← Product Code [CAH](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/CAH) · K131626

# EXCELERON BACTERIAL FILTERS (K131626)

_Exceleron Medical · CAH · Feb 26, 2014 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94therapeutic-devices/CAH/K131626

## Device Facts

- **Applicant:** Exceleron Medical
- **Product Code:** [CAH](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/CAH.md)
- **Decision Date:** Feb 26, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5260
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology

## Indications for Use

Exceleron Medical bacterial filters are single use replacement filters intended for use in oxygen concentrator machines to help remove contaminants, including air borne bacteria and other particulate debris from an air stream. When used with oxygen concentrator machines, the replacement filters may be used in the home, nursing home, hospital, patient care facility.

## Device Story

Exceleron Bacterial Filters are single-use replacement accessories for oxygen concentrators; they function as intake or final patient filters. The device consists of a molded housing containing a glass microfiber filtration bed. As the oxygen concentrator draws room air through its compressor and sieve beds to increase oxygen concentration, the filter removes airborne bacteria and particulate debris from the air stream before delivery to the patient. Used in home, hospital, or nursing home settings; operated by patients or healthcare providers. The filter ensures the delivery of cleaner oxygen-enriched air, reducing potential patient exposure to contaminants.

## Clinical Evidence

No clinical data. Bench testing only. Verification testing included bacterial filtration efficiency (BFE) and air flow resistance. BFE results ranged from 99.9% to 99.999%, comparable to predicate devices. Biocompatibility testing was conducted per ISO 10993-1, including cytotoxicity, sensitization, and intracutaneous reactivity, confirming the device meets safety requirements for limited duration tissue contact.

## Technological Characteristics

Molded housing containing glass microfiber filtration material. Designed as replacement filters for oxygen concentrators (e.g., DeVilbiss, Respironics, Invacare). Principle of operation: mechanical filtration via glass microfiber bed. Biocompatibility evaluated per ISO 10993-1.

## Regulatory Identification

A breathing circuit bacterial filter is a device that is intended to remove microbiological and particulate matter from the gases in the breathing circuit.

## Predicate Devices

- AG Industries Bacterial Filters ([K091363](/device/K091363.md))
- Porous Media Oxygen Concentrator Filters ([K061426](/device/K061426.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

FEB 2 6 2014

# 510(k) Summary Exceleron Bacterial Filter

Date Prepared: May 30, 2013 (Revised Feb 26, 2014)

- Submitter: Exceleron Medical 5000 Township Parkway St Paul, MN 55110 Telephone: 651- 855-1466 Fax: 651-855-1465
- Contact: Mr. Bernard Horwath Regulatory Affairs Consultant 4486 Timberline Ct Vadnais Heights, MN 55127 Telephone: 651-231-1761 Fax: 651-855-1465

Proprietary Name: Exceleron Bacterial Filter (See table below for specific models)

| PROPRIETARY NAME | COMMON NAME              |
|------------------|--------------------------|
| DBX32            | Bacterial Intake Filter  |
| DBX24            | Bacterial Intake Filter  |
| DBX25xx          | Bacterial Intake Filter  |
| DBXEFLO          | Bacterial Intake Filter  |
| BDF47xx          | Bacterial Patient Filter |

Common/Usual Name: Breathing Circuit Bacterial Filter

Classification Name: Breathing Circuit Bacterial Filter, Class II, Product Code CAH 21 CFR 868.5260

# Establishment Registration Number: 3007709321

# Description:

A breathing circuit bacterial filter is a device that is intended to remove microbiological and particulate matter from the gases in the breathing circuit of a respiratory device. The filter is a replaceable accessory device used in oxygen concentrators, either as intake filters or final patient filters. The function of the oxygen concentrator machine is to draw room air into the machine's compressor and concentrate the oxygen content before delivering filtered, oxygen rich air to the patient. From the compressor, the filtered air proceeds to the sieve beds. The sieve beds in the oxygen concentrator machine condition the air, by removing nitrogen from the air stream, which results in a higher concentration of oxygen. The air then passes through the final patient filter before being supplied to the patient. The Exceleron Medical bacterial intake filter and the final patient filter are intended to remove air borne bacteria and other particulate debris from the air stream.

