← Product Code [SCL](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/SCL) · K222385

# Bifurcated Needle (K222385)

_Anhui Tiankang Medical Technology Co., Ltd. · SCL · Dec 6, 2022 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/SCL/K222385

## Device Facts

- **Applicant:** Anhui Tiankang Medical Technology Co., Ltd.
- **Product Code:** [SCL](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/SCL.md)
- **Decision Date:** Dec 6, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5570
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The Bifurcated Needle is intended for use in administering vaccines by the scarification method or administering epidermal allergens.

## Device Story

The Bifurcated Needle is a manual, single-use device featuring two prongs designed to hold a small drop of vaccine or allergen. A healthcare provider performs approximately fifteen punctures into the patient's skin to deliver the substance. The device is constructed from SUS 304 stainless steel and includes a PVC needle tip protector. It is supplied sterile, having undergone ethylene oxide (EO) sterilization to achieve a sterility assurance level (SAL) of 10^-6. The device is intended for manual use by healthcare professionals in clinical settings. It serves as a delivery mechanism for vaccines or allergens, with the bifurcated design facilitating the scarification technique.

## Clinical Evidence

No clinical data. Bench testing only, including biocompatibility (ISO 10993), sterilization validation (ISO 11135), shelf-life/stability (ASTM F1980-16), package integrity (ASTM F88, F1929, F1886), and performance testing (ISO 7864, ISO 9626).

## Technological Characteristics

Materials: SUS 304 stainless steel needle, PVC tip protector. Dimensions: 64mm length, 2mm width. Principle: Manual scarification. Sterilization: Ethylene Oxide (EO) per ISO 11135. Biocompatibility: ISO 10993 compliant. Standards: ISO 7864 (sterile hypodermic needles), ISO 9626 (stainless steel tubing).

## Regulatory Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

## Predicate Devices

- BD Bifurcated Needle ([K020523](/device/K020523.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 6, 2022

Anhui Tiankang Medical Technology Co., Ltd. Zhang Yong Management No.228,Weiyi Road, Economic Development Zone Tianchang, Anhui 239300 China

Re: K222385

Trade/Device Name: Bifurcated Needle Regulatory Class: Unclassified Product Code: LDH Dated: November 10, 2022 Received: November 10, 2022

Dear Zhang Yong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Alan M.
Stevens -
S3

CAPT Alan M. Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K222385

Device Name Bifurcated Needle

Indications for Use (Describe)

The Bifurcated Needle is intended for use in administering vaccines by the scarification method or administering epidermal allergens.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

> Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K222385 510(k) Summary

### l Submitter

Device submitter: Anhui Tiankang Medical Technology Co., Ltd. No.228, Weiyi Road, Economic Development Zone, Tianchang City, Anhui, China.

Contact person: Name: Zhang yong Title: Management representative Phone: +86-13705505106 Fax: +86-550-7309158 E-mail: zy@tkmedical.com

Preparation Date: December 6, 2022

#### II Device

Trade Name of Device: Bifurcated Needle Common Name: System, Delivery, Allergen And Vaccine Regulation Name: None Requlatory Class: Unclassified Product code: LDH Review Panel: General Hospital

### III Predicate Devices

| Trade name:             | BD Bifurcated Needle    |
|-------------------------|-------------------------|
| Common name:            | Bifurcated Needle       |
| Classification:         | Unclassified            |
| Product Code:           | LDH                     |
| Premarket Notification: | K020523                 |
| Manufacturer:           | BECTON, DICKINSON & CO. |

### IV Device description

The proposed Bifurcated Needle have two prongs. A small drop of smallpox vaccine was placed between the prongs and approximately fifteen punctures would be made into the skin. The needles were sterilized by EO to achieve a SAL of 10-6 and supplied sterile in packaging with a shelf life of five years.

{4}------------------------------------------------

## V Indications for use

The Bifurcated Needle is intended for use in administering vaccines by the scarification method or administering epidermal allergens.

## VI Comparison of technological characteristics with the predicate devices

The Bifurcated Needle has the same intended use, technology, design and biocompatibility is either identical or substantially equivalent to existing legally marketed predicate devices. The comparison between the subject device and the predicate devices are listed in below tables:

| Device<br>feature         | Subject Device<br>K222385                                                                                                                            | Predicate Device<br>K020523                                                                                                                             | Comment                 |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| Product                   | Bifurcated Needle                                                                                                                                    | BD Bifurcated Needle                                                                                                                                    | /                       |
| Indications<br>for use    | The Bifurcated Needle is<br>intended for use in<br>administering vaccines by<br>the scarification method or<br>administering epidermal<br>allergens. | The BD Bifurcated Needle is<br>intended for use in<br>administering vaccines by<br>the scarification method or<br>administering epidermal<br>allergens. | Same                    |
| Product code              | LDH                                                                                                                                                  | LDH                                                                                                                                                     | Same                    |
| Principle of<br>operation | For Manual Use Only, For<br>Single Use Only                                                                                                          | For Manual Use Only, For<br>Single Use Only                                                                                                             | Same                    |
| Models                    | BFN001                                                                                                                                               | 301754; 301755; 301756;<br>301757                                                                                                                       | Difference<br>comment 1 |
| Needle Length             | $64mm\pm0.3mm$ Needle                                                                                                                                | 60.325mm                                                                                                                                                |                         |
| Needle<br>dimension       | $2mm\pm0.05mm$                                                                                                                                       | 1.905mm                                                                                                                                                 |                         |
| Materials                 | Bifurcated needle-(SUS<br>304)<br>Needle tip protector-(PVC<br>MT)                                                                                   | Stainless steel                                                                                                                                         | Difference<br>comment 2 |
| Sterilization             | EO Sterilization                                                                                                                                     | Sterilization                                                                                                                                           | Difference<br>comment 3 |

