← Product Code [LJS](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/LJS) · K153487

# CG+ Arrow PICC powered by Arrow VPS Stylet PLUS and Arrow VPS Stylet PLUS (K153487)

_Arrow International, Inc. (Subsidiary of Teleflex, Inc.) · LJS · Feb 4, 2016 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/LJS/K153487

## Device Facts

- **Applicant:** Arrow International, Inc. (Subsidiary of Teleflex, Inc.)
- **Product Code:** [LJS](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/LJS.md)
- **Decision Date:** Feb 4, 2016
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 880.5970
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The intended use of the Arrow VPS Stylet PLUS and VPS Console (together with the VPS System) is to quickly and accurately guide market available central catheters to the desired location, which is the lower third of the superior vena cava (SVC) or at the cavo-atrial junction. The CG+ Arrow PIC Catheters powered by Arrow VPS Stylet PLUS permit venous access to the central circulation through a peripheral vein with the guidance of the VPS system.

## Device Story

Device consists of Arrow VPS Stylet PLUS and CG+ Arrow PICC. Stylet provides stiffness for catheter placement, intravascular ECG sensing, and intravascular ultrasound for guidance. Used in hospital settings by trained clinicians. Stylet contains Doppler transducer and platinum-iridium ECG conductor band at distal tip; connects to VPS Console. Console processes physiological signals (cardiac electrical activity and blood flow) to provide real-time tip location. System displays 'Blue Bullseye' when tip reaches desired location (lower third of SVC or cavo-atrial junction). Provides alternative to fluoroscopy/X-ray for tip confirmation. PICC features Chlorag+ard antimicrobial/antithrombogenic coating on external surface and fluid path to reduce microbial colonization and thrombus accumulation. Benefits include reduced radiation exposure and real-time guidance.

## Clinical Evidence

Bench testing only. Testing included tensile strength (ISO 10555-1), torque strength, tip flexibility, flexing (BS EN ISO 11070), corrosion (BS EN ISO 10555-1), X-ray detectability (ASTM F640-07), electrical performance (Hi-Pot, continuity, Doppler bandwidth, capacitance, sensitivity stability), biocompatibility (cytotoxicity, irritation, sensitization, acute systemic toxicity, hemocompatibility per ISO 10993-1), electrical safety/EMC (IEC 60601-1-2, IEC 60601-2-37), and simulated use insertion/removal forces.

## Technological Characteristics

Stylet: 6 ft length, polyimide body with PTFE coating. Distal tip: 0.4 mm hexagonal Doppler transducer and platinum-iridium (Pt/Ir) ECG conductor band. ECG wire: 304V stainless steel. Connector: 7-prong custom. PICC: Polyurethane, radiopaque, Chlorag+ard (chlorhexidine-based) coating. Sterilization: Ethylene Oxide (10^-6 SAL). Connectivity: Wired to VPS Console. Standards: ISO 10555-1, ASTM F640-07, IEC 60601-1-2, IEC 60601-2-37.

## Regulatory Identification

A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.

## Special Controls

*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”

## Predicate Devices

- Arrow Vascular Positioning System (VPS) stylet PLUS and CG+ Arrow PICC powered by Arrow VPS Stylet PLUS ([K141618](/device/K141618.md))

## Reference Devices

- VPS Console ([K103260](/device/K103260.md))
- VPS Console ([K123813](/device/K123813.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of a human figure, with three overlapping profiles facing to the right. The profiles are rendered in black and create a sense of depth and unity. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 4, 2016

Arrow International, Inc. (subsidiary of Teleflex Inc.) Elizabeth Duncan Sr. Regulatory Affairs Specialist 2400 Bernville Road Reading, PA 19605

Re: K153487

Trade/Device Name: CG+ Arrow PICC powered by Arrow VPS Stylet PLUS and Arrow VPS Stylet PLUS Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: II Product Code: LJS, OBJ Dated: January 6, 2016 Received: January 7, 2016

Dear Ms. Duncan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours, Tina Kiang -S

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

510(k) Number (if known) K153487

#### Device Name

CG+ Arrow PICC powered by Arrow VPS Stylet PLUS

#### Indications for Use (Describe)

The Pressure Injectable PICC with Chlorag+ard Antithrombogenic Technology is indicated for shortterm or long-term peripheral access to the central venous system for intravenous therapy, blood sampling, infusion, pressure injection of contrast media and allows for central venous pressure monitoring. The maximum pressure of pressure injection equipment used with the pressure injectable PICC may not exceed 300 psi.

