MANUAL PD-IO; POWERED PD-IO

K063142 · Vidacare Corporation · FMI · Dec 8, 2006 · General Hospital

Device Facts

Record IDK063142
Device NameMANUAL PD-IO; POWERED PD-IO
ApplicantVidacare Corporation
Product CodeFMI · General Hospital
Decision DateDec 8, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5570
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The Manual PD-IO, and Powered PD-IO provide intraosseous access in the distal tibia of pediatric patients as an alternative to IV access during emergencies.

Device Story

Manual and powered intraosseous (IO) infusion systems provide vascular access in emergency settings. Manual driver uses ergonomic handle to manually insert needle set (stylet and catheter) into bone cortex; powered driver uses battery-operated drill for insertion. Both systems utilize identical needle sets (15G x 15mm). Upon reaching marrow, driver is detached; stylet is removed by counter-clockwise rotation, leaving Luer lock catheter seated in bone. Catheter allows connection of standard syringes/tubing for fluid/drug administration. Used in clinical/emergency settings by healthcare providers. Provides rapid alternative to intravenous access when IV is difficult or impossible to establish in pediatric patients.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on identical technological characteristics and design to previously cleared predicate devices.

Technological Characteristics

Manual driver: pentagon shaft with handle. Powered driver: battery-powered drill. Needle set: 15G x 15mm stylet and catheter. Luer lock interface. Sterile, single-use needle assembly. Biocompatible materials consistent with cleared predicates.

Indications for Use

Indicated for pediatric patients requiring intraosseous access in the distal tibia as an alternative to IV access during emergencies.

Regulatory Classification

Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 5. 510(k) Summary DEC - 8 2006 vidacare 722 Isom Road. San Antonio, TX 78216 Tel (210) 375 8500 Fax (210) 375 8537 Toll Free (within US) 866 479 8500 www.vidacare.com ## SUMMARY Submitter's name: Address: Vidacare Corporation 722 Isom Road San Antonio, TX 78216 210-375-8500 210-375-8537 Name of contact person: Phone: Fax number: Grace Holland Regulatory Specialists, Inc 3722 Ave. Sausalito Irvine, CA 92606 Phone: 949-262-0411 Fax: 949-552-2821 Date the summary was prepared: October 10, 2006 Name of the device: Trade or proprietary name: Common or usual name: Classification name: Manual PD-IO and Powered PD-IO Manual PD-IO and Powered PD-IO Intraosseous Infusion System Hypodermic single lumen needle The legally marketed devices to which we are claiming equivalence [807.92(a)(3)]: | | 510(k)<br>Number | | Trade or Proprietary or<br>Model Name | | Manufacturer | |---|------------------|---|--------------------------------------------|--|----------------| | 1 | K043490 | 1 | Manual PD-IO | | Vidacare Corp. | | 2 | K051992 | 2 | Powered PD-IO | | Vidacare Corp. | | 3 | K913258 | 3 | Disposable Introsseous<br>Infusion Needles | | Cook, Inc. | {1}------------------------------------------------ Description of the device: The following descriptions, for the manual and powered drivers include two different drivers. The same needle set is used for both drivers. The Manual PD-IQ, previously cleared under 510(k) K043490, consists of a proprietary pentagon shaft permanently attached to an ergonomically designed handle. The manual driver is designed to allow the user to manually insert a needle set consisting of a stylet and catheter through the cortex of the bone to a desired depth within the bone marrow to facilitate the infusion of desired fluids. After insertion of the needle set into the bone, the manual driver is detached from the needle set, leaving the stylet and catheter firmly seated in the bone. The stylet is then separated and removed from the catheter by turning the stylet hub counter clockwise leaving a standard Luer lock catheter securely seated in the bone. The catheter Luer lock permits attachment of standard syringes and IV tubing for administration of drugs and fluids (not supplied). The Manual PD-IO system is approved for use in the proximal tibia under 510(k) K043490. The size needle that can be used in the distal tibia utilizing the Manual PD-IO is identical to the predicate, 15G X 15mm. This submission extends the indication for use of the manual device to include the distal tibia in pediatrics utilizing the same technique and device previously cleared for the proximal tibia via 510(k) K043490. The Powered PD-IO, previously cleared under 510(k) K051992 for pediatric use, consists of a reusable battery powered driver connected to a single use disposable intraosseous (IO) needle assembly. Upon activation, the drill supplies power to the needle set in order to penetrate through the cortex of the bone to a desired depth within the bone marrow. After insertion of the needle set, the power driver is detached from the needle set, leaving the stylet and catheter firmly seated in the bone. The stylet is then separated and removed from the catheter by turning the stylet hub counter clockwise leaving a standard Luer lock catheter securely seated in the bone. The catheter Luer lock permits attachment of standard syringes and IV tubing for administration of drugs and fluids (not supplied). The Powered PD-IO intraosseous system is cleared for use in the proximal tibia under 510(k) K051992. The size needle that can be used with the Powered PD-IO is identical to the predicate 15G X {2}------------------------------------------------ 15mm. This submission extends the indication for use of the powered device to include the distal tibia in pediatric patients utilizing the same technique and device previously cleared for the proximal tibia via 510(k) K051992. ### Indications: The Manual PD-IO and Powered PD-IO provide intraosseous access in the distal tibia of pediatric patients as an alternative to IV access during emergencies. Summary of the technological characteristics of our device compared to the predicate devices: The Manual PD-IO for Distal Tibia access has the exact same technology as the Manual PD-IO (K043490). The Powered PD-IO for Distal Tibia access has the exact same technology for use as the Powered PD-IO (K051992). Vidacare also wishes to use as a predicate Disposable Intraosseous Infusion Needles, K913258, by Cook, Inc., as these needles have the exact same indications for use for which we are applying. These needles are used as an alternative to intravenous access during pediatric emergencies. The distal tibia is one of the access sites in their instructions for use. This submission extends the indication for use to include the distal tibia site in pediatrics of the Manual PD-IO and the Powered PD-IO intraosseous systems. There have been no changes to the design or components of the Manual PD-IO and the Powered PD-IO intraosseous systems cleared under 510(k) K043490 and K051992 and therefore the comparison of technological characteristics listed below are identical. Target Population Driver Design Features Needle Design Technique Sterility Biocompatibility Where Used {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized eagle with three stripes representing the three branches of government. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC - 8 2006 Vidacare Corporarion C/O Ms. Grace Holland Regulatory Specialists, Incorporated 3722 Avenue Sausalito Irvine, California 92606 Re: K063142 Trade/Device Name: Manual PD-IO and Powered PD-IO Regulation Number: 21 CFR 880.5570 Regulation Name: Single Lumen Hypodermic Needle Regulatory Class: II Product Code: FMI Dated: October 10, 2006 Received: October 16, 2006 Dear Ms. Holland: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ #### Page 2 - Ms. Holland Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing prestice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Chiu Lin, Ph.D. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # 4. Indications for Use Statement Indications for Use 510(k) Number (if known): _ Device Name: Manual PD-IO and Powered PD-IO The Manual PD-IO, and Powered PD-IO provide intraosseous access in the distal tibia of pediatric patients as an alternative to IV access during emergencies. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) ーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーー Concurrence of CDRH, Office of Device Evaluation (ODE) Jdh All for ABW 12/8/04 shopiclogy, General Hospith .
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...