← Product Code [CAH](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/CAH) · K202459

# Bact-Trap Filter, Bact-Trap Mini, Bact-Trap Midi, Pharma Mini HME/Filter, Bact-HME HME/Filter and Bact-HME Midi HME/Filter (K202459)

_Pharma System AB · CAH · Sep 26, 2021 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94therapeutic-devices/CAH/K202459

## Device Facts

- **Applicant:** Pharma System AB
- **Product Code:** [CAH](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/CAH.md)
- **Decision Date:** Sep 26, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5260
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

Bact Trap filter is a breathing system filter which is designed to reduce possible airborne or liquid-borne cross contamination with micro-organisms and particulate matter via anaesthetic or ventilator breathing systems. The Bact Trap filter may either be used on the patient side of the ventilator anaesthetic device. The Bact HME/Bact-HME Midi/Pharma Mini is a breathing system filter and a Heat and Moisture Exchanger. The combination of a filter and a Heat and Moisture offer the benefit of both product features. Heat and Moisture Exchangers are used as a conditioning system for mechanically ventilated patients whose uppassed. In almost all cases of mechanical ventilation they are a fully valid alternative to heated humidifiers. The Bact-HME Midi/Pharma Mini should be used with patients who have a Tidal Volume between 50 - 1500 ml. The Pharma Mini™HME/Filter, Bact-HME™ filter/HME and Bact-HME™ Midi HME/Filter should be used with patients who have a Tidal Volume as follows: - Pharma Mini™ and Bact-Trap™ Mini 50-900ml - · Bact-HME™ Midi and Bact-Trap™ Midi between 100-1200 ml - · Bact-HMETM and Bact-Trap™ between 250-1500 ml The products mentioned above are designed as disposable single patient use and should be changed at least every 24 hours

## Device Story

Disposable breathing circuit filters and heat/moisture exchangers (HME/HMEF) for use in hospital, sub-acute, and pre-hospital settings. Devices placed at patient end or ventilator/anesthesia machine ports. Filtration via electrostatic charge in polypropylene media; passive humidification via treated paper/foam media capturing exhaled heat/moisture for return during inhalation. Protects patients and equipment from cross-contamination; conditions breathing gases for intubated/tracheostomized patients. Healthcare providers select based on tidal volume requirements (50-1500 ml). Single-patient use; replaced every 24 hours.

## Clinical Evidence

Bench testing only. Performance validated via ISO 9360-1 (gas leakage, moisture loss), ISO 10993 (biocompatibility), ISO 18562 (VOC/particulate matter), ASTM F2101 (filtration efficiency), and EN ISO 23328 (pressure drop). All tests met acceptance criteria.

## Technological Characteristics

Materials: SBC, polypropylene, acrylic/polypropylene fibers, polyethylene, cross-linkable acrylic polymer, silicone antifoam, polyester (PET). Principle: Electrostatic filtration and passive heat/moisture exchange. Connectivity: None (mechanical). Sterilization: Non-sterile. Form factor: Conical 15mm/22mm fittings, optional luer lock ports, straight or angled housings.

## Regulatory Identification

A breathing circuit bacterial filter is a device that is intended to remove microbiological and particulate matter from the gases in the breathing circuit.

## Predicate Devices

- Pharma Systems Bact Trap filters ([K903056](/device/K903056.md))
- Pharma Systems Bact HME ([K903058](/device/K903058.md))

## Reference Devices

- ARC Medical, Inc FilterFlo Filter, ThermoFlo Filter/HME ([K090738](/device/K090738.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 26, 2021

Pharma System AB % Scott Blood Director of Regulatory Affairs MEDIcept Inc. 200 Homer Ave Ashland, Massachusetts 01721

Re: K202459

Trade/Device Name: Bact-Trap Filter, Bact-Trap Mini, Bact-Trap Midi, Pharma Mini HME/Filter, Bact-HME HME/Filter and Bact-HME Midi HME/Filter Regulation Number: 21 CFR 868.5260 Regulation Name: Breathing Circuit Bacterial Filter Regulatory Class: Class II Product Code: CAH Dated: August 19, 2021 Received: August 23, 2021

Dear Scott Blood:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K202459

## Device Name

Bact-Trap™ Filter, Bact-Trap™ Mini Filter, Bact-Trap™ Midi Filter, Pharma Mini™ HME/Filter , Bact-HME™ HME/Filter and Bact-HME™ Midi HME/Filter

## Indications for Use (Describe)

Bact Trap filter is a breathing system filter which is designed to reduce possible airborne or liquid-borne cross contamination with micro-organisms and particulate matter via anaesthetic or ventilator breathing systems. The Bact Trap filter may either be used on the patient side of the ventilator anaesthetic device.