Chlorag+ard Technology treatment on the external surface of the catheter body as well as the entire fluid pathway of the catheter has been shown to be effective in reducing microbial colonization and thrombus accumulation on catheter surfaces. Antimicrobial and antithrombogenic effectiveness were evaluated using in vitro and in vivo test nethods and no correlation between these test methods and clinical outcome has currently been ascertained. It is not intended to be used for the treatment of existing infections or vein thrombosis.

The VPS Stylet and Console are indicated for guidance and tip positioning for central venous catheters. The Stylet provides stiffess for use in placement of the catheter, intravascular capability for ECG detection and recording and intravascular ultrasound for catheter guiding and positioning. The VPS Stylet, when used with the VPS Console, provides real-time catheter tip location information by using the patient's physiological (cardiac electrical activity and blood flow) information. When the VPS System guidance indicator shows a Blue Bullseye, the catherer tip is in the desired location.

The VPS System is indicated for use as alternative method to fluoroscopy or chest x-ray for central venous catheter tip placement confirmation in adult patients when a steady Blue Bullseye is obtained. NOTE: If a steady Blue Bullseve is not obtained, standard hospital practice should be followed to confirm catheter tip location.

Limiting but not contraindicated situations for this technique are in patients where alterations of cardiac rhythm change the presentation of the P-wave as in atrial flutter, severe tachycardia and pacemaker-driven rhythm, and in central venous catheterization procedures performed through femoral vein access which change the presentation of the P-wave. In such patients, who are easily identifiable prior to central venous catheter insertion, the use of an additional method is required to confirm catheter tip location.

| Type of Use (Select one or both, as applicable)                                                |                                             |
|------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

See PRA Statement below.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

510(k) Number (if known) K153487

Device Name Arrow VPS Stylet PLUS

#### Indications for Use (Describe)

The VPS Stylet and Console are indicated for guidance and tip positioning for central venous catheters. The Stylet provides stiffness for use in placement of the catheter, intravascular capability for ECG detection and recording and intravascular ultrasound for catheter guiding and positioning. The VPS Stylet, when used with the VPS Console, provides real-time catheter tip location information by using the patient's physiological (cardiac electrical activity and blood flow) information. When the VPS System guidance indicator shows a Blue Bullseye, the catheter tip is in the desired location.

The VPS System is indicated for use as alternative method to fluoroscopy or chest x-ray for central venous catheter tip placement confirmation in adult patients when a steady Blue Bullseye is obtained. NOTE: If a steady Blue Bullseve is not obtained, standard hospital practice should be followed to confirm catheter tip location.

Limiting but not contraindicated situations for this technique are in patients where alterations of cardiac rhythm change the presentation of the P-wave as in atrial flutter, severe tachycardia and pacemaker-driven rhythm, and in central venous catheterization procedures performed through femoral vein access which change the presentation of the P-wave. In such patients, who are easily identifiable prior to central venous catheter insertion, the use of an additional method is required to confirm catheter tip location.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{5}------------------------------------------------

# K153487

## 510(K) SUMMARY PER 21 CFR 807.92

### Submitter Information

| Name:             | Arrow International, Inc (subsidiary of Teleflex Inc.) |
|-------------------|--------------------------------------------------------|
| Address:          | 2400 Bernville Road<br>Reading, PA 19605-9607 USA      |
| Contact Person:   | Elizabeth Duncan<br>Sr. Regulatory Affairs Specialist  |
| Telephone Number: | (610) 378-0131 Extension 603220                        |
| Fax Number:       | (610) 478-3179                                         |
| Email:            | elizabeth.duncan@teleflex.com                          |
| Date Prepared:    | December 3, 2015                                       |

## Device Name

Device Trade Name: CG+ Arrow PICC powered by Arrow VPS Stylet PLUS and Arrow VPS Stylet PLUS Device Common Name: Catheter: Peripherally Inserted Central Catheter Stylet: Catheter, Ultrasound, Intravascular Device Classification Name: Catheter: Percutaneous Implanted Long-Term Intravascular Catheter Stylet: Diagnostic Intravascular Catheter FDA Classification Regulation: Catheter: 21 CFR: 880.5970 FDA Classification (Catheter and Stylet): Class II FDA Product Code: Catheter: LJS Stylet: OBJ

