Halo Closed Vial Adaptor (Multiple Models based on 13, 20 or 28mm Vial Neck Size), Halo Closed Syringe Adaptor, Halo Closed Line Adaptor, Closed Vial Adaptor

K180574 · J & J Solutions, Inc. D/B/A/ Corvida Medical · ONB · Aug 17, 2018 · General Hospital

Device Facts

Record IDK180574
Device NameHalo Closed Vial Adaptor (Multiple Models based on 13, 20 or 28mm Vial Neck Size), Halo Closed Syringe Adaptor, Halo Closed Line Adaptor, Closed Vial Adaptor
ApplicantJ & J Solutions, Inc. D/B/A/ Corvida Medical
Product CodeONB · General Hospital
Decision DateAug 17, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5440
Device ClassClass 2

Indications for Use

The Halo® system is an airtight and leak proof closed system drug (CSTD) that mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols and spills. The Halo® system also prevents microbial ingress for up to 7 days.

Device Story

Halo® is a sterile, single-use Closed System Transfer Device (CSTD) for handling hazardous drugs. Components include Closed Vial Adaptor (CVA), Closed Syringe Adaptor (CSA), Closed Bag Adaptor (CBA), Closed Line Adaptor (CLA), and Closed Vial Converter (CVC). System integrates with standard luer-lock syringes, IV bags, and infusion sets. Operation relies on dry-to-dry compression fit seals; a single-lumen needle pierces the seals for drug transfer; upon disconnection, the needle retracts and the polymer membrane reseals to prevent leaks and microbial ingress. Used by pharmacists and clinicians in hospitals/clinics to safely compound and administer hazardous drugs. Benefits include minimized exposure to drug vapors, aerosols, and spills, and maintenance of drug sterility for up to 7 days.

Clinical Evidence

No clinical data. Evidence is based on bench testing, including fluorescein leak tests, alcohol vapor leak tests, pressure tests, connection/retention force, and ISO 594-1/2 and ISO 8536-4 compliance. Biocompatibility testing followed ISO 10993-1, -4, -5, -10, and -11. Sterility validated per ISO 11135. Microbial ingress testing demonstrated protection for 7 days after 14 needle penetrations. Particulate testing met USP 788 requirements.

Technological Characteristics

Components: CVA, CSA, CBA, CLA, CVC. Materials: Polypropylene, polyethylene, polymer seals. Connection: Luer-lock, spike, needle-safe, dry-to-dry compression fit. Sterilization: EtO (SAL 10^-6). Connectivity: None (mechanical system).

Indications for Use

Indicated for use by licensed pharmacists and healthcare professionals in hospitals and clinics to reconstitute, transfer, and administer antineoplastic and other hazardous drugs, reducing exposure of personnel and the environment to chemotherapy agents.

