K233284 · Hangzhou Qiantang Longyue Biotechnology Co., Ltd. · LHI · May 1, 2024 · General Hospital
Device Facts
Record ID
K233284
Device Name
Vial Adapter with Filter
Applicant
Hangzhou Qiantang Longyue Biotechnology Co., Ltd.
Product Code
LHI · General Hospital
Decision Date
May 1, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The Vial adapter with filter is intended for use by Healthcare Professionals for the transfer and mixing of drugs contained in vials within clinical, hospital, or healthcare environments.
Device Story
Vial adapter with filter facilitates drug reconstitution and transfer from vials to syringes. Device features integrated piercing spike to penetrate vial elastomeric septum; housing seats securely around vial ferrule. Opposite side contains female luer connector for attachment to standard luer syringe. Used by healthcare professionals in clinical/hospital settings. Device enables closed or filtered fluid transfer, reducing contamination risk during drug preparation. Benefits include efficient, sterile drug handling and mixing. Available in 13mm and 20mm sizes.
Clinical Evidence
No clinical data was conducted. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993), sterilization validation (ISO 11137), shelf-life/aging studies (ASTM F1980), and mechanical/functional performance testing (ISO 22413, ISO 80369-7, ISO 8536-4). All tests passed.
Technological Characteristics
Materials: Polycarbonate, polyethylene, acrylic, polyamide. Single-use, sterile, E-beam sterilized (SAL 10-6). Features: Integrated piercing spike (single lumen), filtration membrane, female luer connector. Sizes: 13mm, 20mm. Standards: ISO 10993 (biocompatibility), ISO 80369-7 (connectors), ISO 22413 (fluid transfer), ISO 8536-4 (infusion equipment).
Indications for Use
Indicated for healthcare professionals for the transfer and mixing of drugs contained in vials in clinical, hospital, or healthcare environments. No specific patient population or contraindications stated.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
K190816 — Vial Adapter Ø20 mm , Vial Adapter Ø13 mm · Avenir Performance Europeenne Medical (Apem) · Nov 19, 2019
K171796 — Vial Adapter 15mm · Medimop Medical Project , Ltd. · Oct 12, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo, which features a stylized caduceus. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 1, 2024
Hangzhou Qiantang Longyue Biotechnology Co., Ltd % Esther Zhang Regulatory Affairs Shanghai Ling Fu Technology Co., Ltd. 4F No. 585-2, Wanyuan Rd. Minhang District Shanghai, Shanghai 201102 China
Re: K233284
Trade/Device Name: Vial Adapter with Filter Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: LHI Dated: April 3, 2024 Received: April 3, 2024
Dear Esther Zhang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Porsche Bennett
Porsche Bennett For David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn,
{2}------------------------------------------------
General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233284
Device Name Vial Adapter with Filter
Indications for Use (Describe)
The Vial adapter with filter is intended for use by Healthcare Professionals for the transfer and mixing of drugs contained in vials within clinical, hospital, or healthcare environments.
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K233284- 510(k) Summary
#### Submitter l
| Device submitter: | Hangzhou Qiantang Longyue Biotechnology Co., LTD<br>104, 301, 302, building 12, building 1,619 WangMei Roadinping<br>street, Linping District, Hangzhou |
|-------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person: | Chunyu Wang<br>Quality Assurance Manager |
| | Phone: 0086-13965638723 |
| | Email: chunyu_wang@nextech-x.com |
| Prepare Date: | April 26, 2024 |
#### ll Correspondent
| Company | Shanghai Ling Fu Technology Co., Ltd. |
|-----------------|-----------------------------------------------------------------------------------------------------|
| | 4F, No.585-2 Wanyuan Road, Minhang District, Shanghai<br>P.R.China |
| Contact person: | Esther ZHANG<br>Regulatory affairs<br>Phone: 0086-13771505757<br>Email: Esther.zhang@llins-tech.com |
#### ��� Subject Device
Trade Name of Device: Vial Adapter with Filter Common Name: Fluid Transfer IV Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product code: LHI Review Panel: General Hospital
#### IV Predicate Device
| Trade name: | Vial Adapter |
|-------------------------|--------------------------------------------------|
| Regulation Name: | Intravascular Administration Set |
| Regulation Number: | 21 CFR 880.5440 |
| Classification: | Class II |
| Product Code: | LHI |
| Premarket Notification: | K222718 |
| Manufacturer: | Hangzhou Qiantang Longyue Biotechnology Co., LTD |
{5}------------------------------------------------
#### > Device description
The sterile device pierces the elastomeric septum of a drug vial with its integrated piercing spike. The device is then pushed fully onto the drug vial and seats securely around the ferrule of the drug vial utilizing the housing of the vial adapter. The connector on opposite side of the vented vial adapter is for the connection of a standard luer syringe for the reconstitution and removal of the contents of the drug vial.
