DOYLE EXTRACTOR
Device Facts
| Record ID | K955748 |
|---|---|
| Device Name | DOYLE EXTRACTOR |
| Applicant | Safetech Intl., Inc. |
| Product Code | FMI · General Hospital |
| Decision Date | Mar 14, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 880.5570 |
| Device Class | Class 2 |
Indications for Use
The Doyle Extractor is a needle extractor intended for use as an accessory to IV administration sets. Specifically, the Doyle Extractor is intended for use during the removal of right-angle needles from implanted IV ports. It is manufactured from polypropylene and provides a suitable leverage for extraction of right-angle needles from the skin and implanted IV port. The device contains a slot on the top blade to secure the IV tubing prior to removal of the needle.
Device Story
Manual needle extraction tool; accessory to IV administration sets. Design utilizes simple lever principle to apply force for withdrawing right-angle needles from skin and implanted IV ports. Features slot on top blade to secure IV tubing during procedure. Used in clinical settings by healthcare providers. Benefits include providing leverage for safe, controlled needle removal.
Clinical Evidence
Bench testing only. Testing determined force required to withdraw right-angle needles from skin and injection ports and confirmed material/design adequacy to provide necessary force.
Technological Characteristics
Material: Polypropylene. Principle: Simple lever mechanism. Form factor: Handheld tool with top blade slot for IV tubing. Non-powered, manual device.
Indications for Use
Indicated for use in the removal of right-angle needles from implanted IV ports in patients requiring IV access.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
Predicate Devices
Related Devices
- K983480 — BPOS-IVAD STABILIZER · Bpos, Inc. · Nov 23, 1998
- K980062 — MONOJECTT BLUNTIP I.V. ACCESS CANNULA WITH VIAL ACCESS PIN DEVICE · Davis & Geck, Inc. · Mar 20, 1998
- K993848 — MILLENNIUM HUBER PLUS SAFETY INFUSION SET · Millennium Medical Distribution, Inc. · Jan 27, 2000
- K080976 — NEXUS I.V. FLUID TRANSFER DEVICE · Nexus Medical, LLC · Nov 10, 2008
- K980098 — VACUTAINER BRAND PRONTO NEEDLES HOLDER · Becton Dickinson Vacutainer Systems · Feb 5, 1998