← Product Code [ONB](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/ONB) · K221513

# Equashield Closed System Transfer Device (K221513)

_Equashield Medical , Ltd. · ONB · Feb 17, 2023 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/ONB/K221513

## Device Facts

- **Applicant:** Equashield Medical , Ltd.
- **Product Code:** [ONB](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/ONB.md)
- **Decision Date:** Feb 17, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5440
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

Closed System Drug Transfer Device (CSTD) for safe preparation, compounding and administration of drugs, including antineoplastic and hazardous drugs. This closed system mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills and also prevents microbial ingress up to 7 days. The system's closed Syringe Unit prevents intended syringe plunger detachment and can be used safely up to its maximal nominal volume with hazardous drugs

## Device Story

EQUASHIELD CSTD is a multi-component, sterile, single-use system for hazardous drug handling; includes syringe units, vial adaptors, spike adaptors, and luer lock connectors. System uses integrated, variable sterile air chamber for pressure equalization; double-membrane method ensures leak-proof, drug-residual-free connections. Syringe unit is fully encapsulated with airtight seal at rear to isolate plunger rod. Used by pharmacists/clinicians in hospitals/clinics to prepare and administer drugs. Mechanical design prevents environmental contamination, drug vapor/aerosol escape, and microbial ingress for up to 7 days. Prevents unintended syringe plunger detachment. Benefits include reduced occupational exposure to hazardous drugs and maintenance of drug sterility.

## Clinical Evidence

Bench testing only. Testing included ISO 8536-4/10, ISO 80369-7, ISO 1135-4, and USP <788>/<790> for particulate matter. Biocompatibility testing (ISO 10993-1) confirmed safety for prolonged contact. Microbial ingress study verified prevention of contamination over 7 days with 10 access cycles. Drug compatibility testing confirmed stability with aggressive hazardous drugs.

## Technological Characteristics

Materials: Polypropylene, PVC, Silicone, 304 SST, POM, ABS, Polyisoprene, Hydrophobic filter. Principle: Multi-component closed system with integrated pressure equalization chamber and double-membrane resealing septa. Sterilization: Ethylene Oxide (EtO) to SAL 10^-6 (plugs Gamma). Connectivity: None (mechanical system).

## Regulatory Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

## Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

## Predicate Devices

- EQUASHIELD® Closed System Transfer Device ([K170706](/device/K170706.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

February 17, 2023

EQUASHIELD Medical Ltd. % Bosmat Friedman Regulatory Consultant ProMedoss, Inc. 3521 Hatwynn Rd. Charlotte, North Carolina 28269

Re: K221513

Trade/Device Name: EQUASHIELD® Closed System Transfer Device Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: ONB Dated: January 16, 2023 Received: January 17, 2023

Dear Bosmat Friedman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Danil Walloscher

David Wolloscheck, PhD Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K221513

#### Device Name

EQUASHIELD® Closed System Transfer Device

#### Indications for Use (Describe)

Closed System Drug Transfer Device (CSTD) for safe preparation, compounding and administration of drugs, including antineoplastic and hazardous drugs. This closed system mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills and also prevents microbial ingress up to 7 days. The system's closed Syringe Unit prevents intended syringe plunger detachment and can be used safely up to its maximal nominal volume with hazardous drugs

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:20px">☑</span> |
|----------------------------------------------|---------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C)  | <span style="font-size:20px">☐</span> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K221513 - 510(k) Summary

### SUBMITTER

Applicant's Name: EQUASHIELD Medical Ltd. Ktalav St.1. POB 12 Migdal Tefen, 2495900, Israel Tel: +972 4 9873737, Fax: +97 4 9873001

## Contact Person:

Bosmat Friedman Regulatory Affairs Consultant 3521 Hatwynn Rd. Charlotte, NC 28269 Phone: 647-975-3974 bosmat.f@promedoss.com

Date Prepared: February 17, 2023

DEVICE

Trade Name: EQUASHIELD® Closed System Transfer Device

Common Name: Closed Antineoplastic and Hazardous Drug Reconstitution And Transfer System Product Code: ONB Regulation Name: Intravascular administration set Regulation No: 880.5440 Class: 2 Review Panel: General Hospital

