AIDERA DIASEND SYSTEM

K101806 · Aidera AB · MRZ · Dec 7, 2010 · General Hospital

Device Facts

Record IDK101806
Device NameAIDERA DIASEND SYSTEM
ApplicantAidera AB
Product CodeMRZ · General Hospital
Decision DateDec 7, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5725
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

Aidera Diasend is indicated for use by individuals or healthcare professionals in the home or health care facilities for transmitting data from home monitoring devices such as glucose meters and insulin pumps to a server database to support diabetes management. The device is indicated for professional use and over-the-counter sales.

Device Story

Diasend system facilitates data transmission from patient home monitoring devices (glucose meters, insulin pumps) to a centralized server database. System comprises a software transmitter, server database, and web interface. Software transmitter runs on standard computing hardware (e.g., desktop, laptop, mobile) to upload device data. Used by patients and healthcare providers in home or clinical environments. Providers access uploaded data via website to support diabetes management decisions. System enables remote monitoring and data aggregation to assist in clinical diabetes care.

Clinical Evidence

Bench testing only. The device was tested for compliance with applicable EMC and FCC requirements and standards. No clinical data provided.

Technological Characteristics

System consists of a software transmitter, server database, and web interface. Transmitter software designed for Windows XP environment. Connectivity via standard computing hardware to server database. No specific materials or sterilization methods applicable as this is a software-based data transmission system.

Indications for Use

Indicated for individuals or healthcare professionals in home or clinical settings to transmit data from glucose meters and insulin pumps to a server database for diabetes management. Suitable for professional and over-the-counter use.

Regulatory Classification

Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K101806 # SUMMARY AND CERTIFICATION # 510(k) SUMMARY DEC - 7 2010 ### Summary of Safety and Effectiveness In accordance with 21 CFR 807.92, the following information constitutes the Aidera summary for the Diasend System. | SUBMITTER'S NAME:<br>ADDRESS: | Aidera<br>Aidera AB<br>Sahlgrenska Science Park<br>Medicinaregatan 8A<br>SE 413 46 Goteborg<br>Sweden | |-----------------------------------------------------|-------------------------------------------------------------------------------------------------------| | CONTACT PERSON:<br>TELEPHONE NUMBER:<br>FAX NUMBER: | Anders Sonesson<br>+46 31 741 17 85<br>+46 31 741 17 01 | | DATE OF SUBMISSION: | June 18, 2010 | #### 1. Identification of device | Proprietary Name: | Diasend | |------------------------|------------------------------------------------------------------------------| | Common Name: | Accessories, Pump, Infusion<br>System, Test, Blood Glucose, Over The Counter | | Classification Status: | Class II according to Sec. 880.2910 and 862.1345 | | Product Codes: | MRZ, NBW | #### 2. Equivalent devices K083221, Aidera AB, Aidera Diasend K072698, Confidant Inc, Confidant 2.5 K032164, Medtronic Inc, Medtronic Minimed DDMS #### 3. Description of the Device Diasend is a system for transmitting data from patients' home monitoring devices and consists of a transmitter, a server database and a website available for the care provider and the patient. The software transmitter is a Diasend software concept developed by Aidera that may run on a computer device, e.g. desktop computer, laptop or mobile phone, designed to transmit data to the Diasend server database. Current implementation is on Windows ХР. #### 4. Intended use Aidera Diasend is indicated for use by individuals or healthcare professionals in the home or health care facilities for transmitting data from home monitoring devices such as glucose meters and insulin pumps to a server database to support diabetes management. The device is indicated for professional use and over-the-counter sales. {1}------------------------------------------------ #### Technological characteristics, comparison to predicate device. ട. The Diasend system is intact except for the addition of the software transmitter compared to previously cleared device. See section 5 for a more elaborate comparison. #### 6. Discussion of performance testing. The Diasend transmitter is tested and found to comply with applicable EMC and FCC requirements and standards. #### 7. Conclusion . Based on comparison with the predicate devices, the Aidera Diasend System is substantially equivalent to the previously cleared devices and presents no new concerns about safety and effectiveness. #### 8. Indications for Use Statement Aidera Diasend is indicated for use by individuals or healthcare professionals in the home or health care facilities for transmitting data from home monitoring devices such as glucose meters and insulin pumps to a server database to support diabetes management. The device is indicated for professional use and over-the-counter sales. 7 {2}------------------------------------------------ Image /page/2/Picture/0 description: The image contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in all caps. The text is written in a simple, sans-serif font and is horizontally oriented. The words are evenly spaced, and the overall appearance is clean and professional. Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake winding around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged in a circular fashion around the symbol. The logo is black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Mr. Anders Sonesson CEO AIDERA AB Medicinaregatan 8A SE 413 46 Goteborg Sweden DEC - 7 2010 Re: K101806 Trade/Device Name: Aidera Diasend System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: MRZ, NBW Dated: November 17, 2010 Received: November 22, 2010 Dear Mr. Sonesson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2- Mr. Sonesson Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ DEC - 7 2010 ## INDICATIONS FOR USE 510(k) Number K101806 ### Device Name: Aidera Diasend System ### Indications for Use: Aidera Diasend is indicated for use by individuals or healthcare professionals in the home or health care facilities for transmitting data from home monitoring devices such as glucose meters and insulin pumps to a server database to support diabetes management. The device is indicated for professional use and over-thecounter sales. (Please do not write below this line - continue on another page if needed) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use X (Per 21 CFR 801.109) OR Over the Counter Use_ X Phil C. Chapman 12/6/10 (Division Sign-Off) (Division Sign-Girl) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: k101806 5
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...