Promisemed X-safety pen needle is intended for use with needle based injector for the injection of drugs.
Device Story
Single-use, double-ended, sterile pen needle for needle-based injection systems (NISs). Device attaches to pen injector via threaded hub. Patient sets dose on injector, inserts needle into skin, and presses button. Safety mechanism automatically retracts needle into shielded housing post-injection to prevent needle stick injuries; some models include secondary shield at cartridge end. Single-use; disposed of after injection. Benefits include reduced risk of accidental needle sticks for patients and healthcare providers.
Clinical Evidence
Bench testing only. No clinical data. Performance verified via ISO 9626 (tubing), ISO 11608-2 (NIS needles), ISO 7864 (sterile needles), and ISO 23908 (sharps protection). Biocompatibility confirmed per ISO 10993 series (cytotoxicity, irritation, sensitization, systemic toxicity, hemolysis, pyrogenicity). Sterility, package integrity, and 5-year shelf life validated per ISO 11135, ASTM F88, ASTM F1929, and ASTM F1980.
Technological Characteristics
Double-ended stainless steel (304, X5CrNi18-10) needle; polypropylene (PP) hub and container; stainless steel trigger/posterior springs; silicone oil lubricant. Dimensions: 29G-32G, 4mm-8mm lengths. EO sterilization (SAL 10^-6). Complies with ISO 9626, ISO 7864, ISO 11608-1/2, ISO 23908, ISO 10993, ASTM F756, ASTM F88, ASTM F1929, ASTM F1980.
Indications for Use
Indicated for use with needle-based injector devices for the injection of drugs in patients requiring subcutaneous or intramuscular injections.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
K210748 — Well-life TM Pen Needles, Well-life TM Safety Pen Needles · W. L. Med Co., Ltd. · May 25, 2022
K223286 — BD AutoShield Duo Pen Needle · Becton, Dickinson and Company · Dec 28, 2022
K110312 — CLICKFINE AUTOPROTECT PEN NEEDLE · Ypsomed AG · Jan 26, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 19, 2022
Promisemed Hangzhou Meditech Co., Ltd. % Wei Hsu Regulatory manager Vee Care (Asia) Limited 17th Chung Pont Commercial Building, 300 Hennessy Road Hong Kong, Hong Kong China
Re: K220129
Trade/Device Name: Promisemed X-Safety Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: August 24, 2022 Received: August 24, 2022
Dear Wei Hsu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220129
Device Name Promisemed X-Safety Pen Needle
Indications for Use (Describe)
Promisemed X-safety pen needle is intended for use with needle based injector for the injection of drugs.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### Date Prepared 1
September 19, 2022
### Submitter's Information 2
Name of Sponsor:
Promisemed Hangzhou Meditech Co., Ltd.
Address:
No. 1388 Cangxing Street, Cangqian Community, Yuhang District, Hangzhou
City, 311121 Zhejiang, China
Contact Name:
Zearou Yang
Telephone No.:
+86 571 88772985
Fax No.:
+86 571 88772985
Email Address:
zearou.yang@promisemed.ca
## ന Trade Name, Common Name, Classification
Trade/Product Name: X-Safety Pen Needle Common Name: Hypodermic single lumen needle Classification name: Needle, Hypodermic, Single Lumen Regulation Number: 21 CFR 880.5570 Device Class: Class II Product Code: FMI
## Identification of Predicate Device(s) ব K163578: EasyTouch Safety Pen Needle
{4}------------------------------------------------
### Description of the Device ഗ
X-Safety Pen Needle is a single-use, double-ended, sterile needles for needlebased injection systems (NISs) that fulfil the specifications of ISO 11608-1.
This device is constructed as a double-ended, stainless steel needle of various sizes that is with a threaded plastic hub.
The device includes needle shielding safety features (at both patient-end and cartridge-end or at patient-end only) to reduce the risk of needle stick injury and its container can be provided as short or long type.
It is packaged in a sealed sterility barrier, and the needle is lubricated.
This is a single-use device and delivered sterilization process is validated according to EN ISO 11135. Sterilization process undergoes routine control.
#### 6 Intended Use
Promisemed X-safety pen needle is intended for use with needle based injector for the injection of drugs.
