K031387 · Hill-Rom Manufacturing, Inc. · FMZ · Aug 1, 2003 · General Hospital
Device Facts
Record ID
K031387
Device Name
ISOLETTE INFANT INCUBATOR, MODEL C2000E
Applicant
Hill-Rom Manufacturing, Inc.
Product Code
FMZ · General Hospital
Decision Date
Aug 1, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5400
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Air Shields ISOLETTE® Infant Incubator is designed to care for the smaller premature baby as well as the healthier full term baby. It does this by providing a controlled environment, one in which the baby can be provided with the necessary care as well as being left undisturbed in the security of the incubator. It is to this end that the product can be used in any department of the hospital that provides neonatal and infant care. One would typically expect the ISOLETTE® to be used in the NICU/SCBU (Neonatal Intensive Care Unit and/or the Special Baby Care Unit). The design lends itself to all levels of care in the NICU making it suitable for use in level I, II, III, and IV where applicable. Other departments would include the Step Down Nursery, Newborn Nursery and Pediatrics. This device, when fitted with the battery backup feature, may be used to move the infant from one area of the hospital to another while maintaining the controlled environment. This device is not intended for home use. This is a prescription device.
Device Story
C2000e Isolette is an enhancement of the C2000 Infant Incubator; provides controlled thermal/environmental support for neonates. Enhancements include a horizontal rail system for accessory attachment and a battery backup system for intra-hospital transport during power failure or movement. Operated by clinical staff in hospital settings (NICU, SCBU, nurseries, pediatrics). Device maintains stable environment; battery backup ensures continuity of care during transit. Benefits include undisturbed care and environmental stability for premature and full-term infants.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Neonatal incubator with horizontal accessory rail and battery backup system. Maintains controlled thermal environment. No specific materials or software algorithms described.
Indications for Use
Indicated for premature and full-term infants requiring a controlled environment for care. Suitable for use in NICU/SCBU (levels I-IV), step-down nurseries, newborn nurseries, and pediatrics. Not for home use; prescription only.
Regulatory Classification
Identification
A neonatal incubator is a device consisting of a rigid boxlike enclosure in which an infant may be kept in a controlled environment for medical care. The device may include an AC-powered heater, a fan to circulate the warmed air, a container for water to add humidity, a control valve through which oxygen may be added, and access ports for nursing care.
K182977 — Isolette 8000 Plus · Draeger Medical Systems, Inc. · Nov 21, 2018
K213551 — Giraffe Omnibed Carestation CS1 · Datex Ohmeda, Inc. · Mar 2, 2022
K980886 — BABYTHERM 8000 LDR · Draeger Medical, Inc. · May 29, 1998
K072512 — MODIFICATION TO: OHMEDA MEDICAL GIRAFFE INCUBATOR (GIRAFFE AND PANDA UNINTERRUPTIBLE POWER SUPPLY) · Datex-Ohmeda, Inc. · Oct 3, 2007
K022876 — MODIFICATION TO INFANT LIFE SUPPORT MODULE-20 · International Biomedical, Inc. · Sep 11, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
# AUG = 1 2003
K031387
#### Section I 510(k) Summary of Safety and Effectiveness
Applicant:
Hill-Rom Air Shields 330 Jacksonville Road Hatboro, Pa. 19040 Registration No: 2510954
Contact Person:
Monica Ferrante Ph 215-682-8691 Fax 215-682-8689
Device trade/proprietary name:
C2000e Isolette® Infant Incubator
Device common/usual/classification name:
Neonatal Incubator
Classification:
General Hospital 21 CFR 862.113 Neonatal Incubator, ???, Class II
Performance Standards:
None applicable
Predicate Device:
K960980 C2000 Isolette® Infant Incubator
Device Description
The C2000e is an enhancement to the currently marketed C2000 Isolette Infant Incubator originally cleared under 510(k) K960980. The features, functions, and performance of the C2000 incubator remain the same. The C2000e enhancements are additions to the C2000 product features. These additions are a horizontal rail system for attachment of accessories and a battery back up for the unit to maintain the controlled environment in
{1}------------------------------------------------
the event of a power failure or to move the infant within the hospital environment.
## Intended Use
The Air Shields ISOLETTE® Infant Incubator is designed to care for the smaller premature baby as well as the healthier full term baby. It does this by providing a controlled environment, one in which the baby can be provided with the necessary care as well as being left undisturbed in the security of the incubator.
It is to this end that the product can be used in any department of the hospital that provides neonatal and infant care. One would typically expect the ISOLETTE® to be used in the NICU/SCBU (Neonatal Intensive Care Unit and/or the Special Baby Care Unit). The design lends itself to all levels of care in the NICU making it suitable for use in level I, II, III, and IV where applicable. Other departments would include the Step Down Nursery, Newborn Nursery and Pediatrics.
This device, when fitted with the battery backup feature, may be used to move the infant from one area of the hospital to another while maintaining the controlled environment.
Description of Modifications
The modifications to the device include a horizontal rail system for attachment of accessories and a battery back up for the unit to maintain the controlled environment in the event of a power failure or to move the infant within the hospital environment.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized caduceus, a symbol often associated with healthcare, featuring a staff with a serpent entwined around it.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG - 1 2003
Ms. Monica Ferrante Regulatory Affairs Hill-Room Manufacturing, Incorporated 330 Jacksonville Road Hatboro, Pennsylvania 19040
Re: K031387
Trade/Device Name: C2000e Isolette® Infant Incubator Regulation Number: 880.5400 Regulation Name: Neonatal Incubator Regulatory Class: II Product Code: FMZ, FPL Dated: June 26, 2003 Received: June 27, 2003
Dear Ms. Ferrante:
We have reviewed your Section 510(k) premarket notification of intent to market the device reterenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
### Page 2 -- Ms. Ferrante
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Patricia Cucurullo
Susan Runner, DDS, MA Interim Direct Division of Anesthesiology, General Hospital Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### SMDA Requirements Section A
### Indication for Use Statement
510(k) Number:
Device Name: C2000e Isolette® Infant Incubator
Indications for Use:
The Air Shields ISOLETTE® Infant Incubator is designed to care for the smaller premature baby as well as the healthier full term baby. It does this by providing a controlled environment, one in which the baby can be provided with the necessary care as well as being left undisturbed in the security of the incubator.
It is to this end that the product can be used in any department of the hospital that provides neonatal and infant care. One would typically expect the ISOLETTE® to be used in the NICU/SCBU (Neonatal Intensive Care Unit and/or the Special Baby Care Unit). The design lends itself to all levels of care in the NICU making it suitable for use in level I, III, III, and IV where applicable. Other departments would include the Step Down Nursery, Newborn Nursery and Pediatrics.
This device, when fitted with the battery backup feature, may be used to move the infant from one area of the hospital to another while maintaining the controlled environment.
This device is not intended for home use.
This is a prescription device.
(Please do not write below this line continue on another page if needed) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109) OR
Over-the-Counter Use
Patricia Cusciente
(Optional Format 1/2/96)
sthesiology. General Hosni
510(k) Number: K031387
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.