## Predicate Devices

- K141618: Arrow Vascular Positioning System (VPS) stylet PLUS and CG+ Arrow PICC . powered by Arrow VPS Stylet PLUS
# Modifications

This Special 510(k) is being submitted for design and material modifications to the predicate Arrow VPS Stylet PLUS and the Arrow VPS Stylet PLUS included in the predicate CG+ Arrow PICC powered by Arrow VPS Stylet PLUS (K141618). The following is a list of modifications being made to the subject device followed by the reasons for the changes:

{6}------------------------------------------------

# Material modification

- Addition of a platinum-iridium conductor band assembly and associated adhesive
# Design Modification

- Increase in stylet body inner diameter (ID); decrease in outer diameter (OD) ●
- Increase in the ECG wire body OD
- Decrease in the co-ax wire OD
- Connector design changes
- Process of PTFE coating

See the "Technological Characteristics and Substantial Equivalence" table below for more detailed information.

The predicate premarket notification for the Arrow VPS Stylet PLUS and the Arrow VPS Stylet PLUS included in the predicate CG+ Arrow PICC powered by Arrow VPS Stylet PLUS (K141618) did not contain information about the VPS Console information was contained within separate cleared 510(k)s (K103260 and K123813). There are no changes to the VPS Consoles associated with this submission.

There are no changes to the catheter or Chlorag+ard technology portion of the predicate Arrow CG+ Arrow PICC powered by the Arrow VPS Stylet (K141618).

# Device Description

The purpose of this premarket notification is to propose design and material modifications to the subject device: the Arrow Vascular Positioning System (VPS) stylet PLUS and CG+ Arrow PICC powered by Arrow VPS Stylet PLUS . The Arrow Vascular Positioning System (VPS) Stylet PLUS (hereafter referred to as Arrow VPS Stylet PLUS) is designed for use with a VPS Console to guide market available central catheters to the desired location, which is the lower third of the superior vena cava (SVC) or at the cavo-atrial junction. The CG+ Arrow PICC powered by Arrow VPS Stylet PLUS provides the user with an Arrow VPS Stylet PLUS already loaded into a central catheter. (The Arrow VPS Stylet PLUS and the Arrow VPS Stylet PLUS included in the CG+ Arrow PICC powered by Arrow VPS Stylet PLUS are exactly the same stylet.)

The subject device, the Arrow VPS Stylet PLUS has the following characteristics:

- 6 ft overall length
- < 0.021" outer diameter over working length of polyimide with polytetrafluoroethylene ● (PTFE)
- Intravascular electrocardiogram (ivECG) signal sensing consists of a platinumiridium (Pt/Ir) conductor band assembly
- Doppler transducer connected to coaxial cable at the distal end
- Coaxial cable and ivECG wire attached to connector at the proximal end to be plugged in ● to VPS Console or extension cable (that in turn connects to the VPS Console)

{7}------------------------------------------------

- Tuohy-Borst adapter
- . Marking accessory

The Arrow VPS Stylet PLUS is for use in a hospital setting by trained clinicians. The Arrow VPS Stylet PLUS is the stylet portion of a vascular positioning system designed to be used with the VPS Console and a market-available catheter. The Arrow VPS Stylet PLUS body is a polyimide tube with a fluoropolymer (PTFE) coating. The tubing contains a Doppler sensor on a coaxial cable and an intravascular electrocardiogram (ivECG) signal sensing platinum-iridium (Pt/Ir) conductor band with an intermediate stainless steel cannula that is welded to the stainless steel ECG wire. The stainless steel portions of the ECG conductor are encased within the platinum-iridium band. The Doppler sensor and the only exposed portion of the ivECG, i.e. the platinum-iridium band, are located at the distal end of the stylet and are used to detect and transmit physiological information to the VPS Console for analysis. The proximal end contains a connector to be plugged into the VPS Console or to an extension cable (that in turn connects to the VPS Console).

The CG+ Arrow PICC Powered by Arrow VPS Stylet PLUS is the Arrow VPS Stylet PLUS preloaded into a 4.5 Fr 1-Lumen, 5.5 Fr 2-Lumen, 40-55 cm pressure injectable Chlorag+ard Peripherally Inserted Central Catheter (PICC).