## 7 Similarities and Differences of the Proposed Device to the Predicate Device
Promisemed X-Safety Pen Needle is substantially equivalent in its technologies and functions to the EasyTouch Safety Pen Needle which cleared under premarket notification number K163578. The following are comparisons between subject device and the predicate device.
A detailed comparison to the predicate is provided in Table 1.
| | Subject Device | Predicate Device<br>(K163578) | |
|-------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Promisemed X-Safety Pen<br>Needle | EasyTouch Safety Pen<br>Needle | Comments |
| Manufacturer | Promisemed Hangzhou<br>Meditech Co., Ltd | MHC Medical Products,<br>LLC. | |
| Device Class | Class II | Class II | Same |
| Product Code | FMI | FMI | Same |
| Regulation<br>number | 880.5570 | 880.5570 | Same |
| Regulation Name | Needle, Hypodermic,<br>Single Lumen | Needle, Hypodermic,<br>Single Lumen | Same |
| Intended Use/<br>Indications for<br>Use | Promisemed X-Safety Pen<br>Needle | The EasyTouch ™ Safety<br>Pen Needle | Same |
| | Needle is intended for use<br>with needle based<br>injector for the injection<br>of drugs. | is intended<br>for use with pen<br>injector devices for the<br>injection of drugs,<br>including insulin and<br>exenatide.<br>Additionally, the<br>attached safety shield<br>automatically locks in<br>place and reduces the<br>occurrence of<br>accidental needle sticks<br>from the patient end of<br>the needle. The shield<br>also serves to hide the<br>needle before and after<br>injection. | The subject device<br>include needle<br>shielding safety<br>features (at both<br>patient end and<br>cartridge end or at<br>patient end only)<br>to reduce the risk<br>of needle stick<br>injury. |
| Operating<br>Principle | Serves as a single use pen<br>needle. The device is<br>removed from its<br>packaging and screwed<br>into a pen injector<br>device. The patient then<br>injects the medicine by<br>first setting the dose on<br>the pen, and then<br>inserting the needle into<br>the skin and then pressing<br>the button on the pen.<br>After the injection the<br>needle automatically<br>retracts into a shielded<br>housing, thus preventing<br>an accidental needle<br>stick.<br>For safety features at<br>both patient end and<br>cartridge end: A second<br>shield covers the needle<br>tail when the needle is<br>removed from the<br>injector.<br>Once the sharps injury<br>prevention feature is<br>activated, it locks in<br>place and the product can<br>no longer be used again. | Serves as a single use<br>pen needle. The device<br>is removed from its<br>packaging and screwed<br>into a pen injector<br>device. The patient<br>then injects the<br>medicine by first<br>setting the dose on the<br>pen, and then inserting<br>the needle into the skin<br>and then pressing the<br>button on the pen.<br>After the injection the<br>needle automatically<br>retracts into a shielded<br>housing, thus<br>preventing an<br>accidental needle stick. | Different<br>An additional<br>protective design is<br>introduced in<br>subject device (for<br>safety features at<br>both patient end<br>and cartridge end)<br>which prevent<br>accidental needle<br>sticks at the<br>cartridge end upon<br>removal of the<br>needle from the<br>injector. |
| Gauge | 29G, 30G, 31G, 32G | 29G, 30G, 31G | Different<br>Subject device has<br>additional gauge<br>size (32G) than<br>predicate device. |
| Needle Length | 4mm, 5mm, 6mm, 8mm | 5mm, 6mm, 12.7mm | Different |
| | | | Subject device has shorter needle length than predicate device. |
| Material (Needle, Hub, Hub protector, Safety feature) | Needle Tube: 304<br>Stainless steel X5CrNi18-10 | Needle Tube: 304<br>Stainless steel needle | Same |
| | Needle Hub:<br>Polypropylene (PP) | Needle Hub:<br>Polypropylene (PP) | Same |
| | Container: Polypropylene (PP) | Hub protector:<br>Polypropylene (PP) | Same |
| | Trigger spring/Posterior spring: Stainless steel | Safety Feature:<br>Stainless steel spring | Same |
| | Lubricant: Silicon oil | Lubricant: Silicon oil | Same |
| Performance | Complied with ISO 9626, ISO 7864, ISO11608-1, ISO 11608-2 | Complied with ISO 9626, ISO 7864, ISO11608-1, ISO 11608-2 | Same |
| Sterilization | EO Sterilization | EO Sterilization | Same |
| | SAL:10-6 | SAL:10-6 | Same |
