SPECTRA-LOR, 10 ML PLASTIC LOSS OF RESISTANCE SYRINGE, L/S, SPECTRA-LOR, 10ML PLASTIC LOSS OF RESISTANCE SYRINGE L/L
K081524 · Spectra Medical Devices, Inc. · FMF · Jul 29, 2008 · General Hospital
Device Facts
Record ID
K081524
Device Name
SPECTRA-LOR, 10 ML PLASTIC LOSS OF RESISTANCE SYRINGE, L/S, SPECTRA-LOR, 10ML PLASTIC LOSS OF RESISTANCE SYRINGE L/L
Applicant
Spectra Medical Devices, Inc.
Product Code
FMF · General Hospital
Decision Date
Jul 29, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Indications for Use
The SPECTRA-LOR, Loss of Resistance Syringe is intended for use, in conjunction with an epidural needle, to verify the needle tip placement is in the epidural space by use of the Loss of Resistance Technique as detailed in medical textbooks and medical journal articles. The LOR Syringe is not intended for injection or aspiration.
Device Story
SPECTRA-LOR is a manual piston syringe used by clinicians during epidural procedures. Device consists of barrel, plunger, and bung; available in 10 ml capacity with Luer Lock or Luer Slip tip. Operates via Loss of Resistance (LOR) technique: clinician advances epidural needle while applying pressure to syringe plunger; sudden decrease in resistance indicates needle tip has entered epidural space. Device provides tactile feedback to clinician to confirm correct needle placement. Not intended for injection or aspiration. Benefits include assisting in accurate epidural space identification.
Clinical Evidence
Bench testing only.
Technological Characteristics
Piston syringe design; 10 ml capacity; Luer Lock or Luer Slip tip. Materials of construction are equivalent to predicate. Manual operation; no energy source. Non-sterile/sterile status not specified, but standard syringe construction.
Indications for Use
Indicated for use in conjunction with an epidural needle to verify needle tip placement in the epidural space via the Loss of Resistance technique. Not for injection or aspiration. Prescription use only.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
K082674 — SPECTRA-GLASS LOR (LOSS OF RESISTANCE SYRINGE, 5CC METAL L/S OR L/L TIP, SPECTRA-GLASS LOR (LOSS OF RESISTANCE) SYRINGE · Spectra Medical Devices, Inc. · Nov 28, 2008
K122416 — ARTIGLASS L. O.R. GLASS SYRINGE · Artiglass Srl · May 3, 2013
{0}------------------------------------------------
# 510(K) SUMMARY KOX (524
## JUL 2 9 2008
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Devices Act of 1990, 21 CFR 807.92.
| Classification Name: | Syringe, Piston |
|----------------------|-----------------------------------------|
| Common Name: | Piston Syringe |
| Proprietary Name: | SPECTRA-LOR, Loss of Resistance Syringe |
| Classification: | Class II |
| Product Code: | 80FMF |
| Regulation Number: | 880.5860 |
### Establishment Name, Contact & Registration Number:
Name:
Spectra Medical Devices, Inc. 260-F Fordham Road Wilmington, MA 01887 Tele: 978-657-0889, Ext. Fax: 978-657-4339
Contact:
Scott Henderson
### Intended Use:
The SPECTRA-LOR, Loss of Resistance Syringe is intended for use, in conjunction with an epidural needle, to verify the needle tip placement is in the epidural space by use of the Loss of Resistance Technique as detailed in medical textbooks and medical journal articles. The LOR Syringe is not intended for injection or aspiration.
Predicate Device: AVID-NIT Loss of Resistance Syringe, FDA 510(K) # K001731
### Device Description:
The construction of the SPECTRA-LOR, Loss of Resistance Syringe is similar to that of a standard piston syringe; it has a barrel, a plunger and a bung. The barrel has graduations which correspond to a capacity of 10 ml and is available in a Luer Lock or Luer Slip tip.
{1}------------------------------------------------
# 510(K) SUMMARY K
## Equivalent Predicate Device:
The SPECTRA-LOR, Loss of Resistance Syringe is equivalent to the predicate device, AVID-NIT Loss of Resistance Syringe in materials of construction, intended use, capacity and bench testing.
## SECTION 5, PAGE 4
: ・ ・
ﺮ ۱۰ - ۱۰ -
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The logo is presented in black and white, giving it a formal and official appearance.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Scott Henderson International Engineering Manager Spectra Medical Devices, Incorporated 4501 Greendale Drive Williamsburg, Virginia 23188
JUL 2 9 2008
Re: K081524
Trade/Device Name: SPECTRA-LOR Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF Dated: May 30, 2008 Received: June 2, 2008
Dear Mr. Henderson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2 - Mr. Henderson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
signature
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## INDICATIONS FOR USE
## 510(k) Number (if known): K081524
Device Name: SPECTRA - LOR
Indications for Use:
The "SPECTRA - LOR", Loss of Resistance Syringe is intended for use, in conjunction with an epidural needle, to verify the needle tip placement in the epidural space by the Loss of Resistance technique as explained in standard medical textbooks. These syringes are not intended for injection or aspiration.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)
Charles Zink
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: k081524
#### SECTION 3, Page 1
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.