BD Intraosseous Vascular Access System EMS Powered Driver (D001003)
K223198 · Bard Access Systems · MHC · Nov 10, 2022 · General Hospital
Device Facts
Record ID
K223198
Device Name
BD Intraosseous Vascular Access System EMS Powered Driver (D001003)
Applicant
Bard Access Systems
Product Code
MHC · General Hospital
Decision Date
Nov 10, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The BD Intraosseous Vascular Access System is intended to provide clinicians and emergency personnel with access to the intraosseous space.
Device Story
The BD Intraosseous Vascular Access System is a hand-held, battery-powered driver used by clinicians and emergency personnel to assist in the insertion of a single-use hypodermic needle through the bone cortex for intraosseous fluid delivery. The system includes a needle set (hypodermic needle, stylet, and drive adapter hub with a ferromagnetic insert), an extension set, and an adhesive-backed securement dressing. The powered driver uses a magnet to attach to the drive adapter hub. The operator manually presses the needle set through soft tissue to the bone, then activates the driver via a button to drill through the bone cortex until a change in pressure is felt. The device provides access for up to 24 hours. The output is the establishment of vascular access, which facilitates resuscitation and lifesaving fluid delivery. The device is designed for use in emergent, urgent, or medically necessary clinical contexts.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including verification and validation activities performed in accordance with 21 CFR 820.30 and relevant consensus standards (e.g., ANSI AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-12).
Technological Characteristics
Reusable battery-powered driver; housing made of co-polyester and 30% glass-fiber reinforced polypropylene. Uses a 32-bit ARM microcontroller. Powered by replaceable lithium-ion batteries. Features LED battery status display and button-operated switch. Needle set uses ferromagnetic attachment. Intermediate level disinfection required. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-12, and ISO 14971.
Indications for Use
Indicated for intraosseous access in the proximal tibia, distal tibia, and humeral head (proximal humerus) of adult and pediatric patients, and the distal femur in pediatric patients when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 10, 2022
Bard Access Systems Nasreen Al-Quaid Staff Regulatory Affairs Specialist 605 North 5600 West Salt Lake City, UT
Re: K223198
Trade/Device Name: BD Intraosseous Vascular Access System EMS Powered Driver (D001003) Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: MHC Dated: October 11, 2022 Received: October 13, 2022
Dear Nasreen Al-Quaid:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the
{1}------------------------------------------------
Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medicaldevice-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@tda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Joyce M. Whang -S
CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223198
#### Device Name
BD Intraosseous Vascular Access System EMS Powered Driver (D001003)
Indications for Use (Describe)
The BD Intraosseous Vascular Access System provides intraosseous access in the proximal tibia, distal tibia and humeral head (proximal humerus) of adult and pediatric patients, and the distal femur in pediatric patients when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Bard Access Systems, Inc. (Bard has joined BD) Special 510(k) Premarket Notification BD Intraosseous Vascular Access System EMS Powered Driver 510(k) Summary
# 510(k) Summary
## 21 CFR 807.92(a)
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (1)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows:
| | Submitter Name: | Bard Access Systems, Inc. (Bard has joined BD) |
|-----------------------|---------------------------------|------------------------------------------------------------|
| | Submitter Address: | 605 North 5600 West<br>Salt Lake City, UT 84116 |
| General<br>Provisions | Contact Person: | Nasreen A. Al-Quaid<br>Staff Regulatory Affairs Specialist |
| | Telephone Number: | 484.569.2865 |
| | Email: | Nasreenara.alquaid@bd.com |
| | Date of Preparation: | 11/07/2022 |
| | | Trade Name: |
| Subject<br>Device | Common Name: | Intraosseous Infusion System |
| | Regulation Number: | 21 CFR §880.5570 |
| | Regulation Classification Name: | Hypodermic, Single Lumen Needle |
| | Regulatory Class: | II |
{4}------------------------------------------------
| Product Code: | MHC | |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|
| Classification Panel: | General Hospital | |
| Trade Name: | BD Intraosseous Infusion System | |
| Common Name: | Intraosseous Infusion System | |
| Regulation Number: | 21 CFR §880.5570 | |
| Predicate<br>Device | Regulation Classification Name: | Hypodermic, Single Lumen Needle |
| Regulatory Class: | II | |
| Product Code: | MHC | |
| Classification Panel: | General Hospital | |
| 510(k) Status: | K203193 (Concurrence date: March 4, 2021) | |
| Device<br>Description | The BD Intraosseous Vascular Access System provides clinicians and emergency personnel with access to the intraosseous space for resuscitation and lifesaving fluid delivery for up to 24 hours. The BD Intraosseous Vascular Access System consists of the following:<br>•a single use hypodermic needle (with needle safety cap),<br>•a powered or manual driver to assist with needle insertion,<br>•an extension set, and<br>•an adhesive-backed securement dressing.<br><br>For insertions using the powered driver, the hypodermic needle includes a needle hub that mates with a stylet connected to a drive adapter hub. The drive adapter hub includes a ferromagnetic material that is attracted by the magnet in the powered driver and attaches to the powered driver prior to needle insertion. The BD Intraosseous Vascular Access System is an easy-grip, hand-held, battery-powered device used to assist in the insertion of the subject device needle through the bone cortex. The assembly of the hypodermic needle and stylet with connected drive adapter hub is referred to as the needle set. | |
