Arrow Endurance Extended Dwell Peripheral Catheter System
K152272 · Arrow International, Inc. (Subsidiary of Teleflex, Inc.) · FOZ · Nov 23, 2015 · General Hospital
Device Facts
Record ID
K152272
Device Name
Arrow Endurance Extended Dwell Peripheral Catheter System
Applicant
Arrow International, Inc. (Subsidiary of Teleflex, Inc.)
Product Code
FOZ · General Hospital
Decision Date
Nov 23, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Arrow Endurance™ Extended Dwell Peripheral Catheter System is intended for short-term use (less than 30 days) to permit delivery of infusion therapies, infusion of blood and blood products, pressure monitoring, high pressure injection and withdrawal of blood.
Device Story
Sterile, single-use peripheral intravascular catheter system; provides vascular access for infusion therapy, blood sampling, pressure monitoring, and high-pressure injection (up to 325 psi). System includes ergonomically designed handle, echogenic needle with passive safety mechanism, guidewire with slider advancer, and radiopaque polyurethane catheter. Features integrated extension tubing, hub seal, and vent plug to contain blood during insertion and prevent exposure. Used in clinical settings by healthcare professionals. Catheter threaded over guidewire into peripheral vessel; clamp used to manage blood containment during component exchange. Benefits include reduced sharps injury risk and minimized blood exposure during insertion.
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 10993 series), mechanical performance (ISO 10555, ISO 11070), Luer compatibility (ISO 594-1/EN 1707), simulated use, insertion force, sharps safety (ISO 23908), and kink resistance (EN 13868).
Technological Characteristics
Radiopaque polyurethane catheter (20% Barium Sulfate); stainless steel/nitinol guidewire. Single-lumen, 18/20/22 gauge, 6-8 cm lengths. Passive needle safety mechanism; bloodless hub seal/extension line. Sterilized via Ethylene Oxide (EO). Complies with ISO 10993, ISO 10555, ISO 11070, ISO 23908, and BS EN 13868.
Indications for Use
Indicated for patients requiring peripheral vascular access for short-term use (<30 days) for blood sampling, blood pressure monitoring, fluid administration, or high-pressure injection. Includes safety feature to minimize sharps injury risk.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
Predicate Devices
Arrow® Endurance™ Extended Dwell Peripheral Catheter System (K151513)
Related Devices
K151513 — Arrow Endurance Extended Dwell Peripheral Catheter System · Arrow Internation, Inc. (Subsidiary of Teleflex Inc.) · Jul 8, 2015
K163513 — Arrow Endurance Extended Dwell Peripheral Catheter System · Arrow International, Inc. (Subsidiary of Teleflex, Inc.) · Feb 24, 2017
K242281 — Arrow Endurance Extended Dwell Peripheral Catheter System (EDC-00620); Arrow Endurance Extended Dwell Peripheral Catheter System (EDC-00622); Arrow Endurance Extended Dwell Peripheral Catheter System (EDC-00820); Arrow Endurance Extended Dwell Peripheral Catheter System (EDC-00822) · Arrow International, LLC (A subsidiary of Teleflex, Inc.) · Dec 20, 2024
K140504 — ACCUCATH BC INTRAVASCULAR CATHETER SYSTEM · Vascular Pathways, Inc. · Jul 17, 2014
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or a stylized human profile, composed of three overlapping shapes.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 23, 2015
Arrow International, Inc. (subsidiary Of Teleflex Inc.) Ms. Tracy Larish Regulatory Affairs Specialist 2400 Bernville Rd Reading, Pennsylvania 19605
Re: K152272
Trade/Device Name: Arrow Endurance™ Extended Dwell Peripheral Catheter System Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II Product Code: FOZ Dated: October 23, 2015 Received: October 23, 2015
Dear Ms. Larish:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Tina Kiang -
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known): _ K152272
Device Name: Arrow Endurance™ Extended Dwell Peripheral Catheter System
#### Indications for Use:
The ARROW Endurance catheter system permits access to the patient's peripheral vascular system for short-term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids. The catheter may be used for high pressure injection. The safety feature is intended to minimize the risk of sharps injuries.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
#### 510(k) SUMMARY
(as required by the Safe Medical Devices Act of 1990 and in accordance with 21 CFR 8807.92(a))
#### FOR
# THE ARROW ENDURANCE™ EXTENDED DWELL PERIPHERAL CATHETER SYSTEM
#### 1. Submitter Information
| Name: | Arrow International, Inc. (subsidiary of Teleflex Inc.) | |
|-------------------|---------------------------------------------------------|--|
| Address: | 2400 Bernville Road | |
| | Reading, PA 19605-9607 | |
| Telephone Number: | (610) 378-0131 | |
| Contact Person: | Tracy Larish | |
| | Regulatory Affairs Specialist | |
| Telephone Number: | (610) 378-0131 Extension 603835 | |
| Fax Number: | (610) 374-5360 | |
| Email: | tracy.larish@teleflex.com | |
| Date Prepared: | August 11, 2015 | |
## 2. Device Name
| Device Trade Name: | Arrow® Endurance™ Extended Dwell Peripheral Catheter System |
|----------------------|--------------------------------------------------------------------------------------------------------|
| Common Name: | Peripheral Intravascular Catheter |
| Classification Name: | Catheter, intravascular, therapeutic, short-term less than 30 days<br>(Class II, FOZ, 21 CFR 880.5200) |
## 3. Predicate Devices
- K151513: Arrow® Endurance™ Extended Dwell Peripheral Catheter System .