The Bact HME/Bact-HME Midi/Pharma Mini is a breathing system filter and a Heat and Moisture Exchanger. The combination of a filter and a Heat and Moisture offer the benefit of both product features. Heat and Moisture Exchangers are used as a conditioning system for mechanically ventilated patients whose uppassed. In almost all cases of mechanical ventilation they are a fully valid alternative to heated humidifiers. The Bact-HME Midi/Pharma Mini should be used with patients who have a Tidal Volume between 50 - 1500 ml.

The Pharma Mini™HME/Filter, Bact-HME™ filter/HME and Bact-HME™ Midi HME/Filter should be used with patients who have a Tidal Volume as follows:

- Pharma Mini™ and Bact-Trap™ Mini 50-900ml
- · Bact-HME™ Midi and Bact-Trap™ Midi between 100-1200 ml
- · Bact-HMETM and Bact-Trap™ between 250-1500 ml

The products mentioned above are designed as disposable single patient use and should be changed at least every 24 hours

| Type of Use (Select one or both, as applicable)                                                           |
|-----------------------------------------------------------------------------------------------------------|
| <div> <span> <b>      ×     </b> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> |
| <div> <span>□ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div>                                   |

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for MEDICEPT. The word "MEDI" is in bold, dark blue letters, while "CEPT" is in a purple, cursive font. Below the logo is the text "Trusted Solutions, Rapid Response," with "Trusted Solutions" in purple and "Rapid Response" in gray.

### 1. 510(K) SUMMARY

In accordance with 21 CFR 807.87(h) and 21 CFR 807.92, the content of this 510(k) summary is provided for the Pharma Systems Bact-Trap™ Filter, Bact-Trap™Mini Filter, Bact-Trap™ Midi Filter, Pharma Mini™ HME/Filter, Bact-HME™ HME/Filter, and Bact-HME™ Midi HME/Filter.

#### 1.1. Submitter's Information

Pharma Systems AB Name: Address: Rubanksgatan 9, 741 71 Knivsta, SWEDEN

+46 18 349500 Phone: +46 18 349530 Fax: Contact: MEDIcept Inc. 200 Homer Ave Ashland, MA 01721 Scott Blood sblood@medicept.com 978.729.5978

Date Prepared: August 18, 2021

#### 1.2. Device Name

Device Name: Bact-Trap™ Filter Bact-Trap™ Mini Filter Bact-Trap™ Midi Filter Pharma Mini™ HME/Filter Bact-HME™ HME/Filter Bact-HME™ Midi HME/Filter Common Name: Breathing Circuit Bacterial Filter Classification Name: Breathing Circuit Bacterial Filter Regulatory Class: Class II per 21 CFR §868.5260 Product Code: CAH - Breathing Circuit bacterial filter and HME

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for MEDICEPT. The word "MEDI" is in large, bold, blue letters, and the word "CEPT" is in large, bold, purple letters. Below the logo is the phrase "Trusted Solutions, Rapid Response" in a smaller font, with "Trusted Solutions" in gray and "Rapid Response" in purple.

### Legally Marketed Predicate Device 1.3.

The following are the identified Predicate Devices:

- Pharma Systems K903056 Bact Trap filters .
	- Intended Use: o
		- Isolation of patients from bacterial and viral contamination, prevention of cross contamination
		- Protection of ventilators, resuscitators, measuring devices, etc. from contamination.
- Pharma Systems K903058 Bact HME ●
	- Intended Use: O
		- Protection of patients from excessive heat and moisture loss.
		- Isolation of patients from bacterial and viral contamination, prevention of cross contamination (HMEF)

The following is the identified Reference Device:

- ARC Medical, Inc K090738- FilterFloFilter, ThermoFlo Filter/HME ●
	- o Intended Use: FilterFloTM filter is a breathing system filter which is designed to reduce possible airborne or liquid-borne cross contamination with micro-organisms and particulate matter via anesthetic or ventilator breathing systems.