Regulatory Classification

Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text. August 17, 2018 J & J Solutions, Inc. d/b/a/ Corvida Medical Dana Schramm Chief Operating Officer 2945 Lone Oak Drive, Suite 150 Eagan, Minnesota 55121 #### Re: K180574 Trade/Device Name: Halo System Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: ONB Dated: July 2, 2018 Received: July 9, 2018 #### Dear Dana Schramm: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Geeta K. Pamidimukkala -S for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K180574 Device Name Halo® Indications for Use (Describe) The Halo® system is an airtight and leak proof closed system drug (CSTD) that mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols and spills. The Halo® system also prevents microbial ingress for up to 7 days. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 510(K) SUMMARY K180574 #### SUBMITTER J & J Solutions, Inc. d/b/a Corvida Medical 2945 Lone Oak Drive, Suite 150 Eagan, Minnesota 55121 #### ESTABLISHMENT REGISTRATION NUMBER 3012235045 #### CONTACT Dana Schramm Chief Operating Officer Ph. 319-335-2547 x103 Fax. 319-250-4135 dana.schramm@corvidamedical.com ### DATE PREPARED August 8, 2018 ## NAME OF MEDICAL DEVICE | Trade Name: | Halo® | |----------------------|--------------------------------------------------------------------------------| | Common/Usual Name: | Closed Antineoplastic and Hazardous Drug Reconstitution<br>and Transfer System | | Classification Name: | Intravenous Administration Set | #### DEVICE CLASSIFICATION | Classification Panel: | General Hospital | |-----------------------|------------------| | Regulatory Class: | II | | Product Code: | ONB | | Regulation Number: | 21 CFR 880.5440 | ## MANUFACTURER Corvida Medical 2945 Lone Oak Drive, Suite 150 Eagan, Minnesota 55121 {4}------------------------------------------------ ## PREDICATE DEVICE | Trade Name: | Halo® (K150486) | |----------------------|--------------------------------------------------------------------------------| | Common/Usual Name: | Closed Antineoplastic and Hazardous Drug Reconstitution<br>and Transfer System | | Classification Name: | Intravenous Administration Set | ### DEVICE DESCRIPTION The Halo® is a Closed System Transfer Device (CSTD) for the safe handling of hazardous drugs, especially for the compounding and administering of hazardous drugs according to the National Institute for Occupational Safety and Health (NIOSH) definition of an airtight and leak proof closed system transfer device. It is a sterile singleuse device. There are five components of the Halo® system, Closed Vial Adaptor (CVA), Closed Syringe Adaptor (CSA), Closed Bag Adaptor (CBA), Closed Line Adaptor (CLA), and Closed Vial Converter (CVC). These components integrate with industry standard luer-lock syringes, IV bags, infusion sets, and other patient connections to form a complete closed system. This system prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills. In addition, the components are designed to prevent microbial ingress into the system, including maintaining sterility of drugs in the vial for up to 7 davs. The ability to prevent microbial ingress for up to 7 days should not be interpreted as modifying, extending, or superseding a manufacturer labeling recommendations for the storage and expiration dating. Refer to drug manufacturer's recommendations and USP compounding guidelines for shelf life and sterility information. The system uses industry compatible luer locks, bag spikes and spike ports, dual lumen spikes, single lumen needles, and dry to dry compression fit seals when connecting Halo® components together. A single lumen needle perforates the dry-to-dry compression fit seals for the transfer of drugs between Halo® components. Upon separation the needle is retracted and the seal membrane prevents transfer of environmental contaminants into the system and/or escape of drug or vapor. #### INDICATION FOR USE The Halo® system is an airtight and leak proof closed system drug transfer device (CSTD) that mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols and spills. The Halo® system also prevents microbial ingress for up to 7 days. ## TECHNOLOGICAL COMPARISON TO PREDICATE DEVICES Equivalence was determined using a side by side tabular comparison between the predicate and proposed devices which included: Features, Intended Use, Labeling, Materials, Specifications, Performance Data, and Technological Aspects. The proposed modified device is substantially equivalent to the predicate device and does not raise {5}------------------------------------------------ different questions of safety or effectiveness. {6}------------------------------------------------ | | Proposed Device (K180574) | Predicate Device (K150486) | |----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for Use | The system is an airtight and<br>leak proof closed system drug<br>transfer device (CSTD) that<br>mechanically prohibits the<br>transfer of environmental<br>contaminants into the system<br>and the escape of drug or<br>vapor concentrations outside<br>the system, thereby<br>minimizing individual and<br>environmental exposure to<br>drug vapor, aerosols and<br>spills. The Halo® system also<br>prevents microbial ingress for<br>up to 7 days. | The system is an airtight and<br>leak proof closed system drug<br>transfer device (CSTD) that<br>mechanically prohibits the<br>transfer of environmental<br>contaminants into the system<br>and the escape of drug or<br>vapor concentrations outside<br>the system, thereby<br>minimizing individual and<br>environmental exposure to<br>drug vapor, aerosols and<br>spills. The Halo® system also<br>prevents microbial ingress for<br>up to 168 hours. | | Classification | Class II | Class II | | Regulation<br>Number | 888.5440 | 888.5440 | | Product Code | ONB | ONB | ## Regulatory Classification, Risk, and Indications for Use No differences in Regulatory Classification, Risk, or Indications for Use except for the conversion of 168 hours to 7 days for microbial ingress prevention claim. # Technological Characteristics | | Proposed Device (K180574) | Predicate Device (K150486) | |----------------------|-----------------------------------|---------------------------------| | System<br>Components | 13mm Vial Adaptor (CVA130) | 13mm Vial Adaptor (CVA130) | | | 20mm Vial Adaptor (CVA200) | 20mm Vial Adaptor (CVA200) | | | 28mm Vial Adaptor (CVA280) | 28mm Vial Adaptor (CVA280) | | | Closed Syringe Adaptor (CSA100) | Closed Syringe Adaptor (CSA100) | | | Closed Bag Adaptor (CBA100) | Closed Bag Adaptor (CBA100) | | | Closed Line Adaptor (CLA100) | Closed Line Adaptor (CLA100) | | | Closed Line Adaptor (CLA200G) | NA | | | Closed Line Adaptor (CLA200B) | NA | | | Closed Vial Converter<br>(CVC130) | NA | {7}------------------------------------------------ | Characteristics | Closed System used to reconstitute, transfer, and administer antineoplastic and other hazardous drugs in healthcare setting. Indicated to reduce exposure of healthcare personnel and the surrounding environment to chemotherapy agents. | Closed System used to reconstitute, transfer, and administer antineoplastic and other hazardous drugs in healthcare setting. Indicated to reduce exposure of healthcare personnel and the surrounding environment to chemotherapy agents. | |-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Principles of Operation | Multi-components intended to be used as a system. | Multi-components intended to be usedas a system. | | Technological Characteristics | The proposed system consists of four main components that attach to standard drug vials, syringes, patient lines or secondary sets, and standard IV bags. | The predicate system consists of four main components that attach to standard drug vials, syringes, patient lines or secondary sets, and standard IV bags. | | | The proposed CSTD system uses industry compatible luer lock, spike, and needle safe connections to form the closed systems for drug transfer. | The predicate CSTD system uses industry compatible luer lock, spike, and needle safe connections to form the closed systems for drug transfer. | | | The proposed CSTD system uses polymer seals that prevent environmental contaminants from entering into the system and/or escape of drug or vapor. When system components are joined together the seals are pressed together and then pierced by needles. When disconnected, the polymer reseals to create a leak-proof and drug residual-free connection. | The proposed CSTD system uses polymer seals that prevent environmental contaminants from entering into the system and/or escape of drug or vapor. When system components are joined together the seals are pressed together and then pierced by needles. When disconnected, the polymer reseals to create a leak-proof and drug residual-free connection. | | Device Type | Rx/Single Use | Rx/Single Use | | Target Users | Licensed Pharmacists/Health Care Professionals | Licensed Pharmacists/Health Care Professionals | | Environment | Hospitals and clinics | Hospitals and clinics | | Sterilization | EtO / SAL 10-6 | EtO / SAL 10-6 | No differences in technological characteristics, system function, user populations, use environment, or sterilization method. {8}------------------------------------------------ | Changed<br>Component | Change Description | Change Discussion on SE | |------------------------------------|--------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Closed Line<br>Adaptor (CLA) | Material -<br>Polyethylene to<br>Polypropylene | Changed component material to take advantage of<br>molding and assembly productivity.<br>Polypropylene's use in medical devices and<br>CSTD's is well documented. This material change<br>has been fully tested for liquid escape, vapor<br>escape, and biocompatibility. No different<br>questions of safety or effectiveness are raised. | | | Components –<br>Added<br>Locking Line<br>Adaptors CLA200G CLA200B | The CLA200 models incorporate a locking clip to<br>provide added retention to a patient's line. All other<br>dimensions are the same as the existing predicate<br>CLA. No different questions of safety or<br>effectiveness are raised. | | | Assembly CLA100 CLA200G CLA200B | The luer lock threads are now created by an<br>injection molded insert that is ultrasonically<br>welded into the body. The predicate device<br>consisted of a single piece injection molded body<br>with luer lock threads. No different questions of<br>safety or effectiveness are raised. | | Closed Syringe<br>Adaptor (CSA) | Assembly | To take advantage of productivity and<br>manufacturing economies, changed CSA needle<br>bond from injection over mold to UV curable<br>adhesive. This design and manufacturing change<br>has been tested for liquid escape, vapor escape,<br>and adhesive bond tensile and compression<br>strength. No different questions of safety or<br>effectiveness are raised. All other features and<br>dimension remain the same as the existing<br>predicate CSA. | | 13mm Vial<br>Converter<br>(CVC130) | Component Addition | Added 13mm Vial Converter as an accessory<br>component that enables the use of the 20mm<br>Closed Vial Adaptor with a 13mm capped vial.