## PREDICATE DEVICE

Predicate device:

- EQUASHIELD® Closed System Transfer Device, manufactured by . EQUASHIELD Medical Ltd., cleared under K170706; Product Code: ONB.
## DEVICE DESCRIPTION

The EQUASHIELD® Closed System Drug Transfer Device (CSTD) is intended for safe preparation, reconstitution, compounding and administration of drugs, including antineoplastic and hazardous drugs. This closed system mechanically prohibits the

{4}------------------------------------------------

transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills and also prevents microbial ingress up to 7 days. The system's closed Syringe Unit prevents intended and unintended syringe plunger detachment and can be used safely up to its maximal nominal volume with hazardous drugs.

EQUASHIELD® CSTD is a sterile, single use system. The various system components are listed in the table below:

| Name                                    | Description                                                             |
|-----------------------------------------|-------------------------------------------------------------------------|
| Syringe Unit                            | Syringe for drug transfer                                               |
| Vial Adaptor                            | Adaptor to the drug vial                                                |
| Spike Adaptor & Spike Adaptor<br>180    | Adaptor for the IV bag for injection and for infusion<br>administration |
| Spike Adaptor W & Spike Adaptor<br>W180 | Adaptor for the IV bag for withdrawal                                   |
| Luer Lock Adaptor 1                     | Adaptor for injection into IV lines                                     |
| Luer Lock Adaptor 2                     | Adaptor for injection and withdrawal for IV lines                       |
| Female Luer Lock connector              | Connector for standard IV tubing set ports                              |
| Protective Plug                         | A plug to protect connectors during transportation                      |
| Tubing sets                             | Accessory for injection into an IV line                                 |
| Reconstitution Set Accessory            | Accessory for reconstituting powdered drugs                             |
| Luer Lock Adaptor 1C                    | Adaptor for injection into catheters                                    |
| Luer Lock Adaptor 1DC                   | Adaptor for medication transfer between<br>EQUASHIELD® Syringe Units    |
| Male Priming Connector                  | Connector for priming of IV line                                        |

The above components are combined to create a system and are not intended to be used individually.

The variable sterile air chamber integrated into the encapsulated syringe provides selfcontained pressure equalization. The connector unit is welded to the syringe and uses the double-membrane method as high efficiency microbial barrier and for leak-proof and drug residual-free connections to the adaptors of the system. The double membrane prevents the transfer of environmental contaminants into the system and/or escape of drug or vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, spills and also prevents microbial ingress up to 7 days. The purpose of this submission is to obtain FDA clearance for new system components and material changes for previously cleared components and to revise the indication for use for clarity with respect to the functionality of the Syringe Unit.

{5}------------------------------------------------

| Characteristic                              | Subject Device<br>Equashield Closed System Drug<br>Transfer Device<br>K221513                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Predicate Device<br>Equashield Closed System Drug<br>Transfer Device<br>K170706                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Comments                                                                                                                                                         |
|---------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use                          | Closed System Drug Transfer Device<br>(CSTD) for safe preparation,<br>reconstitution, compounding and<br>administration of drugs, including<br>antineoplastic and hazardous drugs.<br>This closed system mechanically<br>prohibits the transfer of environmental<br>contaminants into the system and the<br>escape of drug or vapor concentrations<br>outside the system, thereby<br>minimizing individual and<br>environmental exposure to drug vapor,<br>aerosols, and spills and also prevents<br>microbial ingress up to 7 days. The<br>system's closed Syringe Unit prevents<br>intended and unintended syringe<br>plunger detachment and can be used<br>safely up to its maximal nominal<br>volume with hazardous drugs | Closed System drug Transfer Device<br>(CSTD) for preparation,<br>reconstitution, compounding and<br>administration of drugs, including<br>antineoplastic and hazardous drugs.<br>This closed system mechanically<br>prohibits the transfer of<br>environmental contaminants into the<br>system and the escape of drug or<br>vapor concentrations outside the<br>system, thereby minimizing<br>individual and environmental<br>exposure to drug vapor, aerosols,<br>and spills and also prevents<br>microbial ingress up to 7 days. | Different. Please see<br>discussions of<br>differences in<br>Indications for Use<br>statement, below the<br>Summary of<br>Technological<br>Characteristics table |
| Prescription (Rx)<br>or Over the<br>Counter | Rx                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Rx                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                             |

