K970737 · Bd Becton Dickinson Vacutainer Systems Preanalytic · FMI · Mar 28, 1997 · General Hospital
Device Facts
Record ID
K970737
Device Name
B-D ULTRA-FINE II SHORT 31G X 5/16 PEN NEEDLE
Applicant
Bd Becton Dickinson Vacutainer Systems Preanalytic
Product Code
FMI · General Hospital
Decision Date
Mar 28, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use for the modified device remains the same as the predicate device; for use with a pen injector device for the subcutaneous injection of insulin.
Device Story
Single-use, double-pointed 31G stainless steel needle; epoxy-bonded to polypropylene hub; protected by polyethylene shield and outer container with laminate seal. Device attaches to pen injector; penetrates insulin cartridge septum; enables subcutaneous insulin delivery. Used by patients or clinicians in home or clinical settings. Sterilized via gamma irradiation.
Clinical Evidence
Bench testing only. Evaluated needle pull-out, penetration, and insulin injection forces. Results demonstrate performance equivalent to predicate device.
Indicated for subcutaneous insulin injection using a pen injector device.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
Predicate Devices
B-D 30g x 5/16” pen needle
Related Devices
K120690 — INSULIN PEN NEEDLE · Tiger Medical Products, Ltd. · Jun 22, 2012
K152877 — SureFine Pen Needle · Shina Med Corporation · Jun 7, 2016
Submission Summary (Full Text)
{0}
K970757
# SUMMARY OF SAFETY AND EFFECTIVENESS
MAR 28 1997
## 1.0 Submitted By:
Peter Zurlo
Manager, Regulatory Affairs
BECTON DICKINSON CONSUMER PRODUCTS
1 Becton Drive
Franklin Lakes, NJ 07417-1883
Phone: 201-847-6447
Fax: 201-848-0457
## 2.0 Device Name:
B-D 31g x 5/16” pen needle
## 3.0 Predicate Device:
B-D 30g x 5/16” pen needle
## 4.0 Device Description:
The double-pointed Type 304 stainless steel needle is epoxy bonded to a plastic hub (polypropylene) and covered with a plastic needle shield (polyethylene). This needle/hub/shield assembly is placed in a plastic outer container (polyethylene), the open end of which is sealed closed with a laminate seal for subsequent protection. Gamma irradiation sterilization will be employed and the product will meet $10^{-6}$ sterility assurance levels. During use, the laminate seal is peeled away and the exposed needle hub is screwed onto the Pen injector device. This exposed needle hub will penetrate the rubber septum of an insulin containing cartridge (supplied by others). After the needle shield is removed, it is ready for a subcutaneous insulin injection. The needle assembly is a single-use disposable device.
## 5. Intended Use:
The intended use for the modified device remains the same as the predicate device; for use with a pen injector device for the subcutaneous injection of insulin.
021897.PZ1
19
{1}
6. Technological Characteristics:
The B-D 31g x 5/16” pen needle and the predicate device (the B-D 30g x 5/16” pen needle) have the same technological characteristics.
SEE ITEM 4 ABOVE FOR A DESCRIPTION.
N.B. The only change to the predicate device is the gauge size of the needle.
7. Performance Summary:
The B-D 31g x 5/16” pen needle has been bench tested, which included needle pull-out, penetration and insulin injection forces and has proven to function in an equivalent manner as the predicate device.
Based on the results of bench testing the B-D 31g x 5/16” pen needle is considered safe and effective when used as intended.
011195.pz1
20
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.