SWABFLSUH (3ML, 5ML, 10ML) MODEL 10331, 10351, 10311
K101270 · Excelsior Medical Corp. · NGT · Feb 10, 2011 · General Hospital
Device Facts
Record ID
K101270
Device Name
SWABFLSUH (3ML, 5ML, 10ML) MODEL 10331, 10351, 10311
Applicant
Excelsior Medical Corp.
Product Code
NGT · General Hospital
Decision Date
Feb 10, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The flush syringe is intended for flushing of IV catheters and IV tubing. SwabCap® is intended for use on a swab-able luer access valve as a disinfecting cleaner prior to line access and to act as a physical barrier to contamination between the line accesses. SwabCap® will disinfect the valve five (5) minutes after application and act as a physical barrier to contamination for up to ninety-six (96) hours under normal conditions if not removed. The SwabFlush™ product is provided as: 1. 10 ml fill in 10 ml saline flush syringe with integrated SwabCap®. 2. 5 ml fill in 10 ml saline flush syringe with integrated SwabCap®. 3. 3 ml fill in 10 ml saline flush syringe with integrated SwabCap®.
Device Story
SwabFlush combines a pre-filled saline flush syringe with an integrated SwabCap disinfectant cap. Syringe contains 0.9% Sodium Chloride; SwabCap contains 70% isopropyl alcohol. Modifications to standard syringe include a plunger thumb-press cup to hold the SwabCap and rounded barrel recesses to prevent plunger rotation, allowing the syringe to serve as a tool for applying the cap to luer access valves. Used by healthcare professionals in clinical settings. After IV line flushing, the clinician uses the syringe as a handle to install the SwabCap onto the valve. The cap disinfects the valve within 5 minutes and provides a physical barrier for up to 96 hours. Benefits include immediate access to disinfectant post-flush and reduced contamination risk.
Clinical Evidence
Bench testing only. Verification included ISO 7886-1 (sterile syringes), ISO 594-1/2 (luer tapers), and ISO 10993 (biocompatibility). Sterilization validation achieved required sterility assurance levels. Functional testing confirmed saline compliance with USP monograph, shelf-life stability, and mechanical integrity (drop, vibration, compression) per ISTA/ASTM standards. Design validation performed in a simulated clinical setting.
Technological Characteristics
Single-use, disposable prefilled saline syringe with integrated 70% isopropyl alcohol disinfectant cap. Materials meet ISO 10993. Standard male luer lock fitting. Features anti-rotational barrel recesses and plunger cup for cap integration. Sterilized via validated methods. Complies with ISO 7886-1 and ISO 594-1/2.
Indications for Use
Indicated for flushing IV catheters and IV tubing in patients requiring vascular access. SwabCap component indicated for disinfection of swab-able luer access valves and as a physical barrier to contamination.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
Predicate Devices
Excelsior Disposable Syringe w/ Normal Saline (0.9% Sodium Chloride) (K962938)
Excelsior Saline pre-filled Syringe(s) in Sterile Field Packaging (K082837)
{0}------------------------------------------------
K101270 p1of3
Excelsior Medical Corporation Traditional 510(k)
SwabFlush
# 510(k) Summary
FEB 1 n 2011
| Manufacture Name: | Excelsior Medical Corporation |
|------------------------------|-------------------------------------------------|
| Contact Name: | John Linfante |
| Postal Address: | 1933 Heck Avenue<br>Neptune, NJ 07753 |
| Phone Number: | 732-643-6088 |
| Fax: | 732-776-7600 |
| Title: | VP, Regulatory Affairs and Quality<br>Assurance |
| Date: | February 4, 2011 |
| Device Proprietary Name: | SwabFlush™ |
| Device Common or Usual Name: | Flush Syringe with Device Disinfectant<br>Cap |
| Classification Name: | Saline Vascular Access Flush |
| Classification Code: | NGT/LKB |
| Regulation Number: | General Hospital |
#### Predicate Devices:
Substantial equivalence is claimed to the following devices as related to intended use, technology and design characteristics:
- . Excelsior Disposable Syringe w/ Normal Saline (0.9% Sodium Chloride), K962938 (Excelsior Medical Corporation)
- Excelsior Saline pre-filled Syringe(s) in Sterile Field Packaging, K082837 . (Excelsior Medical Corporation)
- SwabCap®, K083508 (Excelsior Medical Corporation) .
#### Description of the Device
SwabFlush" is a saline intravenous flush syringe that is combined with an integrated SwabCap®.
Each SwabFlush" syringe contains a sterile, non-pyrogenic isotonic solution of 0.9% Sodium Chloride Injection USP, 9 mg NaCl per ml, with an osmolarity of 0.3 l mOsm/mL, pH 4.5 - 7.0.
A SwabCap® is a luer access valve cap that contains 70% isopropyl alcohol as a disinfectant. A SwabCap® is designed to securely fit on swab-able luer access valves. The product is intended for single-use and is provided sterile.
The SwabFlush" provides the healthcare professional with immediate access to the SwabCap luer access valve disinfectant cap after an IV flush.
{1}------------------------------------------------
K101270 p.2 of 3
SwabFlush"
### Intended Use/Indications for Use
The flush syringe is intended for flushing IV catheters and IV tubing. The integrated SwabCap® is intended for use on swab-able luer access valves as a disinfecting cleaner prior to line access and to act as a physical barrier to contamination between the line accesses. SwabCap® will disinfect the valve five (5) minutes after application and act as a physical barrier to contamination for up to ninety-six (96) hours under normal conditions if not removed.