#### Table 6-1 Substantial equivalence discussion for Bifurcated Needle

{5}------------------------------------------------

| Device<br>feature | Subject Device<br>K222385                                                                                                                  | Predicate Device<br>K020523                                                                                                                | Comment |
|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Biocompatibility  | Contact level: blood path,<br>indirect, limited contact (<24<br>hours).<br>Conforms to the<br>requirement of ISO 10993<br>series Standards | Contact level: blood path,<br>indirect, limited contact (<24<br>hours).<br>Conforms to the<br>requirement of ISO 10993<br>series Standards | Same    |
| Labeling          | Complied with 21 CFR part<br>801                                                                                                           | Complied with 21 CFR part<br>801                                                                                                           | Same    |

# Difference comment 1

The subject device is similar to the predicate device. There are minor differences in needle length and dimension. These differences do not affect intended use. In addition, differences were addressed through ISO 7864 and ISO 9626. Therefore, the differences on models and the specification of needle length and needle width will not raise different question of safety and effectiveness.

# Difference comment 2

The material of the needle of the subject device is similar to the predicated device. They are both stainless steels. The differences on materials do not raise new questions about safety and effectiveness.

# Difference comment 3

The sterilization of the predicate device was unknown while the subject device was sterilized by EO to achieve a SAL of 10-6 and it has been validated to ISO11135. Both devices are sterile.

# Discussion:

From clinical aspect: The subject device has the same indication for use as the predicate.

From technological aspect: The subject device has the same principle of operation, and configuration.

From product performance. The main difference is that specification of models and l the specification of needle length and needle width. However, all of them will be selected by health care provider per injection requirement and this difference does not affect indication for use. Additionally, the performance of needle has been evaluated and the test results met the requirements of ISO 7886-1 and ISO 7864. Therefore, this difference does not affect substantially equivalency on safety and effectiveness.

From material perspective: The biocompatibility of the subject device was evaluated through tests according to the requirements of ISO10993 series standards. Therefore, this difference does not affect substantially equivalency on safety and effectiveness.

{6}------------------------------------------------

From sterilization: The subject device was sterilized by EO to achieve a SAL of 10-6 and it has been validated to ISO11135. Therefore, this difference does not affect substantially equivalency on safety and effectiveness.

### VII Performance data

The following performance data were provided in support of the substantial equivalence determination.

#### Biocompatibility testing

Biocompatibility of the Bifurcated Needle was evaluated in accordance with ISO 10993-1:2018 for the body contact category of "External communication device – Blood path indirect" with a contact duration of "Limited (< 24 hours)". The following tests were performed, as recommended:

| In Vitro Cytotoxicity Test     | ISO 10993-5: 2009  |
|--------------------------------|--------------------|
| Skin Sensitization Test        | ISO 10993-10: 2010 |
| In Vitro Hemolysis Test        | ISO 10993-4: 2017  |
| Intracutaneous Reactivity Test | ISO 10993-10: 2010 |
| Acute Systemic Toxicity Test   | ISO 10993-11: 2017 |
| Pyrogen Test                   | ISO 10993-11: 2017 |

#### Sterilization and shelf life testing

The sterilization method has been validated to ISO11135, which has thereby determined the routine control and monitoring parameters.

The testing is performed according to the following standards:

| EO residue  | ISO 10993-7:2008 |
|-------------|------------------|
| ECH residue | ISO 10993-7:2008 |

The shelf life of 5 year is determined based on stability studies which include ageing test according to FDA recognized standard ASTM F1980-16.

Package integrity testing was conducted on the final, packaged, and sterile devices after environmental conditioning and simulated transportation. All packaging deemed acceptable for protection of product and sterility maintenance.

The testing is performed according to the following standards:

| Seal strength                      | ASTM F88/F88M-15        |
|------------------------------------|-------------------------|
| Blue Dye Penetration               | ASTM F 1929-2015        |
| Seal Integrity (Visual Inspection) | ASTM F 1886/ F 1886M-16 |

### Performance testing

Performance testing is performed according to the following standards:

- A ISO 7864: 2016, Sterile hypodermic syringes for single use
- A ISO 9626:2016, Stainless Steel Needle Tubing For The Manufacture of Medical Devices-Requirements and test methods.

{7}------------------------------------------------

# VIII Clinical Test Conclusion

No clinical study is included in this submission.

# IX Conclusion

The Bifurcated Needle IS substantially equivalent to its predicate device. The non-clinical testing demonstrates that the device is as safe, as effective and performs as well as the legally marketed device.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/SCL/K222385](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/SCL/K222385)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
