K080904 · Feel Tech · FMI · Jun 4, 2008 · General Hospital
Device Facts
Record ID
K080904
Device Name
FEEL-FINE INSULIN PEN NEEDLE
Applicant
Feel Tech
Product Code
FMI · General Hospital
Decision Date
Jun 4, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
These disposable sterile insulin pen needles are intended for subcutaneous injection of insulin in the treatment of diabetes.
Device Story
Feel Fine Insulin Pen Needle is a sterile, single-use, disposable device designed for subcutaneous insulin delivery. It consists of a needle hub, needle, needle cap, and sterile cap with blister paper. The device attaches to an insulin pen to facilitate injection. It operates on the principle of a common piston syringe. The device is intended for use by patients or healthcare providers for diabetes management. It provides a sterile fluid path and is non-toxic and non-pyrogenic.
Clinical Evidence
Bench testing only. Testing included biocompatibility, mechanical performance, and sterility validation (including EO residues). No clinical data was required or provided.
Technological Characteristics
Sterile, single-use, disposable hypodermic needle. Components include needle hub, needle, needle cap, and sterile cap. Sterilized via ethylene oxide (EO). Operates as a common piston syringe. No software or electronic components.
Indications for Use
Indicated for subcutaneous insulin injection in patients with diabetes.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
page 1 of 1
JUN - 4 2008
510(K) Summary for K080904 Feel Tech # 92-4 Haengjung-ri, Gwangdcok-myeon, Cheonan-si, Choong-nam Korea Tel. : 82-41-522-2446-8 Fax .: 82-41-522-2449 Contact person: Boo Sool Kim Date prepared: June 2, 2008
Trade Name: Feel Fine Insulin Pen Ncedle 1.
- Common Name: Pen needle 2.
- Classification Name: Syringe, piston, product code FMI, Regulation: 880.5570 3.
- 4. Class of device: Class II.
- The legally marketed device to which we are claiming equivalence [807.92(a)(3)) : 5. B.Braun "one.clickTM needle", K033575.
- Description of device: Feel Fine Insulin Pen Needle consists of a sterile cap, needle cap 6. needle hub, which can be fixed with needle and blister paper. The sterile cap functions to maintain the sterility of the needle because sterile cap covers the needle hub and needle cap with blister paper sealed on the opening hole of sterile cap. The needle hub can be connected with pen. The necdle cap covers intended to provide physical protection to the needle tube. They are supplied with a sterile fluid path, (EO), nontoxic, and non pyrogenic, for single use only, disposable. The devices operate on the principles of common piston syringes.
- Intended use: These disposable sterilc insulin pen needles are intended for 7. subcutaneous injection of insulin in the treatment of diabetes.
- Technological characteristics: Feel Fine Insulin Pen Needles and the predicate devices 8. have identical technological characteristics and perform the same way as common piston syringes. These needles are EO sterilized.
- Performance: Bench tests were performed. Bench testing included biocompatibility, 9. mechanical testing, sterility testing including EO residues. The tests demonstrated that the device is as safe, as effective, and performs in a substantially equivalent manner to the predicate device.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circle around the eagle. The eagle is black, and the text is also black.
Food and Drug Administration 9200 Corporate Bouievard Rockville MD 20850
JUN - 4 2008
Feel Tech C/O Mr. Daniel Kamm Principal Consultant Kamm & Associates P.O. Box 7007 Deerfield, Illinois 60015
Re: K080904
Trade/Device Name: Feel Fine Insulin Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: March 28, 2008 Received: April 1, 2008
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register,
{2}------------------------------------------------
Page 2 - Mr. Kamm
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Clare
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
K080904
## Indications for Use
510(k) Number (if known): K080904
Feel Fine Insulin Pen Needle Device Name:
Indications For Use:
These disposable sterile insulin pen needles are intended for subcutaneous injection of insulin in the treatment of diabetes.
V Prescription Us (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anthony O. Natan
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of 1
510(k) Number: K4849994
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.