← Product Code [MEG](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MEG) · K063755

# PORTEX HYPODERMIC NEEDLE-PRO FIXED NEEDLE SYRINGE (K063755)

_Smiths Medical Asd, Inc. · MEG · Feb 27, 2007 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MEG/K063755

## Device Facts

- **Applicant:** Smiths Medical Asd, Inc.
- **Product Code:** [MEG](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MEG.md)
- **Decision Date:** Feb 27, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5860
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

This device is intended for aspiration and injection of fluids. The needle protection device covers the needle after use to help prevent needle sticks. This device is intended for aspiration and injection of U-100 Insulin Only. The needle protection device covers the needle after use to help prevent needle sticks.

## Device Story

Portex® Hypodermic Needle-Pro® Fixed Needle Syringe consists of a graduated syringe barrel with a permanently affixed needle and an integral, hinged safety sheath. Used for fluid aspiration and injection (including U-100 Insulin). After injection, user manually rotates the sheath to the desired position and presses the needle into the sheath using a one-handed technique to contain the needle, reducing risk of accidental needle sticks. Device is then discarded in a sharps container. Operated by clinicians or patients in clinical or home settings. Provides physical needle containment to prevent sharps injuries.

## Clinical Evidence

Simulated clinical use studies were conducted for both Insulin and TB syringe configurations. Studies confirmed the device could be used effectively for fluid injection and that the needle could be successfully shielded within the protective mechanism after use. No human clinical trials were required; evidence based on bench testing and simulated use.

## Technological Characteristics

Graduated syringe with permanently affixed needle and hinged polypropylene safety sheath. Manual activation mechanism. Materials include polypropylene. Device is a piston syringe with single lumen needle. No electronic components, software, or energy sources.

## Regulatory Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

## Predicate Devices

- Terumo SurGuard™ Insulin, Allergy and General Use Safety Syringe ([K024249](/device/K024249.md))

## Reference Devices

- Needle-Pro® ([K011925](/device/K011925.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the text "K063755 1 of 2 smiths". The text appears to be handwritten, with the exception of the word "smiths", which is in a bold, sans-serif font. The numbers "K063755" are at the top of the image, followed by "1 of 2" and then "smiths" at the bottom. The image is black and white.

FEB 2 7 2007

## SECTION 5, 510(k) Summary

## Company Information:

Smiths Medical ASD, Inc. 10 Bowman Drive Keene, NH 03431 (603) 352-3812, prompt 4, ext 2493 Contact: Brian D. Farias Regulatory Affairs Manager

Summary Prepared: December 18, 2006

## Product Name:

## Trade Name: Portex® Hypodermic Needle-Pro® Fixed Needle Syringe

Common Name: Fixed Needle Hypodermic Syringe with attached needle protection

Classification Name: 880.5860 Piston Syringe with Single Lumen Needle with antistick

## Predicate Device(s):

K024249 (Terumo Medical Corporation) Terumo SurGuard™ Insulin, Allergy and General Use Safety Syringe and alternate brand name Portex "Hypodermic Needle-Pro® Fixed Needle Syringe (referred to as the "Terumo SurGuard™ Safety Syringe" throughout the remainder of this submission).

Please note that the attached needle protection (brand name Needle-Pro®) of the predicate device is supplied to Terumo Medical Corporation by Smiths Medical holds a 510(k) for this device, K011925. The needle protection component of the proposed device is essentially the same as that cleared above except it is made from a different polypropylene.

### Device Description:

The Portex® Hypodermic Needle-Pro® Fixed Needle Syringe is a graduated hypodermic syringe with a permanently affixed needle and integral needle safety sheath. The safety sheath rotates so it can be adjusted to the desired position relative to the needle bevel and syringe graduations. The syringe is used to inject fluids into the body. The Insulin syringe barrel is graduated in U-100 Insulin units and the TB syringe is graduated in milliliters (ml). After the procedure is completed, the needle is pressed into the sheath using a one-handed technique. The needle enters the protective sheath and is contained within the sheath. The device is then discarded into a sharps container.

#### Smiths Medical ASD, Inc.

#### Anesthesia and Safety Devices Division

10 Bowman Drive Keene NH 03431 USA Tel: +1 603 352 3812 Fax: +1 603 352 3703 www.smiths-medical.com

{1}------------------------------------------------

Kc 63755-
2 of 2

### Indications for Use:

This device is intended for aspiration and injection of fluids including Insulin. The needle protection device covers the needle after use to help prevent needle sticks. [See pages 9 and 10 for the specific indication for Insulin and TB (General Use) syringes)

### Technological Characteristics:

The proposed and predicate devices have a permanently affixed needle and a hinged style protective sheath that is manually activated after use.

#### Non-Clinical Data:

Bench testing confirms that the proposed device and the predicate device have similar performance specifications based on the applicable standards for this device and FDA guidance for devices with sharps injury prevention features.

#### Clinical Data:

Simulated clinical use studies were conducted for Insulin and TB syringes which confirmed that the device could be used effectively with the needle shielded inside the protection device after use.

#### Conclusion:

The bench testing and simulated clinical use studies conducted demonstrate that the proposed device is safe and effective and is substantially equivalent to the predicate device.

Very truly yours,

SMITHS MEDICAL ASD, INC.

Brian D. Farias Regulatory Affairs Manager

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - U.S.A." is arranged in a circular pattern around the eagle. The eagle is black, and the text is also black. The background is white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Brian D. Farias Regulatory Affairs Manager Smiths Medical ASD, Incorporated 10 Bowman Drive Keene, New Hampshire 03431

FEB 2 7 2007

Re: K063755

Trade/Device Name: Portex® Hypodermic Needle-Pro® Fixed Needle TB Syringe, Portex® Hypodermic Needle-Pro® Fixed Needle Insulin Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Il Product Code: MEG Dated: December 18, 2006 Received: December 19, 2006

Dear Mr. Farias:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Farias

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

く

Sincerely yours,

Susan Purnas

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## SECTION 4B, Indications for Use Statement

# Indications for Use

| 510(k) Number (if known): | K063755                                                |
|---------------------------|--------------------------------------------------------|
| Device Name:              | Portex® Hypodermic Needle-Pro® Fixed Needle TB Syringe |

Indications for Use:

This device is intended for aspiration and injection of fluids. The needle protection device covers the needle after use to help prevent needle sticks.

Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

c. Intyre U.M.

K463755

{5}------------------------------------------------

# SECTION 4A, Indications for Use Statement

# Indications for Use

| 510(k) Number (if known): | K063755                                                     |
|---------------------------|-------------------------------------------------------------|
| Device Name:              | Portex® Hypodermic Needle-Pro® Fixed Needle Insulin Syringe |

Indications for Use:

This device is intended for aspiration and injection of U-100 Insulin Only. The needle protection device covers the needle after use to help prevent needle sticks.

Prescription Use Over-The-Counter Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page _1 of 1 _

Clm 2

大�6375;

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MEG/K063755](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/MEG/K063755)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
