K120642 · Draeger Medical Systems, Inc. · FMT · Nov 2, 2012 · General Hospital
Device Facts
Record ID
K120642
Device Name
RESUSCITAIRE WITH AUTOBREATH
Applicant
Draeger Medical Systems, Inc.
Product Code
FMT · General Hospital
Decision Date
Nov 2, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5130
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Resuscitaire® Radiant Warmer is intended for thermoregulation, skin temperature monitoring, APGAR timing and resuscitation of newly born infants up to 10kg. It is not intended for long term resuscitation or home use.
Device Story
Resuscitaire Radiant Warmer provides thermoregulation, skin temperature monitoring, and APGAR timing for newborns. Modified version includes AutoBreath feature for automated resuscitation. Device consists of bassinet, warmer, and controller module. AutoBreath feature uses pneumatically driven logic circuit to provide gas-powered, time-cycled respiratory support via face mask. Operates in labor and delivery settings; used by clinicians. Provides consistent I:E ratio and respiratory rate, reducing human variation compared to manual resuscitation. Benefits include standardized, automated respiratory support during critical neonatal care. Output includes thermal regulation and controlled gas delivery for resuscitation.
Clinical Evidence
Bench testing only. Performance testing, functional/operation testing, verification and validation, biocompatibility (ISO 10993), and risk analysis were conducted to demonstrate substantial equivalence. No clinical data presented.
Technological Characteristics
Infant radiant warmer with integrated pneumatically driven logic circuit for AutoBreath. Materials biocompatible per ISO 10993. Connectivity: None. Sterilization: Not applicable. Software: None (pneumatic logic). Conical connectors per ISO 5356-1.
Indications for Use
Indicated for thermoregulation, skin temperature monitoring, APGAR timing, and resuscitation of newly born infants up to 10kg. Not for long-term resuscitation or home use.
Regulatory Classification
Identification
The infant radiant warmer is a device consisting of an infrared heating element intended to be placed over an infant to maintain the infant's body temperature by means of radiant heat. The device may also contain a temperature monitoring sensor, a heat output control mechanism, and an alarm system (infant temperature, manual mode if present, and failure alarms) to alert operators of a temperature condition over or under the set temperature, manual mode time limits, and device component failure, respectively. The device may be placed over a pediatric hospital bed or it may be built into the bed as a complete unit.
Special Controls
*Classification.* Class II (Special Controls):(1) The Association for the Advancement of Medical Instrumentation (AAMI) Voluntary Standard for the Infant Radiant Warmer;
(2) A prescription statement in accordance with § 801.109 of this chapter (restricted to use by or upon the order of qualified practitioners as determined by the States); and
(3) Labeling for use only in health care facilities and only by persons with specific training and experience in the use of the device.
K201628 — Panda iRes Warmer, Giraffe Warmer · GE Healthcare · Feb 12, 2021
K090697 — MODIFICATION TO: GIRAFFE AND PANDA WARMER · Ohmeda Medical · Apr 16, 2009
K213551 — Giraffe Omnibed Carestation CS1 · Datex Ohmeda, Inc. · Mar 2, 2022
K123309 — GIRAFFE AND PANDA WARMERS · Ohmeda Medical · Feb 5, 2013
K230278 — Babyroo TN300 · Draeger Medical Systems, Inc. · May 30, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
K12064'2
510(k) Summary acc. to 807.92
Drager Medical Systems, Inc.
NOV 2 2012 3135 Quarry Road Telford, PA 18969 Establishment Registration Number: 2510954 Contact Person: Gale Winarsky Manager, Regulatory Affairs Phone: 215-660-2239 215-721-5424 Fax: Date summary was prepared: 2012-10-04 Device Name: Trade Name: Resuscitaire Radiant Warmer
Classification Name: Regulation Number: Product Code: Class:
Manufacturer Name and Address:
Infant Radiant Warmer 21 CRF 880.5130 -FMT ll
Legally Marketed Device Identification: Substantial equivalence is claimed to the Resuscitaire® Radiant Warmer K003335, Neopuff Infant Resuscitator K892885, and MVP-10 Infant Ventilator (MVP10) K89638.
