K080077 · Icu Medical, Inc. · FMG · Feb 13, 2008 · General Hospital
Device Facts
Record ID
K080077
Device Name
CLAVE STOPCOCK
Applicant
Icu Medical, Inc.
Product Code
FMG · General Hospital
Decision Date
Feb 13, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The ICU Medical Clave Stopcock is a device used to administer or withdraw fluids that flow from a container to a patient's vascular system through a needle or catheter inserted into a vein or artery. This device is a four-way stopcock with an integral CLAVE side-port for closed, needleless access to the patient tubing or other fluid container.
Device Story
The Clave Stopcock combines a 4-way flow valve stopcock with an integrated needleless connector (Clave). It facilitates fluid administration or withdrawal from containers to a patient's vascular system. The device features three fluid path positions: Flush, Flow, and Port Access. The integrated Clave side-port allows for closed, needleless access to patient tubing. It is intended for use by clinicians in clinical settings. The device simplifies access by merging two established technologies into a single unit, maintaining the same operational functionality as the predicate devices.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on design, material, and operational similarity to predicate devices. Bench testing and adherence to international standards and cGMPs are cited to support safety and performance.
Technological Characteristics
Materials: Polycarbonate (housing, plug, spike), HD Polyethylene (handle), Valox (Clave body), Silicone (Clave plug). 4-way flow valve with integrated needleless port. Sterilization: EtO or E-Beam. No software or electronic components.
Indications for Use
Indicated for patients requiring administration or withdrawal of fluids via vascular access (vein or artery) using a needle or catheter. No specific age or gender restrictions provided.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
K222889 — FlowArt® Needle-Free Split Septum Valve · Asset Medikal Tasarim San Tic As. · Jan 20, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
FEB 1 3 2008
ICU MEDICAL INC. 4455 Atherton Drive Salt Lake City, Utah (801) 264 - 1332, Phone (801) 264 - 1755, Fax Tracy S. Best, Sr. Regulatory Affairs Specialist Preparation Date: January 09, 2008
# SPECIAL 510(K) Summary of Safety and Effectiveness for the:
| Trade Name: | Clave® Stopcock |
|----------------------|----------------------------------------------------------------------|
| Common Name: | Accessory, Set, Administration, I.V. |
| Classification Name: | Set, Administration, Intravascular, 21 CFR 880.5440, Class II Device |
| Product Code: | Primary Code: FMG, Secondary Code: LHI |
# Legally Marketed Predicate Devices for Substantial Equivalence:
* K 941190 - Clave Connector - ICU Medical, Inc. * K970855 - Clave Connector - ICU Medical, Inc. *K953573 - ICU Stopcock with Extension Tubing -- ICU Medical, Inc.
# Rationale for SE:
This submission is a combination of two devices that have been in the field for more than 15 years. These two devices are regarded highly within the industry and valued by our customers. This device will integrate the spike portion of the Clave into the molded portion of the Stopcock or the spike portion will function separately as it does now and be used as the base plug for the Stopcock portion. Then the other portions of the device remain the same as does the indications for use. There are no functional differences between any of the predicate devices or the proposed devices in terms of use.
# Description of Submitted Device:
The Clave® Stopcock is a device that incorporates the functionality of needleless connectivity with the 4-way flow valve of a stopcock. The fluid path has three positions: Flush Position; Flow Position; and Port Access Position. The proprietary technology of the Clave enables the device to have a needleless port for administration of medications or other IV fluids as directed by the physician.
# Intended Use:
The ICU Medical Clave Stopcock is a device used to administer or withdraw fluids that flow from a container to a patient's vascular system through a needle or catheter inserted into a vein or artery. This device is a four-way stopcock with an integral CLAVE side-port for closed, needleless access to the patient tubing or other fluid container.
{1}------------------------------------------------
| Component: | ICU Medical, Inc.<br>Clave Stopcock | ICU Medical, Inc.<br>Clave | ICU Medical, Inc.<br>Stopcock with ext. Tubing |
|-----------------------|-------------------------------------|----------------------------|------------------------------------------------|
| Stopcock Housing: | Polycarbonate | | Polycarbonate |
| Stopcock Plug: | Polycarbonate | | Polycarbonate |
| Stopcock Handle: | HD Polyethylene | | Polyethylene & HDPE |
| Clave Body: | Valox | Valox | N/A |
| Clave Plug: | Silicone | Silicone | N/A |
| Clave Spike | Polycarbonate | Polycarbonate | N/A |
| Package: | Peel Blister pak | Peel Blister Pak | Peel Blister Pak |
| Sterilization method: | EtO or E-Beam | E-Beam | EtO or E-Beam |
| 510(k) Approval | This submission | K941190 or K970855 | K953573 |
# Technological Characteristics and Substantial Equivalence Table:
The operational characteristics are identical as they have only changed to morph two devices together. Operation is the same for all Clave or Stopcocks, including the predicate devices .
### Safety and Performance:
ICU Medical Clave Stopcock conform to the requirements of published international standards as well as those FDA recognized standards and/or published guidelines prior to marketing the device, Additionally, ICU Medical's Sterility Assurance Level, (SAL) has an established and validated history of meeting the 10° level. These devices will be packaged in a way as to ensure conformity with that SAL level. The manufacturing of these devices will be assembled in a quality environmy that is certified independently and complies with cGMPs.
# Conclusion:
The materials, performance, and operational features of both the submitted device and the predicate devices are substantially equivalent and safe and effective for their intended use.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three curved lines representing its wings. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the left side of the logo.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 3 2008
Mr. Tracy S. Best Senior Regulatory Affairs Specialist ICU Medical, Incorporated 4455 Atherton Drive Salt Lake City, Utah 84123
Re: K080077
Trade/Device Name: ICU Medical Clave® Stopcock Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FMG, LHI Dated: January 9, 2008 Received: January 15, 2008
Dear Mr. Best:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Best
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other I ederal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Clu S
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
ררמנגסט X 1 of )
# Indications for Use
# 510(k) Number (if known):
Device Name:
ICU Medical Clave® Stopcock
#### Indications for Use:
The ICU Medical Clave Stopcock is a device used to administer or withdraw fluids that flow from a container to a patient's vascular system through a needle or catheter inserted into a vein or artery. This device is a four-way stopcock with an integral CLAVE side-port for closed, needleless access to the patient tubing or other fluid container.
Prescription Use XX (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
China man
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: ____ ├ ψ ふゆゅ2つ
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.