← Product Code [LZG](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/LZG) · K140439

# OMNIPOD INSULIN MANAGEMENT SYSTEM (K140439)

_Insulet Corporation · LZG · Nov 13, 2014 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/LZG/K140439

## Device Facts

- **Applicant:** Insulet Corporation
- **Product Code:** [LZG](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/LZG.md)
- **Decision Date:** Nov 13, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5725
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The OmniPod® Insulin Management System is intended for subcutaneous (below the skin) delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin and for the quantitative measurement of glucose in fresh capillary whole blood (in vitro) from the fingertip. The glucose measurements should not be used for the diagnosis of or screening for diabetes or for neonatal use. The PDM glucose meter is intended for single-patient use and should not be shared. OneTouch® Verio™ Test Strips are used with the built-in blood glucose meter with One Touch Verio technology for the quantitative measurement of blood glucose in fresh capillary whole blood drawn from the fingertips. OneTouch Verio Control Solutions are used to check that the meter and test strips are working together properly and that the test is performing correctly.

## Device Story

Tubeless insulin pump system comprising a wearable Pod and a Personal Diabetes Manager (PDM) remote controller. PDM integrates a blood glucose meter using OneTouch Verio technology. Patient applies Pod subcutaneously; PDM controls insulin delivery rates and performs glucose measurements from capillary blood samples. System designed for diabetes management; provides automated insulin delivery based on user-defined settings. Healthcare providers use PDM data to adjust therapy; patients use system for daily glucose monitoring and insulin administration. Benefits include simplified insulin delivery and integrated glucose monitoring.

## Clinical Evidence

Clinical evaluation conducted per ISO 15197:2003(E) Section 8 (User Performance Evaluation) involving lay users and healthcare professionals. Study confirmed system accuracy and performance equivalence to the predicate device. Non-clinical testing included repeatability, intermediate precision, method comparison, linearity, environmental stress (temperature/humidity), hematocrit interference, and altitude testing.

## Technological Characteristics

Tubeless insulin pump; PDM with integrated blood glucose meter. Operates on wireless communication between Pod and PDM. Tested to IEC 60601-1 (electrical safety) and IEC 60601-2-24 (infusion pump performance). Software classified as Major level of concern. Environmental operating range: 50°F to 104°F (10°C to 40°C); 20% to 90% relative humidity.

## Regulatory Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

## Predicate Devices

- OmniPod Insulin Management System ([K122953](/device/K122953.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 13,2014

Insulet Corporation Antonette M. DeLeo, MS, RAC Senior Regulatory Affairs Specialist 600 Technology Park Drive, Suite 200 Billerica, MA 01821

Re: K140439

Trade/Device Name: OmniPod Insulin Management System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: LZG, NBW Dated: October 10, 2014 Received: October 14, 2014

Dear Ms. DeLeo,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Susan Runno DDS, MA

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K140439

Device Name OmniPod Insulin Management System

#### Indications for Use (Describe)

The OmniPod® Insulin Management System is intended for subcutaneous (below the skin) delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin and for the quantitative measurement of glucose in fresh capillary whole blood (in vitro) from the fingertip.

The glucose measurements should not be used for the diagnosis of or screening for diabetes or for neonatal use. The PDM glucose meter is intended for single-patient use and should not be shared.

OneTouch® Verio™ Test Strips are used with the built-in blood glucose meter with One Touch Verio technology for the quantitative measurement of blood glucose in fresh capillary whole blood drawn from the fingertips.

OneTouch Verio Control Solutions are used to check that the meter and test strips are working together properly and that the test is performing correctly.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

EF

{3}------------------------------------------------

# 510(k) Summary

| Submitter:                        | Insulet Corporation<br>600 Technology Park Drive<br>Suite 200<br>Billerica, MA 01821<br>USA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                   | Antonette M. DeLeo, MS, RAC<br>Senior Regulatory Affairs Specialist<br>Telephone: 978-600-7443<br>Fax: 781-600-4329<br>E-mail: adeleo@insulet.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Date Prepared:                    | November 7, 2014                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Trade Name:                       | OmniPod Insulin Management System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Common Name:                      | Insulin Infusion Pump                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Classification<br>Name:           | Pump, Infusion, Insulin                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Product Code<br>(Classification): | LZG                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Product Code<br>(Subsequent):     | NBW                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Product Code<br>(Subsequent):     | LFR                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Predicate Device:                 | OmniPod Insulin Management System<br>K122953                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Device<br>Description:            | The OmniPod® Insulin Management System is a tubeless<br>insulin pump intended for subcutaneous delivery of insulin at<br>set and variable rates for the management of diabetes mellitus<br>in persons requiring insulin and for the quantitative<br>measurement of glucose in fresh capillary whole blood (in<br>vitro) from the fingertip. This System is comprised of two<br>primary components, the insulin pump (Pod) and the remote<br>controller (Personal Diabetes Manager). The Personal<br>Diabetes Manager incorporates LifeScan OneTouch Verio<br>blood glucose measurement capability through a built in<br>blood glucose meter. The blood glucose meter uses One |

