← Product Code [QNQ](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/QNQ) · K231856

# TopFine® LDS (Low Dead Space) Syringe (2 models (TS-LDS2334, TS-LDS2534)) (K231856)

_Medexel Co.,Ltd · QNQ · Mar 15, 2024 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/QNQ/K231856

## Device Facts

- **Applicant:** Medexel Co.,Ltd
- **Product Code:** [QNQ](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/QNQ.md)
- **Decision Date:** Mar 15, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5860
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

TopFine® LDS (Low Dead Space) Syringe is intended for use to inject fluid into or withdraw fluids from the parts of body below the surface of the skin.

## Device Story

TopFine® LDS Syringe is a single-use, manually operated piston syringe with a permanently attached hypodermic needle. It consists of a calibrated hollow barrel, plunger, and needle. The device is used by clinicians to inject or withdraw fluids from the body. The low dead space design (≤ 0.03 ml) minimizes fluid retention. The device is supplied sterile via Ethylene Oxide gas. It is intended for use in clinical settings where subcutaneous or intramuscular fluid delivery or aspiration is required.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-1), physical performance (ISO 7864, ISO 7886-1, ISO 9626), sterility, and packaging integrity (ASTM D4169, ASTM F88, ASTM F1929, ASTM F1980).

## Technological Characteristics

Materials: SUS 304 needle, Polypropylene barrel/plunger/cap, Polyisoprene gasket. Manual operation. Low dead space (≤ 0.03 ml). Sterilization: Ethylene Oxide (SAL 10^-6). Standards: ISO 7864:2016, ISO 7886-1:2017, ISO 9626:2016, ISO 10993-1, USP <788>, USP <161>, USP <85>, ASTM D4169-22, ASTM F88/F88-15, ASTM F1929-15, ASTM F1980-21.

## Regulatory Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

## Predicate Devices

- FEELject LDV (Low dead volume) syringe ([K213010](/device/K213010.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 15, 2024

MedExel Co., Ltd. % Peter Chung, President PlusGlobal 300 Atwood Street Pittsburgh, Pennsylvania 15213

Re: K231856

Trade/Device Name: TopFine® LDS (Low Dead Space) Syringe (2 models (TS-LDS2334, TS-LDS2534)) Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: QNQ Dated: February 14, 2024 Received: February 14, 2024

Dear Peter Chung:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

2

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

# Sincerely, Shruti N. Mistry -S

Shruti Mistry Assistant Director, Injection Team Division of Drug Delivery and General Hospital Devices, and Human Factors Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality

Center for Devices and Radiological Health

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K231856

Device Name

TopFine® LDS (Low Dead Space) Syringe (2 models (TS-LDS2334, TS-LDS2534))

Indications for Use (Describe)

TopFine® LDS (Low Dead Space) Syringe is intended for use to inject fluid into or withdraw fluids from the parts of body below the surface of the skin.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K231856- 510(K) Summary

| I. Submitter                                                                                                                     |
|----------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Submitter: MedExel Co., Ltd.<br>252, Geumgwangosan-ro, Geumgwang-myeon<br>Anseong Korea, South<br>Phone: +82-10-9992-9912 |
| Primary Contact: Peter Chung<br>US Agent<br>contact@plusglobal.net                                                               |
| Secondary Contact: Juntaek Seo<br>RA manager<br>jtseo@medexel.co.kr                                                              |
| Date prepared: 15 March, 2024                                                                                                    |

{4}------------------------------------------------

#### II. Device Name

| Device Trade Name   | TopFine® LDS (Low Dead Space)<br>Syringe<br>(2 models (TS-LDS2334, TS-LDS2534)) |
|---------------------|---------------------------------------------------------------------------------|
| Common Name         | Piston syringe                                                                  |
| Classification Name | Low Dead Space Piston Syringe                                                   |
| Regulation Number   | 21 CFR 880.5860                                                                 |
| Product Code        | QNQ                                                                             |

- III. Legally Marketed Predicate Device

| Predicate #          | K213010                                |
|----------------------|----------------------------------------|
| Predicate Trade Name | FEELject LDV (Low dead volume) syringe |

#### Device Description IV.

The TopFine® LDS (Low Dead Space) Syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a hypodermic single lumen needle which is a permanently attached. The device is used to inject fluids into, or withdraw fluids from, the body. It is made of plastic and silicone materials that allows for smooth plunger movement, and is manually operated. This is a single-use device. This product is packed by sterile paper and sterilized by Ethylene Oxide gas.