## Device Description:
The modified Resuscitaire Radiant Warmer is for newbom infants and consists of a bassinet, warmer and a controller module which provides heat control, monitoring of the skin temperature and an APGAR timer. It also includes optional manual and automated resuscitation with suction and oxygen delivery and a patient gas supply breathing circuit.
### Intended Use:
The Resuscitaire Radiant Warmer is intended for thermoregulation, skin temperature monitoring, APGAR timing and resuscitation of newly born infants up to 10kg. It is not intended for long term resuscitation or home use.
{1}------------------------------------------------
Comparison of Technological Characteristics with Predicate Devices:
·
. "
| Specification | Predicate | Predicate | Predicate | Device<br>Under<br>Review | Comments |
|------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Resuscitaire®<br>Radiant Warmer | MVP-<br>10K1Ventilator<br>(MVP10)<br>Predicate for<br>AutoBreath<br>Feature Only | Neopuff In-<br>fant Resusci-<br>tator<br>Predicate for<br>AutoBreath<br>Feature Only | Resuscitaire®<br>Radiant Warmer | |
| Manufacturer | Hill-Rom Air<br>Shields | Bio-Med<br>Devices Inc. | Fisher &<br>Paykel | Draeger Medical<br>Systems, Inc. | |
| 510(k) Number | K003335 | K896381 | K892885 | | |
| Regulation Number | 880.5130 | 868.5895 | 868.5925 | 880.5130 | |
| Product Code | FMT | CBK | BTL | FMT | |
| Classification | II | II | II | II | |
| Intended Use | Intended for<br>thermoregulation,<br>skin temperature<br>monitoring,<br>APGAR timing<br>and resuscitation<br>of newborn in-<br>fants in the labor<br>and delivery<br>room. | Intended for<br>respiratory<br>support of neo-<br>natal and pedi-<br>atric patients<br>both in hospital<br>and during<br>transport. Pri-<br>marily for use<br>in applications<br>requiring tidal<br>volume up to<br>660 ml. May<br>be used with a<br>wide range of<br>I:E ratios in-<br>cluding inverse<br>ratios. | Intended for<br>resuscitation<br>of infants in<br>labor & deliv-<br>ery, postnatal<br>wards, ORs,<br>transport,<br>special care<br>baby units,<br>and NICUs | Thermoregulation,<br>skin temperature<br>monitoring,<br>APGAR timing<br>and resuscitation<br>of newborn in-<br>fants. It is not<br>intended for home<br>use or long-term<br>resuscitation | The Intended Use<br>has been modi-<br>fied to clarify the<br>use of the device<br>with or without<br>the AutoBreath<br>option. |
| Indications for Use | Thermoregulation,<br>skin temperature<br>monitoring,<br>APGAR timing<br>and resuscitation<br>of newborn in-<br>fants. | Not Found on<br>FDA Site | Not Found on<br>FDA Site | Thermoregulation,<br>skin temperature<br>monitoring,<br>APGAR timing<br>and resuscitation<br>of newly born<br>infants up to 10<br>kg. It is not in-<br>tended for long-<br>term resuscitation<br>or home use. | The Indications<br>have been modi-<br>fied to clarify the<br>use of the device<br>with or without<br>the AutoBreath<br>option. |
| Target<br>Population/Patient<br>Population | Newborn infants | Newborn and<br>Pediatrics | Newborns<br>and<br>Pediatrics | Newly born in-<br>fants up to 10 kg. | The 10 kg is in<br>reference to the<br>weight the Re-<br>suscitaire Radiant<br>Warmer bed is<br>made to hold. |
| Environment of<br>Use | Labor and deliv-<br>ery room in a<br>Healthcare Facil-<br>ity | | | Labor and deliv-<br>ery setting in a<br>Healthcare Facil-<br>ity | |
| AutoBreath™<br>Infant Resuscita-<br>tor Feature (AB) | | | | | |