The CG+ Arrow PICC is a short-term or long-term, single use catheter designed to provide access to the central venous system. It consists of a non-tapered, radiopaque polyurethane extruded catheter body with a softer, contoured Blue Flex Tip. The catheter can be used for the iniection of contrast media. The external catheter body and the internal fluid path of the device are treated with Chlorag+ard, a Chlorhexidine-based coating technology.

# Intended Use

The intended use of the subject and predicate (K141618) devices are identical and are listed below.

The intended use of the Arrow VPS Stylet PLUS and VPS Console (together with the VPS System) is to quickly and accurately guide market available central catheters to the desired location. which is the lower third of the superior vena cava (SVC) or at the cavo-atrial junction.

The CG+ Arrow PIC Catheters powered by Arrow VPS Stylet PLUS permit venous access to the central circulation through a peripheral vein with the guidance of the VPS system.

## Indications for Use

The indications for use of the subject and predicate (K141618) devices are identical and are listed below.

## Arrow VPS Stylet PLUS:

{8}------------------------------------------------

The VPS Stylet and Console are indicated for guidance and tip positioning for central venous catheters. The Stylet provides stiffness for use in placement of the catheter, intravascular capability for ECG detection and recording and intravascular ultrasound for catheter guiding and positioning. The VPS Stylet, when used with the VPS Console, provides real-time catheter tip location information by using the patient's physiological (cardiac electrical activity and blood flow) information. When the VPS System guidance indicator shows a blue bullseye, the catheter tip is in the desired location.

The VPS System is indicated for use as an alternative method to fluoroscopy or chest x-ray for central venous catheter tip placement confirmation in adult patients when a steady blue bullseye is obtained. NOTE: If a steady blue bullseye is not obtained, standard hospital practice should be followed to confirm catheter tip location.

Limiting but not contraindicated situations for this technique are in patients where alterations of cardiac rhythm change the presentation of the P-wave as in atrial fibrillation, atrial flutter, severe tachycardia and pacemaker-driven rhythm, and in central venous catheterization procedures performed through femoral or saphenous vein access which change the presentation of the Pwave. In such patients, who are easily identifiable prior to central venous catheter insertion, the use of an additional method is required to confirm catheter tip location.

# CG+ Arrow PIC Catheters powered by Arrow VPS Stylet PLUS:

The Pressure Injectable PICC with Chlorag+ard Antimicrobial and Antithrombogenic Technology is indicated for short-term or long-term peripheral access to the central venous system for intravenous therapy, blood sampling, infusion, pressure injection of contrast media and allows for central venous pressure monitoring. The maximum pressure of pressure injection equipment used with the pressure injectable PICC may not exceed 300 psi.

Chlorag+ard Technology treatment on the external surface of the catheter body as well as the entire fluid pathway of the catheter has been shown to be effective in reducing microbial colonization and thrombus accumulation on catheter surfaces. Antimicrobial and antithrombogenic effectiveness were evaluated using in vitro and in vivo test methods and no correlation between these test methods and clinical outcome has currently been ascertained. It is not intended to be used for the treatment of existing infections or vein thrombosis.

The VPS Stylet and Console are indicated for guidance and tip positioning for central venous catheters. The Stylet provides stiffness for use in placement of the catheter, intravascular capability for ECG detection and recording and intravascular ultrasound for catheter guiding and positioning. The VPS Stylet, when used with the VPS Console, provides real-time catheter tip location information by using the patient's physiological (cardiac electrical activity and blood flow) information. When the VPS System guidance indicator shows a blue bullseye, the catheter tip is in the desired location.

{9}------------------------------------------------

The VPS System is indicated for use as an alternative method to fluoroscopy or chest x-ray for central venous catheter tip placement confirmation in adult patients when a steady blue bullseye is obtained. NOTE: If a steady blue bullseye is not obtained, standard hospital practice should be followed to confirm catheter tip location.

