Midline Catheter

K252402 · Terumo Medical Products (Hangzhou) Co., Ltd. · PND · Dec 18, 2025 · General Hospital

Device Facts

Record IDK252402
Device NameMidline Catheter
ApplicantTerumo Medical Products (Hangzhou) Co., Ltd.
Product CodePND · General Hospital
Decision DateDec 18, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5200
Device ClassClass 2

Indications for Use

The Midline Catheter is indicated for insertion into a patient's vascular system for short-term use (<30 days) to sample blood or administer fluids intravenously. It may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of the procedure. The Midline Catheter is suitable for use with power injectors.

Device Story

Midline Catheter is a peripherally inserted, radiopaque polyurethane catheter used for short-term (<30 days) intravenous fluid administration or blood sampling. Device includes catheter assembly, slider, hub, and 3D-Shin needle with passive safety mechanism to prevent needlestick injuries. Operated manually by clinicians in clinical settings. After blood return, needle is withdrawn and safety shield covers tip. Compatible with power injectors. Benefits include reliable vascular access and reduced needlestick risk. Device is MR Conditional.

Clinical Evidence

No clinical data included. Substantial equivalence supported by bench testing, including safety mechanism activation, tensile strength, tip flexural fatigue, burst pressure, MRI compatibility (ASTM F2052, F2213, F2119), and biocompatibility (ISO 10993).

Technological Characteristics

Radiopaque polyurethane catheter; 18G-22G gauges; 80-100mm lengths. Components: stainless steel needle/rivet, polycarbonate/polypropylene hubs/sliders, isoprene valve. Manual operation. EO sterilized (ISO 11135, SAL 10^-6). MR Conditional. Power injection compatible (up to 7mL/sec). Complies with ISO 10555-1.

Indications for Use

Indicated for insertion into vascular system for short-term (<30 days) blood sampling or IV fluid administration. Suitable for any patient population, contingent on vascular anatomy and procedure appropriateness.