The proposed device has 2 models depending on the diameter of the needle.

| No. | Model name | Diameter of needle     | Length of needle         |
|-----|------------|------------------------|--------------------------|
| 1   | TS-LDS2334 | 23G (0.600 ~ 0.673 mm) | 26.5 mm (+1.5 / -2.5 mm) |
| 2   | TS-LDS2534 | 25G (0.500~ 0.530 mm)  | 26.5 mm (+1.5 / -2.5 mm) |

The difference between the above 2 models is only the needle diameter.

#### V. Indications for Use

TopFine® LDS (Low Dead Space) Syringe is intended for use to inject fluid into or withdraw fluids from the parts of body below the surface of the skin.

{5}------------------------------------------------

| Item                        | Proposed Device                                                                                                                                                     | Predicate Device                                                                                                                                                    | Remark                                                                                                                         |                         |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| 510(k) No.                  | K231856                                                                                                                                                             | K213010                                                                                                                                                             | -                                                                                                                              |                         |
| Regulation No.              | 21 CFR 880.5860                                                                                                                                                     | 21 CFR 880.5860                                                                                                                                                     | Same                                                                                                                           |                         |
| Product Code                | QNQ                                                                                                                                                                 | QNQ, FMI                                                                                                                                                            | Same                                                                                                                           |                         |
| Indications for Use         | TopFine® LDS (Low Dead Space) Syringe is intended for use to inject fluid into or withdraw fluids from the parts of body below the surface of the skin.             | FEELject LDV (Low dead volume) Syringe is intended for use to inject fluid into or withdraw fluids from the parts of body below the surface of the skin.            | Same                                                                                                                           |                         |
| Type of use                 | Prescription Use                                                                                                                                                    | Prescription Use                                                                                                                                                    | Same                                                                                                                           |                         |
| Mechanism of action         | The plunger of syringe can be pulled and pushed along inside the barrel, allowing the syringe to take in and expel the fluids through the connector to the patient. | The plunger of syringe can be pulled and pushed along inside the barrel, allowing the syringe to take in and expel the fluids through the connector to the patient. | Same                                                                                                                           |                         |
| Number of uses              | Single Use Only                                                                                                                                                     | Single Use Only                                                                                                                                                     | Same                                                                                                                           |                         |
| Configuration and materials | Needle<br>- Needle: SUS 304<br>- Cap: Polypropylene<br>- Barrel: Polypropylene<br>- Plunger: Polypropylene<br>- Gasket: Polyisoprene                                | Needle<br>- Needle: SUS 304<br>- Cap: Polypropylene<br>- Barrel: Polypropylene<br>- Plunger: Polypropylene<br>- Gasket: Polyisoprene                                | Same                                                                                                                           |                         |
|                             | Needle Gauge                                                                                                                                                        | 23G, 25G                                                                                                                                                            | 23G, 25G                                                                                                                       | Same                    |
|                             | Needle dimension – OD                                                                                                                                               | 23G: 0.600 ~ 0.673 mm<br>25G: 0.500 ~ 0.530 mm                                                                                                                      | 23G: 0.600 ~ 0.673 mm<br>25G: 0.500 ~ 0.530 mm                                                                                 | Same                    |
|                             | Needle dimension – needle length                                                                                                                                    | 26.5 mm<br>(+1.5 / -2.5 mm)                                                                                                                                         | 25.0 mm<br>25.4 mm<br>38.1 mm                                                                                                  | Different<br>See note 1 |
|                             | Syringe nozzle type                                                                                                                                                 | Permanently attached                                                                                                                                                | Permanently attached                                                                                                           | Same                    |
|                             | Syringe capacity                                                                                                                                                    | 1 ml                                                                                                                                                                | 1 ml                                                                                                                           | Same                    |
| Dead space                  | Low dead volume ≤ 0.03 ml                                                                                                                                           | Low dead volume ≤ 0.03 ml                                                                                                                                           | Same                                                                                                                           |                         |
| Biocompatibility            | According to ISO 10993-1                                                                                                                                            | According to ISO 10993-1                                                                                                                                            | Same                                                                                                                           |                         |
| Performance Needle          | ISO 7864:2016 Sterile hypodermic needles for single use – Requirements and test methods                                                                             | ISO 7864:2016 Sterile hypodermic needles for single use – Requirements and test methods                                                                             | Same                                                                                                                           |                         |
|                             |                                                                                                                                                                     | ISO 9626:2016 Stainless<br>steel needle tubing for the<br>manufacture of medical<br>devices - Requirements<br>and test methods                                      | ISO 9626:2016 Stainless<br>steel needle tubing for the<br>manufacture of medical<br>devices – Requirements<br>and test methods |                         |
|                             | Syringe                                                                                                                                                             | ISO 7886-1:2017 Sterile<br>hypodermic syringes for<br>single use - Part 1:<br>Syringes for manual use                                                               | ISO 7886-1:2017 Sterile<br>hypodermic syringes for<br>single use - Part 1:<br>Syringes for manual use                          | Same                    |
|                             | Sterility                                                                                                                                                           | Sterilized by EO Gas<br>SAL=10-6                                                                                                                                    | Sterilized by EO Gas<br>SAL=10-6                                                                                               | Same                    |
|                             | Endotoxin limit                                                                                                                                                     | 20 EU/Device                                                                                                                                                        | 20 EU/Device                                                                                                                   | Same                    |
|                             | Labeling                                                                                                                                                            | Complies with 21 CFR<br>part 801                                                                                                                                    | Complies with 21 CFR<br>part 801                                                                                               | Same                    |
|                             | Physical<br>specification                                                                                                                                           | Tip type<br>: Slip Tip<br>Image: Syringe tip                                                                                                                        | Tip type<br>: Slip Tip<br>Image: Syringe tip                                                                                   | Same                    |
|                             | Mechanical<br>specification                                                                                                                                         | Hub/needle bond strength<br>(ISO 7864)<br>1) 23 Gauge: ≥ 34 N<br>2) 25 Gauge: ≥ 22 N                                                                                | Hub/needle bond strength<br>(ISO 7864)<br>1) 23 Gauge: ≥ 34 N<br>2) 25 Gauge: ≥ 22 N                                           | Same                    |