#### 4. Device Description
The purpose of this Traditional 510(k) submission is a product line extension to the Arrow Endurance™ Extended Dwell Peripheral Catheter System (K151513) to add 18 gauge and 22 gauge catheters in both 6 and 8 cm lengths. Arrow International's legally marketed predicate device, the Arrow Endurance™ Extended Dwell Peripheral Catheter System (K151513), was cleared with a 20 gauge catheter in lengths of 6 cm and 8 cm. The below device description has been modified to include the entire product offering.
{4}------------------------------------------------
The Arrow® Endurance™ Extended Dwell Peripheral Catheter System is a sterile, single use peripheral intravascular device designed to permit access to the peripheral vascular system. The insertion device consists of an ergonomically designed handle with an integral echogenic needle with a passively-activated needle protection mechanism, guide wire with slider advancer, catheter release tab, and single-lumen catheter is advanced over the needle and threaded over a guidewire into a peripheral vessel. Throughout catheter insertion. blood is contained within the device to aid in prevention of blood exposure. The catheter system consists of a translucent radiopaque polyurethane catheter, a needle with openings to enhance flashback visibility, a seal in the catheter hub designed to reduce blood exposure, a stabilization with a strain relief nose designed to reduce kinking at the hub of the catheter, integrated extension tubing with Luer hub, vent plug to prevent blood from leaking out during insertion, and a clamp to eliminate blood exposure when the vent plug is removed and replaced with a mating Luer component such as an infusion set or Luer access device.
The catheter is intended for short-term use (less than 30 days) to permit delivery of infusion therapies, infusion of blood and blood products, pressure monitoring, high pressure injection at a maximum of 325 psi, and withdrawal of blood.
The Arrow Endurance™ Extended Dwell Peripheral Catheter System is available in single lumen, 18, 20 and 22 gauge configurations with usable lengths of 6 cm (2.36") and 8 cm (3.15").
## 5. Intended Use
The Arrow Endurance™ Extended Dwell Peripheral Catheter System is intended for short-term use (less than 30 days) to permit delivery of infusion therapies, infusion of blood and blood products, pressure monitoring, high pressure injection and withdrawal of blood.
## 6. Indications for Use
The indications for use for the subject device are identical to the predicate and are listed below:
The Arrow Endurance catheter system permits access to the patient's peripheral vascular system for short-term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids. The catheter may be used for high pressure injection. The safety feature is intended to minimize the risk of sharps injuries.
## 7. Technological Characteristics
The Arrow Endurance™ Extended Dwell Peripheral Catheter System is substantially equivalent to the predicate Arrow Endurance™ Extended Dwell Peripheral Catheter System (K151513) in terms of indications for use, intended use, design, functional performance and materials of construction.
{5}------------------------------------------------
| Design Feature | Subject device:<br>Arrow Endurance™ Extended<br>Dwell Peripheral Catheter System | Predicate device:<br>Arrow Endurance™ Extended Dwell<br>Peripheral Catheter System (K151513) |
|-----------------------------|----------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| Catheter body OD | 18 Ga. and 22 Ga. | 20 Ga. |
| Catheter body ID | 0.039" (18 Ga.) and 0.027" (22 Ga.) | 0.032" (20 Ga.) |
| Catheter body material | Polyurethane | Polyurethane |
| Catheter usable length | 6 cm (2.36")<br>8 cm (3.15") | 6 cm (2.36")<br>8 cm (3.15") |
| Catheter body Radiopacifier | 20% Barium Sulfate | 20% Barium Sulfate |
| Integrated Guide wire | Yes | Yes |
| Guide wire material/size | Stainless Steel and Nitinol/0.010" | Stainless Steel and Nitinol/0.010" |
| Needle safety feature | Yes | Yes |
| Blood safety feature | Bloodless (seal and extension line) | Bloodless (seal and extension line) |
| Pressure Injection Limits | 325 psi | 325 psi |
| Sterilization Method | EO | EO |
| Flashback visualization | Yes | Yes |
## 8. Nonclinical Testing
Bench testing performed on the Arrow Endurance™ Extended Dwell Peripheral Catheter System supports substantial equivalence of the subject devices. The following testing has been completed for the subject devices:
- Biocompatibility in accordance with ISO 10993-1, 10999-3, 10993-4, 10993-5, o 10993-6, 10993-7, 10993-10, 10993-11 and 10993-12
- Requirements from ISO 10555 and ISO 11070 including: O
- l Radio-Detectability
- l Surface
- l Corrosion Resistance
- I Freedom from leakage (Air and Liquid)
- I Flow rate
- Pressure Injection
- I Tensile Testing
- Blood Containment
- Luer testing in accordance with BS EN 20594-1(ISO 594-1) & BS EN 1707 O
- Simulated Use testing o
- Penetration/Insertion force O
- Sharps Safety Feature performance in accordance to CDRH Sharps Guidance and O ISO 23908
- Catheter Kink resistance in accordance to BS EN 13868 o
## 9. Conclusions
The described intended use, results of verification testing performed, comparison of design and fundamental technology and comparison testing to the predicate device demonstrated that the subject 18 gauge and 22 gauge devices are substantially equivalent to the legally marketed predicate device, Arrow Endurance™ Extended Dwell Peripheral Catheter System (K151513).
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.