TSK STERIJEKT PREMIUM DISPOSABLE HYPODERMIC NEEDLE
Device Facts
| Record ID | K970370 |
|---|---|
| Device Name | TSK STERIJEKT PREMIUM DISPOSABLE HYPODERMIC NEEDLE |
| Applicant | Tsk Laboratory, Japan |
| Product Code | FMI · General Hospital |
| Decision Date | Apr 10, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 880.5570 |
| Device Class | Class 2 |
| Attributes | Pediatric |
Indications for Use
To inject fluids into, or withdraw fluids from parts of the body below the surface of the skin.
Device Story
TSK STERUEKT is a sterile, single-use hypodermic needle; consists of stainless steel beveled cannula bonded to plastic hub; compatible with standard luer lock or luer slip syringe tips. Used by clinicians for fluid injection or withdrawal from body tissues. Device provides physical conduit for fluid transfer; ensures patient safety through sterile, disposable design; prevents cross-contamination.
Clinical Evidence
Bench testing only. Results: Protector removal and needle bond strength met specifications. Biocompatibility testing passed for sensitization, intracutaneous reactivity, hemolysis, systemic toxicity, and cytotoxicity. LAL pyrogen test passed.
Technological Characteristics
Stainless steel cannula; plastic hub; luer lock/luer slip compatible; sizes 14G-31G; sterile, single-use; removable plastic protector.
Indications for Use
Indicated for patients requiring fluid injection or withdrawal via hypodermic needle; applicable for all ages and genders requiring subcutaneous, intramuscular, or intravenous access.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
Predicate Devices
- Becton Dickinson & Company, 'B-D PrecisionGlide' needles
Related Devices
- K210232 — Sterile Hypodermic Needles for Single Use · Zhejiang Kangkang Medical-Devices Co., Ltd. · Sep 23, 2021
- K191643 — TK Sterile Hypodermic Needle · Anhui Tiankang Medical Technology Co., Ltd. · Dec 19, 2019
- K132233 — K-PACK II NEEDLE - 27G X 3/4 THIN WALL · Terumo Europe N.V. · Sep 6, 2013
- K062608 — 29G AND 30G K-PACK II NEEDLE · Terumo Europe N.V. · Sep 15, 2006
- K231720 — Sterile Hypodermic Needles for Single Use · Wepon Medical Technology Co., Ltd. · Sep 8, 2023