<br>Test results establish that the CVC130 connection<br>performs in a similar fashion to the reference<br>devices (CVA130 and CVA200). Assemblies were<br>tested for liquid escape, vapor escape, attachment<br>and detachment force. No different questions of<br>safety or effectiveness are raised. | | Labeling | Labels and IFU's | Added labels and directions for the new<br>components. No different questions of safety or<br>effectiveness are raised. | # Discussion of Design/Manufacturing Differences {9}------------------------------------------------ ## SUMMARY OF PERFORMANCE TESTING Results from tests completed on the Halo® components demonstrate it to be an easy to use device that prevents microbial ingress and/or escape of drug or vapor through multiple reconnections of components up to 14 times and are substantially equivalent with respect to operational performance. Product testing consisted of the following: | Product Functional Testing | Results | |---------------------------------------------------------|----------| | • Fluorescein Leak Test | No Leaks | | • Alcohol Vapor Leak Test | No Leaks | | • Pressure Test | No Leaks | | • Insertion (Connection) and Retention Force | Pass | | • ISO594-1 Part 1: General Requirements | Pass | | • ISO 594-2 Part 2: Lock Fittings | Pass | | • ISO 8536-4 Infusion equipment for medical use: Part 4 | Pass | ### Package Integrity and Shelf Life Packaging and Shelf Life validation was performed on aged Halo® packaging according to ASTM F2096: Standard Test Method for Detecting Gross Leaks in Medical Packaging by Internal Pressurization, ASTM F1886: Standard Test Method for Determining Integrity of Seals for Medical Device Packaging by Visual Inspection, and ASTM F88 standard test method for seal strength of flexible barrier materials. All testing passed. #### Biocompatibilitv Biocompatibility testing was performed on the Halo® materials according to FDA Guidance and ISO 10993-1 Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process. Testing included cytotoxicity, sensitization, irritation, systemic toxicity, and hemocompatibility according to standards set forth in ISO 10993- 4, Biological evaluation of medical devices -- Part 4: Selection of tests for interactions with blood, ISO 10993- 5, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity, ISO 10993-10 Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization, and ISO 10993-11 Biological evaluation of medical devices -- Part 11: Tests for systemic toxicity. All testing passed. A chemical characterization and toxicological risk assessment was used to evaluate the subchronic systemic toxicity endpoint (see Extractables Screening section below). {10}------------------------------------------------ ## Sterility Sterilization validation was performed on the finished Halo® components according to ISO 11135 Sterilization of health-care products -- Ethylene oxide -- Requirements for the development, validation and routine control of a sterilization process for medical devices. Testing included pyrogenicity, bioburden, and EO residuals to standards set forth in ISO 10993-7 Biological evaluation of medical devices -- Part 7: Ethylene oxide sterilization residuals, ISO 11737-1 Sterilization of medical devices --Microbiological methods -- Part 1: Determination of a population of microorganisms on products, and AAMI/ANSI ST72 bacterial endotoxins - test methods, routine monitoring, and alternatives to batch testing. All testing passed. ## DMA Compatibility Compatibility testing was performed to validate Halo® compatibility with antineoplastic drugs and DMA(N,N-dimethylacetamid). Halo® was found to be compatible. ## Extractables Screening Extractables screening was completed to determine what compounds and their estimated concentrations are extracted from the Halo® Closed System Transfer Device for hazardous drugs (CSTD) under the conditions of the extractions (72 hours at 37°C) when extracted in the following solvents: hexane, 50% ethanol, pH 3 (HCI adjusted) 0.9% saline, and 33% aqueous dimethylacetamide (DMA). A toxicological risk assessment was performed on the extracted compounds. ## Microbial Ingress Protection Microbial ingress testing was performed on Halo® to validate microbial ingress protection after repetitive needle penetration of the seal. These test results were originally reported in K150486. Testing results demonstrate Halo® was protected against microbial ingress for a period of 7 days after breaching the seal 14 times with the CSA needle. All sealing features remain the same as the predicate devices and the design changes do not modify any of the sealing properties from the original predicate devices. The ability to prevent microbial ingress for up to 7 days should not be interpreted as modifying, extending, or superseding a manufacturer labeling recommendations for the storage and expiration dating. Refer to drug manufacturer's recommendations and USP compounding guidelines for shelf life and sterility information. ## Particulate Testing Particulate contamination testing was performed on the Halo® system according to USP 788 to demonstrate lack of contamination. Testing results demonstrated particulate levels in the Halo system are low and meet USP 788 requirements. {11}------------------------------------------------ ### Substantial Equivalence Conclusion Performance testing established that the proposed Halo® devices performed substantially equivalent (SE) to the performance testing on the predicate Halo® devices. Equivalence was determined using a side by side tabular comparison between the predicate and proposed Features, Intended Use, Labeling, Materials, Specifications, Performance Data, and Technological Aspects. The proposed devices are Substantially Equivalent to the predicate devices. Based on the analysis of the comparison between the predicate and proposed devices regarding risk analysis, design controls, and performance evaluation the data shows the modifications do not raise different questions of safety or efficacy and demonstrates substantial equivalence to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...