Sterile Hypodermic Needles for Single Use

K210232 · Zhejiang Kangkang Medical-Devices Co., Ltd. · FMI · Sep 23, 2021 · General Hospital

Device Facts

Record IDK210232
Device NameSterile Hypodermic Needles for Single Use
ApplicantZhejiang Kangkang Medical-Devices Co., Ltd.
Product CodeFMI · General Hospital
Decision DateSep 23, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5570
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Sterile Hypodermic Needles for Single Use are intended to be used with a luer lock or luer slip syringe and injection devices for general purpose fluid injection/aspiration.

Device Story

Sterile hypodermic needle for single use; consists of stainless steel needle tube, polypropylene hub, and protective cover. Used in hospitals and clinics by medical professionals or trained caregivers. Connects to luer lock or luer slip syringes for manual fluid injection or aspiration. Device facilitates delivery of medications or withdrawal of fluids. Benefits include sterile, single-use design to prevent cross-contamination. Performance relies on standard needle geometry and ISO-compliant hub connectivity.

Clinical Evidence

Bench testing only. No clinical data provided. Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, sensitization, hemolysis, intracutaneous reactivity, acute systemic toxicity, pyrogenicity). Sterilization validated per ISO 11135. Performance testing conducted per ISO 7864, ISO 9626, and ISO 80369-7.

Technological Characteristics

Materials: Stainless steel (needle tube), Polypropylene (hub, cover). Dimensions: 18G-30G gauges; 1/2"-1 1/2" lengths. Features: Normal/thin wall, short/long bevel. Connectivity: Luer lock/slip per ISO 80369-7. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Standards: ISO 7864, ISO 9626, ISO 10993-1, ISO 11135, ISO 11607, ASTM F1980, ASTM D4169, ASTM F88, ASTM F1140, ASTM F1929, ASTM F2096.

Indications for Use

Indicated for general purpose fluid injection or aspiration in patients when used with a luer lock or luer slip syringe by medical professionals and trained caregivers.