{1}------------------------------------------------

## Indications for Use:

Exceleron Medical bacterial filters are single use replacement filters intended for use in oxygen concentrator machines to help remove contaminants, including air borne bacteria and other particulate debris from an air stream. When used with oxygen concentrator machines, the replacement filters may be used in the home, nursing home, hospital, patient care facility.

## Substantial Equivalence:

The Exceleron Bacterial Filter is substantially equivalent to the following predicate devices:

- AG Industries Bacterial Filters K091363 .
- Porous Media Oxygen Concentrator Filters K061426 .

## Technological Characteristics:

Technically from a design and mechanism of action standpoint, the Exceleron Bacterial Filters are substantially equivalent to the predicate devices. All utilize a dimensionally equivalent molded housing containing glass microfiber filtration material with appropriate connector fittings for attachment to the applicable breathing circuit. Exceleron Bacterial Filters and the predicate devices have the same principle of operation and mechanism of action, namely air flows through the glass microfiber bed to remove contaminants, including air borne bacteria and other particulate debris.

All the filters are designed as replacement filters for use on oxygen concentrator machines, such as DeVilbiss, Respironics, and Invacare.

## Biocompatibility:

The bacterial filter is a tissue contact device that filters the air delivered to the patient. As a result, a systematic risk analysis and biological evaluation were conducted in accordance with ISO 10993-1. Biological Evaluation of Medical Devices. The device was categorized as a tissue contact device with limited duration contact due to the dry gas pathway. As such the following tests were conducted per ISO 10993-1: cytotoxicity, sensitization and intracutaneous reactivity. This testing showed that the Exceleron Bacterial Filter devices meet the biocompatibility requirements for their intended use.

#### Performance Bench Testing:

Design verification testing was performed for the Exceleron Bacterial Filter to demonstrate physical and functional requirements were met. Testing included bacterial filtration efficiency (BFE) at an increased challenge and air flow resistance. All tests were successful. The Exceleron filters demonstrated BFE capabilities from 99.9% to 99.999%, which are in the same range as the predicate BFE numbers, and air flow resistance was comparable or less than the comparable products, demonstrating substantial equivalence.

#### Conclusion:

Through the data and information presented, Exceleron Medical, considers the Bacterial Filters substantially equivalent to the predicate devices already on the market (cleared by the 510(k) process) in terms of indications for use, scientific technology, design, materials, principle of operation and functional performance and present no new concerns about safety and effectiveness.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center = WO66-G609 Silver Spring, MD 20993-0002

February 26, 2014

Exceleron Medical C/O Mr. Bernard Horwath HRG 4486 Timberline Ct Vadnais Heights MN 55127

Re: K131626

Trade/Device Name: Exceleron Bacterial Filter Regulation Number: 21 CFR 868.5260 Regulation Name: Breathing Circuit Bacterial Filter Regulatory Class: II Product Code: CAH Dated: January 24, 2014 Received: January 28, 2014

Dear Mr. Horwath:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{3}------------------------------------------------

Page 2 - Mr. Horwath

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Image /page/3/Picture/7 description: The image shows the text "Kwame Ulmers ODA". The text is arranged in two rows, with "Kwame" on the top row and "Ulmers" on the bottom row. The letters "ODA" are stylized and appear to be part of a logo. The image is black and white and has a slightly grainy texture.

for

Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known) K131626

Device Name Exceleron Bacterial Filter

Indications for Use (Describe)

Exceleron Medical bacterial filters are single use replacement filters intended for use in oxygen concentrator machines to help remove contaminants, including air borne bacteria and other particulate debris from an air stream. When used with oxygen concentrator machines, the replacement filters may be used in the home, norsing home, hospital, patient care facility.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

J Over-The-Counter Use (21 CFR 801 Subpart C)

# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

A COLLECT COLLECT FOR FOR FOR FOR FOR FOR FOR THE FOR AND AND AND AND AND AND AND AND Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Image /page/4/Picture/12 description: The image shows a stylized logo of the FDA. The letters 'FDA' are formed by a series of parallel lines, creating a layered effect. The 'F' and 'D' are blocky and angular, while the 'A' is triangular with a horizontal line running through its center, also made of parallel lines.

Anva C. Harry -S 2014.02.26 16:25:56 -05'00'

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94therapeutic-devices/CAH/K131626](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94therapeutic-devices/CAH/K131626)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