{5}------------------------------------------------
K223198 Page 3 of 11
| Intended Use | The BD Intraosseous Vascular Access System is intended to provide clinicians and emergency personnel with access to the intraosseous space. |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | The BD Intraosseous Vascular Access System provides intraosseous access in the proximal tibia, distal tibia and humeral head (proximal humerus) of adult and pediatric patients, and the distal femur in pediatric patients when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases for up to 24 hours. |
| Technological<br>Characteristics | At a high level, the subject device, BD Intraosseous Vascular Access System EMS Powered Driver and the cited predicate device are based on the following same technological and use elements:<br>Intended Use, Indications for Use, Target Patient Population Reusable Powered Intraosseous Driver- used to provide intraosseous access Power Driver Attachment to Needle - at distal end of device and same insertion technique |
{6}------------------------------------------------
# K223198
# Page 4 of 11
| • | Use of the same Needle Set Configurations – choice of use based on tissue depth | | | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| • | Use of LED Battery Display - indication the status of the battery | | | |
| • | Use of an Operating Switch to run the device | | | |
| • | Use of a Motor and Circuit Board Assembly and Microcontroller ARM Processor | | | |
| • | Use of Lithium-ion Batteries to run the device | | | |
| The modifications made to the predicate device include modifications to the design and technology, software and labeling. The following technological differences exist between the subject and predicate devices: | | | | |
| 1. | Design and Technology; | | | |
| • | Change in driver shape, placement of LED battery indicator lights &operating switch. | | | |
| • | Housing and Rear Cap material difference due to shape modification | | | |
| • | Hardware and battery changes to facilitate replaceable battery | | | |
| 2. | Software: Modification to structure of software (firmware). | | | |
| 3. | Labeling: Separate IFU to support charging differences of the device | | | |
| The following tables provides a summary comparison between the subject and predicate device. | | | | |
| Attribute | Subject Device<br>BD Intraosseous Vascular Access<br>System EMS Powered Driver | Predicate device<br>BD Intraosseous Infusion System<br>(K203193) | Discussion of Substantial<br>Equivalence | |
| Owner | Bard Access Systems, Inc. | Bard Access Systems, Inc. | Identical | |
| 510(k) status | Subject of this 510(k) | K203193 | Identical | |
| Intended Use | The BD Intraosseous Vascular Access<br>System is intended to provide clinicians | The BD Intraosseous Vascular Access<br>System is intended to provide clinicians | Identical | |
| Indications for<br>use | and emergency personnel with access<br>to the intraosseous space<br>The BD Intraosseous Vascular Access<br>System provides intraosseous access in<br>the proximal tibia, distal tibia and<br>humeral head (proximal humerus) of<br>adult and pediatric patients, and the<br>distal femur in pediatric patients when<br>intravenous access is difficult or<br>impossible to obtain in emergent, urgent,<br>or medically necessary cases for up to<br>24 hours. | and emergency personnel with access to<br>the intraosseous space<br>The BD Intraosseous Vascular Access<br>System provides intraosseous access in<br>the proximal tibia, distal tibia and<br>humeral head (proximal humerus) of<br>adult and pediatric patients, and the<br>distal femur in pediatric patients when<br>intravenous access is difficult or<br>impossible to obtain in emergent, urgent,<br>or medically necessary cases for up to<br>24 hours. | Identical | |
| Powered<br>Driver Use | Reusable | Reusable | Identical | |
| Intended<br>Patient<br>Population | Adults and Pediatrics who require short-<br>term (less than 30 days intraosseous<br>infusion therapy | Adults and Pediatrics who require short-<br>term (less than 30 days intraosseous<br>infusion therapy | Identical | |
| General Device<br>Description | The BD Intraosseous Vascular Access<br>System provides clinicians and<br>emergency personnel with access to the<br>intraosseous space for resuscitation and<br>lifesaving fluid delivery for up to 24<br>hours. The BD Intraosseous Vascular<br>Access System consists of the following:<br>•a single use hypodermic needle (with<br>needle safety cap),<br>•a powered or manual driver to assist<br>with needle insertion,<br>•an extension set, and<br>•an adhesive-backed securement<br>dressing.<br>For insertions using the powered driver,<br>the hypodermic needle includes a<br>needle hub that mates with a stylet<br>connected to a drive adapter hub. The<br>drive adapter hub includes a magnetic<br>insert that attaches to the powered<br>driver prior to needle insertion. The BD<br>Intraosseous Vascular Access System is<br>an easy-grip, hand-held, battery-<br>powered device used to assist in the<br>insertion of the subject device needle<br>through the bone cortex. The assembly<br>of the hypodermic needle and stylet with<br>connected drive adapter hub is referred<br>to as the needle set. | The BD Intraosseous Vascular Access<br>System provides clinicians and<br>emergency personnel with access to the<br>intraosseous space for resuscitation and<br> | Identical | |
| | | | | |
| Insertion Site…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.