The FilterFloTM filter may either be used on the patient side or on the device side of the ventilator anesthetic device and is used as a hygienic measure alternatively to decontamination of breathing system and / or breathing gas conveying parts of the ventilator.

The TherrmoFlo™ filter/HME is a breathing system filter and a Heat and Moisture Exchanger. The combination of a filter and a Heat and Moisture Exchanger offer the benefit of both product features. Heat and Moisture Exchangers are used as a conditioning system for mechanically ventilated patients whose upper airways are bypassed. In almost all cases of mechanical ventilation they are a fully valid alternative to heated humidifiers. The product is the only conditioning opportunity of breathing gases in cases of emergency ventilation or during transport since heated humidified are almost impossible to use. The TherrnoFlo™ filter/HME should be used with patients who have a Tidal Volume between 250 - 1500 ml.

The products mentioned above are designed as disposable single patient use and should be changed at least every 24 hours.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for MEDICEPT. The word "MEDI" is in large, bold, blue letters, and the word "CEPT" is in a stylized purple font. Below the logo is the text "Trusted Solutions, Rapid Response" in a smaller, gray font.

The predicates referenced above have not been subject to a design-related recall.

### Device Description 1.4.

The device is a standard breathing circuit bacterial filter used for respiratory management. A breathing circuit bacterial filter is a device that is intended to remove microbiological and particulate matter from the gases in the breathing circuit. The function of any breathing circuit is to deliver oxygen and anesthetic gases, and eliminate carbon dioxide. These devices can be placed at the following locations on a breathing circuit:

- At the inhalation port of the CO2 absorber of the anesthesia gas machine ●
- At the exhalation port of the CO2 absorber of the anesthesia gas machine
- At the patient end - at the mask or ET tube
- These placements are also common on critical ventilator circuits

The Bact-Trap™ filter is a breathing filter which can be used on the patient's side or on the device side of the ventilator/anesthetic device. It is used as a hygienic measure to decontaminate the breathing circuit.

The Pharma Mini™ HME/Filter , Bact-HME™ HME/Filter and Bact-HME™ Midi HME/Filter is a breathing system filter and a Heat and Moisture Exchanger (HME).

Heat and Moisture Exchangers and HME Filters (HMEF) are designed for patients with a compromised upper airway. HME and HMEFs are used with intubated or tracheostomized patients to collect moisture and heat from the expired gases and return the inhalation cycle. HME and HMEF contain either plastic fibers or open cell polyurethane foam which offer sufficient surface area to mimic or replace the patient's upper airways while intubated. HMEFs have an additional bacterial/viral filter to help prevent the transmission of bacteria and viruses. Bacterial/Viral filter media consists of electrostatically charged fibers and water repellent cover web on both sides. They also work to prevent cross infection to and from the patient during mechanical ventilation of lungs.

The proposed devices listed below are single patient use and should be changed every 24 hours in the environment of use. The patient population applicable to these devices are specified by the tidal volumes for the filter/Heat and Moisture Exchanger.

The Pharma Mini™HME/Filter, Bact-HME™ filter/HME and Bact-HME™ Midi HME/Filter should be used with patients who have a Tidal Volume as follows:

- Pharma Mini™ and Bact-Trap™ Mini 50-900ml
- · Bact-HME™ Midi and Bact-Trap™ Midi between 100-1200 ml
- · Bact-HMETM and Bact-Trap™ between 250-1500 ml

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image is a logo for MEDICEPT. The word "MEDI" is in large, bold, blue letters, and the word "CEPT" is in a stylized, purple font. Below the words is the phrase "Trusted Solutions, Rapid Response" in a smaller, gray font. The logo is simple and professional, and the colors are eye-catching.

