← Product Code [MRZ](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MRZ) · K141273

# WELLDOC DIABETESMANAGER SYSTEM AND DIABETESMANAGER- RX SYSTE (K141273)

_Welldoc, Inc. · MRZ · Jul 8, 2014 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MRZ/K141273

## Device Facts

- **Applicant:** Welldoc, Inc.
- **Product Code:** [MRZ](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MRZ.md)
- **Decision Date:** Jul 8, 2014
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 880.5725
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Software as a Medical Device

## Indications for Use

DiabetesManager® (OTC Use): The WellDoc DiabetesManager® System is indicated for use by healthcare providers (HCPs) and their adult patients - aged 21 years and older -who have type 2 diabetes. The DiabetesManager® System is intended to provide secure capture, and transmission of blood glucose data as well as information to aid in diabetes self-management. The DiabetesManager® System analyzes and reports blood glucose test results and supports medication adherence. It includes software intended for use on mobile phones or personal computers in the home or in professional healthcare settings. The software also allows for entry of other diabetes-related healthcare information and provides educational information. The DiabetesManager® System is not intended to replace the care provided by a licensed healthcare professional. including prescriptions, diagnosis, or treatment. DiabetesManager®-Rx (Prescription Use): The WellDoc DiabetesManager@-Rx System is indicated for use by healthcare providers (HCPs) and their adult patients - aged 21 years and older - who have type 2 diabetes. The DiabetesManager®)-Rx System is intended to provide secure capture, storage, and transmission of blood glucose data as well as information to aid in diabetes self-management. The DiabetesManager®-Rx System analyzes and reports blood glucose test results and supports medication adherence. In addition, the DiabetesManager®-Rx System provides coaching messages (motivational, and educational) based on real-time blood glucose values and trends. It includes software intended for use on mobile phones or personal computers in the home or in professional healthcare settings. The software also allows for entry of other diabetes-related healthcare information and provides educational information. The DiabetesManager®-Rx System is not intended to replace the care provided by a Licensed healthcare professional, including prescriptions, diagnosis, or treatment.

## Device Story

WellDoc BlueStar (DiabetesManager/DiabetesManager-Rx) is a software system for type 2 diabetes management. It captures, stores, and transmits blood glucose data; analyzes results; supports medication adherence; and provides educational/motivational coaching messages based on real-time glucose trends. Used by patients and healthcare providers in homes or clinical settings via mobile phones or PCs. Output assists patients in self-management and provides HCPs with data to support clinical decision-making. Does not replace professional medical care, diagnosis, or treatment.

## Technological Characteristics

Software-based system for mobile phones and personal computers. Functions include data capture, transmission, analysis, and reporting of blood glucose values. Provides educational and motivational coaching messages.

## Regulatory Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 17, 2024

WellDoc, Incorporated Divya Kallamadi Senior Director of Regulatory Affairs and Quality System 1501 Saint Paul Street, Suite 118 Baltimore, Maryland 21202

Re: K141273

Trade/Device Name: WellDoc BlueStar (WellDoc DiabetesManager® System and DiabetesManager®- Rx System) Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion pump Regulatory Class: Class II Product Code: MRZ, NDC

Dear Divya Kallamadi:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated July 8, 2014. Specifically, FDA is updating this SE Letter because FDA has better categorized your device technology under product codes MRZ and NDC.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Juliane Lessard, OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices, 240-402-6126, Juliane.Lessard(@fda.hhs.gov.

Sincerely,

Juliane C. Lessard -S

Juliane C. Lessard, Ph.D. Director DHT3C: Division of Drug Delivery, General Hospital and Human Factors Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Onig Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

## July 8, 2014

WellDoc Incorporated Ms. Divya Kallamadi Senior Director of Regulatory Affairs and Quality System 1501 St. Paul Street, Suite 118 Baltimore, MD 21202

Re: K141273

Trade/Device Name: WellDoc BlueStar (WellDoc DiabetesManager® System and DiabetesManager@-Rx System) Regulation Number: 21 CFR 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: MRZ, LNX Dated: June 25, 2014 Received: June 25, 2014

Dear Ms. Kallamadi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Page 2 - Ms. Kallamadi

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office

of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mary S. Runner -S

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known) K100066, K112370, K120314

Device Name WellDoc DiabetesManager® System and DiabetesManager®-Rx System WellDoc BlueStar

### Indications for Use (Describe)

DiabetesManager® (OTC Use): The WellDoc DiabetesManager® System is indicated for use by healthcare providers (HCPs) and their adult patients - aged 21 years and older -who have type 2 diabetes. The DiabetesManager® System is intended to provide secure capture, and transmission of blood glucose data as well as information to aid in diabetes self-management. The DiabetesManager® System analyzes and reports blood glucose test results and supports medication adherence. It includes software intended for use on mobile phones or personal computers in the home or in professional healthcare settings. The software also allows for entry of other diabetes-related healthcare information and provides educational information.

The DiabetesManager® System is not intended to replace the care provided by a licensed healthcare professional. including prescriptions, diagnosis, or treatment.

DiabetesManager®-Rx (Prescription Use): The WellDoc DiabetesManager@-Rx System is indicated for use by healthcare providers (HCPs) and their adult patients - aged 21 years and older - who have type 2 diabetes. The DiabetesManager®)-Rx System is intended to provide secure capture, storage, and transmission of blood glucose data as well as information to aid in diabetes self-management. The DiabetesManager®-Rx System analyzes and reports blood glucose test results and supports medication adherence. In addition, the DiabetesManager®-Rx System provides coaching messages (motivational, and educational) based on real-time blood glucose values and trends. It includes software intended for use on mobile phones or personal computers in the home or in professional healthcare settings. The software also allows for entry of other diabetes-related healthcare information and provides educational information.

The DiabetesManager®-Rx System is not intended to replace the care provided by a Licensed healthcare professional, including prescriptions, diagnosis, or treatment.

| Type of Use (Select one or both, as applicable)                                  |                                                                                 |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span>      Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span>      Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

|  |  | FOR FDA USE ONLY |
|--|--|------------------|

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{4}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MRZ/K141273](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MRZ/K141273)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
