K141473 · Smc Intl., Inc. · MQX · Aug 1, 2014 · General Hospital
Device Facts
Record ID
K141473
Device Name
SMC ACUPUNCTURE NEEDLE
Applicant
Smc Intl., Inc.
Product Code
MQX · General Hospital
Decision Date
Aug 1, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5580
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This acupuncture needle is intended to pierce the skin in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states.
Device Story
SMC Acupuncture Needle is a sterile, single-use, hair-thin stainless steel needle; inserted into specific acupuncture points on the skin by qualified practitioners. Device facilitates acupuncture therapy. Available in various diameters (0.16-0.70 mm) and lengths (15-135 mm).
Clinical Evidence
No clinical trials performed. Substantial equivalence supported by bench testing (dimensional conformance, visual inspection, design verification) and biocompatibility testing per ISO 10993 standards. Post-market experience and literature cited to support safety and performance.
Technological Characteristics
Material: 304 stainless steel; Coating: Polydimethylsiloxane; Sterilization: Gamma radiation and/or Ethylene oxide; Dimensions: 0.16-0.70 mm diameter, 15-135 mm length; Biocompatibility: ISO 10993 compliant.
Indications for Use
Indicated for use by qualified acupuncture practitioners to pierce the skin for acupuncture therapy.
Regulatory Classification
Identification
An acupuncture needle is a device intended to pierce the skin in the practice of acupuncture. The device consists of a solid, stainless steel needle. The device may have a handle attached to the needle to facilitate the delivery of acupuncture treatment.
Special Controls
*Classification.* Class II (special controls). The device, when it is an acupuncture point locator or a single use acupuncture needle, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9. Acupuncture needles must comply with the following special controls:(1) Labeling for single use only and conformance to the requirements for prescription devices set out in 21 CFR 801.109,
(2) Device material biocompatibility, and
(3) Device sterility.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 1, 2014
SMC Mr. Jong-man Lee President (Jisan It Venture Center 601) 15 Beotkkot-ro 120gil Geumcheon-gu Seoul 153-829 KOREA
Re: K141473
Trade/Device Name: SMC Acupuncture Needle Regulation Number: 21 CFR 880.5580 Regulation Name: Acupuncture needle Regulatory Class: II Product Code: MQX Dated: June 10, 2014 Received: June 12, 2014
Dear Mr. Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Lee
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mary S. Runner -S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K141473
Device Name SMC ACUPUNCTURE NEEDLE
Indications for Use (Describe)
SMC ACUPUNCTURE NEEDLE is intended to pierce the skin in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/2/Picture/11 description: The image shows the logo for the Food and Drug Administration (FDA). The logo is a stylized version of the letters "FDA", with the "F" and "D" connected and the "A" placed to the right. The logo is in gray color. The logo is simple and recognizable.
Digitally signed by Richard C. Chapman -S Date: 2014.08.01 11:19:39 -04'00'
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary 807.92(c)
Submitter's Name, address, telephone number, a contact person and date the summary was prepared: 807.92(a)(1)
| Submitter's Name: | SMC. |
|-------------------------------|----------------------------------------------------------------------------------------------|
| Submitter's Address: | (Jisan IT Venture Center 601) 15, Beotkkot-ro 12-gil,<br>Geumcheon-gu, Seoul, 153-829, Korea |
| Submitter's Telephone: | 82-2-803-5800 |
| Submitter's Fax: | 82-2-803-5900 |
| Submitter's Contact: | Jong-man Lee |
| Date 510(k) summary prepared: | March 04, 2014 |
Name of the device, including the trade or proprietary name if applicable, the common or usual name and the classification name, if known: 807.92(a)(2)
| Proprietary Name: | SMC ACUPUNCTURE NEEDLE |
|----------------------------|---------------------------------|
| Common or Usual Name: | needle, acupuncture, single use |
| Classification Name: | Acupuncture needle |
| Regulation Number: | 21 CFR 880.5580 |
| Review Panel: | General Hospital |
| Product Code: | MQX |
| Classification Code: | 2 |
| Type of 510(k) submission: | Traditional |
# Predicate Device:
### 807.92(a)(3)
The summary identifies the legally marketed device to which your firm is claiming equivalence;
WOOJEON ACUPUNCTURE NEEDLE - K111392 (1) Woojeon Co.
{5}------------------------------------------------
## Description of the Device
The Acupuncture Needles are sterile, hair thin needles which are inserted into specific points on the skin, called "acupuncture points". The Acupuncture Needles are manufactured from stainless steel and sterilized with gamma irradiation and/or ethylene oxide.
The Acupuncture Needles are available in nine needle diameters (0.16 to 0.70 mm), ten needle lengths (15-135 mm), and ten tube lengths (30-160 mm).
### Statement of the Intended Use
This acupuncture needle is intended to pierce the skin in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states.
#### Summary of the technological characteristics of the device compared to the predicate devices 807.92(a)(6)
The SMC ACUPUNCTURE NEEDLE is substantially equivalent to WOOJEON ACUPUNCTURE NEEDLE. Table 1 summarizes the technological characteristics of the SMC ACUPUNCTURE NEEDLE vs. the predicate device.
# Table 1:
# Comparison of SMC ACUPUNCTURE NEEDLE and WOOJEON ACUPUNCTURE NEEDLE
| Attribute | SMC ACUPUNCTURE<br>NEEDLE<br>K- | WOOJEON ACUPUNCTURE<br>NEEDLE<br>K111392 |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | This acupuncture needle is<br>intended to pierce the skin in the<br>practice of acupuncture by<br>qualified practitioners of<br>acupuncture as determined by<br>the states. | This acupuncture needle is<br>intended to pierce the skin in the<br>practice of acupuncture by<br>qualified practitioners of<br>acupuncture as determined by<br>the states. |
| Available in needle<br>diameters | 0.16 to 0.70 mm | 0.16 to 0.50 mm |
| Available in needle<br>lengths | 15-135 mm | 15-60 mm |
| Available in tube | 30-160 mm | 30-145 mm |
| Material<br>(part of needle body) | 304 stainless steel | 304 stainless steel |
| Coating | Polydimethylsiloxane | Polydimethylsiloxane |
### 807.92(a)(4)
807.92(a)(5)
{6}------------------------------------------------
| Attribute | SMC ACUPUNCTURE<br>NEEDLE<br>K- | WOOJEON ACUPUNCTURE<br>NEEDLE<br>K111392 |
|----------------------|-----------------------------------------------------------------|------------------------------------------|
| Sterilization method | Radiation (Gamma) for Pouch<br>and/or Ethylene oxide of Blister | Radiation (Gamma) for Pouch |
### Non-clinical Performance Data
Performance testing was conducted to evaluate and characterize the performance of the SMC Preclinical testing conducted included dimensional conformance evaluation, visual inspections, and design verification to confirm airway passage equivalency and biocompatibility testing based on the applicable elements of ISO 10993 series.
### Clinical Data
Comparison with the predicate indicates they are similar in functions and efficiency, and the post market experience proves that it is substantially equivalent.
### Conclusion
Biocompatibility tests and performance tests show that the device meets the requirements of those standards. Literatures and post market experience show that the device is substantially equivalent. Comparison with the predicate device shows that the device has similar specification and performance. Thus, we conclude that WOOJEON ACUPUNCTURE NEEDLE is substantially equivalent to the predicate device.
# 807.92(b)(1)
### 807.92(b)(3)
807.92(b)(2)
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.