# INDICATIONS FOR USE

# SUMMARY OF TECHNOLOGICAL CHARACTERISTICS

| Characteristic                    | CSTD System<br>Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | CSTD System<br>Predicate device (K170706)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Comments                              |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|
| System<br>components              | - Vial Adaptors<br>- Syringe Unit<br>- Spike Adaptors for injection<br>- Spike Adaptor for withdrawal<br>- Luer Lock Adaptors<br>- Female Luer Lock connectors<br>- Luer Lock Adaptor DC<br>- Luer Lock Adaptor C<br>- Protective Plug<br>- Y-Site Tubing (Accessory)<br>- Secondary Tubing (Accessory)<br>- Male Priming Connector                                                                                                                                                                                                                                                                                                                                                                                                                                                          | - Vial Adaptor<br>- Syringe Unit<br>- Spike Adaptor for injection<br>- Spike Adaptor W- for withdrawal<br>- Luer Lock Adaptor<br>- Male Luer Lock Connector<br>- Female Luer Lock Connector<br>- Protective Plug<br>- Y-Site Tubing (Accessory)<br>- Secondary Tubing (Accessory)<br>- Reconstitution Tubing (Accessory)<br>- Catheter Luer Lock Adaptor<br>- Syringe-Syringe Luer Lock Adaptor                                                                                                                                                                                                                                                                                                                                                                                              | Different. Please see<br>comment # 1  |
| Characteristic                    | CSTD System<br>Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | CSTD System<br>Predicate device (K170706)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Comments                              |
| Characteristics                   | Closed System used for<br>antineoplastic and hazardous drug<br>reconstitution, transfer and<br>administration, in order to prevent<br>contamination of the surrounding<br>environment and of the drug                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Closed System used for<br>antineoplastic and hazardous drug<br>reconstitution, transfer and<br>administration, in order to prevent<br>contamination of the surrounding<br>environment and of the drug                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                  |
| Principles of<br>Operation        | Multi-component system.<br>Components are intended to be<br>used as a system.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Multi-component system.<br>Components are intended to be<br>used as a system.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                  |
| Technological<br>Characteristics  | A leak-proof connector with a<br>single use syringe permanently<br>attached to it as part of the system.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | A leak-proof connector with a<br>single use syringe permanently<br>attached to it as part of the system.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                  |
|                                   | All system components are sealed<br>with resealing membranes<br>(Septum). When components are<br>joined together the two membranes<br>are pressed together and then<br>pierced by needles. System has<br>integrated closed pressure<br>equalization.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | All system components are sealed<br>with resealing membranes<br>(Septum). When components are<br>joined together the two membranes<br>are pressed together and then<br>pierced by needles. System has<br>integrated closed pressure<br>equalization.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                       |
|                                   | The system syringe is closed from<br>all sides. The syringe barrel is<br>sealed airtight also at its rear end,<br>thereby isolating the plunger rod<br>and the interior of the barrel.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | The system syringe is closed from<br>all sides. The syringe barrel is<br>sealed airtight also at its rear end,<br>thereby isolating the plunger rod<br>and the interior of the barrel.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                       |
| System                            | A fully encapsulated Syringe Unit<br>is the active transfer device of this<br>closed system.<br>The syringe barrel is sealed airtight<br>also at its rear end, thereby<br>isolating the plunger rod and the<br>interior of the barrel. A leak-proof<br>connector is permanently welded<br>to the syringe.<br>The closed pressure equalization<br>system with a chamber containing<br>sterile air is built-in the Syringe<br>Unit and makes the system airtight<br>consequently containing all<br>aerosols, particles and vapors.<br>For transfer of fluids the Syringe<br>Unit connects to the passive Vial<br>Adaptors, infusion bag and<br>infusion tubing adaptors of the<br>system, using the double<br>membrane method to create a leak-<br>proof and drug residual-free<br>connection. | A fully encapsulated Syringe Unit<br>is the active transfer device of this<br>closed system.<br>The syringe barrel is sealed airtight<br>also at its rear end, thereby<br>isolating the plunger rod and the<br>interior of the barrel. A leak-proof<br>connector is permanently welded<br>to the syringe.<br>The closed pressure equalization<br>system with a chamber containing<br>sterile air is built-in the Syringe<br>Unit and makes the system airtight<br>consequently containing all<br>aerosols, particles and vapors.<br>For transfer of fluids the Syringe<br>Unit connects to the passive Vial<br>Adaptors, infusion bag and<br>infusion tubing adaptors of the<br>system, using the double<br>membrane method to create a leak-<br>proof and drug residual-free<br>connection. | Difference. Please see<br>comment # 2 |
| Materials within<br>the drug path | Polypropylene, PVC, Silicone, 304<br>SST, POM, ABS, Polyisoprene,<br>Hydrophobic filter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Polypropylene, PVC, Silicone, 304<br>SST, POM, ABS, Polyisoprene,<br>Hydrophobic filter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                  |
| Drug compatibility                | Compatible with hazardous drug<br>usage                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Compatible with hazardous drug<br>usage                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                  |
| Characteristic                    | CSTD System<br>Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | CSTD System<br>Predicate device (K170706)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Comments                              |
| Device Type                       | Rx/Single Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Rx/Single Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                  |
| Target Users                      | Licensed Pharmacists/Health Care<br>Professionals                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Licensed Pharmacists/Health Care<br>Professionals                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                  |
| Environment                       | Hospitals and clinics                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Hospitals and clinics                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                  |
| Sterilization                     | EO/SAL 10-6<br>(Plug - Gamma)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | EO/SAL 10-6<br>(Plug - Gamma)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                  |