Regulatory Classification

Identification

An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION December 18, 2025 Terumo Medical Products (Hangzhou) Co., Ltd. Kimberly Jacob-Paredes Regulatory Affairs Specialist M4-9-5, Hangzhou Economic &amp; Technological Development Zone Hangzhou, 310018 China Re: K252402 Trade/Device Name: Midline Catheter Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: Class II Product Code: FOZ Dated: November 19, 2025 Received: November 19, 2025 Dear Kimberly Jacob-Paredes: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1} K252402 - Kimberly Jacob-Paredes Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K252402 - Kimberly Jacob-Paredes Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Colleen J. Lawrimore -S Colleen Lawrimore, Ph.D. *For David Wolloscheck, Ph.D.* Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K252402 | | | Device Name Midline Catheter | | | Indications for Use (Describe) The Midline Catheter is indicated for insertion into a patient's vascular system for short-term use (<30 days) to sample blood or administer fluids intravenously. It may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of the procedure. The Midline Catheter is suitable for use with power injectors. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} TERUMO TERUMO MEDICAL Products (Hangzhou) Co., Ltd. M4-9-5, Hangzhou Economic &amp; Technological Development Zone Hangzhou, Zhejiang, 310018 China TELEPHONE: (86) 571-8731-8100 FAX: (86) 571-8691-0293 Midline Catheter 510(k) Summary # 510(K) SUMMARY – K252402 # SUBMITTER INFORMATION ## Contact Details: Owner/Operator Terumo Medical Products (Hangzhou) Co., Ltd. Medical M4-9-5, Hangzhou Economic &amp; Technological Development Zone Hangzhou, Zhejiang China 310018 Registration Number: 3004102031 ## Contact Person: Kimberly Jacob-Paredes Regulatory Affairs Specialist Terumo Medical Corporation 265 Davidson Ave, Suite 320 Somerset, NJ 08873 Tel. (732) 995-6675 Fax (410) 398-6079 E-mail: kimberly.jacob@terumo.com Date prepared: December 18th, 2025 ## SUBJECT DEVICE Trade Name: Midline Catheter Common Name: Catheter, intravascular, therapeutic, short-term less than 30 days Classification Name: Intravascular Catheter Classification Panel: General Hospital Regulation Number: 21 CFR 880.5200 Product Code: FOZ Device Class: Class II Terumo Medical Corporation, USA {5} TERUMO TERUMO MEDICAL Products (Hangzhou) Co., Ltd. M4-9-5, Hangzhou Economic &amp; Technological Development Zone Hangzhou, Zhejiang, 310018 China TELEPHONE: (86) 571-8731-8100 FAX: (86) 571-8691-0293 Midline Catheter 510(k) Summary ## PREDICATE DEVICE The legally marketed device to which substantial equivalence is claimed is: Predicate Device: K162377-PowerGlide Pro™ Midline Catheter Manufacturer: C. R. Bard, Inc. Regulation: 21 CFR 880.5200 Product Code: FOZ Classification: Class II ## DEVICE DESCRIPTION The Midline Catheter is a peripherally inserted catheter composed of a radiopaque polyurethane catheter, a catheter slider, a catheter hub, needle hub, and 3D-Shin™ needle featuring an engineered three-dimensional geometry of the needle tip, with a passive safety mechanism. After observing the blood return, the 3D-Shin™ needle is withdrawn, and the needle tip is fully covered by the safety mechanism, reducing the risk of accidental needlestick injury. ## Design/Construction The subject device and the predicate device PowerGlide Pro™ Midline Catheter (K162377) have similar construction, except for the Guidewire component and the Tip configuration. Terumo Medical Corporation, USA {6} TERUMO TERUMO MEDICAL Products (Hangzhou) Co., Ltd. M4-9-5, Hangzhou Economic &amp; Technological Development Zone Hangzhou, Zhejiang, 310018 China TELEPHONE: (86) 571-8731-8100 FAX: (86) 571-8691-0293 Midline Catheter 510(k) Summary ## Materials The Materials for the subject device are provided in Table 1. Table 1: List of Materials | No. | Name of Component | Raw Material | | --- | --- | --- | | 1** | Catheter | Polyurethane | | 2** | Catheter Tip | Polyurethane | | 3* | Inner Needle | Stainless steel | | 4* | Rivet pin | Stainless steel | | 5* | Hemostasis valve | Isoprene rubber | | 6 | Catheter protector | Polypropylene | | 7 | Grip | Polycarbonate | | 8 | Needle hub | Polycarbonate | | 9* | Catheter hub | Polypropylene | | 10 | Catheter slider | Polypropylene | | 11* | Plug | Polycarbonate | | 12* | Plug Holder | Polypropylene | | 13 | Safety shield | Polycarbonate | * Blood Path, Indirect, **Circulating Blood Terumo Medical