| Specification | Predicate | Predicate | Predicate | Device Under Review | Comments |
| Device Name | Resuscitaire®<br>Radiant Warmer | MVP-<br>10K1Ventilator<br>(MVP10)<br>Predicate for<br>AutoBreath<br>Feature Only | Neopuff In-<br>fant Resusci-<br>tator<br>Predicate for<br>AutoBreath<br>Feature Only | Resuscitaire®<br>Radiant Warmer | |
| Operating Principle | N/A | Gas powered,<br>continuous<br>flow, time cy-<br>cled BPM,<br>pneumatically<br>driven logic<br>circuit | Gas powered,<br>continuous<br>flow, manu-<br>ally cycled<br>BPM | Same as<br>Predicate MVP-10 | |
| I:E Ratio | N/A | 1:8 to 3:1<br>(adjustable) | Manually<br>estimated | Fixed internally at<br>1:2 nominal (1:1.6<br>to 1:2.2) (non<br>adjustable) | The fixed I:E ratio<br>of the AB falls<br>within the normal<br>range for neona-<br>tal ventilation as<br>published in the.<br>Assisted Ventila-<br>tion of the Neo-<br>nate 4th Edition<br>copyright 2003,<br>(1:1 to 1:3). Addi-<br>tionally the 2010<br>AHA Guidelines<br>for neonatal re-<br>suscitation rec-<br>ommend breath<br>rates of 40 -60<br>BPM.<br><br>All three devices<br>can provide I:E<br>ratios within the<br>recommended<br>range<br><br>The Neopuff is<br>manually esti-<br>mated and con-<br>trolled by the user<br>therefore both the<br>I:E ratio and BPM<br>will vary.<br><br>The preset I:E<br>ratio of the RW82<br>w/ AB eliminates<br>human variation<br>by guaranteeing<br>a consistent I:E<br>ratio.<br><br>The MVP-10<br>incorporates a |
| Specification | Predicate | Predicate | Predicate | Device Under Review | Comments |
| Device Name | Resuscitaire®<br>Radiant Warmer | MVP-<br>10K1Ventilator<br>(MVP10)<br>Predicate for<br>AutoBreath<br>Feature Only | Neopuff In-<br>fant Resusci-<br>tator<br>Predicate for<br>AutoBreath<br>Feature Only | Resuscitaire®<br>Radiant Warmer | |
| Adjustable PEEP | N/A | Variable up to<br>$18 ± 3$ cm H2O<br>at a flow of 6<br>l/min | @ 5 LPM 1-5<br>cm H20/mbar<br>@ 8 LPM 1-9<br>cm H20/mbar<br>@ 10 LPM 2-<br>15 cm<br>H20/mbar<br>@ 15 LPM 3-<br>25 cm<br>H20/mbar | @ 5 LPM<br>minimum PEEP<br><2 cm H2O<br>@ 10 LPM<br>minimum PEEP ≤<br>4 cm H2O<br>@15 LPM<br>maximum PEEP<br>> 14 cm H2O | ratios including<br>inversion ratios<br>for use outside<br>L&D and with<br>larger patients.<br><br>The RW82 w/AB<br>is only used for<br>clinician adminis-<br>tered short term<br>resuscitation in<br>L&D settings.<br>Although not<br>identical, the<br>PEEP ranges are<br>similar. Based on<br>the application of<br>use the differ-<br>ences are not<br>significant.<br><br>For newborns the<br>recommendations<br>are: minimum<br>PEEP from 3 to 4<br>cmH2O, and flow<br>rate 5 to 8 LPM<br>for low birth<br>weight infants<br>and flow rate 6 to<br>10 LPM for term<br>infants |
| Adjustable<br>Respiratory Rate<br>Range | Achieved via bag<br>and mask resusci-<br>tation, user de-<br>pendant | Variable from 0<br>to 120 BPM | Achieved via<br>t-piece<br>breathing<br>circuit, user<br>dependant | 18 to 60 BPM ±<br>10% of setting | Both the RW82<br>w/ AB and the<br>MVP10 allow the<br>user to set the<br>respiratory rate.<br>The advantage of<br>a set respiratory<br>rate is to elimi-<br>nate the human<br>variation associ-<br>ated with manu-<br>ally controlled<br>rates (RW82 &<br>Neopuff).<br><br>The MVP10 has<br>a larger respira-<br>tory rate range. |
| | Predicate | | | | |
| evice Name | Resuscitaired | 10K1Ventilato<br>MVP10)<br>Predicate fo<br>AutoBreath<br>Feature Only | Neopuff In-<br>fant Resusci-<br>aror<br>Predicate for<br>AutoBreath<br>Feature Only | Radiant Warmer | |