{6}------------------------------------------------

{7}------------------------------------------------

### EQUASHIELD® CSTD – 510(k) Summary

Discussions of differences in Indications for Use statement:

The EQUASHIELD® CSTD System has the same intended use and the same indications for use as the predicate, the previously cleared EQUASHIELD® CSTD System (K170706) except for the addition of a sentence to the indications for use to clarify the functionality of the Syringe Unit with respect to syringe plunger detachment. This is not a new feature to the product and was described in the predicate 510(k), even though it was not included in the indication for use of the predicate device.

Discussion of differences in technological characteristics:

Comment #1 The new components added to the CSTD family are (by Ref#):

- Vial Adaptor: VA-13C/2 i
- Spike Adaptors: SA-180 & SA-W180
- Luer Lock Adaptors: LL-2S & LL-2
- Male Priming Connector: MC-2
- Female Luer Lock connectors: FC-1S &FC-180
- Protective Plug: PP-2

The design and principle of operation of all new system components are based on the predicate components. All components have been tested and validated and perform as intended. Risk assessment identified no new or modified risks relative to predicate components. Differences do not raise new or different questions of safety or effectiveness

Comment #2 The subject device includes a number of new components that are each based on a variant of a component included in the cleared predicate system. All the new components have been validated and perform as intended. Risk assessment identified no new or modified risks of the proposed CSTD relative to the predicate. Differences do not raise new or different questions of safety or effectiveness

The proposed changes, which include changes relating to the materials used in some components and the inclusion of several additional components, have been assessed through bench testing, biocompatibility testing, drug compatibility testing, usability testing and a biological risk assessment. None of the proposed additions and modifications were found to raise new safety or effectiveness concerns and the results of the testing support the substantial equivalence claim. With respect to principles of operation, both systems are identical. Consequently, the EQUASHIELD® CSTD System is as safe and effective as its predicate without raising any new safety and/or effectiveness concerns.