| Shelf Life | 5 years | 5 years | Same |
| Single use | Yes | Yes | Same |
| Biocompatibility | Complied with ISO10993 series standards, and the following tests are performed<br>- Cytotoxicity: No cytotoxicity<br>- Skin Irritation: No evidence of skin irritation<br>- Skin Sensitization: No evidence of sensitization<br>-Acute and Subacute Systemic Toxicity: No systemic toxicity<br>-Hemolysis: No evidence of hemolysis<br>-Pyrogen: Non-pyrogenic | Complied with ISO10993 series standards, and the following tests are performed and passed<br>- Cytotoxicity: No cytotoxicity<br>- Skin Irritation: No evidence of skin irritation<br>- Skin Sensitization: No evidence of sensitization<br>-Acute Systemic Toxicity: No systemic toxicity<br>- Hemolysis: No evidence of hemolysis<br>- Pyrogen: Non-pyrogenic | Same |
{5}------------------------------------------------
{6}------------------------------------------------
The Promisemed X-Safety Pen Needle has the same intended use and technological characteristics as the predicate device. The basic technological and operating principles are the same for both devices. Both the subject and predicate devices are disposable, sterile, single patient use devices.
The subject device has additional safety feature at the cartridge end which has been validated through ISO 23908. The subject device has wider gauge size and
{7}------------------------------------------------
shorter needle length than predicate device. The specification of needle length and gauze size fulfil the requirement of ISO 9626:2016 and ISO 7864:2016. The differences between the subject device and its predicate do not alter the intended use and do not raise new questions of safety and effectiveness.
### 8 Performance Testing Summary
The bench testing performed verifies that the performance of the subject devices is substantially equivalent in terms of critical performance characteristics to the predicate device. These tests include:
- ISO 9626: Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods
- · ISO11608-2: Needle-based injection systems for medical use -Requirements and test methods — Part 2: Needles
- · ISO 7864: Sterile hypodermic needles for single use Requirements and test methods
- · ISO 23908: Sharps injury protection Requirements and test methods
- Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling
## Biocompatibility
- ISO 10993-1:2018 Biological Evaluation of Medical Devices -- Part 1: Evaluation and testing within a risk management process
- ISO 10993-5:2009 Biological Evaluation of Medical Devices -- Part 5: Tests for in Vitro Cytotoxicity
- ISO 10993-10:2010 Biological Evaluation of Medical Devices -- Part 10: Tests for Irritation and Skin Sensitization
- · ISO 10993-11:2006, Biological evaluation of medical devices Part 11: Tests for systemic toxicity-Acute systemic toxicity
- · ISO 10993-11:2017, Biological evaluation of medical devices Part 11: Tests for systemic toxicity-Subacute systemic toxicity
{8}------------------------------------------------
- ASTM F756-13 Standard Practice for Assessment of Hemolytic Properties of Materials
- USP42-NF37<151> pyrogen test
- USP39 NF34<85> Endotoxin test (LAL)
- · Particulate matter testing was conducted in accordance with USP <788> Particulate Matter in Injections and met the USP acceptance criteria.
- Sterility, Shipping, and Shelf-Life
- ISO 11135:2014, Sterilization of health-care products Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices
- Package integrity testing, after environmental conditioning and . simulated transportation in accordance with ISTA-3A 2008, was conducted on the final, packaged, and sterile devices. All packaging deemed acceptable for protection of product and sterility maintenance.
- · Sterile Barrier Packaging Testing performed on the proposed device:
- Seal strength ASTM F88/F88-15
- Dye penetration ASTM F1929-15
- Sterility test USP38-NF33_C71
- Shelf life of 5 years is validated using ASTM F1980-07 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
- Conclusion g
Based on the information provided within this 510(k) submission, the proposed subject device is substantially equivalent to the predicate device and is as safe, as effective and performs as well as the legally marketed predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.