The Pharma Mini, Bact-HME Midi and Bact-HME HME Filters and Bact-Trap Mini, Bact-Trap Midi and Bact-Trap filters are offered in several housing configurations, all of which are typical of filters. These configurations include:

- All have standard conical 15 mm / 22 mm fittings for connections ●
- Some models have a port version Female luer lock port for gas sampling for end-tidal . CO2
- Some models have a straight version .
- Some models have an angled version eliminates the need for a mask elbow .
- Straight version packaged with a standard mask elbow for convenience of the user ●

#### 1.4.1. Environment of Use

Hospital, sub-acute institutions, and pre-hospital

#### Principles of Operation 1.4.2.

There are two principles of operation with the subject devices:

- . Filtration is via the principle of electrostatic charges that attract and capture the microbes in the polypropylene media.
- Passive humidification is the use of a "treated" media (paper or foam) which absorbs the patient's exhaled heat and moisture and upon inhales the retained heat and humidity is released to the dry inhalation gases.

#### 1.4.3. Materials of Use

The proposed devices contain the following biocompatible materials:

- The Bacterial/viral filters Bact-Trap, Bact-Trap Midi, Bact-Trap Mini, Bact-Trap Gas Return 7010, 7011, 7015, 7050, 7054, 7055, 7061, 7110, 7120 are composed of the following biocompatible materials:
	- . SBC
	- . Polypropylene
	- Acrylic and Polypropylene fibers .
	- . Polyethylene
- The Heat- and Moisture Exchange Filters Bact-HME, Bact-HME Midi, Pharma Mini 6000, 6020, 6100, 6120, 6121, 6130. 6310, 6320 are composed of the following biocompatible materials:
	- . SBC

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for MEDICEPT. The first part of the logo, "MEDI", is in large, bold, dark blue letters. The second part of the logo, "CEPT", is in a cursive, purple font. Below the logo is the text "Trusted Solutions, Rapid Response" in a smaller, gray font.

- . Polypropylene
- Acrylic and Polypropylene fibers .
- . Polyethylene
- Cross linkable acrylic polymer .
- . Silicone antifoam emulsion
- . Polyethylene
- Thermally bound Polyester(Polyethylene Terephthalate) .

# Duration and Type of Contact:

The proposed devices are intended for limited or less than 24 hours of use. When in use, the devices come in indirect contact with the human body (i.e Mucosal membrane of mouth and nose).

### 1.5. Indications for Use

Bact Trap filter is a breathing system filter which is designed to reduce possible airborne or liquid-borne cross contamination with micro-organisms and particulate matter via anaesthetic or ventilator breathing systems. The Bact Trap filter may either be used on the patient side or on the device side of the ventilator anaesthetic device.

The Bact HME/Bact-HME Midi/Pharma Mini is a breathing system filter and a Heat and Moisture Exchanger. The combination of a filter and a Heat and Moisture Exchanger offer the benefit of both product features. Heat and Moisture Exchangers are used as a conditioning system for mechanically ventilated patients whose upper airways are bypassed. In almost all cases of mechanical ventilation they are a fully valid alternative to heated humidifiers. The Bact HME/Bact-HME Midi/Pharma Mini should be used with patients who have a Tidal Volume between 50 - 1500 ml.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for MEDICEPT. The word "MEDI" is in bold, dark blue letters, while "CEPT" is in a lighter purple color. Below the logo is the text "Trusted Solutions, Rapid Response", with "Trusted Solutions" in purple and "Rapid Response" in gray.