| | | | | | is due to the pa-<br>tient population of<br>the MVP10 which<br>includes pediat-<br>rics. The RW82<br>w/ AB is only<br>intended for the<br>resuscitation of<br>newborns in L&D<br>environments. |
| Adjustable Airway<br>Pressure Relief<br>(range of working<br>pressure) (PIP) | None | 10 cm H20 + 5<br>to 80 cm H2O<br>+ 10 | 2 to 80 cm<br>H2O | 0 to 50 cmH2O +<br>5 cm H2O (0 to<br>4.9 kPa + 0.5<br>kPa) | The MVP10 and<br>the Neopuff have<br>a larger range<br>based on patient<br>populations for<br>ventilation, while<br>RW82 w/ AB is<br>only intended for<br>the resuscitation<br>of newborns in<br>L&D environ-<br>ments. |
| Fixed Max Pres-<br>sure (P Lim min) | 160 cm H2O<br>(15.7 kPa ) ± 10% | Variable up to<br>70 cm H2O ±<br>10 | Variable up to<br>80 cm H2O | 50 cm H2O +<br>10% (5.0 kPa +<br>10%) | The differences<br>between the fixed<br>max pressures of<br>the MVP-10,<br>Neopuff & Auto-<br>Breath are re-<br>lated to the envi-<br>ronment in which<br>resuscitation is<br>provided. |
| | | | | | The Neopuff uses<br>a T-piece circuit<br>with manually<br>adjustable PEEP<br>& PIP with higher<br>fixed max. pres-<br>sure for use in<br>the NICU. |
| | | | | | Like the Neopuff,<br>the RW82 w/ AB<br>also provides<br>manually adjust-<br>able PEEP & PIP.<br>It differs in that it<br>provides a lower<br>fixed max. pres-<br>sure because the |
| Specification | Predicate | Predicate | Predicate | Device Under Review | Comments |
| Device Name | Resuscitaire®<br>Radiant Warmer | MVP-<br>10K1Ventilator<br>(MVP10)<br>Predicate for<br>AutoBreath<br>Feature Only | Neopuff In-<br>fant Resusci-<br>tator<br>Predicate for<br>AutoBreath<br>Feature Only | Resuscitaire®<br>Radiant Warmer | |
| Fixed Min Pressure<br>(P Lim min) | 0 cmH2O | 0 cmH2O | 0 cm H2O ±2 | Same as<br>predicate RW82 | intended use is<br>limited to the<br>L&D.<br><br>The RW82 is not<br>a predicate for<br>this function. |
| Logic Circuit | None | Approx 4 LPM<br>at 50 BPM | None | 5 LPM | Like the MVP-10<br>the RW82 w/ AB<br>is a pneumatically<br>controlled device.<br>Both circuits con-<br>trol the time inter-<br>vals between<br>exhalation and<br>inspiration and<br>the pressure in<br>the patient circuit. |
| Dead Space | N/A | 0.5 ml max | 6 ml | 3.4 ml | Dead space will<br>vary based on the<br>make up/mfg. of<br>the patient<br>breathing circuit. |
| Body Weight<br>Range | N/A | Applications<br>requiring TV up<br>660 mls. | Up to 10 kg | Same as Neopuff | |
| Patient Interface | N/A | Endotracheal<br>Tube | Face Mask or<br>Endotracheal<br>Tube | Face Mask | |
| Infant<br>Resuscitator<br>Feature | | | | | |
| Patient Gas Supply<br>Airway Pressure<br>Limit, Operator<br>Adjustable | 0 to 50 cm H2O<br>(0 to 4.9 kPa) ±<br>10% of max set-<br>ting | N/A | N/A | Same as<br>predicate RW82 | |
| Patient Outlet Flow<br>Control Range | 0 to 15 LPM ± 3%<br>of full scale or ±<br>10% of setting,<br>whichever is<br>greater | N/A | N/A | Same as<br>predicate RW82 | |
| Fixed Airway Pres-<br>sure Limit, Preset | 60 cm H2O (4.9<br>kPa) ± 20 % of<br>setting | N/A | N/A | Same as<br>predicate RW82 | |
| Auxiliary Supply<br>Pressure Limit | 160 cm H2O<br>(15.7 kPa) ± 10%<br>of setting | N/A | N/A | Same as<br>predicate RW82 | |
| Specification | Predicate | Predicate | Predicate | Device<br>Under<br>Review | Comments |
| Device Name | Resuscitaire®<br>Radiant Warmer | MVP-<br>10K1Ventilator<br>(MVP10)<br>Predicate for<br>AutoBreath<br>Feature Only | Neopuff In-<br>fant Resusci-<br>tator<br>Predicate for<br>AutoBreath<br>Feature Only | Resuscitaire®<br>Radiant Warmer | |