Limiting but not contraindicated situations for this technique are in patients where alterations of cardiac rhythm change the presentation of the P-wave as in atrial fibrillation, atrial flutter, severe tachycardia and pacemaker-driven rhythm, and in central venous catheterization procedures performed through femoral or saphenous vein access which change the presentation of the Pwave. In such patients, who are easily identifiable prior to central venous catheter insertion, the use of an additional method is required to confirm catheter tip location.

| Characteristic                                                                             | Predicate Device                                                                                                                                                                                                                                    | Subject Device                                                                                                                                                                                       |
|--------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Design                                                                                     | 6 foot long stylet which contains a Doppler<br>sensor and an intravascular electrocardiogram<br>(ivECG) signal sensing wire at the distal tip                                                                                                       | SAME: 6 foot long stylet which contains a<br>Doppler sensor and an intravascular<br>electrocardiogram (ivECG) signal sensing wire at<br>the distal tip                                               |
| Signal Conductor                                                                           | Two conductor stylet wires (ECG wire and<br>coaxial cable)                                                                                                                                                                                          | SAME: Two conductor stylet wires (ECG wire<br>and coaxial cable)                                                                                                                                     |
| ECG Conductor pathway                                                                      | Exposed conductive surface at distal end of<br>ECG wire                                                                                                                                                                                             | SAME: Exposed conductive surface at distal end<br>of ECG wire                                                                                                                                        |
| ECG wire OD (nominal)                                                                      | .006"                                                                                                                                                                                                                                               | .007"                                                                                                                                                                                                |
| Connector                                                                                  | Single, 7 prong custom, 2 barbs                                                                                                                                                                                                                     | Single, 7 prong custom, with the addition of a<br>barb, two clips and grooves                                                                                                                        |
| Stylet body OD and ID (nominal)                                                            | ID: 0.0134"<br>OD: 0.0187"                                                                                                                                                                                                                          | ID: 0.0144"<br>OD: 0.0174"                                                                                                                                                                           |
| Coaxial cable OD (nominal)                                                                 | OD: .0063"                                                                                                                                                                                                                                          | OD: .0062"                                                                                                                                                                                           |
| Stylet Patient Contacting Materials<br>(Circulating Blood contact for limited<br>duration) | Stylet body: Polyimide with Fluoropolymer<br>(PTFE) heat shrink<br>ECG Conductor: 304V SST                                                                                                                                                          | Stylet body: SAME material, different process,<br>Polyimide coated with Fluoropolymer (PTFE)<br>ECG Conductor: Platinum-Iridium band welded<br>to 304V SST                                           |
|                                                                                            | Tuohy-Borst: Polycarbonate, Polypropylene,<br>silicone, PTFE, TPE                                                                                                                                                                                   | Tuohy-Borst: SAME, Polycarbonate,<br>Polypropylene, silicone, PTFE, TPE                                                                                                                              |
|                                                                                            | Adhesive encased over transducer: Loctite<br>3311 UV Cured Acrylic                                                                                                                                                                                  | SAME: Loctite 3311 UV Cured Acrylic                                                                                                                                                                  |
|                                                                                            | Adhesive used to bond ECG assembly to<br>stylet body: Loctite 3311 UV cured acrylic                                                                                                                                                                 | Adhesive used to bond ECG assembly to stylet<br>body: Loctite 4014 Cyanoacrylate acrylic adhesive                                                                                                    |
| Operating Frequency of Stylet Doppler<br>Transducer                                        | 11.667 MHz                                                                                                                                                                                                                                          | SAME: 11.667 MHz                                                                                                                                                                                     |
| Size of Stylet Doppler Transducer                                                          | 0.4 mm                                                                                                                                                                                                                                              | SAME: 0.4 mm                                                                                                                                                                                         |
| Stylet Doppler Transducer Shape                                                            | Hexagonal                                                                                                                                                                                                                                           | SAME: Hexagonal                                                                                                                                                                                      |
| Marking Accessory                                                                          | Twist-locking collet                                                                                                                                                                                                                                | SAME: Twist-locking collet                                                                                                                                                                           |
| PIC Catheters                                                                              | All characteristics: materials, diameter (4.5,<br>5.5, 6 Fr.), lengths, manufacturing processes,<br>conditions and aids                                                                                                                             | SAME: All characteristics: materials, diameter<br>(excluding the 6 Fr, since this sized catheter is not<br>included in the subject device), lengths,<br>manufacturing processes, conditions and aids |
| Chlorhexidine Technology                                                                   | All characteristics: Identity, formulation,<br>concentration (content per surface area),<br>manufacturing processes, conditions and aids,<br>method of application to the device and<br>mechanism by which the agent is released<br>from the device | SAME: All characteristics                                                                                                                                                                            |

