HR POLYURETHANE PICC AND MIDLINE CATHETERS
Device Facts
| Record ID | K964556 |
|---|---|
| Device Name | HR POLYURETHANE PICC AND MIDLINE CATHETERS |
| Applicant | Johnson & Johnson Medical, Inc. |
| Product Code | LJS · General Hospital |
| Decision Date | Apr 16, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 880.5970 |
| Device Class | Class 2 |
Indications for Use
Peripherally inserted central venous (PICC) and midline catheters placed for venous access or prolonged intravenous therapy.
Device Story
Polyurethane catheters for venous access/prolonged IV therapy; single or dual lumen configurations; optional flushable guidewire; sterile individual units or insertion kits. Insertion via introducer sheath. PICC tip location: superior vena cava; midline tip location: axilla. Used by clinicians in clinical settings for vascular access. Benefits: provides reliable venous access for therapy administration.
Clinical Evidence
Bench testing only. Testing included flow rate, tensile strength (body and hub), catheter stiffness, elongation, leakage, burst pressure, catheter collapse, biocompatibility (ISO 10993), and sterility (LAL procedure for non-pyrogenicity). All tests passed.
Technological Characteristics
Polyurethane catheter material; single/dual lumen; sterile, pyrogen-free; includes introducer sheath and optional flushable guidewire. Biocompatibility per ISO 10993.
Indications for Use
Indicated for patients requiring venous access or prolonged intravenous therapy. PICC catheters are indicated for tip placement in the superior vena cava; midline catheters are indicated for tip placement below the shoulder at the level of the axilla.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
Predicate Devices
- L-Cath PICC and midline
- One-Cath PICC and midline
- Arrow PICC
- SoloPICC and midline
- Per-Q-Cath PICC
- V-Cath PICC and midline
- SiL-Cath PICC and midline catheters
Related Devices
- K964447 — PICC CATHETER/MIDLINE CATHETER · Menlo Care, Inc. · Apr 16, 1997
- K951620 — MIDMARK MIDLINE CATHETER · Menlo Care, Inc. · Apr 4, 1996
- K963257 — PERIPHERALLY INSERTED MEDLINE CATHETER · Arrow Intl., Inc. · Mar 4, 1997
- K091586 — VASCU-PICC AND MIDLINE CATHETERS, SINGLE, DOUBLE AND TRIPLE LUMEN · Medcomp · Jul 23, 2009
- K252373 — BioFlo Hybrid PICC with ENDEXO Technology and PASV Valve Technology; BioFlo PICC with ENDEXO Technology and PASV Valve Technology; BioFlo PICC with ENDEXO Technology; BioFlo Midline Catheter; Xcela Power Injectable PICC; Xcela PICC with PASV Valve Technology; Xcela Hybrid PICC with PASV Valve Technology · Spectrum Vascular · Dec 16, 2025