K961008 · Baxter Healthcare Corp · LHI · Oct 10, 1997 · General Hospital
Device Facts
Record ID
K961008
Device Name
AUTOMIX 3+3/AS COMPOUNDER SYSTEM
Applicant
Baxter Healthcare Corp
Product Code
LHI · General Hospital
Decision Date
Oct 10, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
This device is an electromechanical positive displacement fluid compounding device that provides for the compounding of fluids via gravimetric weighing.
Device Story
Electromechanical positive displacement fluid compounding system; utilizes gravimetric weighing for accurate fluid transfer. Modification includes 'Positive Identification Solution Family Monitor' to verify source solution identity and placement. Used in clinical settings for compounding fluids; operated by trained personnel. System provides automated verification of source solutions to ensure correct formulation. Benefits include increased safety, speed, and accuracy in fluid compounding processes.
Clinical Evidence
No clinical data. Functional acceptability of hardware and software modifications demonstrated via bench testing, verification, and validation procedures.
Technological Characteristics
Electromechanical positive displacement compounding system; gravimetric weighing mechanism; integrated computer software for solution identification; Class II device (21 CFR 880.5440).
Indications for Use
Indicated for the compounding of fluids via gravimetric weighing using an electromechanical positive displacement system. Prescription use only.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
Clintec Automix® 3 + 3 Compounder System (K894827)
Related Devices
K973593 — BRIDGE SENTRY BEDSIDE UNIT, BRIDGE SENTRY ADMINISTRATION SET-STANDARD CONFIGURATION, BRIDGE SENTRY ADMINISTRATION SET 5 · Bridge Medical, Inc. · Aug 25, 1998
K222048 — Spectrum IQ Infusion System with Dose IQ Safety Software · Baxter Healthcare Corporation · Sep 6, 2022
K220417 — Spectrum IQ Infusion System with Dose IQ Safety Software · Baxter Healthcare Corporation · Apr 29, 2022
K031185 — ABBOTT PLUM A+ INFUSION PUMP WITH A HPL/RS PLUG-AND-PLAY MODULE · Abbott Laboratories · Apr 30, 2003
Submission Summary (Full Text)
{0}
1. 1008
510(k) Premarketing Notification
Clintec Automix® 3 + 3 / AS Compounder System
March 11, 1996
OCT 10 1997
Attachment 9.
# 510(k) Summary
(as required by 21 CFR 807.92)
Submitted by: Dennis A. Ocwieja
Vice President
Regulatory Affairs and Quality Assurance
Clintec Nutrition Company
Three Parkway North, Suite 200
Deerfield, IL 60015
phone: (847) 317-3058
fax: (847) 317-3139
Date Prepared: 03/08/96
Trade/Proprietary Name: Clintec Automix® 3 + 3 / AS Compounder System
Common/Usual Name: Electromechanical Positive Displacement Fluid Compounding System
Classification Name: Compounding System with Integrated Computer Software
Classification: Class II in 21 CFR § 880.5440, Set, I.V., Fluid Transfer
Predicate Device: Clintec Automix® 3 + 3 Compounder System
The Automix® 3 + 3 / AS Compounder System is an electromechanical positive displacement fluid compounding device that provides for the safe, fast, and accurate compounding of a wide variety of fluids via gravimetric weighing. It has the same operational and functional intended use and remains the same in all respects to the currently marketed system, except for the proposed modifications to incorporate a system to aid in the verification of source solution identity and placement.
The Positive Identification Solution Family Monitor, an equipment and software upgrade designed to aid in the verification of the identity of the source solutions being compounded, is a modification and upgrade to the compounding system currently marketed which is found substantially equivalent, for purposes of the Federal Food, Drug and Cosmetic Act only, to the Automix® 3+3 Compounder System under K894827, dated October 6, 1989.
Performance testing for the Clintec Automix® 3 + 3 / AS Compounder System described in the submission demonstrates the functional acceptability of the proposed hardware and software modifications and enhancements.
The design changes are documented and their verification and validation procedures are described in the application.
000315
MAR 11 1996
{1}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20856
Ms. Marcia Marconi
Vice President Regulatory Affairs
Baxter Healthcare Corporation
Route 120 & Wilson Road
Roound Lake, Illinois 60073
OCT 10 1997
Re: K961008
Trade Name: Automix® 3+3/AS Compounder System
Regulatory Class: II
Product Code: LHI
Dated: September 26, 1997
Received: September 30, 1997
Dear Ms. Marconi:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531
{2}
Page 2 - Ms. Marconi
through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Director
Division of Dental, Infection Control and General Hospital Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{3}
OCT 9 '97 14:49 FROM REGULATORY AFFAIRS 10214803002-344771 PAGE.002/002
510(k) Number (if known): K961008
Device Name: Automix® 3+3/AS Compounder System
Indications For Use:
This device is an electromechanical positive displacement fluid compounding device that provides for the compounding of fluids via gravimetric weighing.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Dental, Infection Control, and General Hospital Devices
510(k) Number K961008
Prescription Use ☑
(Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐
(Optional Format 1-2-96)
10/09/97
**TOTAL PAGE.002**
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.