{4}------------------------------------------------

|                                                 | Touch Verio Test Strips. OneTouch Verio Level 3 (Mid) and<br>OneTouch Verio Level 4 (High) Control Solutions are used to<br>check that the meter and test strips are working together<br>properly. OneTouch Verio Test Strips are included in the<br>Starter Kit. OneTouch Verio Control Solutions are not<br>included in the Starter Kit and must be purchased separately.                                                     |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Statement of<br>Intended Use:                   | The OmniPod Insulin Management System is intended for use<br>in the management of insulin therapy and optional blood<br>glucose monitoring by patients with diabetes mellitus.                                                                                                                                                                                                                                                  |
| Statement of<br>Indications for Use:            | The OmniPod® Insulin Management System is intended for<br>subcutaneous delivery of insulin at set and variable rates for<br>the management of diabetes mellitus in persons requiring<br>insulin and for the quantitative measurement of glucose in<br>fresh capillary whole blood (in vitro) from the fingertip.                                                                                                                |
|                                                 | The glucose measurements should not be used for the<br>diagnosis of or screening for diabetes or for neonatal use. The<br>PDM glucose meter is intended for single-patient use and<br>should not be shared.                                                                                                                                                                                                                     |
|                                                 | OneTouch® Verio™ Test Strips are used with the built-in<br>blood glucose meter with One Touch Verio technology for the<br>quantitative measurement of blood glucose in fresh capillary<br>whole blood drawn from the fingertips.                                                                                                                                                                                                |
|                                                 | OneTouch Verio Control Solutions are used to check that the<br>meter and test strips are working together properly and that<br>the test is performing correctly.                                                                                                                                                                                                                                                                |
| Summary of<br>Technological<br>Characteristics: | The proposed device has the same technological<br>characteristics and is similar in design and configuration as<br>compared to the predicate device.                                                                                                                                                                                                                                                                            |
| Summary of Non-<br>Clinical Data:               | Insulet completed the appropriate validation and verification<br>activities required by the Guidance for Industry and FDA<br>Staff – Total Product Life Cycle: Infusion Pump – Premarket<br>Notification [510(k)] Submissions Draft Guidance and other<br>guidance, as applicable. The following performance and<br>safety testing has confirmed the proposed device to be<br>substantially equivalent to the predicate device: |
|                                                 | ● Physical Characteristics: the proposed device has<br>been tested and successfully met physical dimension<br>and weight requirements;                                                                                                                                                                                                                                                                                          |
|                                                 | ● Drop and Vibration: the proposed device has been                                                                                                                                                                                                                                                                                                                                                                              |

{5}------------------------------------------------

tested and successfully met all of the relevant requirements for drop and vibration testing per IEC 60601-2-24;

- Hardware: the proposed device has been tested meets . all of its associated hardware specifications;
- . Software: documentation was prepared and submitted for a MAJOR level of concern device in accordance with FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices:
- . Electrical Safety: the proposed OmniPod Insulin Management System has been tested and successfully passed all of the relevant sections of IEC 60601-1 Medical electrical equipment, General requirements for Safety:
- . RF Wireless Safety and Performance: the proposed device has been tested and verified to ensure proper wireless communication between the Pod and PDM;
- . Electromagnetic Interference: the proposed device has been tested and successfully met all of the relevant sections (Radiated emissions, Electrostatic discharge immunity frequency. electromagnetic field immunity, and Power frequency magnetic field immunity test) to satisfy compliance:
- . Alarm Sound Level: the proposed device has been tested and successfully met requirements for sound pressure level;
- Environmental: the proposed device has been tested and successfully met IEC 60601-2-24 requirements for temperature, humidity and atmospheric pressure;
- . Battery Life: the proposed device has been tested and successfully met the requirement of sufficient battery power for expected use:
- . Shipping: the proposed device has been tested and successfully met requirements for aesthetic and functional integrity after shipping.

LifeScan completed appropriate testing to verify the product design and the laboratory/non-clinical blood glucose meter performance accuracy. The following evaluations were performed and have confirmed the proposed device to be substantially equivalent to the predicate device:

{6}------------------------------------------------

|                           | Repeatability Precision Intermediate Precision Method Comparison (System Accuracy Capillary Sample) Device Comparison (Accuracy Equivalence) Linearity Effect of Temperature and Relative Humidity  Operating temperature: 50°F to 104°F (10°C to 40°C) Operating relative humidity range: 20% to 90%   Hematocrit Sample Identification Sample Volume Interferences: Exogenous and Endogenous Altitude Pre and Post Cleaning and Disinfection (System Accuracy)                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary of Clinical Data: | LifeScan completed a clinical study, Clinical Evaluation – Lay User and HCP Comparison to a Validated Method, Instructions for Use and System Use Evaluation, to fulfill the requirements of ISO 15197:2003 (E) – Section 8 User Performance Evaluation. This evaluation confirmed the proposed device to be substantially equivalent to the predicate device:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Conclusion from Data:     | Insulet believes that the information and data provided in this submission clearly describes the proposed device and demonstrates that the device is adequately designed for the labeled indication for use. Performance, verification and validation testing was conducted to characterize performance of the proposed device and the predetermined acceptance criteria was met. Results of this testing have documented that the proposed device is substantially equivalent to the predicate device and is suitable for the labeled indication for use. Therefore, the proposed OmniPod Insulin Management System is substantially equivalent to the identified predicate.<br><br>Insulet Corporation has demonstrated that the proposed OmniPod Insulin Management System is substantially equivalent to the predicate device based upon indications for use, design, test results and the same fundamental scientific technology. |

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/LZG/K140439](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/LZG/K140439)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