#### Technological Characteristics
The SwabFlush" syringe has the same technological characteristics as the predicates. SwabFlush" syringe and predicate syringes are all single use, disposable prefilled saline syringes.
The SwabFlush" syringe is similar in design and uses the same types of materials as the Excelsior Medical Disposable Syringe with Normal Saline (K962938) and Saline prefilled Syringe(s) in Sterile Field Packaging (K082837). The SwabFlush" syringe, like the predicate syringes, has a standard male luer lock fitting which is intended to connect to standard female luer lock fittings, including luer access valves. The saline contained in the SwabFlush" syringe and predicate syringes meet the same requirements for 0.9% Sodium Chloride Injection, USP. Both the SwabFlush™ syringe and predicate syringes meet the same requirements for syringe functional characteristics such as, but not limited to, injection force and liquid/air leakage. The SwabFlush" syringe is filled, sterilized, and packaged using the same equipment and processes of the Disposable Syringe with Normal Saline (K962938). The SwabFlush " syringe and the Disposable Syringe with Normal Saline (K962938) are intended for the same user and indicated for the same use.
The integrated SwabCap® within the SwabFlush™ syringe is identical to the predicate device. There have been no changes to the intended use, design, materials or processing of SwabCap® as cleared under K083508.
#### Technological Modifications
The SwabFlush" has two unique technological modifications from the predicate devices. The first is a modification of the thumb press end of the plunger. Instead of a flat surface, the end has an open cup feature that allows a single fully assembled and packaged SwabCap® to be integrated with the syringe. The second modification is in the form of rounded recesses molded into the opening of the syringe barrel. This feature prevents a fully injected syringe plunger from rotating. This anti-rotational feature allows the fully injected syringe to become a tool for SwabCap® application and facilitates its installation to the luer access device.
#### Substantial Equivalence Discussion
The subject device is similar to the predicate devices based on intended use, technology and design. No changes have been made to the design of the SwabCap® as cleared by the FDA for integration into the syringe plunger.
{2}------------------------------------------------
K101270
SwabFlush™ p³-g
## Non Clinical Testing Data to Demonstrate Safety
All of the non clinical testing was conducted in accordance to the requirements of design control. The following is a summary of the non clinical data that demonstrate the SwabFlush" syringe is safe and effective. It consists of the following:
## DESIGN VERIFICATION
Excelsior has demonstrated that the SwabFlush"" syringe meets all ISO 7886-1: Sterile Hypodermic Syringes for single use standards and ISO 594-1&-2: Conical fittings with a 6 % (Luer) taper for syringes, needles and certain other medical equipment.
The device was confirmed to meet all biocompatibility requirements of ISO 10993: Biological evaluation of medical devices.
Sterilization validations were performed to validate the effectiveness of the sterilization methods to achieve the desired sterility assurance level.
Additional tests performed ensure the SwabFlush" syringe functions to hold and apply a SwabCap®, while not affecting the disinfectant content of the SwabCap®. The saline flush was verified to comply with 0.9% Sodium Chloride Injection USP monograph requirements. The verification also included the controlled aging of SwabFlush" to verify all product requirements including sterility throughout the shelf life. Ship testing was performed on a representative final packaged product for drop, vibration, and compression as specified by applicable ISTA or ASTM standards.
All results met the criteria established in the testing.
## DESIGN VALIDATION
A Design Validation was done to verify that the SwabFlush™ meets the clinical need expected for a typical flush procedure in a simulated clinical setting.
## Conclusion
Based on the information provided in this 510(k) premarket notification, SwabFlush" is substantially equivalent in terms of safety and effectiveness to the predicate devices identified above.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features the HHS logo, which is a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the logo.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Excelsior Medical Corporation C/O Mr. John Linfante Vice President Regulatory Affairs/Quality Assurance Excelsior Medical Corporation 1933 Heck Avenue Neptune. New Jersey 07753
FEB 1 0 2011
· Re: K101270
Trade/Device Name: SwabFlush™ Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II Product Code: NGT Dated: January 25, 2011 Received: January 27, 2011
Dear Mr. Linfante:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. :
{4}------------------------------------------------
### Page 2- Mr. Linfante
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices
Office of Device Evaluation Canter for Devices and Radiological Health
{5}------------------------------------------------
: : : ·
## Indications for Use
| 510(k) Number: | K101270 |
|----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | SwabFlush™ |
| Indication for Use: | The flush syringe is intended for flushing of IV catheters and IV tubing. |
| | SwabCap® is intended for use on a swab-able luer access valve as a disinfecting cleaner prior to line access and to act as a physical barrier to contamination between the line accesses. SwabCap® will disinfect the valve five (5) minutes after application and act as a physical barrier to contamination for up to ninety-six (96) hours under normal conditions if not removed. |
| | The SwabFlush™ product is provided as:<br>1. 10 ml fill in 10 ml saline flush syringe with integrated SwabCap®.<br>2. 5 ml fill in 10 ml saline flush syringe with integrated SwabCap®.<br>3. 3 ml fill in 10 ml saline flush syringe with integrated SwabCap®. |
| Prescription Use (21 CFR Part 801 Subpart D) | X And/Or Over the Counter Use (21 CFR Part 801 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign-Off
Anesthesiology, General Hospital, Infection Control and Dental Devices
510(k)

(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices
Page 1 of 1
510(k) Number: K101270
、「
。
. '
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.