← Product Code [LJS](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/LJS) · K073451

# PRESSURE INJECTABLE PICC (K073451)

_Arrow International, Inc. · LJS · Jan 15, 2008 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/LJS/K073451

## Device Facts

- **Applicant:** Arrow International, Inc.
- **Product Code:** [LJS](/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/LJS.md)
- **Decision Date:** Jan 15, 2008
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 880.5970
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The Arrow International Pressure Injectable PICC is intended for short-term or long-term peripheral access to the central venous system for intravenous therapy and blood sampling.

## Device Story

Radiopaque polyurethane peripherally inserted central catheter (PICC); available in 4 Fr and 5 Fr sizes; includes BlueFlex and non-BlueFlex tip configurations. Device provides peripheral access to central venous system for IV therapy, blood sampling, and power injection of contrast media. Luer hubs labeled for specific pressure injection rates (4ml/sec or 5ml/sec). Used by clinicians in clinical settings. Enables safe delivery of contrast media via power injectors up to 300 psi. Bench testing confirms structural integrity under repetitive bending, cleaning, and high-pressure injection conditions.

## Clinical Evidence

Bench testing only. Testing included flow rate qualification, repeat injection, and catheter static burst pressure (post-sterilization, simulated shipping, cleaning, repetitive bending, and vesicant fluid exposure). Results demonstrate performance criteria met based on FDA-recognized standards.

## Technological Characteristics

Radiopaque polyurethane catheter; 4 Fr and 5 Fr diameters; 40cm to 60cm lengths. Features BlueFlex and non-BlueFlex tip designs. Sterile kit configuration. Pressure injection rated up to 300 psi. No software or electronic components.

## Regulatory Identification

A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.

## Special Controls

*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”

## Predicate Devices

- Arrow International's Peripherally Inserted Central Catheter ([K061289](/device/K061289.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Kø73451 (P.1 of A2)

:

## 510(k) Summary

.

.

|                                                                              | Arrow International, Inc.<br>2400 Bernville Road<br>Reading, PA 19605-9607 USA<br>JAN 15 2008                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person:                                                              | Kevin Lentz<br>Regulatory Affairs Project Manager<br>Phone: 610-378-0131<br>Fax: 610-478-3172<br>Email: kevin.lentz@teleflexmedical.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Date summary prepared:                                                       | December 7, 2007                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Device trade name:                                                           | Pressure Injectable Peripherally Inserted Central Catheter<br>(21 CFR 880.5970, Product Code LJS)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Device common name:                                                          | PICC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Device classification name:                                                  | Catheter, intravascular, therapeutic, long-term greater than<br>30 days                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Legally marketed devices<br>which the device is<br>substantially equivalent: | Arrow International's Peripherally Inserted Central<br>Catheter (K061289)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Description of the device:                                                   | The Pressure Injectable Peripherally Inserted Central Catheters<br>have the following characteristics:<br>Radiopaque polyurethane catheters4 Fr single-lumen BlueFlex® Tip and non-BlueFlex® Tip5 Fr double lumen BlueFlex® Tip and non-BlueFlex® TipUsable length of 4 Fr catheters are 40cm to 60cmUsable length of 5 Fr catheters are 40cm to 60cmCatheters are provided in sterile kit configurationsThe 4 Fr BlueFlex® Tip, 4Fr non-BlueFlex® Tip and 5 Fr<br>BlueFlex® Tip catheters are labeled for "4ml/sec, pressure<br>injectable" on the Luer hub(s) to facilitate the proper use of<br>the deviceThe 5 Fr non-BlueFlex® Tip catheter is labeled for<br>"5ml/sec, pressure injectable" on the Luer hubs to facilitate<br>the proper use of the device |

. .

{1}------------------------------------------------

## K473451 (0.2002)

Intended use of the device: The Arrow International Pressure Injectable PICC is intended for short-term or long-term peripheral access to the central venous system for intravenous therapy and blood sampling.

## Performance tests: The following tests were performed to demonstrate substantial equivalence:

- . Flow Rate Qualification Test
- . Repeat Injection Test
- Catheter Static Burst Pressure (after sterilization, simulated . shipping conditions, cleaning solutions, repetitive bending of catheter, and use of high vesicant fluid on catheter)

The Arrow International Pressure Injectable PICC met performance data: performance criteria of the safety and effectiveness tests performed based on FDA recognized standards and guidance. The results of bench tests demonstrate that Arrow's PICC's are as safe and effective as compared to the predicate Arrow International PICC (K061289).

Summary: Arrow International's Pressure Injectable Peripherally Inserted Central Catheter (PICC) has the same intended use as the predicate device, Arrow International's PICC, (K061289). Based on the assessment of non-clinical performance data, Arrow International's Pressure Injectable Peripherally Inserted Central Catheter (PICC) is substantially equivalent to the legally marketed predicate device.

> The Arrow International predicate device (K061289) was used as a reference device to show the primary design has not changed significantly and that proper verification was done to prove that the catheter can be used safely for its intended use.

Assessment of non-clinical

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three distinct head profiles, suggesting a sense of community or collaboration. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" is arranged in a circular pattern around the emblem.

JAN 15 2008

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Kevin Lentz Regulatory Affairs Project Manager Arrow International, Incorporated 2400 Bernville Road Reading, Pennsylvania 19605-9607

Re: K073451

Trade/Device Name: Pressure Injectable PICC Regulation Number: 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: II Product Code: LJS Dated: January 4, 2008 Received: January 7, 2008

Dear Mr. Lentz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Lentz

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use Statement

510(k) Number : K073451

Device Name: Pressure Injectable PICC

Indications For Use: The Pressure Injectable PICC is indicated for short or long term peripheral access to the central venous system for intravenous therapy, blood sampling, infusion and power injection of contrast media. The maximum pressure of power injector equipment used with the pressure injectable PICC may not exceed 300 psi.

X Prescription Use AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE)

signature

on of Ancementology, General Houpes), caon Clumbl, Chrisal Devices

15473451

---

**Source:** [https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/LJS/K073451](https://fda.innolitics.com/submissions/HO/subpart-f%E2%80%94general-hospital-and-personal-use-therapeutic-devices/LJS/K073451)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
