← Product Code [LJS](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/LJS) · K963806

# CLAMPLESS VALVED CATHETER - TUNNELED (CVC-T) (K963806)

_Bmw Medical, Inc. · LJS · Dec 12, 1996 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/LJS/K963806

## Device Facts

- **Applicant:** Bmw Medical, Inc.
- **Product Code:** [LJS](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/LJS.md)
- **Decision Date:** Dec 12, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5970
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

BMW’s Clampless Valved Catheter-Tunneled is designed for patients who require long-term access to the central venous system for administration of fluids including, but not limited to, hydration fluids, antibiotics, chemotherapy, analgesics, nutritional therapy, and blood products. It is also indicated for blood specimen withdrawal. This product is effective for venous access in adults, children, and infants who require intravenous therapy.

## Device Story

Multi-lumen tunneled central venous catheter; provides long-term venous access. Features external three-way silicone rubber slitted disc valve housed within rigid plastic proximal adapter; valve remains closed when inactive; opens distally for infusion; opens proximally for aspiration. Catheter tubing composed of extruded silicone rubber mixed with barium sulfate for radiopacity. Device allows fluid administration and blood withdrawal; valve design reduces risk of air embolism and bleedback while avoiding direct blood contact associated with distal-tip valves. Used by clinicians in clinical settings for patients requiring intravenous therapy.

## Clinical Evidence

Bench testing only. Performance data evaluated valve function, hemolysis, aspiration pressures, stiffness, elongation at break, radiopacity, flow rate, and tensile/burst strength. Biocompatibility assessed via published requirements for hub materials. Results indicate performance characteristics within range of predicate Hickman and Groshong catheters.

## Technological Characteristics

Multi-lumen tunneled catheter; extruded silicone rubber with barium sulfate (radiopacity). Three-way valve: slitted silicone rubber disc in rigid medical-grade plastic adapter. Proximal hub configuration. Mechanical operation (pressure-activated valve).

## Regulatory Identification

A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.

## Special Controls

*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”

## Predicate Devices

- Bard Access Systems, Inc. Groshong Multi-Lumen Central Venous Catheters
- Bard Access Systems, Inc. Hickman Multi-Lumen Central Venous Catheters

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

BMW Medical, Inc.'s Multi-Lumen Clampless Valved Catheter-Tunneled (CVC-Tunneled) Premarket Notification
510 K SUMMARY
Prepared: Sept. 11, 1996
K963806
Dec. 12, 1996

1. Submitted by:
Phil Triolo, Ph. D.
V. P. Research and Development
BMW Medical, Inc.
3598 West, 1820 South
Salt Lake City, UT 84104-4959
Tel: (801) 954-8444, (800) 418-2828
Fax: (801) 954-8484.

2. Contact Person:
Phil Triolo, Ph. D.
V. P. Research and Development

3. Device Identification:
Trade Name: Clampless Valved Catheter-Tunneled (CVC-T
Common Name: Multi-lumen Central Venous Catheter
Classification Name: Intravascular Catheter, Long-term

4. Predicate Device(s):
Bard Access Systems, Inc., (Bard) Groshong Multi-Lumen Central Venous Catheters and Hickman Multi-Lumen Central Venous Catheters.

12

{1}

13

# BMW Medical, Inc.'s Multi-Lumen Clampless Valved Catheter-Tunneled (CVC-Tunneled) Premarket Notification

## 5. Device Description:

BMW Medical has developed a catheter that provides the benefits of the valve incorporated at the distal tip of the Groshong catheter on an open-ended catheter. The BMW valve is external to the central venous system and is protected inside the rigid catheter adapter.

BMW Medical, Inc.’s multi-lumen catheters are composed of extruded silicone rubber tubing which is homogeneously mixed with barium sulfate (BaSO₄) prior to extrusion to impart radiopacity. The catheters vary in diameter and number of lumens to accommodate specific applications.

A silicone rubber joint permanently attaches the multi-lumen catheter tubing to the 2 or 3 legs required to access each lumen of a dual or a triple lumen catheter. A 2-piece rigid plastic adapter is attached to the proximal end of each single lumen leg. Each leg is mechanically locked onto the barbed end of the adapter with a silicone rubber sleeve. The other end of the adapter terminates in a female luer which allows for attachment of the catheter to a male luer fitting.

The adapter houses a slitted disc of silicone rubber which serves as a three-way valve. The valve remains closed when the catheter is not in use. The valve opens toward the distal tip of the catheter when fluids are infused into the patient via the catheter, and it opens toward the female luer of the proximal hub during aspiration of blood samples.

## 6. Intended Use:

BMW’s multi-lumen central venous catheters have the same intended use as Bard Access Systems’ Groshong long-term central venous catheters, and Bard’s Hickman preamendment long-term central venous catheters.

BMW’s Clampless Valved Catheter-Tunneled is designed for patients who require long-term access to the central venous system for administration of fluids including, but not limited to, hydration fluids, antibiotics, chemotherapy, analgesics, nutritional therapy, and blood products. It is also indicated for blood specimen withdrawal. This product is effective for venous access in adults, children, and infants who require intravenous therapy.

## 7. Summary of Technological Characteristics of Device in relation to Predicate Device(s):

BMW Medical, Inc. has developed a multi-lumen tunneled central venous catheter that provides the benefits of the valve incorporated at the distal tip of Bard’s multilumen Groshong catheters on an open-ended catheter similar to Bard’s Hickman multi-lumen central venous catheters. The BMW valve is external to the central venous system and is protected inside the catheter adapter. This configuration provides the advantages of a normally closed three-way valve while avoiding the influences of continuous direct blood contact that can potentially interfere with the function of valves positioned in the vasculature.

{2}

14

# BMW Medical, Inc.'s Multi-Lumen Clampless Valved Catheter-Tunneled (CVC-Tunneled) Premarket Notification

BMW’s Multi-lumen tunneled CVCs are trimmed from their distal ends whereas the Groshong legs may be trimmed from the proximal ends, necessitating final attachment of the Groshong catheters to their hubs by the user.

BMW’s catheter adapter (hub) is fabricated from a medical grade plastic that is not identical to that of the predicate devices. This medical grade material has been used on legally marketed devices for purposes identical to those intended for BMW’s Multi-lumen CVCs.

## 8. Assessment of Performance Data used to justify Substantial Equivalence Claim

The primary advantages afforded by valved catheters are that they reduce the potential for air embolism and bleedback. Performance data indicate that the valve included in the hub of the BMW Multi-lumen CVC-Tunneled is at least as effective in minimizing these potentials as are the predicate devices. In addition, no hemolysis of blood is observed when aspirating blood through BMW’s valve, and aspiration pressures are low.

Published biocompatibility flow chart requirements indicate that BMW’s hub material meets biocompatibility requirements, and is therefore acceptable for use in this catheter application.

Further, test data collected on the stiffness, elongation at break, radiopacity, flow rate, and tensile and burst strength of the catheter, valve function, and leak test data demonstrate that the BMW Multi-lumen CVC-Tunneled is substantially equivalent to the predicate Hickman and Groshong multi-lumen central venous catheters manufactured by C. R. Bard.

Based on the above and physical comparisons, BMW’s Multi-lumen Clampless Valved Catheter-Tunneled products were found to exhibit physical features and performance and safety characteristics which fall within the range of values typically found for Bard’s open-ended Hickman catheters and valved Groshong catheters.

## 9. Conclusion

The BMW Clampless Valved Catheter-Tunneled products are substantially equivalent to Bard’s valved Groshong and open-ended Hickman catheters.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/LJS/K963806](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/LJS/K963806)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