# Technological Characteristics Comparison with the Predicate 1.6. Devices

| Device<br>Feature                              | Proposed Device: Bact-<br>Trap, Pharma Mini,<br>Bact-HME, Backt-<br>HME Filter and<br>HME/Filter<br>K202459                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Primary<br>Predicate:<br>Pharma<br>Systems<br>K903056 - Bact<br>Trap filters                                                                                                                                             | Primary<br>Predicate:<br>Pharma<br>Systems<br>K903058 -<br>Bact HME                                                                                                                                         | Reference Predicate:<br>ARC Medical<br>K090738-<br>ThermoFlo™                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use                         | Bact Trap filter is a<br>breathing system filter<br>which is designed to<br>reduce possible airborne<br>or liquid-borne cross<br>contamination with<br>micro-organisms and<br>particulate matter via<br>anaesthetic<br>or ventilator breathing<br>systems.<br>The Bact Trap filter may<br>either be used on the<br>patient side or on the<br>device side of the<br>ventilator anaesthetic<br>device.<br>The Bact HME/Bact-<br>HME Midi/Pharma Mini<br>is a breathing system<br>filter and a Heat and<br>Moisture Exchanger.<br>The combination of a<br>filter and a Heat and<br>Moisture Exchanger<br>offer the benefit of both<br>product features. Heat<br>and Moisture<br>Exchangers are used as a<br>conditioning system for<br>mechanically ventilated<br>patients whose upper<br>airways are bypassed. In<br>almost all cases of<br>mechanical ventilation<br>they are a fully valid<br>alternative to heated | Isolation of<br>patients from<br>bacterial and<br>viral<br>contamination,<br>prevention of<br>cross<br>contamination.<br>Protection of<br>ventilators,<br>resuscitators,<br>measuring<br>devices, from<br>contamination. | Protection of<br>patients from<br>excessive heat<br>and moisture<br>loss.<br>Isolation of<br>patients from<br>bacterial and<br>viral<br>contamination<br>prevention of<br>cross-<br>contamination<br>(HMEF) | FilterFlo™ filter is a<br>breathing system filter<br>which is designed to<br>reduce possible<br>airborne or liquid-<br>borne cross<br>contamination with<br>micro-organisms and<br>particulate matter via<br>anaesthetic or<br>ventilator breathing<br>systems.<br>The FilterFlo™ filter<br>may either be used on<br>the patient side or on<br>the device side of the<br>ventilator/<br>anaesthetic device and<br>is used as a hygienic<br>measure alternatively<br>to decontamination of<br>breathing system and /<br>or breathing gas<br>conveying parts of the<br>ventilator.<br>The ThermoFlo™<br>filter/HME is a<br>breathing system filter<br>and a Heat and<br>Moisture Exchanger.<br>The combination of a<br>filter and a Heat and<br>Moisture Exchanger<br>offer the benefit of<br>both product features.<br>Heat and Moisture<br>Exchangers are used<br>as a conditioning<br>system for<br>mechanically |
| Device<br>Feature                              | Proposed Device: Bact-<br>Trap, Pharma Mini,<br>Bact-HME, Backt-<br>HME Filter and<br>HME/Filter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Primary<br>Predicate:<br>Pharma<br>Systems<br>K903056 - Bact<br>Trap filters                                                                                                                                             | Primary<br>Predicate:<br>Pharma<br>Systems<br>K903058 –<br>Bact HME                                                                                                                                         | Reference Predicate:<br>ARC Medical<br>K090738-<br>ThermoFlo™                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                                | humidifiers. The Bact<br>HME/Bact-HME Midi/<br>Pharma Mini should be<br>used with patients who<br>have a Tidal Volume<br>between 50 - 1500 ml.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                          |                                                                                                                                                                                                             | ventilated patients<br>whose upper airways<br>are bypassed. In<br>almost all cases of<br>mechanical ventilation<br>they are a fully valid<br>alternative to heated<br>humidifiers. The<br>product is the only<br>conditioning<br>opportunity of<br>breathing gases in<br>cases of emergency<br>ventilation or during<br>transport since heated<br>humidified are almost<br>impossible to use. The<br>ThermoFlo™<br>filter/HME should be<br>used with patients who<br>have a Tidal Volume<br>between 250 - 1500<br>ml.                                                                                                                                                                                                                                                                                                                                                                                             |
| Classification                                 | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                     | Same                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Target<br>Population                           | Adult Tidal Volume<br>50-1500 ml                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                     | Same                                                                                                                                                                                                        | Tidal Volume 250-<br>1500ml                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Clinical<br>Setting                            | Hospital, sub-acute,<br>pre-hospital and<br>home                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                     | Same                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Principle of<br>Operation                      | Filtration via the<br>principle of<br>electrostatic charges<br>in the media.<br>Passive                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                     | Same                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Device<br>Feature                              | Proposed Device: Bact-<br>Trap, Pharma Mini,<br>Bact-HME, Backt-<br>HME Filter and<br>HME/Filter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Primary<br>Predicate:<br>Pharma<br>Systems<br>K903056 - Bact<br>Trap filters                                                                                                                                             | Primary<br>Predicate:<br>Pharma<br>Systems<br>K903058 -<br>Bact HME                                                                                                                                         | Reference Predicate:<br>ARC Medical<br>K090738-<br>ThermoFlo™                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                                | functioning as heat<br>and moisture<br>exchangers.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                          |                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Compatibility<br>with<br>Environment<br>of Use | Intended for use<br>with ventilators,<br>anesthesia gas<br>machines                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                     | Same                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Prescriptive                                   | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                      | Yes                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Sterile/Non-<br>Sterile                        | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Non-Sterile                                                                                                                                                                                                              | Non-Sterile                                                                                                                                                                                                 | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Single Patient<br>Use                          | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                      | Yes                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Maximum<br>Use                                 | 24 hours                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                     | Same                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Shelf-Life                                     | 3 years                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                     | Same                                                                                                                                                                                                        | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Filtration<br>Efficiency                       | BFE: >99.999%<br>VFE: >99.99%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same                                                                                                                                                                                                                     | Same                                                                                                                                                                                                        | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for MEDICEPT. The first part of the logo, "MEDI" is in large, bold, blue letters. The second part of the logo, "CEPT" is in a stylized purple font. Underneath the logo is the text "Trusted Solutions, Rapid Response" in a smaller font.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for MEDICEPT. The first part of the name, "MEDI", is in bold, dark blue letters. The second part of the name, "CEPT", is in a cursive, purple font. Below the name is the phrase "Trusted Solutions, Rapid Response" in a smaller font.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for MEDI-CEPT. The word "MEDI" is in large, bold, blue letters, while "CEPT" is in a stylized purple font. Below the company name is the tagline "Trusted Solutions, Rapid Response" in a smaller font, with "Trusted Solutions" in purple and "Rapid Response" in gray.

