SUNSET HEALTHCARE SOLUTIONS BACTERIAL/VIRAL HME FILTER, MODEL BHF 104

K102483 · Sunset Healthcare Solutions, LLC · CAH · Sep 13, 2011 · Anesthesiology

Device Facts

Record IDK102483
Device NameSUNSET HEALTHCARE SOLUTIONS BACTERIAL/VIRAL HME FILTER, MODEL BHF 104
ApplicantSunset Healthcare Solutions, LLC
Product CodeCAH · Anesthesiology
Decision DateSep 13, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5260
Device ClassClass 2

Indications for Use

The BHF 104 is a disposable single-use filter for patients who require humidification during the delivery of ventilator gases and provide filtration for reducing possible cross-contamination between patient and equipment. The BHF 104 is for use in hospital, ICU, anesthesia, and respiratory therapy.

Device Story

BHF 104 is a disposable, single-use, barrier-type heat and moisture exchanger (HME) and bacterial/viral filter. Device consists of a translucent plastic housing (K-Resin) with 15mm/22mm connector ports and a gas sampling luer port. It is placed in-line within a breathing circuit to provide humidification of inspired gases and filtration of pathogens. Operated by clinicians in hospital, ICU, anesthesia, and respiratory therapy environments. Device captures moisture from exhaled breath to humidify subsequent inhaled gases; filter media traps bacteria and viruses to prevent cross-contamination between patient and ventilator equipment. Output is passive; clinical decision-making relies on standard ventilator management protocols. Benefits include reduced risk of cross-contamination and maintenance of airway humidity for ventilated patients.

Clinical Evidence

Bench testing only. Performance data includes Bacterial Filtration Efficiency (>99.9999%), Viral Filtration Efficiency (99.999%), flow resistance at 30/60/90 L/min, HME moisture output (37.5 mg H2O/L at 500ml VT), dead space (53ml), and mechanical integrity of conical connectors.

Technological Characteristics

Housing: K-Resin KR03 (Styrene Butadiene Block Copolymer). Filter: 3M Filtrete Air Media Filter Type G. Passive HME/filtration principle. Dimensions: 90mm length, 55mm diameter. Connectors: 15mm ID x 22mm OD and 22mm ID x 15mm OD. Gas sampling luer port. Non-sterile delivery.

Indications for Use

Indicated for patients requiring humidification during ventilator gas delivery and filtration to reduce cross-contamination between patient and equipment. For use in hospital, ICU, anesthesia, and respiratory therapy settings.