Corporation, USA {7} TERUMO TERUMO MEDICAL Products (Hangzhou) Co., Ltd. M4-9-5, Hangzhou Economic &amp; Technological Development Zone Hangzhou, Zhejiang, 310018 China TELEPHONE: (86) 571-8731-8100 FAX: (86) 571-8691-0293 Midline Catheter 510(k) Summary # Specifications The specifications for the subject device are provided in Table 2. Table 2: Midline Catheter Specifications | Product code | Lumen | Needle type | Catheter gauge | Catheter length | Safety cover | Color of catheter hub | | --- | --- | --- | --- | --- | --- | --- | | MCS1810D00F | Single lumen | 3D needle | 18G | 100mm | Type 1 | Green | | MCS1808D00F | | | 18G | 80mm | Type 2 | | | MCS2010D00F | | | 20G | 100mm | Type 1 | Pink | | MCS2008D00F | | | 20G | 80mm | Type 2 | | | MCS2208D00F | | | 22G | 80mm | Type 2 | Blue | # INDICATIONS FOR USE The Midline Catheter is indicated for insertion into a patient's vascular system for short-term use (&lt;30 days) to sample blood or administer fluids intravenously. It may be used for any patient population, with consideration given to adequacy of vascular anatomy and appropriateness of the procedure. The Midline Catheter is suitable for use with power injectors. # TECHNOLOGICAL CHARACTERISTICS A comparison of the technological characteristics is summarized in Table 3 below. Terumo Medical Corporation, USA {8} TERUMO TERUMO MEDICAL Products (Hangzhou) Co., Ltd. M4-9-5, Hangzhou Economic &amp; Technological Development Zone Hangzhou, Zhejiang, 310018 China TELEPHONE: (86) 571-8731-8100 FAX: (86) 571-8691-0293 Midline Catheter 510(k) Summary Table 3: Comparison of Technological Characteristics | Device Characteristic | Subject Device Specifications: Midline Catheter | Predicate Device Specifications: PowerGlide Pro™ Midline Catheter, K162377 | Same/Different Assessment | | --- | --- | --- | --- | | Indications for Use | The Midline Catheter is indicated for insertion into a patient's vascular system for short-term use (<30 days) to sample blood or administer fluids intravenously. It may be used for any patient population, with consideration given to the adequacy of vascular anatomy and appropriateness of the procedure. The Midline Catheter is suitable for use with power injectors | The PowerGlide Pro™ Midline Catheter is inserted into a patient's vascular system for short-term use (<30 days) to sample blood or administer fluids intravenously. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of the procedure. The PowerGlide Pro™ Midline Catheter is suitable for use with power injectors. | Same | | Operation Principle | Manual | Same | Same | | Design / Construction | • Catheter assembly (catheter, catheter slide, catheter hub, inner needle, needle hub) In accordance with ISO 10555-1 | • Catheter assembly (catheter and hub, introducer needle, needle hub, and Guidewire) In accordance with ISO 10555-1 | The Midline Catheter is not intended for use with a guidewire. To address this concern, Terumo conducted a Tip Flexural Fatigue test to simulate cyclic kinking | Terumo Medical Corporation, USA {9} TERUMO TERUMO MEDICAL Products (Hangzhou) Co., Ltd. M4-9-5, Hangzhou Economic &amp; Technological Development Zone Hangzhou, Zhejiang, 310018 China TELEPHONE: (86) 571-8731-8100 FAX: (86) 571-8691-0293 Midline Catheter 510(k) Summary | Device Characteristic | Subject Device Specifications: Midline Catheter | Predicate Device Specifications: PowerGlide Pro ™ Midline Catheter, K162377 | Same/Different Assessment | | --- | --- | --- | --- | | | | | and evaluate burst pressure performance. In this test, the catheter tip underwent 30 cycles of bending and releasing, followed by a burst pressure test in accordance with ISO 10555-1:2023. The results demonstrated that the product remained intact throughout the testing process. Based on these findings, Terumo believes that the Midline Catheter does not introduce any new safety or effectiveness concerns, even when used without a guidewire. | | Materials | • **Catheter – Polyurethane • *Catheter Hub – Polypropylene • *Inner Needle – Stainless Steel • **Catheter Tip – Polyurethane • Catheter Slider - | • Catheter – Polyurethane • Catheter Hub – Isoplast Polyurethane • Needle – Stainless Steel • Catheter Tip – Isoplast Polyurethane Guidewire - Nitinol | Although the materials differ, Terumo has conducted chemical characterization and biocompatibility tests on the Midline Catheter, and the results confirm that the materials do not raise different questions | Terumo Medical Corporation, USA {10} TERUMO TERUMO MEDICAL Products (Hangzhou) Co., Ltd. M4-9-5, Hangzhou Economic &amp; Technological Development Zone Hangzhou, Zhejiang, 310018 China TELEPHONE: (86) 571-8731-8100 FAX: (86) 571-8691-0293 Midline Catheter 510(k) Summary | Device Characteristic | Subject Device Specifications: Midline Catheter | Predicate Device Specifications: PowerGlide Pro ™ Midline Catheter, K162377 | Same/Different Assessment | | --- | --- | --- | --- | | | Polypropylene | | of safety or effectiveness compared to the predicate. | | Materials | • Guidewire - Not Applicable • Needle Hub - Polycarbonate • *Rivet Pin - Stainless Steel • *Hemostasis valve - Polyisoprene • Catheter Protector - Polypropylene • Grip - Polycarbonate • Catheter Slider - Polypropylene • *Plug - Polycarbonate • *Plug Holder - Polypropylene • Safety Shield - Polycarbonate | • Guidewire Nitinol • Needle Hub - Information not publicly known • Rivet Pin - Not Applicable • Hemostasis Valve - Not Applicable • Catheter Protector - Information not publicly known • Grip - Information not publicly known • Catheter Slider - Information not publicly known • Plug - Not Applicable • Plug holder - Not Applicable • *Safety Shield - Information not publicly known | Rivet Pin: The rivet pin serves as a mechanical function by securely connecting the catheter to the catheter hub. This connection is critical to prevent the catheter from detaching during use. Risks associated with the rivet pin have been addressed through biocompatibility testing and MRI safety evaluation. Additionally, per ISO 10555-1, tensile strength testing was conducted on the catheter and the catheter hub connection for both the Midline Catheter and the predicate device. Results confirmed equivalent | Terumo Medical Corporation, USA {11} TERUMO TERUMO MEDICAL Products (Hangzhou) Co., Ltd. M4-9-5, Hangzhou Economic &amp; Technological Development Zone Hangzhou, Zhejiang, 310018 China TELEPHONE: (86) 571-8731-8100 FAX: (86) 571-8691-0293 Midline Catheter 510(k) Summary | Device Characteristic | Subject Device Specifications: Midline Catheter | Predicate Device Specifications: PowerGlide Pro ™ Midline Catheter, K162377 | Same/Different Assessment | | --- | --- | --- | --- | | | | | mechanical performance. Hemostasis Valve and Plug/Plug Holder: The hemostasis valve is designed to prevent blood leakage during catheter insertion. When a male Luer is connected, it pushes the plug forward, which in turn opens the valve to establish a fluid path. The plug and plug holder are components that support this function. Biocompatibility testing has been completed for all materials, and no new risks were identified. The valve and plug do not introduce new safety or effectiveness concerns compared to the predicate. | Terumo Medical Corporation, USA {12} TERUMO TERUMO MEDICAL Products (Hangzhou) Co., Ltd. M4-9-5, Hangzhou Economic &amp; Technological Development Zone Hangzhou, Zhejiang, 310018 China TELEPHONE: (86) 571-8731-8100 FAX: (86) 571-8691-0293 Midline Catheter 510(k) Summary | Device Characteristic | Subject Device Specifications: Midline Catheter | Predicate Device Specifications: PowerGlide Pro ™ Midline Catheter, K162377 | Same/Different Assessment | | --- | --- | --- | --- | | | | | Needle and Needle Hub: Tensile strength testing was conducted per in-house standards based on the predicate's performance. Testing was conducted on the needle and the needle hub for both the Midline Catheter and the predicate device. Results demonstrated equivalent mechanical performance. Catheter Slider and Grip: Slider resistance testing was conducted per in-house standards based on the predicate's performance. The catheter slider and grip are integrated components on both the Midline catheter and the predicate device. | Terumo Medical Corporation, USA {13} TERUMO TERUMO MEDICAL Products (Hangzhou) Co., Ltd. M4-9-5, Hangzhou Economic &amp; Technological Development Zone Hangzhou, Zhejiang, 310018 China TELEPHONE: (86) 571-8731-8100 FAX: (86) 571-8691-0293 Midline Catheter 510(k) Summary | Device Characteristic | Subject Device Specifications: Midline Catheter | Predicate Device Specifications: PowerGlide Pro ™ Midline Catheter, K162377 | Same/Different Assessment | | --- | --- | --- | --- | | | | | Protector Fitting Strength: Fitting strength testing of the protector was conducted per in-house standards based on the predicate's performance. Results confirmed equivalent retention and reliability. Safety Shield Performance: Per ISO 23908, the Safety shield performance was validated for both the Midline