K152824 · Shandong Caremed Medical Products Co., Ltd. · FMI · Mar 7, 2016 · General Hospital
Device Facts
Record ID
K152824
Device Name
MEDT PEN NEEDLE
Applicant
Shandong Caremed Medical Products Co., Ltd.
Product Code
FMI · General Hospital
Decision Date
Mar 7, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MEDT Pen Needle is intended for use with pen injector devices for the subcutaneous injection of insulin.
Device Story
MEDT Pen Needle is a single-use, sterile, manual hypodermic needle designed for subcutaneous insulin delivery. It consists of a stainless steel needle, polypropylene hub, and polyethylene inner sheath, packaged with an outer sheath and Tyvek paper. The device is intended for use by patients or clinicians with compatible pen injectors (e.g., NovoPen Echo, HumaPen Luxura). Operation involves attaching the needle to the injector, performing the injection, and disposing of the needle after a single use. The device facilitates insulin administration, aiding in glycemic control for patients with diabetes. It is provided sterile via ethylene oxide (EO) sterilization.
Clinical Evidence
Bench testing only. No clinical data. Testing included physical, mechanical, and chemical evaluations per ISO 7864, ISO 9626, and ISO 11608-2. Biocompatibility testing (cytotoxicity, intracutaneous reactivity, skin sensitization, acute systemic toxicity, hemolysis, and pyrogenicity) followed ISO 10993 and ASTM F756-13. Sterility and shelf-life were validated per USP and ISO 10993-7 standards.
Technological Characteristics
Materials: Stainless steel (needle), polypropylene (hub/outer sheath), polyethylene (inner sheath). Energy: Manual. Dimensions: 29G-32G gauges; lengths 4mm-12mm. Sterilization: Ethylene oxide (SAL 10^-6). Standards: ISO 7864, ISO 9626, ISO 11608-2, ISO 10993, ASTM F88/F88M-09, ASTM F756-13.
Indications for Use
Indicated for subcutaneous injection of insulin using pen injector devices in patients requiring insulin therapy.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
K120690 — INSULIN PEN NEEDLE · Tiger Medical Products, Ltd. · Jun 22, 2012
K192464 — Injection Pen Needle · Gangan Medical Technology Jiangsu Co., Ltd. · Aug 25, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring a stylized human figure in profile, with three faces overlapping to represent the department's focus on health and human well-being. The emblem is simple and modern in design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 7, 2016
Shandong Caremed Medical Products Co., Ltd. % Ms. Diana Hong General Manager Mid-link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 CHINA
Re: K152824
Trade/Device Name: MEDT Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: January 28, 2016 Received: February 3, 2016
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -
for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K152824
Device Name MEDT Pen Needle
Indications for Use (Describe)
The MEDT Pen Needle is intended for use with pen injector devices for the subcutaneous injection of insulin.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92. The assigned 510(k) Number: K152824
- 1. Date of Preparation: 03/03/2016
- Sponsor Identification 2.
## Shandong Caremed Medical Products Co., Ltd.
(Yantai) Industrial Park, DingTao County (The East of Zhanqian Road), Shandong, 274100, China
Contact Person: Lin Yanyan Position: Management representative Tel: 86-530-7397700 Fax: 86-530-7397711 Email: lyyan@shanghaicarelife.com
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Mr. Lee Fu (Alternative Contact Person)
## Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 240-238-7587 Email: info@mid-link.net
- 4. Identification of Proposed Device
Trade Name: MEDT Pen Needle Common Name: Insulin Pen Needle
Regulatory Information Classification Name: Needle, Hypodermic, Single Lumen; Classification: II; Product Code: FMI; Regulation Number: 21 CFR 880.5570;
{4}------------------------------------------------
Review Panel: General Hospital;
Intended Use Statement:
The MEDT Pen Needle is intended for use with pen injector devices for the subcutaneous injection of insulin.
Device Description
The proposed device, MEDT Pen Needle, is a single-use device, which is designed for used with a pen injector for the subcutaneous injection of insulin. It consist of three components, which are (1) needle, (2) hub and (3) inner sheath; it is sealed in a package consists of (1) a outer sheath and (2) Tyvek Paper.
