MODIFICATION TO AUTOFUSER

K090300 · Ace Medical Us, LLC · MEB · Apr 30, 2009 · General Hospital

Device Facts

Record IDK090300
Device NameMODIFICATION TO AUTOFUSER
ApplicantAce Medical Us, LLC
Product CodeMEB · General Hospital
Decision DateApr 30, 2009
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.5725
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Autofuser family of ambulatory infusion pumps with integrated administration set, either separately or as part of a convenience kit, is intended for general infusion use. Routes of administration include intravenous, percutaneous, subcutaneous, intra-arterial and epidural, and into intra-operative (soft tissue/body cavity) sites. Within the Autofuser family are pump models intended for patient-controlled infusion using the integrated bolus button. General infusion uses include continuous infusion of a local anesthetic near a nerve for regional anesthesia and pain management for pre-operative, perioperative and postoperative surgery.

Device Story

Disposable elastomeric infusion pump system; consists of balloon-style medication reservoir and integrated flow-rate controlling administration set. Used for continuous infusion of medications, including local anesthetics for pain management. Some models include integrated bolus button for patient-controlled infusion. Packaged as standalone pump or within convenience kits containing filling syringe, catheter, introducer, and carrying pouch. Operated by clinicians or patients in ambulatory or clinical settings. Provides controlled medication delivery via mechanical pressure; no electronic components or software.

Clinical Evidence

Bench testing only. Performance testing verified and validated modifications to the system, demonstrating that the device meets predetermined performance specifications.

Technological Characteristics

Disposable elastomeric infusion pump; balloon-style medication reservoir; integrated flow-rate controlling administration set. Non-electronic, mechanical operation. Kit components include filling syringe, catheter, catheter introducer, and carrying pouch.

Indications for Use

Indicated for patients requiring general infusion therapy, including regional anesthesia and pain management (pre-operative, perioperative, postoperative). Routes: intravenous, percutaneous, subcutaneous, intra-arterial, epidural, and intra-operative soft tissue/body cavity sites.

Regulatory Classification

Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K090300 APR 30 2009 #### 510(k) SUMMARY #### Ace Medical Autofuser Elastomeric Infusion Pump System # Submitted by: Ace Medical US, LLC 9102 Turfway Bend Drive, Powell, OH 43065-8497 Contact Person: Phil Marsh Phone: (614) 733-3601 Fax: (614) 733-3602 Establishment Registration: 3005627389 Date Prepared: January 31, 2009 # Device Information: Name of Device: Autofuser Elastomeric Infusion Pump System Common Name: Elastomeric Infusion Pump & Procedure Kit Classification Name: Elastomeric Infusion Pump Product code: MEB - Elastomeric Infusion Pump Device Classification: Class II Regulation Number: 880.5725 Classification Panel: 80, General Hospital and Personal Use Device # Predicate Device Ace Medical Autofuser Elastomeric Infusion Pump System (K060258) #### Intended Use / Indications for Use The Autofuser family of ambulatory infusion pumps with integrated administration set, either separately or as part of a convenience kit, is intended for general infusion use. Routes of administration include intravenous, percutaneous, subcutaneous, intra-arterial and epidural, and into intra-operative (soft tissue/body cavity) sites. Within the Autofuser family are pump models intended for patient-controlled infusion using the integrated bolus button. General infusion uses include continuous infusion of a local anesthetic near a nerve for regional anesthesia and pain management for pre-operative, perioperative and postoperative surgery. {1}------------------------------------------------ # Technological Characteristics The Autofuser system consists of a family of disposable infusion pumps and associated procedure kits. The pump is comprised of a balloon-style medication reservoir and an integrated flowrate-controlling administration set. When the Autofuser pump is packaged in a procedure kit, the kit includes legally marketed components such as filling syringe, catheter, catheter introducer, and pump carrying pouch. # Performance Data Performance testing was performed to verify/validate the modifications to the system. Test results demonstrate that the modified Autofuser system met its predetermined performance specifications. # Substantial Equivalence The modified Autofuser system is as safe and effective as the previously-cleared Autofuser system. The modified device has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor technological differences between the modified device and its predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the modified Autofuser is as safe and effective as the predicate device. Thus, the modified Autofuser system is substantially equivalent to the previously-cleared predicate Autofuser system. {2}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing wings or feathers. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is in all caps and appears to be in a sans-serif font. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # APR 3 0 2009 Mr. Andrew N. Lamborne Ace Medical US, LLC 9102 Turfway Bend Drive Powell, Ohio 43065 Re: K090300 Trade/Device Name: Autofuser Elastomeric Infusion Pump System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: MEB Dated: March 24, 2009 Received: March 31, 2009 Dear Mr. Lamborne: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {3}------------------------------------------------ Page 2- Mr. Lamborne You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Antan O. hum for Susan Runner, D.D.S., MA Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use Statement 510(k) Number (if known): K090300 Device Name: Autofuser Elastomeric Infusion Pump System Indications for Use: The Autofuser family of ambulatory infusion pumps with integrated administration set, either separately or as part of a convenience kit, is intended for general infusion use. Routes of administration include intravenous, percutaneous, intra-arterial and epidural, and into intra-operative (soft tissue/body cavity) sites. Within the Autofuser family are pump models intended for patient-controlled infusion using the integrated bolus button. General infusion uses include continuous infusion of a local anesthetic near a nerve for regional anesthesia and pain management for pre-operative, perioperative and postoperative surgery. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: k093000 Page of of
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...