Catheter and the predicate device. The test results demonstrated equivalent protection against needlestick injury. | | Package Specifications | • Unit Box • Shelf box • Shipping carton • Patient Label | • Unit Box • Shelf box • Shipping carton • Patient Label | Same | Terumo Medical Corporation, USA {14} TERUMO TERUMO MEDICAL Products (Hangzhou) Co., Ltd. M4-9-5, Hangzhou Economic &amp; Technological Development Zone Hangzhou, Zhejiang, 310018 China TELEPHONE: (86) 571-8731-8100 FAX: (86) 571-8691-0293 Midline Catheter 510(k) Summary | Device Characteristic | Subject Device Specifications: Midline Catheter | Predicate Device Specifications: PowerGlide Pro ™ Midline Catheter, K162377 | Same/Different Assessment | | --- | --- | --- | --- | | | 18G, 20G, 22G | 18G, 20G, 22G | Same | | Sterilization | EO (validated in accordance with ISO 11135 to achieve SAL 10^{-6}) | Provided Sterile (EO) | Same | | Shelf life | 3 Years | Not Publicly Available | Although the predicate device does not publicly disclose its shelf life duration, the proposed 3 year shelf life for the Midline Catheter does not raise new questions of safety or effectiveness based on testing conducted in accordance with ASTM F1980 and ISO 10993. These studies confirmed that the device maintains its physical integrity, sterility, and functional performance throughout the proposed shelf life. | Terumo Medical Corporation, USA {15} TERUMO TERUMO MEDICAL Products (Hangzhou) Co., Ltd. M4-9-5, Hangzhou Economic &amp; Technological Development Zone Hangzhou, Zhejiang, 310018 China TELEPHONE: (86) 571-8731-8100 FAX: (86) 571-8691-0293 Midline Catheter 510(k) Summary | Device Characteristic | Subject Device Specifications: Midline Catheter | Predicate Device Specifications: PowerGlide Pro ™ Midline Catheter, K162377 | Same/Different Assessment | | --- | --- | --- | --- | | MRI Safety | MR Conditional | Not Applicable | Although the predicate device does not claim MRI compatibility, the Midline Catheter has the MR conditional symbol on the label, which is supported by a comprehensive MR safety test in accordance with ASTM F2052-21, ASTM F2213-2017, and ASTM F2119-07(2013). These tests confirmed that the device meets the criteria for MR Conditional. | | Power Injection Maximum Flow Rate | 18G = 7mL/sec 20G = 5mL/sec 22G = 2mL/sec | 18G = 7mL/sec 20G = 5mL/sec 22G = 2mL/sec | Same | * Blood Path, Indirect, **Circulating Blood** Terumo Medical Corporation, USA {16} TERUMO TERUMO MEDICAL Products (Hangzhou) Co., Ltd. M4-9-5, Hangzhou Economic &amp; Technological Development Zone Hangzhou, Zhejiang, 310018 China TELEPHONE: (86) 571-8731-8100 FAX: (86) 571-8691-0293 Midline Catheter 510(k) Summary # NON-CLINICAL TESTS ## Performance Performance testing was conducted to verify the safety and effectiveness of the Midline Catheter. This device is intended for short-term use (less than 30 days) for blood sampling or intravenous fluid administration. It is suitable for all patient populations, with consideration given to the adequacy of vascular anatomy and appropriateness of the procedure. The Midline catheter is compatible with power injectors throughout its shelf life. Testing was performed to confirm compliance with relevant internal and external standards, demonstrate MR Conditional status, and establish substantial equivalence to the predicate device. Performance tests were carried out on both non-aged and accelerated-aged samples. Table 4 below outlines the specific performance tests conducted on the Midline Catheter. Table 4: Summary of Performance Tests | Test Item | | --- | | Activation of Safety Mechanism | | Safety Cover Tensile Strength | | Needle Tip Exposure in Safe Mode | | Catheter Tip Adhesion Break Force | | Swelling Outer Diameter (OD) | | Catheter Collapse | | Tip Location During and After Power Injection | | Tip Flexural Fatigue (Cyclic Kinking and Burst Test) | | Tip Separation Force | | Simulated Clinical Use | | MRI Compatibility Test | | Radiopacity Test | | Particle Contamination | Terumo Medical Corporation, USA {17} TERUMO TERUMO MEDICAL Products (Hangzhou) Co., Ltd. M4-9-5, Hangzhou Economic &amp; Technological Development Zone Hangzhou, Zhejiang, 310018 China TELEPHONE: (86) 571-8731-8100 FAX: (86) 571-8691-0293 Midline Catheter 510(k) Summary # MRI Compatibility Testing MRI compatibility testing was conducted on the finished Midline Catheter, excluding the needle component. To remove the needle, the device's safety mechanism must be activated; therefore, only the catheter and catheter