Regulatory Classification

Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. September 23, 2021 Zhejiang Kangkang Medical-Devices CO., Ltd. Chun Guo Ouality Manager Longwang Industrial District, Chumen Town Yuhuan, Zhejiang 317605 China Re: K210232 Trade/Device Name: Sterile Hypodermic Needles for Single Use Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: August 23, 2021 Received: August 27, 2021 Dear Chun Guo: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K210232 Device Name Sterile Hypodermic Needles for Single Use Indications for Use (Describe) The Sterile Hypodermic Needles for Single Use are intended to be used with a luer slip syringe and injection devices for general purpose fluid injection/aspiration. Type of Use (Select one or both, as applicable) | <span style="font-size: 10pt;"></span> <label><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | |--------------------------------------------------------------------------------------------------------------------------------------------| | <span style="font-size: 10pt;"></span> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # K210232 510(k) summary ## l Submitter Device submitter: Zhejiang Kangkang Medical-Devices CO., Ltd. Longwang Industrial District, Chumen Town, Yuhuan, Zhejiang, 317605, China Contact person: Chun Guo General Manager Phone: 13586180908 / 89902728 Fax: 0576-87427630 Email: guochun@vip.126.com Date: 09/23/2021 #### II Device Trade Name of Device: Sterile Hypodermic Needles for Single Use Common Name: Hypodermic Single Lumen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product code: FMI Review Panel: General Hospital #### III Predicate Devices | Trade name: | Self-destruction Safety Syringes for Single Use;<br>Sterile Hypodermic Syringes for Single Use;<br>Sterile Hypodermic Needles for Single Use (used as the<br>predicate device);<br>Sterile Safety Hypodermic Needles for Single Use | |-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Common name: | Hypodermic single lumen needle | | Classification: | Class II, 21 CFR 880.5570 | | Product Code: | FMI | | Premarket Notification: | K180417 | | Manufacturer: | Berpu Medical Technology Co., Ltd | #### IV Device description The Sterile Hypodermic Needles for Single Use is composed of a needle hub, protective cover, needle tube and jointing, connected with a syringe for general purpose fluid injection/aspiration. {4}------------------------------------------------ | Device | Needle length | Needle gauge | Type of wall | Blade angle | |----------------------------------------------------|-----------------------------------|----------------------------------------------------------------|------------------------------|-------------------------------| | Sterile<br>Hypodermic<br>Needles for<br>Single Use | 1/2", 5/8", 1",<br>1 1/4", 1 1/2" | 30G, 27G, 26G,<br>25G, 24G, 23G,<br>22G, 21G, 20G,<br>19G, 18G | Normal wall<br>and thin wall | Short bevel<br>and long bevel | ## V Indications for use The Sterile Hypodermic Needles for Single Use are intended to be used with a luer lock or luer slip syringe and injection devices for general purpose fluid injection/aspiration. ## VI Comparison of technological characteristics with the predicate devices The Sterile Hypodermic Needles for Single Use have the same or equivalent intended use, technology, design and performance specifications to existing legally marketed predicate devices. The differences between the Sterile Hypodermic Needles for Single Use and predicate devices do not alter suitability of the proposed device for its intended use. | Device<br>feature | Subject Device | Predicate Device K180417 | Comments | |---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------| | Indications for use | The Sterile Hypodermic<br>Needles for Single Use are<br>intended to be used with a luer<br>lock or luer slip syringe and<br>injection devices for general<br>purpose fluid<br>injection/aspiration. | The Sterile Hypodermic<br>Needles for<br>Single Use are intended to<br>be used with a luer slip or<br>luer slip syringe and<br>injection devices for general<br>purpose fluid<br>injection/aspiration. | Same | | Product code | FMI | FMI | Same | | Regulation<br>number | 21 CFR 880.5570 | 21 CFR 880.5570 | Same | | Class | II | II | Same | | Principle of<br>operation | For manual use only | For manual use only | Same | | Intended user | Medical professionals and<br>trained care givers | Medical professionals and<br>trained care givers | Same | | Environment<br>of use | Hospitals and clinics | Hospitals and clinics | Same | | Needle gauge | 30G, 27G, 26G, 25G, 24G,<br>23G, 22G, 21G, 20G, 19G,<br>18G | 14G, 15G, 16G, 17G, 18G,<br>19G, 20G, 21G, 22G, 23G,<br>24G, 25G, 26G, 27G, 29G,<br>30G | Difference<br>1 | | Length | $1/2", 5/8", 1", 1 1/4", 1 1/2"$ | 6-60mm | | | Type of wall | normal wall or thin wall | not provided | | | blade angle | short bevel or long bevel | not provided | | | Needle hub | Needle hub<br>Polypropylene | Needle hub<br>Polypropylene | Same | {5}------------------------------------------------ | Device<br>feature | Subject Device | | Predicate Device K180417 | | Comment<br>s | |---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|--------------| | main structure<br>and materials | Needle<br>tube | Stainless steel | Needle<br>tube | Stainless steel | | | | protective<br>cover | Polypropylene | protective<br>cap | Polypropylene | | | Needle hub | Color-coded per ISO 6009 | | Color-coded per ISO 6009 | | Same | | Single use | Yes | | Yes | | Same | | Performance<br>specifications | Complies with ISO 7864; ISO<br>9626; ISO 80369-7 | | Complies with ISO 7864;<br>ISO 9626; ISO 80369-7 | | Same | | Sterilization | EO | | EO | | Same | | SAL | 10-6 | | 10-6 | | Same | | Pyrogen | Non-pyrogenic | | Non-pyrogenic | | Same | | Biocompatibili<br>ty | The biocompatibility evaluation<br>for the subject device was<br>conducted in accordance with<br>the International Standard ISO<br>10993-1 "Biological Evaluation<br>of Medical Devices - Part 1:<br>Evaluation and Testing Within<br>a Risk Management Process,"<br>as recognized by FDA and the<br>"Use of International Standard<br>ISO 10993-1 "Biological<br>evaluation of medical devices-<br>Part 1: Evaluation and testing<br>within a risk management<br>process", June 16, 2016. The<br>syringe of testing included the<br>following tests:<br>Cytotoxicity;<br>Skin sensitization:<br>Hemolysis:<br>Intracutaneous reactivity;<br>Acute systemic toxicity;<br>Pyrogenicity.<br>The evaluation of the above<br>testing items meets the<br>requirements | | Comply with ISO 10993.<br>The test is as follows:<br>The devices meet<br>biocompatibility endpoints<br>for cytotoxicity, irritation,<br>sensitization, systemic<br>toxicity,<br>hemolysis and material-<br>mediated pyrogens. | | Same | | Labeling | Meet the requirements of 21<br>CFR Part 801 | | Meet the requirements of 21<br>CFR Part 801 | | Same | # Difference 1 The needle gauge and length of subject devices are different from the predicate device, and the type of wall and needle bevel are unknown. However, this difference is just in dimension. Different needle specifications will be selected by the HCP per the patient's condition. This difference does not affect intended use. In addition, differences were addressed through ISO 7864, ISO 9626 and ISO 80369-7. Therefore, these differences do not affect safety and effectiveness {6}------------------------------------------------ ## VII Performance data The following performance data were provided in support of the substantial equivalence determination. ## Biocompatibility testing Biocompatibility of the Sterile Hypodermic Needles for Single Use, Sterile Safety Hypodermic Needles for Single Use were evaluated in accordance with ISO 10993-1:2018 for the body contact category of "External communication device – Blood path indirect" with a contact duration of "Limited (< 24 hours)". The following tests were performed, as recommended: USP <788> Particulate Matter Testing ## Sterilization and Shelf-life testing The sterilization method has been validated to ISO11135, which has thereby determined the routine control and monitoring parameters. The shelf life of the Sterile Hypodermic Needles for Single Use, Sterile Safety Hypodermic Needles for Single Use is determined based on stability study which includes ageing test. The testing is performed according to the following standards: - △ ISO 11135:2014 Sterilization of health-care products - Ethylene oxide -Requirements for the development, validation and routine control of a sterilization process for medical devices - A ISO 10993-7:2008 Biological evaluation of medical devices — Part 7: Ethylene oxide sterilization residuals - > ISO 11607-1: 2019 Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems - > ISO 11607-2: 2019 Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes - > ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices - > ASTM D4169-16 Standard Practice for Performance Testing of Shipping Containers and Systems - > ASTM F88/F88M-15 – Standard Test Method for Seal Strength of Flexible Barrier Materials {7}------------------------------------------------ - ASTM F1140/F1140M-13 Standard Test Methods for Internal Pressurized Failure A of Unrestrained Packages - > ASTM F1929-15 – Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by De Penetration - > ASTMF2096-11 – Standard Test Method for Detecting Gross Leakes in Packaging by Internal Pressurization (Bubble Test) ## Performance testing Performance testing is performed according to the following standards: - > ISO 7864:2016 Sterile hypodermic needles for single use — Requirements and test methods - A ISO 9626:2016, Stainless Steel Needle Tubing for the Manufacture of Medical Devices - > ISO 80369-7:2016 Small-bore connectors for liquids and gases in healthcare applications — Part 7: Connectors for intravascular or hypodermic applications ## VIII Conclusion The Sterile Hypodermic Needles for Single Use are substantially equivalent to its predicate device. The non-clinical testing demonstrates that the device is as safe, as effective and performs as well as the legally marketed device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...