Regulatory Classification

Identification

A breathing circuit bacterial filter is a device that is intended to remove microbiological and particulate matter from the gases in the breathing circuit.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K102483 ال # 510(k) Summary for the Bacterial/Viral HME Filter (BHF 104) #### Name and Address of Sponsor Sunset Healthcare Solutions, Inc 2201 S. Halsted St. Suite 1344 Chicago, IL 60608 #### Name and Address of Manufacturer Altera Tibbi Malzeme San Ve Tic As. Turan Mah Tire Organize Sanayi Bolgesi. Tire/Izmir Aegean Region, 35900 Turkey Phone: 90-232-2375949 #### Establishment Registration Number 3006534321 ### Name and Address of Official Correspondent Regulatory Insight, Inc. 5401 S. Cottonwood Ct. . Greenwood Village, Colorado 80121 Contact: Mr. Kevin Walls, RAC Tel: 720-962-5412 Fax: 720-962-5413 Email: kevin@reginsight.com #### Date Summary was Prepared December 17, 2010 #### Device Name Trade Name: Bacterial/Viral HME Filter (BHF 104) Common Name: Disposable Heat and Moisture Exchanger and Bacteria/Viral Filter Classification Name: Breathing circuit bacterial filter #### Classification, Panel and Product Code Class 2, Anesthesiology, Regulation Number: 868.5260 Product Code: CAH {1}------------------------------------------------ ### Indications for Use The BHF 104 is a disposable single-use filter for patients who require humidification during the delivery of ventilator gases and provide filtration for reducing possible crosscontamination between patient and equipment. The BHF 104 is for use in hospital, ICU, anesthesia, and respiratory therapy. ## Legally Marketed Device to Which Substantial Equivalence is Claimed DATEX-OHMEDA HMEF 500 (K021265) ## Device Description The BHF 104 filter is a single-use, barrier type filter, fabricated from a plastic resin, and delivered to the customer in a sealed, non-sterile package. The translucent housing has aligned connector ports (15mm ID x 22mm OD, 22mm ID x 15mm OD), a perpendicular filter media, and a gas sampler luer port, protruding at 45 degrees from the base of the filter housing. The entire filter is 90mm (3.54 inches) in length, with a maximum cross section diameter of 55mm (2.16 inches), and a mass of 24g (.85 ounces). The surface area is 26.4cm2. The BHF 104 has 15mm ID x 22mm OD and 22mm ID x 15mm OD male and female connectors, which allows compatibility with various compressors, ventilators, masks, and aerosol tubes. ### Specifications | Tidal Volume: | 150-1500ml | |--------------------------|------------------------------------------------------------| | Flow rate resistance: | 30lt/min 5mm/H20<br>60lt/min 13mm/H20<br>90lt/min 28mm/H20 | | Bacterial filtration: | >99.9999% | | Viral filtration: | 99.999% | | Dead Space: | 53ml | | HME Output per ISO 9360: | 37.5mg/L H2O at 500ml VT | ### Performance Testing Performance was based on the following tests, which are contained in the body of the 510(k) application: - . Bacterial Filtration Efficiency (BFE) - Viral Filtration Efficiency (VFE) - Dead Space 트 - I Resistance to Flow - Heat and Moisture Exchange (HME) l - . Conical Connectors {2}------------------------------------------------ ## Substantial Equivalence | | DATEX-OHMEDA HMEF 500 | Bacterial/Viral HME Filter (BHF 104) | |---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) # | K021265 | N/A | | Indications for use | The HMEF500 is a disposable single-use device indicated for patients who require humidification during the delivery of ventilator gases and provide filtration for reducing possible cross contamination between patient and equipment. The HMEF 500 is for use in hospital, ICU, anesthesia, respiratory therapy, during transport and with resuscitators. The device can be used on adult and pediatric patients. The device is indicated for use by qualified medical personnel only. | The BHF 104 is a disposable single-use filter for patients who require humidification during the delivery of ventilator gases and provide filtration for reducing possible cross-contamination between patient and equipment. The BHF 104 is for use in hospital, ICU, anesthesia, and respiratory therapy. | | Diameter | 45mm | 55mm | | Length | 74mm | 90mm | | BFE | 99.999% | 99.9999% | | VFE | 99.98% | 99.999% | | Housing Material | PP Polypropylene | K-Resin KR03 - Styrene Butadiene Block Copolymer | | Filter Material | PP and Acrylic Fibers | 3M Filtrete Air Media Filter - Type G | | Fittings | 15mm ID x 22mm OD<br>15mm OD x 22mm ID | 15mm ID x 22mm OD<br>15mm OD x 22mm ID | | Moisture Output | 30mg H2O/L @ 500ml VT | 37.5 mg H2O/L @ 500ml VT | | Tidal Volume Range | 120-500 | 150-1500 | | Dead Space | 30ml | 53ml | | Weight | 15g | 24g | | Sampling Port | Luer lock | Luer lock | None of the differences listed in the table above have any affect on safety or effectiveness. . . {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake entwined around it, and three horizontal bars above it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the caduceus symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Sunset Healthcare Solutions. LLC C/O Mr. Kevin Walls, RAC Principal Consultant Regulatory Insight, Incorporated 5401 S. Cottonwood Court Greenwood Village, Colorado 80121 SEP 13 2011 Re: K102483 Trade/Device Name: Sunset Healthcare Solutions Bacterial/Viral HME Filter (BHF 104) Regulation Number: 21 CFR 868.5260 Regulation Name: Breathing Circuit Bacterial Filter Regulatory Class: II Product Code: CAH Dated: August 19, 2011 Received: August 22, 2011 Dear Mr. Walls: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. Walls Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem /default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industrv/default.htm. Sincerely yours. hr for Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {5}------------------------------------------------ Indications for Use 510(k) Number (if known): K102483 Device Name: Sunset Healthcare Solutions Bacterial/Viral HME Filter (BHF 104) Indications for Use: The BHF 104 is a disposable single-use filter for patients who require humidification during the delivery of ventilator gases and provide filtration for reducing possible cross-contamination between patient and equipment. The BHF 104 is for use in hospital, ICU, anesthesia, and respiratory therapy. Prescription Use × AND/OR (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) L Schutter (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices Page 1 of 1 K 102483 510(k) Number: _
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...