### VI. Comparison of technological characteristics with the predicate device

{6}------------------------------------------------

### Note 1: Needle length

The needle length of proposed device is different from the needle of predicate device. In order to confirm the performance between the different needle specifications, the proposed device was verified by measuring the needle length at each time point of accelerated aging test. The proposed device was found to be substantially equivalent to the predicate device.

{7}------------------------------------------------

### VII. Performance data

### 1. Biocompatibility Testing

In accordance with ISO 10993-1 the subject device is classified as: Externally Communicating Device, Blood Path Indirect, Limited Contact (<24hours). The following testing was conducted.

- 1) Cytotoxicity
- 2) Sensitization
- 3) Irritation
- 4) Acute Systemic Toxicity
- 5) Material Mediated Pyrogenicity
- 6) Hemolysis

Particulate matter testing was conducted in accordance with USP <788> Particulate Matter in Injection and met the USP acceptance criteria.

### 2. Physical Performance Testing

- (1) ISO 7864:2016-Sterile hypodermic needles for single use requirements and test methods
- (2) ISO 7886-1:2017-Sterile hypodermic syringes for single use Part 1: Syringes for manual use
- (3) ISO 9626:2016-Stainless steel needle tubing for the manufacture of medical devices

### 3. Sterility, Shipping and Shelf-Life

The sterilization method is Ethylene Oxide. EO not detected (Limit of Detection: 0001), ECH- Not detected. (Limit of Detection: 0.02), validation method: Overkill Approach (e.g., Half-Cycle method), pyrogen test: Bacterial endotoxin USP <161>Medical devices -Bacterial Endotoxin and pyrogen test, USP <85> Bacterial endotoxins test.

- Package integrity testing, after environmental conditioning and simulated transportation using ASTM D4169-22, was conducted on final packaged, and sterile devices. All packaging is acceptable for protection of final finished product and sterility maintenance.
- Sterile Barrier Packaging Testing performed on the proposed device:
	- Seal peel test ASTM F88/F88 -15 O
	- Dye migration- ASTM F1929-15 o
- Shelf life of 3 years is validated using the FDA recognized standard ASTM F1980-21 Standard Guide for Accelerated Aging of Sterile Barrier System for Medical Devices.

### 4. Clinical Studies

Clinical testing was not required to support substantial equivalence.

{8}------------------------------------------------

## VIII. Conclusion

The differences between the predicate device and the subject device do not raise any new or different questions of safety or effectiveness. The TopFine® LDS (Low Dead Space) Syringe (2 models (TS-LDS2334, TS-LDS2534) is substantially equivalent to the FEELject LDV (low dead volume) syringe with respect to the indications for use, target populations, treatment method, and technological characteristics.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/QNQ/K231856](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/QNQ/K231856)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