hub were included in the evaluation. Testing was performed in accordance with the recommendations outlined in the FDA Guidance Document for Industry and FDA Staff, "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment." The objective was to determine MR Conditional safety information for the device. Table 5 below summarizes the MRI tests performed on the Midline Catheter. Table 5: Summary of MRI Tests Performed | Test Items | Applicable Standards | | --- | --- | | Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment | ASTM F2052-21 | | Standard Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment | ASTM F 2213-2017 | | Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants | ASTM F2119-07(2013) | Performance testing for $t = 0$ and $t = 36AA$ (3-year accelerated age) and the MRI Compatibility testing confirmed that all predetermined acceptance criteria were met. The results support that the Midline Catheter is suitable for clinical use throughout its labeled shelf life and meets the criteria for MR Conditional. Terumo Medical Corporation, USA {18} TERUMO TERUMO MEDICAL Products (Hangzhou) Co., Ltd. M4-9-5, Hangzhou Economic &amp; Technological Development Zone Hangzhou, Zhejiang, 310018 China TELEPHONE: (86) 571-8731-8100 FAX: (86) 571-8691-0293 Midline Catheter 510(k) Summary ## Biocompatibility Classification In accordance with ISO 10993-1, the Midline Catheter is classified as an Externally Communicating Device, Circulating Blood, Prolonged Exposure (24 hours to 30 days). ## Chemical Characterization Chemical Characterization was conducted on the finished sterilized device in accordance with ISO 10993-18: 2020. Testing results demonstrate that no extractables at concentrations higher than the toxicological screening limit (TSL) were detected. ## Biocompatibility Testing The finished device's blood/body contacting parts were tested in accordance with the tests recommended in the FDA Guidance: Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a risk management process". Screening tests were performed on non-aged and accelerated-aged devices to show that biocompatibility is maintained throughout the shelf life of the product. Table 6 below provides a list of biocompatibility testing conducted on the Midline Catheter. Terumo Medical Corporation, USA {19} TERUMO TERUMO MEDICAL Products (Hangzhou) Co., Ltd. M4-9-5, Hangzhou Economic &amp; Technological Development Zone Hangzhou, Zhejiang, 310018 China TELEPHONE: (86) 571-8731-8100 FAX: (86) 571-8691-0293 Midline Catheter 510(k) Summary Table 6: Biocompatibility Tests | Non-aged, Sterile, Whole Device | Applicable Standard | | --- | --- | | Genotoxicity | ISO 10993-3:2014 | | Implantation | ISO 10993-6:2016 | | Hemocompatibility | ISO 10993-4:2017 | | Cytotoxicity | ISO 10993-5:2009 | | Sensitization | ISO 10993-10:2021 | | Irritation or Intracutaneous Reactivity | ISO 10993-23:2021 | | Acute System Toxicity | ISO 10993-11:2017 | | Subacute Toxicity | ISO 10993-11:2017 | | Complement Activation | ISO 10993-4:2017 | | Pyrogenicity | USP, General Chapter <151> | | Accelerated age (36-months (3 years)), Sterile, Whole Device | Applicable Standard | | Cytotoxicity | ISO 10993-5:2009 | | Irritation or Intracutaneous Reactivity | ISO 10993-23:2021 | | Acute System Toxicity | ISO 10993-11:2017 | The testing results demonstrate that the device is biocompatible throughout the product's shelf life. ## Sterilization The sterility of the device is assured using a sterilization method validated in accordance with ISO 11135:2014, Sterilization of health-care products — Ethylene oxide — Requirements for the development, validation, and routine control of a sterilization process for medical devices. The Midline Catheter is sterilized to provide a Sterility Assurance Level (SAL) of $10^{-6}$. Terumo Medical Corporation, USA {20} TERUMO TERUMO MEDICAL Products (Hangzhou) Co., Ltd. M4-9-5, Hangzhou Economic &amp; Technological Development Zone Hangzhou, Zhejiang, 310018 China TELEPHONE: (86) 571-8731-8100 FAX: (86) 571-8691-0293 Midline Catheter 510(k) Summary ## CLINICAL TESTS This 510(k) does not include data from clinical tests. ## CONCLUSION In summary, the Midline Catheter, which is the subject of this Traditional 510(k) submission, is substantially equivalent in its intended use, technological characteristics, materials, and performance to the PowerGlide Pro™ Midline Catheter (K162377), manufactured by C. R. Bard, Inc. Terumo Medical Corporation, USA
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...