CELSITE PEDIATRIC VENOUS SYSTEM
Device Facts
| Record ID | K952435 |
|---|---|
| Device Name | CELSITE PEDIATRIC VENOUS SYSTEM |
| Applicant | B.Braun Medical, Inc. |
| Product Code | LJT · General Hospital |
| Decision Date | Apr 17, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 880.5965 |
| Device Class | Class 2 |
| Attributes | Therapeutic, Pediatric |
Indications for Use
The CELSITE® Pediatric Venous Port is a implanted catheter system which allows safe, repeated access to the patient’s bloodstream. The chamber and catheter design can be used for the administration of medications and fluids.
Device Story
Implantable venous access system; consists of port chamber and catheter. Used for repeated administration of medications and fluids into patient bloodstream. Intended for pediatric population. Device implanted surgically; allows clinician access via percutaneous needle puncture of port septum. Facilitates long-term therapy; reduces need for repeated venipuncture. Design similar to existing adult ports but scaled for pediatric use.
Clinical Evidence
No clinical data provided. Safety and effectiveness supported by established clinical history of implanted ports, literature review, and MDR analysis indicating low risk profile relative to high implantation volume.
Technological Characteristics
Implantable port and catheter system. Materials tested per Tripartite Guidance for Plastics. Design features a chamber and catheter for intravascular infusion. Sterilization method not specified.
Indications for Use
Indicated for pediatric patients requiring safe, repeated access to the bloodstream for the administration of medications and fluids.
Regulatory Classification
Identification
A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.
Predicate Devices
- CELSITE™ Implantable Drug Delivery System (K902401)
- PORT-A-CATH LOW PROFILE IMPLANTABLE VENOUS ACCESS (K942024)
Related Devices
- K962230 — CELSITE PORT WITH PRECONNECTED CATHETER · B.Braun Medical, Inc. · Dec 16, 1996
- K061424 — CELSITE/ CELSITE CONCEPT, ACCESS PORTS · Aesculap, Inc. · Jul 31, 2006
- K031844 — VAXCEL WITH PASV PORT (LOW PROFILE), MODELS 45-236 AND 45-238 · Boston Scientific Corp · Jul 15, 2003
- K960350 — THERAPORT VASCULAR ACCESS SYSTEM; MODEL 1001, 1002, 1601, 1602 · Biocontrol Technology, Inc. · Apr 15, 1996
- K093414 — PORT-A-CATH POWER P.A.C. IMPLANTABLE VENOUS ACCESS SYSTEMS · Smiths Medical Asd, Inc. · Mar 1, 2010