusion set; 23G and 25G needle sizes with 350mm microbore tubing. Device accesses peripheral vascular system for fluid infusion or blood withdrawal via syringe or luer adapter. Features hinged, transparent shield cover attached to wing hub; manually activated by user post-withdrawal to cover needle; audible click confirms latching. Designed to keep user fingers behind needle point to minimize accidental stick risk. Used in clinical settings by healthcare professionals. Output is physical access to vascular system; safety mechanism reduces sharps injury risk.

Clinical Evidence

Bench testing only. No clinical data. Verification activities included flow rate, dead space volume, air/liquid leakage, fitting strength, bonding strength, needle penetration resistance, and safety feature activation force. All tests met acceptance criteria.

Technological Characteristics

Materials: standard medical-grade components. Design: winged hub, microbore tubing, luer adapter, hinged shield cover. Sterilization: Ethylene oxide (EN ISO 11135-1:2007) to SAL 10^-6. Connectivity: None. Software: None.

Indications for Use

Indicated for patients requiring peripheral vascular access for intravenous fluid administration or blood specimen withdrawal. No specific age or gender restrictions. Contraindications not specified.

Regulatory Classification

Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K100946 ### 510(k) SUMMARY as required by 807.92 Summary of Safety & Effectiveness Information AUS 0 2 2010 Submitter information | Prepared for: | TERUMO EUROPE N.V.<br>Interleuvenlaan 40,<br>3001 Leuven,<br>BELGIUM | |-----------------------------|-------------------------------------------------------------------------------------------------| | Prepared by/Contact person: | Mrs. M.J. Aerts - Manager Regulatory Affairs<br>Tel. (+32) 16 38 13 53<br>Fax (+32) 16 40 02 49 | | Date prepared: | June 2010 | ### .1. Device Name ### Proprietary Name SURFLO® Winged Infusion Set with Needle Protection (Surshield™) Classification Name Intravascular Administration Set (-FPA) 21 CFR, Section 880.5440 Classification: Class II #### 2. Reason for Submission This 510k is being submitted to extend Terumo Europe's cleared Surflo Winged Infusion Set with Needle Protection (Surshield) (K052887 and K072894) product line to include 23G and 25G needle sizes with microbore tubing taking into consideration the potential issues of safety and effectiveness specific for a thinner tubing. This 510k will provide supporting information that the 23G Surflo Winged Infusion Sets with Needle Protection (Surshield) and microbore tubing design are safe and effective and an acceptable extension of the current cleared Surflo Winged Infusion Set with Needle Protection (Surshield) product line (K052887 and K072894). - ે. Intended Use The Surflo Winged Infusion Sets with Needle Protection (Surshield) are intended to access the peripheral vascular system, for intravenous administration of fluids and/or withdrawal of blood specimens using a syringe, luer adapter, or other compatible/appropriate devices, Additionally, after withdrawal of the needle from the patient's vein. the shield cover can be manually activated to cover the needle to minimize risk of accidental needle stick. Note: This is the same intended use as the predicate device. Surflo Winged Infusion Set with Needle Protection (Surshield), cleared under KU52887 and KU72894. Page 10 {1}------------------------------------------------ ### 4. Description The 23G and the 25G Terumo Surflo Winged Infusion Sets with Needle Protection (Surshield) are sterile, single use devices consisting of a needle attached to a winged hub, microbore tubing, adapter and adapter cap, und a hinged shield cover that attaches to the wing just below the needle-to-wing junction. The shield cover can be turned 180 degrees on the hinge. As the needle is removed from the patient's vessel, the user's finger actively pushes the shield cover until it latches onto needle using a one- or twohanded technique. An audible click is noted upon activation. The shield cover is designed to allow the user's finger to remain behind the needle point so that the risk injury is minimized. The shield cover is transparent for easy confirmation of the needle held in it. The device possesses a 350 min length microbore tubing, #### Substantjal Equivalence 5. The 23G and the 25G Surflo Winged Infusion Sets with Needle Protection (Surshield) with microbore tubing, manufactured by Terumo Europe N:V., submitted in this 510(k) file are substantially equivalent in intended use, descriptions, technology/principles of operation, materials and performance to the following cleared devices: - 1. Surflo Winged Infusion Set with Needle Protection (Surshield) (K052887), manufactured by Teruno Europe N.V. - 2. Surflo Winged Infusion Set with Needle Protection (Surshield) (K072894), manufactured by Terumo Europe N.V. - 3. Surflo' Winged Infusion Set with Filter and Needle Protection (Surshield) (K070547), manufactured by Terumo Europe N.V. Any differences between the devices do not raise any significant issues of safety and effectiveness. Note: For performance testing reference will also be made to the Surflo Winged Infusion Set with Filher and Needle Protection (K070547), as these sets are identical as the sets to be cleared by this submission except for the presence of the fitter, consequently their use is limited to infusion. #### ઉં: Summary of Technological Modifications compared to Predicate Devices: The technological modifications made to the subjected device compared to the predicate devices are summarized in the following table: | Predicate device | Modification summary (Proposed device) | |-------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Surflo Winged Infusion Set with<br>Needle Protection (Surshield)<br>(K052887) & (K072894) | a) Introduction of microbore tubing<br>b) Adapted design Surshield protector (the connection with<br>the tube remains the same) | | Surflo Winged Infusion Set with<br>Filter and Needle Protection<br>(K070547) | a) Use of the same luer adaptor as predicate device BUT<br>without filter<br>b) Adapted design Surshield protector (the connection with<br>the tube remains the same) | Page 11 {2}------------------------------------------------ ## 7. Summary of Verification Activities: All necessary verification and validation tests have been performed for the The Surflo Winged Infusion Sets with Needle Protection (Surshield) to assure substantial equivalence to the predicate devices. Summary of all verification activities including acceptance criteria is given in the following table: | | ACCEPTANCE CRITERIA | |------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 1. Flow rate | The flow rate for the set with a 23 G needles is ≥ 1.7ml/min & for a set with<br>a 25 G needles is ≥ 1.5 ml/min | | 2. Dead space volume | Dead space volume for sets with 23 G & 25G needle ≤ 0.25 ml | | 3. Air flow choke test | No obstruction of fluid through the set | | 4. Air leakage set<br>(= Integrity of the set) | No air leakage from the set<br>(European Pharmacopoeia.3.2.6.) | | 5. Air leakage adapter | No air leakage at the luer fitting (EN 1707) | | 6. Liquid leakage<br>adapter | No liquid leakage at the luer fitting (EN 1707) | | 7. Conical fitting | 6% luer (EN 20594-1) | | 8. Fitting strength.<br>protector | The force to pull the protector from the wing hub is ≥ 0.3 N and ≤ 12N | | 9. Torque resistance<br>cap/adapter | The torque force required to unscrew the cap from the adapter is ≤ 9N.cm. | | 10.Bonding strength<br>Cannula | The bonding strength between the cannula and the wing hub is ≥ 20N | | 11.Force at break | The force to separate the the tubing from the wing/connector is ≥ 15N | | 12.Needle penetration<br>resistance | Needle point ≤ 0.14N<br>Drag ≤ 0.03N | | 13.Break strength of<br>shield cover joint | The force to detach the Surshield protector from the wing hub is ≥ 4N | | 14.Force to unlock the<br>safety feature | The force required to unlock the Surshield protector from the cannula is ≥ 2N | | 15.Misalignment<br>Surshield Protector | The angle measured axially from the cannula between the horizontally<br>positioned wings and the vertically positioned Surshield protector does not<br>exceed 15° | The Surflo Winged Infusion Sets with Needle Protection (Surshield) met all acceptance criteria as indicated in table above. None of the obtained data raises any new issue of safety and effectiveness. Page 12 {3}------------------------------------------------ #### 8. Additional Safety Information The sterility of the Surflo Winged Infusion Sets with Needle Protection (Surshield) is assured by using a validated sterilization method qualified in accordance with in accordance with the requirements of EN ISO 11135-1:2007 "Sterilization of health care products - Ethylene oxide - Part 1:Requirements for development, validation and routine control of a sterilization process for medical devices" to a sterilizy assurance level (SAL) of 10-6 as required by EN 556-1: "Sterilization of Medical Devices -Requirements for medical devices to be designated "STERILE" - Part - !: Requirements for terminally sterilized medical devices". Ethylene oxide residual levels resulting from EtO sterilization are in compliance with EN ISO 10993-7: "Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals ... The 23G and the 25G Surflo Winged Infusion Sets with Needle Protection (Surshield) are Externally Communicating devices, Circulating Blood, Limited Exposure (24 hrs). The devices' blood contacting materials were tested in accordance with the tests recommended in the FDA General Program Mennorandum #G95-1 (5/1/95): Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part-1: Evaluation and testing". The expiration dating for the 25G and the 25G Surflo Winged Infusion Sets with Needle Protection (Surshield) have been established at 5 years which.is the same as the cleared Surflo Winged Infusion Sets with Needle Protection (Surshield) (K052887 and K072894). - Conclusion 9. The Surflo Winged Infusion Sets with Needle Protection (Surshield) manufactured by Terumo Europe N.V. and submitted in this 510(k) file are substantially equivalent in intended use. description, specifications, and technology/principles of operation, materials, and performance to the following cleared devices: - 1. Surflo Winged Infusion Set with Needle Protection (Surshield) (K052887), manufactured by Terumo Europe N.V. - 2. Surflo Winged Infusion Set with Needle Protection (Surshield) (K072894), manufactured by Terumo Eurone N.V. - 3. Surflo Winged Infusion Set with Filter and Needle Protection (Surshield) (K070547), manufactured by Terumo Europe N.V. Differences between the devices do not raise any new issues of safely or effectiveness. {4}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The bird is positioned to the right of the text, which is arranged in a circular fashion around the bird. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA". ### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Ms. M.J. Aerts Regulatory Affairs Manager Terumo Europe N.V. Interleuvenlaan 40 B-3001 Leuven-Belgie Belgium AUG-0 2 2010 Re: K100946 Trade/Device Name: Surflo Winged Infusion Set with Needle Protection (Surshield™) Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: July 1, 2010 Received: July 6, 2010 Dear Ms. Aerts: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ ### Page 2- Ms. M.J. Aerts Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/ CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Susan Pease Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indication for Use 510(k) Number (if known): K100946 Device Name: Surflo Winged Infusion Set with Needle Protection (Surshield") Indication For Use: The Surflo Winged Infusion Set with Needle Protection (Surshield) is intended to access the peripheral vascular system, for intravenous administration of fluids and/or withdrawal of blood specimens using a syringe, luer adapter, or other compatible/appropriate devices. Additionally, after withdrawal of the needle from the patient's vein, the shield cover can be manually activated to cover the needle to minimize risk of accidental needle stick. Prescription Use __ X (21 CFR Part 801 Subpart D) And/Or Over the Counter Use (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Richard C. Chapman (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K100946
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...