← Product Code [MQX](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MQX) · K090910

# HANSO ACUPUNCTURE NEEDLE (K090910)

_Hansol Medical Co. · MQX · Jun 30, 2009 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MQX/K090910

## Device Facts

- **Applicant:** Hansol Medical Co.
- **Product Code:** [MQX](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MQX.md)
- **Decision Date:** Jun 30, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5580
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

An acupuncture needle is a device intended to pierce the skin in the practice of acupuncture. The device consists of a solid, stainless steel needle. The device may have a handle attached to the needle to facilitate the delivery of acupuncture treatment.

## Device Story

Hansol Acupuncture Needle is a sterile, single-use, solid stainless steel needle; includes handle to facilitate treatment delivery. Used by acupuncture practitioners to pierce skin during acupuncture therapy. Device provides mechanical skin penetration; no electronic or software components. Benefits include sterile, single-use application for acupuncture practice.

## Clinical Evidence

Bench testing only; included biocompatibility and sterility testing. No clinical data provided.

## Technological Characteristics

Material: surgical stainless steel. Form factor: solid needle with handle. Sterilization: sterile. Single-use only. No energy source, connectivity, or software.

## Regulatory Identification

An acupuncture needle is a device intended to pierce the skin in the practice of acupuncture. The device consists of a solid, stainless steel needle. The device may have a handle attached to the needle to facilitate the delivery of acupuncture treatment.

## Special Controls

*Classification.* Class II (special controls). The device, when it is an acupuncture point locator or a single use acupuncture needle, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9. Acupuncture needles must comply with the following special controls:(1) Labeling for single use only and conformance to the requirements for prescription devices set out in 21 CFR 801.109,
(2) Device material biocompatibility, and
(3) Device sterility.

## Predicate Devices

- ASIA-MED Single use, Acupuncture Needle ([K052085](/device/K052085.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### 510(k) Summary ನ.

## See 510(k) Summary, below.

- 1. Trade Name: . Hansol Acupuncture Needle Common Name: Single use, Acupuncture Needle. Classification Name: Single use, Acupuncture Needle. Product Code: MQX Regulation: 880.5580 Class of device : ClassII.
JUN 300 2009

13

- 2. The legally marketed device to which we are claiming equivalence: ASIA-MED Single use, Acupuncture Needle. K052085
- 3. Description of device: Hansol Acupuncture Needle consists of components commonly found on Single use, Acupuncture Needle.

The Hansol acupuncture needle is a sterile surgical stainless steel, single use only acupuncture needle. The Hansol acupuncture needle meets the general specification and criteria for an acupuncture needle and is effective for the practice of acupuncture.

- 4. Intended Use: An acupuncture needle is a device intended to pierce the skin in the practice of acupuncture. The device consists of a solid, stainless steel needle. The device may have a handle attached to the needle to facilitate the delivery of acupuncture treatment.
- Technological Characteristics: Hansol Acupuncture Needle and the predicate device have identical technological characteristics and perform the same way as common acupuncture Needle They are sterilized
- 6. Performance: Bench tests were performed.

Bench testing included biocompatibility, sterility testing.

The tests demonstrated that the device is as safe , as effective and performs in a substantially equivalent manner to the predicate device.

{1}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

### Public Health Service

JUN 30 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Hansol Medical Company C/O Mr. Peter Chun 300 Atwood Street Pittsburgh, Pennsylvania 15213

Re: K090910

Trade/Device Name: Hansol Acupuncture Needle Regulation Number: 21 CFR 880.5580 Regulation Name: Acupuncture Needle Regulatory Class: II Product Code: MOX Dated: June 9, 2009 Received: June 22, 2009

Dear Mr. Chun:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

# Page 2- Mr. Chun

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Anthony Q. anton for
Susan Runner, D.D.S., M.A.

Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

## Enclosure.

{3}------------------------------------------------

510(k) Number (if known): _K0909 1 0 _________________________________________________________________________________________________________________________________________

Device Name: Single use, Acupuncture Needle

Indications For Use:

An acupuncture needle is a device intended to pierce the skin in the practice of acupuncture. The device consists of a solid, stainless steel needle. The device may have a handle attached to the needle to facilitate the delivery of acupuncture treatment.

Prescription Use X (Pr 21 CFR 801 Subpart D)

AND/OR .

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign=Off) Division of Sign-Olly
Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K090910

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MQX/K090910](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MQX/K090910)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