- ડ. Identification of Predicate Device(s)
510(k) Number: K133059 Product Name: Insulin Pen Needle Manufacturer: Wenzhou Beipu Science & Technology Co., Ltd
- 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
## Physical, Mechanical and Chemical Tests performed on the proposed device
- > ISO 7864: 1993 Sterile hypodermic needles for single use
- > ISO 9626:1991 AMD 2001 Stainless steel needle tubing for the manufacture of medical devices
- A ISO 11608-2:2012 Needle-based injection systems for medical use- Requirements and test methods-Part 2: Needles
### Test Item
- Materials Surface finish Cleanliness Limits for acidity and alkalinity Size designation Dimensions Stiffness Resistance to breakage Resistance to corrosion
{5}------------------------------------------------
Cleanliness Limits for acidity or alkalinity Limits for extractable metals Size designation Colour coding Needle hub Sheath Needle tube Needle point Performance Materials Dimensions Determination of flow rate through the needle Bond between hub and needle tube Needle points Freedom form defects Lubrication Dislocation of measuring point at patient end Determination of functional compatibility with needle-based injection systems Ease of assembly and disassembly Sterility
Sterile Barrier Packaging Testing performed on the proposed device:
Seal Strength: ASTM F88/F88M-09 Standard test method for seal strength of flexible barrier materials
Sterilization and Shelf Life Testing performed on the proposed device:
| EO residue | ISO 10993-7:2008 Biological evaluation of medical devices- Part<br>7: Ethylene oxide sterilization residuals |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| ECH residue | ISO 10993-7:2008 Biological evaluation of medical devices- Part<br>7: Ethylene oxide sterilization residuals |
| Bacteria Endotoxin Limit | USP 37-NF 32 <85> Bacterial Endotoxins Limits |
| Shelf Life Evaluation | Physical, Mechanical, Chemical, Package were performed on<br>accelerated aging samples to verify the claimed shelf life of the<br>device |
| | Sterility Test was performed per USP 38-NF 33 <71> Sterility<br>Tests |
Biocompatibility test on the proposed device:
{6}------------------------------------------------
| Cytotoxicity test | ISO 10993-5:2009 Biological evaluation of medical devices- Part 5: Tests for in vitro cytotoxicity |
|--------------------------------|------------------------------------------------------------------------------------------------------------------|
| Intracutaneous Reactivity test | ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Tests for irritation and skin sensitization |
| Skin Sensitization Test | ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Tests for irritation and skin sensitization |
| Acute Systemic Toxicity Test | ISO 10993-11:2006 Biological evaluation of medical devices- Part 11: Tests for systemic toxicity |
| Hemolysis Test | ASTM F 756-13 Standard practice for assessment of hemolytic properties of materials |
| Pyrogen Test | USP 38-NF 33 <151> Pyrogen test |
- 7. Substantially Equivalent (SE) Comparison
| Item | Proposed Device<br>K152824 | Predicate Device<br>K133059 |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | FMI | Same |
| Regulation Number | 880.5570 | Same |
| Intended Use | The MEDT Pen Needle is intended for<br>use with pen injector devices for the<br>subcutaneous injection of insulin. | Same |
| Configuration and<br>material | Needle tube Stainless Steel<br>hub Polypropylene<br>Inner sheath Polyethylene<br>Outer sheath Polypropylene<br>Tyvek paper Paper | Needle tube Stainless Steel<br>hub Polypropylene<br>Tube sheath Polypropylene<br>Hub sheath Polypropylene<br>Sealed paper Paper |
| Operation mode | Manual | Same |
| Needle Gauge | 29G, 30G, 31G, 32G | Same |
| Needle Dimension<br>(mm) | $0.23 \times 4, 0.23 \times 5, 0.23 \times 6,$<br>$0.25 \times 4, 0.25 \times 5, 0.25 \times 6, 0.25 \times 8,$<br>$0.30 \times 5, 0.30 \times 6, 0.30 \times 8,$<br>$0.33 \times 10, 0.33 \times 12$ | $0.23 \times 4,$<br>$0.25 \times 4, 0.25 \times 5, 0.25 \times 6, 0.25 \times 8,$<br>$0.30 \times 8, 0.30 \times 10,$<br>$0.33 \times 12$ |
| Performance | Comply with ISO 7864, ISO 9626, and<br>ISO 11608-2 | Same |
| Compatible pen<br>injectors | NovoPen Echo<br>HumaPen Luxura | Same |
| Sterile | EO sterilized, SAL: 10-6 | Same |
| Single Use | Single use | Same |
| Endotoxin Limit | 20 EU per product | Same |
- Table 1 Comparison of Technology Characteristics
{7}------------------------------------------------
| Labeling | Conform with 21 CFR 801 | Same |
|------------------------------|------------------------------------------|------|
| Biocompatibility | Conform with ISO 10993 standards | Same |
| Cytotoxicity | No cytotoxicity | |
| Intracutaneous<br>Reactivity | No Intracutaneous reactivity | |
| Skin Sensitization | No significant evidence of sensitization | |
| Acute Systemic<br>Toxicity | No systemic toxicity | |
| Hemolysis | No evidence of hemolysis | |
| Pyrogen | No pyrogen | |
## Comparison discussion:
The subject device has the same indications for use and the same principle of operation as the predicate device. Both subject and predicate devices are provided as sterile, nonpyrogenic and individually packaged.
The differences between the subject and the predicate devices include:
- . The material used for Inner sheath of the subject device is Polyethylene. The material used for Tube sheath of the predicate device is Polypropylene.
- . The subject and predicate devices have same needle gauges but different needle lengths. The subject device contains twelve (12) needle sizes. The predicate device contains eight (8) needle sizes.
The performance testing of the subject device met all acceptance criterial. These differences do not have impacts to product technology and performance to the subject device, and do not raise new issues of product performance.
#### 8. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.