{8}------------------------------------------------

# PERFORMANCE DATA

Non-clinical testing:

Non-clinical tests were conducted to verify that the proposed device meets all design specifications and complies with applicable standards as follows:

- ISO 8536-4 Infusion equipment for medical use - Part 4: Systems for single use, gravity feed
- ISO 8536-10 Infusion Equipment for Medical Use Part 10: Accessories for ● Fluid Lines for Single Use with Pressure Infusion Equipment
- ISO 80369-7 Small-bore connectors for liguids and gases in healthcare ● applications - Part 7: Connectors for intravascular or hypodermic applications
- . ISO 1135-4 Transfusion equipment for medical use - Part 4: Transfusion sets for single use, gravity feed
- USP<788> and USP<790> Visible and Sub-visible Particulate Matters in ● Injections

# Biocompatibility:

Biocompatibility testing was performed on the EQUASHIELD® CSTD components (finished and sterilized) as an Externally Communicating Device. Blood Path Indirect. Prolonged Contact (>24hrs to 30days) according to FDA Guidance "Use of International Standard ISO 10993-1. "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and ISO 10993-1 Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process. The following tests were conducted:

- . Cytotoxicity
- Sensitization
- Irritation ●
- Acute systemic toxicity ●
- Material mediated pyrogenicity ●
- Hemocompatibility ●
- Subacute Toxicity (dual route repeated exposure method)

Sterility Shelf life and Shipping Simulation:

- The EQUASHIELD® CSTD is a sterile medical device. The sterilization process ● uses Ethylene Oxide (EtO) to achieve a sterility assurance level (SAL) of at least 10-6. The validation process followed the "overkill" approach according to ISO 11135:2014 Sterilization of health-care products -- Ethylene oxide --Requirements for the development, validation and routine control of a sterilization process for medical devices.
- Bacterial Endotoxin test (LAL) was performed and passed the acceptance criteria of USP<161>.
- Ethylene Oxide (EtO) and Ethylene Chlorhydrine (ECH) residuals were found ● to comply with ISO 10993-7:2008 Biological evaluation of medical devices part 7: Ethylene oxide sterilization residuals for prolonged exposure devices.
- The shelf life of the EQUASHIELD® CSTD is 3 years, established using ASTM . F1980-16 "Standard Guide for Accelerated Aging of Sterile Barrier Systems for

{9}------------------------------------------------

Medical Devices." The study included sterilized samples of the CSTD which was accelerated aged to the equivalent of 3 years. Performance, functional and package integrity tests were conducted following accelerated aging and transportation simulation and passed acceptance criteria.

- Packaging integrity was conducted after sterilization, accelerated aging and . simulated transportation in accordance with ASTM D4169-16 under DC13. The tests included final, packed, and sterile samples of the EQUASHIELD CSTD which were tested for
	- Visual inspection ASTM F1886 o
	- Dye test ASTM F1929 o
	- Pell test ASTM F88 o
	- Burst test ASTM F2054 o
	- Bubble emission test ASTM D3078 o
	- Sterility USP<71> o
	- o Microbiological Barrier Test (for the package) ASTM F1608

## Microbial Ingress:

Microbial Ingress study was conducted per the 2008 FDA guidance, "Intravascular Administration Sets Premarket Notification Submissions [510(k)]." The study included sterile and aged EQUASHIELD® CSTD samples which were repeatedly accessed in a total of 10 times over the period of 7 days. None of the tested samples showed growth, therefore, the samples met the acceptance criteria and the EQUASHIELD® CSTD was verified to prevent microbial ingress after aging as labeled.

## Drug Compatibility:

Hazardous drug compatibility testing was performed on the EQUASHIELD® CSTD using hazardous drugs commonly used clinically with CSTDs and that may react aggressively with polymers (such as N.N-dimethylacetamid, known as DMA). Compatibility was verified using chemical analytic experiments and mechanical, functional and performance testing. Study results showed that the EQUASHIELD® CSTD does not affect the drugs' stability and the drugs do not negatively impact the devices' mechanical, functions, or performance characteristics.

## CONCLUSION

The EQUASHIELD® CSTD System has the same indications for use as the predicate, the previously cleared CSTD System (K170706) except for the addition of clarification text associated with a previously cleared feature of the syringe component. The proposed design changes, which include changes relating to the materials used in some components and the inclusion of several additional components, have been assessed through bench testing and biocompatibility testing. None of the proposed additions and modifications were found to raise new safety or effectiveness concerns and the results of the testing support substantial equivalence to the predicate system. With respect to principles of operation, both systems are identical. Consequently, the EQUASHIELD® CSTD System is substantially equivalent to its predicate.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/ONB/K221513](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/ONB/K221513)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