| Auxiliary Flow<br>Control Range | 0 to 15 LPM ± 3%<br>of full scale or ±<br>10 % of setting<br>whichever is<br>greater | N/A | N/A | Same as<br>predicate RW82 | |
:
・
:
{2}------------------------------------------------
・
-・
:
Page 3 of 8
·
·
:
{3}------------------------------------------------
· .
. . . .
.
Page 4 of 8
{4}------------------------------------------------
1
:
Page 5 of 8
·
·
{5}------------------------------------------------
1. September 2017
{6}------------------------------------------------
### Discussion of Non-clinical Studies:
The modification of the RW82 to include AutoBreath was tested in accordance with applicable standards, guidance and internal design control procedures including performance testing, functional/operation testing, verification and validation, biocompatibility, risk analysis and verification of risk control measures and was determined to be as safe and effective for its intended use as the predicates. Testing performed indicate that the modifications as described in this submission have not altered the fundamental application of the device in its intended environment.
### Biocompatibility:
Testing was performed to ISO 10993. The results show the relevant components to be biocompatible.
### Sterilization:
Not applicable
Standards and Guidance:
Performance Standards:
None
International Standards:
ISO 5356-1 Anaesthetic and respiratory equipment - Conical connectors - Part 1: Cones and sockets
ISO 10651-5:2006 -- Lung ventilators for medical use -- Particular requirements for basic safety and essential performance - Part 5
AAMI/ANSI/ISO 10993-1:2009, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
AAMI/ANSI/ISO 10993-1:2009, Biological evaluation of medical devices - Part 5, Tests for in vitro Cytotoxicity
AAMI/ANSI/ISO 10993-1:2009, Biological evaluation of medical devices - Part 10, Tests for irritation and delayed hypersensitivity
AAMI/ANSI/ISO 10993-1:2009, Biological evaluation of medical devices - Part 12, Sample preparation and reference materials
ISO 5356-1 Anaesthetic and respiratory equipment - Conical connectors - Part 1: Cones and sockets
Page 7 of 8
{7}------------------------------------------------
**Dräger**
Guidance:
Draft Emergency Resuscitator Guidance, April 1993 Draft Reviewer Guidance for Ventilators, July 1995
Conclusion Drawn from Non-Clinical Studies:
The results of the non-clinical testing, and comparison to the predicate devices show that the modified Resuscitaire Radiant Warmer meets the performance requirements of the standards and guidance mentioned above and is substantially equivalent to the predicate devices.
{8}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Dräger Medical Systems, Incorporated Ms. Gale Winarsky Manager, Regulatory Affairs 3135 Quarry Road Telford, Pennsylvania 18969
November 2, 2012
Re: K120642
Trade/Device Name: Resuscitaire Radiant Warmer Regulation Number: 21 CFR 880.5130 Regulation Name: Infant Radiant Warmer Regulatory Class: II Product Code: FMT Dated: October 5, 2012 Received: October 9, 2012
Dear Ms. Winarsky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{9}------------------------------------------------
## Page 2 - Ms. Winarsky
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go'to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Sm…
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.