← Product Code [MQX](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MQX) · K980471

# KHT GUAM NEEDLE, ACUPUNCTURE NEEDLE (K980471)

_American Institute of Koryo Hand Therapy · MQX · Apr 9, 1998 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MQX/K980471

## Device Facts

- **Applicant:** American Institute of Koryo Hand Therapy
- **Product Code:** [MQX](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MQX.md)
- **Decision Date:** Apr 9, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5580
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

To pierce the skin in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states.

## Device Story

KHT Guam Needle is a sterile, single-use acupuncture needle. Used by qualified acupuncture practitioners to pierce skin during acupuncture therapy. Device design, materials, and sterility meet general specifications for single-use acupuncture needles. Intended to provide a safe, sterile alternative to reusable needles used historically.

## Clinical Evidence

No clinical data. Bench testing only regarding sterility, biocompatibility, and design specifications.

## Technological Characteristics

Sterile, single-use acupuncture needle. Class II device. Materials and design meet general specifications for single-use acupuncture needles.

## Regulatory Identification

An acupuncture needle is a device intended to pierce the skin in the practice of acupuncture. The device consists of a solid, stainless steel needle. The device may have a handle attached to the needle to facilitate the delivery of acupuncture treatment.

## Special Controls

*Classification.* Class II (special controls). The device, when it is an acupuncture point locator or a single use acupuncture needle, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9. Acupuncture needles must comply with the following special controls:(1) Labeling for single use only and conformance to the requirements for prescription devices set out in 21 CFR 801.109,
(2) Device material biocompatibility, and
(3) Device sterility.

## Predicate Devices

- Seirin Brand Acupuncture Needle ([K963769](/device/K963769.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K980471

## APR - 9 1998

## American Institute of Koryo Hand Therapy, Inc.

3435 Wilshire Blvd., Suite 690, Los Angeles, CA 90010 Tel: (213) 380-5878 Fax: (213) 380-5876

Premarket Notification (510(k) Number):

FOA/CDRH/ODE/DMC

6 FEB 98 10 54

RECEWED

|                      | 510(k) SUMMARY                                                |
|----------------------|---------------------------------------------------------------|
|                      | As Required by 21 CFR 807.92(c)                               |
| Trade Name:          | KHT Guam Needle                                               |
| Common Name:         | Acupuncture Needles                                           |
| Classification Name: | Class II, Single use acupuncture needle                       |
|                      | Legally Marketed Device to Which We Are Claiming Equivalence: |
|                      | Seirin Brand Acupuncture Needle                               |
|                      | 510(k) Document Number: K963769                               |

Acupuncture needles are defined as prescription devices intended to pierce the skin in the practice of acupuncture by qualified practitioners of acupuncture as determined by the States.

Acupuncture needles have been used for the general practice of acupuncture in the United States for over 30 years. Since that time, we are not aware of any serious or life threatening accidents involving acupuncture needles.

Acupuncture needles which were sold through commercial interstate distribution prior to May 28, 1976 were non-sterile, reusable acupuncture needles. Acupuncture needles which are currently being marketed through interstate distribution (i.e., 1997) offer greater safety since they are sterilzed, single use only acupuncture needles.

The subject of this 510(k) Notification - the KHT Guam Needle - is a sterile, single use only acupuncture needle. The design, material used, sterility and biocompatibility of this acupuncture needle meet the general specifications and criteria for an single use acupuncture needle and is effective for the practice of acupuncture.

In conclusion, based on the information provided with this 510(k) Notification, the KHT Guam Needle meets the criteria for 510(k) acceptance. The KHT Guam Needle is equivalent to acupuncture needles which were in commercial distribution prior to May 28, 1976. Also, the KHT Guam Needle is equivalent to other single use acupuncture needles which are currently being sold through interstate commerce.

Jin Hae Lew, Director

1-22-98
Date

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three curved lines representing its body and wings. The eagle is enclosed in a circle, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the circle's perimeter.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

American Institute of Koryo Hand Therapy, Incorporated .C/O Mr. George Su APR - 9 1998 Requlatory Consultant Crosslinks International 17870 Castleton Street, Suite 265 Industry, California 91748

Re : K980471 KHT Guam Needle, Acupuncture Needles Trade Name: Requlatory Class: II Product Code: MOX Dated: January 22, 1998 Received: February 6, 1998

Dear Mr. Su:

We have reviewed your Section 510 (k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਜੋ substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in Please note: this response to your the Federal Reqister. premarket notification submission does not affect any obligation you might have under sections 531 through 542 of

{2}------------------------------------------------

Page 2 - Mr. Su

the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other qeneral information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Timothy A. Ulatowski

Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

K980471

## Indications For Use

510(k) Number ( if known ) _

KHT Guam Needle, Acupuncture Needle Device Name:

Indications For Use:

To pierce the skin in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states.

## ( PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation ( ODE )

Patruca Cassesite
Division of Oil & Gas

(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 510(k) Number _

OR

Prescription Use

Over-The-Counter Use __

( Optional Format 1-2-96 )

Page 13 of 19

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MQX/K980471](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MQX/K980471)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