## Technological Characteristics and Substantial Equivalence

{10}------------------------------------------------

| Characteristic                                                            | Predicate Device                                                                                                  | Subject Device                                                                                                            |
|---------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Instruction s for Use                                                     | Content                                                                                                           | Stylet: SAME<br>Catheter: SAME                                                                                            |
| Stylet Shelf life                                                         | 1 year                                                                                                            | SAME: 1 year                                                                                                              |
| Sterilization: All characteristics,<br>including: method, assurance level | All characteristics, including:<br>Method: Ethylene Oxide<br>Assurance Level: 10-6                                | SAME: All characteristics, including:<br>Method: Ethylene Oxide<br>Assurance Level: 10-6                                  |
| Packaging                                                                 | Standalone kits: PET tray sealed in a Tyvek<br>header pouch<br>Preloaded kits: PETG tray with a Tyvek<br>lidstock | Standalone kits: Polystyrene (HIPS) tray with<br>Tyvek sterile barrier<br>Preloaded kits: PETG tray with a Tyvek lidstock |

The Arrow YPS Stylet PLUS and CG+ Arrow PICC powered by Arrow VPS Stylet PLUS is substantially equivalent to the predicate Arrow VPS Stylet PLUS and the CG+ Arrow PICC powered by Arrow VPS Stylet PLUS (K141618) in terms of: intended use, indications for use, patient population, contraindications, operating principle, functional performance, sterilization, fundamental scientific technology, features, material and design.

There are no changes to the catheter or Chlorag+ard Technology portion of the predicate CG+ Arrow PICC catheters powered by Arrow VPS Stylet PLUS (K141618). There are no changes being made to the VPS Consoles (K103260 and K123813), as part of this submission.

# Nonclinical Testing

The modifications proposed to the subject device were evaluated using risk management. The risk analysis method used to assess the impact of the modifications was per ISO 14971 and included a Risk Management Plan. Hazards Analysis, and Risk Management Report. As a result of the risk assessment, the testing listed in the table below was determined to be required. The resultant testing listed in the table below was performed on the subject device: the Arrow VPS Stylet PLUS and the CG+ Arrow PICC powered by Arrow VPS Stylet PLUS after being preconditioned with EtO sterilization, ISTA simulated environmental conditions and one year accelerated aging:

| Test Device | Test                                                                                                 | Results |
|-------------|------------------------------------------------------------------------------------------------------|---------|
| Stylet      | Tensile per ISO 10555-1                                                                              | PASS    |
| Stylet      | Torque strength Per the Coronary and Cerebrovascular Guidewire guidance                              | PASS    |
| Stylet      | Tip Flexibility Per the Coronary and Cerebrovascular Guidewire guidance                              | PASS    |
| Stylet      | Flexing BS EN ISO 11070                                                                              | PASS    |
| Stylet      | Corrosion BS EN ISO 10555-1                                                                          | PASS    |
| Stylet      | Component Compatibility: Marking accessory grip strength                                             | PASS    |
| Stylet      | Component Compatibility: Tuohy-Borst Adapter grip strength                                           | PASS    |
| Stylet      | Tuohy-Borst Leak per 594-1                                                                           | PASS    |
| Stylet      | Physical characteristics                                                                             | PASS    |
| Stylet      | X-ray Detectability ASTM F640-07                                                                     | PASS    |
| Stylet      | Electrical Performance: Hi-Pot, Continuity, Doppler Bandwidth, Capacitance,<br>Sensitivity stability | PASS    |
| Stylet      | Biocompatibility: Cytotoxicity, Irritation, Sensitization, Acute Systemic                            | PASS    |

{11}------------------------------------------------

| Test Device     | Test                                                                                           | Results |
|-----------------|------------------------------------------------------------------------------------------------|---------|
|                 | Toxicity and Hemocompatibility per ISO 10993-1                                                 |         |
| Combined device | Electrical Safety and Electromagnetic Compatibility Testing (IEC 60601-1-2,<br>IEC 60601-2-37) | PASS    |
|                 | Catheter Compatibility Simulated Use Insertion/ Removal                                        | PASS    |
|                 | Catheter Compatibility Force to Remove stylet from catheter                                    | PASS    |

# Conclusions

The subject device is substantially equivalent to the predicate device in terms of intended use, indications for use, patient population, contraindications, operating principle, functional performance, fundamental scientific technology, features, design and material. The results of the risk assessment and resultant testing performed have demonstrated that the design and material modifications is as safe, as effective and performs as well as the predicate device and therefore the subject device is considered substantially equivalent to the cited predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/LJS/K153487](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/LJS/K153487)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