### 1.7. Summary of Non-Clinical Testing

| Test Method                          | Test Performed                                                                                                                                | Summary                                                                                                                  | Results<br>(pass/fail) |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|------------------------|
| ISO 9360-1                           | Gas Leakage and compliance test<br>Moisture loss test                                                                                         | No leak @ 1 psi for 2 min<br>Equivalent to Predicate device                                                              | Pass                   |
| ISO 10993                            | Cytotoxicity test<br>Sensitization test<br>Irritation test                                                                                    | Non-cytotoxic<br>Non-sensitizer<br>Non-irritant                                                                          | Pass                   |
| ISO 18562-2 and<br>-3                | VOC and Particulate Matter<br>Testing                                                                                                         | All test method acceptance<br>criteria were met.                                                                         | Pass                   |
| ISO 18562-4                          | Leachables evaluation in polar and<br>nonpolar solvents                                                                                       | All test method acceptance<br>criteria were met.                                                                         | Pass                   |
| ISO 10993-17<br>and ISO 10993-<br>18 | Toxicological Risk Assessment                                                                                                                 | All test method acceptance<br>criteria were met.                                                                         | Pass                   |
| ASTM 1980-16                         | Accelerated Aging                                                                                                                             | Device meets its performance<br>specification post-<br>conditioning                                                      | Pass                   |
| ASTM F2101                           | Viral Filtration Efficiency<br>Bacterial Filtration Efficiency                                                                                | Filtration efficiency is<br>equivalent to Predicate device<br>Filtration efficiency is<br>equivalent to Predicate device | Pass                   |
| EN ISO 23328-2<br>EN ISO 23323-2     | Pressure Drop                                                                                                                                 | Device meets its performance<br>specification                                                                            | Pass                   |
| N/A                                  | Housing Burst Strength                                                                                                                        | The measured values are<br>sufficient for ensuring safe<br>use of the measured devices.                                  | Pass                   |
| ISO 23328-1                          | Filtration Performance of<br>Technostat Plus filter media for<br>Bact-Trap bacterial/viral filters and<br>HME filters shelf life confirmation | The measured values are<br>sufficient for ensuring safe<br>use of the measured devices.                                  | Pass                   |
| ISO 9360-1                           | Bact-Trap bacterial/viral filters and<br>HME filters shelf life confirmation                                                                  | The measured values are<br>sufficient for ensuring safe<br>use of the measured devices.                                  | Pass                   |

#### 1.8. Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the device (Bact-Trap, Pharma Mini, Bact-HME, Bact-HME Filter and HME/Filter) is as safe, as effective, and performs as well as or better than the legally marketed device

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94therapeutic-devices/CAH/K202459](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94therapeutic-devices/CAH/K202459)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
