← Product Code [MEG](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MEG) · K141640

# AUTOMATICALLY RETRACTABLE SAFETY SYRINGES WITH FIXED NEEDLE (K141640)

_Shantou Wealy Medical Instrument Co.,Ltd · MEG · Nov 19, 2014 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MEG/K141640

## Device Facts

- **Applicant:** Shantou Wealy Medical Instrument Co.,Ltd
- **Product Code:** [MEG](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MEG.md)
- **Decision Date:** Nov 19, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5860
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The Automatically Retractable Safety Syringes with Fixed Needle devices are indicated for use where a safe and reliable method for intramuscular and subcutaneous injection of medication in a patient is desired.

## Device Story

Automatically Retractable Safety Syringe with Fixed Needle; sterile, single-use, disposable piston syringe. Consists of calibrated hollow barrel and movable plunger. Operates like conventional hypodermic syringe; features spring-action mechanism that automatically retracts contaminated needle into barrel immediately following injection. Used by healthcare professionals for intramuscular and subcutaneous medication delivery. Retraction mechanism ensures needle remains safely inside barrel post-use, reducing risk of needlestick injuries during disposal.

## Clinical Evidence

Bench testing only. Performance testing included design verification, design validation, biocompatibility, and sterilization. Results confirmed the 1ml, 3ml, and 10ml devices are equivalent to the 5ml predicate device in design, materials, and function.

## Technological Characteristics

Piston syringe with integrated spring-actuated needle retraction mechanism. Sterile, single-use, disposable. Available in 1ml, 3ml, 10ml volumes with various needle gauges (20G-30G). Fixed needle design. No electronic components or software.

## Regulatory Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

## Predicate Devices

- Automatically Retractable Safety Syringes with Fixed Needle (5 ml) ([K113587](/device/K113587.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, resembling a bird-like shape.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 19,2014

Shantou Wealy Medical Instrument Company Limited Ms. Charmaine Henderson Regulatory Consultant 570 Silverado Drive Lafayette, CA 94549

Re: K141640

Trade/Device Name: Automatically Retractable Safety Syringe with Fixed Needle (1ml, 3ml, 10ml) Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: MEG Dated: August 25, 2014 Received: August 29, 2014

Dear Ms. Henderson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Susan Runco DDS, MA

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{2}------------------------------------------------

## Indications for Use

#### 510(k) Number (if known) K141640

#### Device Name

Device Name
Automatically Retractable Safety Syringe with Fixed Needle (Iml, 3ml, 10ml with 30G, 29G, 27G, 23G, 23G, 21G, or 20G)

#### Indications for Use (Describe)

Indically Retractable Safety Syringe with Fixed Needle (Iml, 3ml, 10ml with 30G, 29G, 27G, 26G, 25G, 24G, 23G, 22G, 21G, or 20G) is a sterile, single-use, disposable and non-reusable, retractable safety syringe which is intended to provide a safe and reliable method for intramuscular and subcutaneous injection of medication into a patient.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the The burden time for this onloor on existing data sources, gather and maintain the data needed and complete time to review instruction. Send commonts regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### 510(k) Summary Shantou Wealy Medical Instrument Co. Ltd Automatically Retractable Safety Syringes with Fixed Needle

#### Date of Summary Preparation: May 2, 2014

- Submitter: Shantou Wealy Medical Instrument Co, Ltd North Jin Huan Road Shantou, Guangdong, China 515064 Tel: +86-754-88218123 Fax: +86-754-82121654
Establishment Registration Number: 3005202235

| <b>Company Contact:</b> | Charmaine Henderson           |
|-------------------------|-------------------------------|
|                         | Regulatory Affairs Consultant |
|                         | Office: 925.212.5264          |
|                         | Email: crhinlaf@gmail.com     |

- Trade name: Automatically Retractable Safety Syringes with Fixed Needle (1ml, 3ml, 10ml)
- Device name (Common Name): Piston syringe

Device Classification: Class II

Product Code: MEG

Regulation Number:

Requlation Description: piston syringe

Predicate Device:

Shantou Wealy Medical Instrument Co. Ltd Automatically Retractable Safety Syringes with Fixed Needle (5 ml) 510(k) K113587 Concurrence date: February 24, 2012

### General Device Description:

The Automatically Retractable Safety Syringes with Fixed Needle devices are piston syringes, intended for medical purposes and consist of a calibrated hollow barrel and a movable plunger. The syringe works like a conventional hypodermic syringe except that the contaminated needle is retracted inside the syringe immediately after patient injection. The needle retracting mechanism is activated by a spring action mechanism after injection is completed. The exposed needle remains safely inside the empty syringe barrel for disposal. The Automatically Retractable Safety Syringes with Fixed Needle are sterile, single use, disposable and non-reusable.

{4}------------------------------------------------

# Indications for Use:

The Automatically Retractable Safety Syringes with Fixed Needle devices are indicated for use where a safe and reliable method for intramuscular and subcutaneous injection of medication in a patient is desired.

# Summary of Technological Characteristics:

The modified Automatically Retractable Safety Syringes with Fixed Needle share the same technological characteristics as its predicate device. The characteristics include the same design, same syringe type, same materials, same principle of operation, same safety features, and same intended use.

# Summary of Performance Testing:

The modified Automatically Retractable Safety Syringes with Fixed Needle (1ml, 3ml and 10 ml) devices are substantially equivalent to the predicate device in design, materials, function and intended use. The performance testing of the modified Automatically Retractable Safety Syringes with Fixed Needle devices were reviewed and tested appropriately for design verification, design validation, biocompatibility and sterilization. The test results concluded that the modified Automatically Retractable Safety Syringes with Fixed Needle devices (1ml, 3ml and 10 ml) are substantially equivalent to the predicate device; Automatically Retractable Safety Syringes with Fixed Needle (5ml).

# Substantial Equivalence:

Shantou Wealy Medical Instrument Co, Ltd considers the modified Automatically Retractable Safety Syringes with Fixed Needle (1ml, 3ml, and 10ml) product performance to be substantially equivalent to its predicate device, Automatically Retractable Safety Syringes with Fixed Needle (5ml) because there are no changes to the product performance specifications or device functional technology.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MEG/K141640](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MEG/K141640)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
