PRO-PICC (BASIC TRAY, LONG WIRE TRAY, BASIC NURSING TRAY, FULL NURSING TRAY)
Applicant
Medcomp
Product Code
LJS · General Hospital
Decision Date
Sep 23, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5970
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PRO-PICC ™ C catheter is indicated for short term or long term peripheral access to the central venous system for intravenous therapy and power injection of contrast media. For blood sampling, infusion, or therapics, use a 4F or larger catheter. The maximum recommended infusion rate varies by catheter French size and is printed on the catheter.
Device Story
PRO-PICC CT is a triple-lumen, open-ended peripherally inserted central catheter (PICC) designed for IV therapy and power injection of contrast media. Constructed from radiopaque polyurethane with a reverse-taper lumen design; features a soft hub with suture wings for securement. Includes extension tubes with clamps and female luer connectors. Purple-colored extension line and lumen identify power-injectable capability; clamp contains infusion rate and flushing instructions. Inserted via stylet and adaptor sideport. Used in clinical settings by healthcare professionals for central venous access. Benefits include reliable power injection of contrast media and long-term venous access.
Clinical Evidence
No clinical data. Safety and effectiveness demonstrated via in vitro bench testing, including air/liquid leakage, force at break, elongation, gravity flow, static burst pressure, high-pressure injection flow rate, and chemical testing per ISO 10555-1 and 10555-3.
Technological Characteristics
Triple-lumen, open-ended PICC; radiopaque polyurethane material. Reverse-taper lumen design. 6F size, 60cm length. Purple color-coding for power injection identification. Biocompatibility per ISO 10993. Sterilized product.
Indications for Use
Indicated for short or long-term peripheral access to the central venous system for IV therapy and power injection of contrast media in patients requiring venous access. Requires 4F or larger catheter for blood sampling, infusion, or therapies.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
Predicate Devices
C.R. Bard, Inc. 6Fr. TL PowerPICC Catheter (K053501)
K131687 — 6F X 60CM TRIPLE PRO-LINE CT BASIC IR SET · Medical Components, Inc. · Aug 16, 2013
K123617 — 3F PRO-PICC · Medical Components, Inc. · May 15, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
## A. Submitter Information: Submitteri
| Submitter: | MEDCOMP®<br>1499 Delp Drive<br>Harleysville, PA 19438<br>Tel: (215) 256-4201<br>Fax: (215) 256-9191 |
|-------------------|-----------------------------------------------------------------------------------------------------|
| Contact: | Jean Callow<br>Regulatory Specialist |
| Date Prepared: | September 7, 2007 |
| Trade Name: | Medcomp® PRO-PICC TM CT |
| Common Name: | Peripherally Inserted Central Catheter |
| Classification: | LJS |
| Regulation Name: | Percutaneous, implanted, long-term<br>intravascular catheter |
| C.F.R. Section: | 880.5970 Class II |
| Predicate Device: | C.R. Bard, Inc. K053501 6Fr. TL<br>PowerPICCTM Catheter<br>Medcomp K053345 Pro-LineTM CT Press |
## D. Device Description:
B.
ﺰ
The PRO-PICC m C catheter is a triple lumen open-ended picc designed for power injection through one designated lumen. The catheter is an extension of the Medcomp® Pro-Line™ CT Power Injectable CVC (K053345) and Power Injectable Implantable Port (K070003) catheter line. The PRO-PICC m C catheter is comprised of a soft radiopaque polyurethane material. The lumen has a reverse taper design and is connected to the extensions via a soft pliable hub with suture wing for secure placement. Clamps are provided on the extension tubes to prevent air/fluid communication. Female luer connectors provide the connection for intravenous administration. The designated power injectable extension line and clamp material are purple in color to differentiate it from the non-power injectable extensions and the purple lumen identifies it as a power injectable catheter. The center extension also is printed with the words power injectable. The I.D. Ring within the clamp on the power extension contains information regarding checking for blood return and flushing along with rate of infusion for power injection.
Injectable CVC.
The PRO-PICC m ™ catheter is available in 6F triple lumen. The catheter has a usable length of 60 cm with depth markings in 5 cm increments. Stylet and adaptor sideport are provided to assist in catheter insertion.
The catheter is packaged sterile in two radiology versions and two nursing configurations with the necessary accessories to facilitate catheter insertion.
> Medcomp PRO-PICC ru CT catheter Summary Page 1 of 2
{1}------------------------------------------------
#### E. Intended Use:
'The PRO-PICC ™ C catheter is indicated for short term or long term peripheral access to the central venous system for intravenous therapy and power injection of contrast media. For blood sampling, infusion, or therapics, use a 4F or larger catheter. The maximum recommended infusion rate varies by catheter French size and is printed on the catheter.
#### મું. Performance Data:
In vitro testing was performed on the PRO-PICC m C catheter to assure reliable design and performance in accordance with ISO 10555-1 and 10555-3 and internal engineering protocol. Testing includes air/liquid leakage, force at break, elongation, gravity flow, static burst pressure, high pressure injection flow rate and chemical testing.
Clinical studies were not deemed necessary since in vitro testing was sufficient to demonstrate safety and effectivencss by way of comparison to legally marketed predicate devices.
All materials used for the proposed device were previously cleared in the Medcomp K053345 Pro-Line™ CT Pressure Injectable CVC. Biocompatibility testing on the Pro-Line™ CT Pressure Injectable CVC demonstrated that the materials used meets the requirements of ISO 10993 for a permanent external communicating blood contact device.
#### G. Comparison to Predicate Device:
The PRO-PICC m C catheter is substantially equivalent to the predicate devices in terms of intended use, insertion method, anatomical location, design, materials, performance, labeling, packaging and method of sterilization.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter of the circle.
SEP 2 3 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Jean Callow Regulatory Specialist Medcomp 1499 Delp Drive Harleysville, Pennsylvania 19438
Re: K081904
Trade/Device Name: PRO-PICCru CT Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted Long-Term Intravascular Catheter Regulatory Class: II Product Code: LJS Dated: July 2, 2008 Received: July 8, 2008
Dear Ms. Callow:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
# Page 2 - Ms. Callow
Please be advised that FDA's issuance of a substantial equivalence determination docs not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: _____ PRO-PICC 7m CT
Indications for Use:
The PRO-PICC m C catheter is indicated for short term or long term peripheral access to the central venous system for intravenous therapy and power injection of contrast media. For blood sampling, infusion, or therapies, use a 4F or larger catheter. The maximum recommended infusion rate varies by catheter French size and is printed on the catheter.
Prescription Use X_____________________________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anthony Di nome
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control. Dental Devices
Page 1_of 1
510(k) Number: _ Ka81944
- 4
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.