The device is intended for use by Healthcare Professionals (HCPs) in a clinical, hospital, or other healthcare environment. The subject device is available by prescription use only and has no known contraindications.
The proposed vented vial adapter is available in 13mm, 20mm diameter to accommodate respective sizes of drug vials.
#### VI Indications for use
The Vial adapter with filter is intended for use by Healthcare Professionals for the transfer and mixing of drugs contained in vials within clinical, hospital, or healthcare environments.
#### VII Comparison of technological characteristics with the predicate devices
The Vial Adapter intended use, technological characteristics, design and performance specifications are either identical or substantially equivalent to the existing legally marketed predicate device. The differences between the subject and predicate devices do not raise different questions of safety and effectiveness.
| Device feature | Subject Device<br>K233284 | Predicate Device<br>K222718 | Comments |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|-----------|
| Indications for use | The Vial adapter with filter<br>is intended for use by<br>Healthcare Professionals<br>for the transfer and mixing<br>of drugs contained in vials<br>within clinical, hospital, or<br>healthcare environments. | The Vial Adapter is<br>indicated for the transfer<br>and mixing of drugs<br>contained in vials. | Identical |
| Product code | LHI | LHI | Identical |
| Regulation<br>number | 21 CFR 880.5440 | 21 CFR 880.5440 | Identical |
{6}------------------------------------------------
| Class | Class II | Class II | Identical |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Principle of operation | Single use | Single use | Identical |
| Size | 13mm, 20mm | 13mm, 20mm, 28mm | Different<br>Comment 1 |
| Material | Polycarbonate,<br>Polyethylene, Acrylic,<br>Polyamide | Polycarbonate | Different<br>Comment 2 |
| Connector | Female Luer fitting | Female Luer fitting;<br>Male Luer fitting | Different<br>Comment 3 |
| Piercing Spike | Plastic - Single Lumen | Plastic - Single Lumen | Identical |
| Filter | Yes | No | Different<br>Comment 4 |
| Performance | Performance test of:<br>Penetration force.<br>Detachment<br>force from Drug<br>Vial.<br>Spike Tip Ductility.<br>Fluid Leakage.<br>Liquid Filtration | Performance test of:<br>Penetration force;<br>Detachment<br>force from Drug<br>Vial;<br>Spike Tip Ductility;<br>Fluid Leakage. | Different<br>Comment 5 |
| Vial Adapter Fit<br>(Vial Side) | Snap Fit to Vial | Snap Fit to Vial | Identical |
| Sterilization<br>Method | Electron beam Irradiation | Electron beam Irradiation | Identical |
| Sterility<br>Assurance Level | SAL 10-6 | SAL 10-6 | Identical |
| Biocompatibility | Conforms to ISO 10993 | Conforms to ISO 10993 | Identical |
| Labeling | Proposed device<br>labeling (IFU) includes<br>transfer and mixing<br>instructions | Proposed device<br>labeling (IFU) includes<br>transfer and mixing<br>instructions | Identical |
| Shelf life | 4 years | 3 years | Different<br>Comment 6 |
{7}------------------------------------------------
### Discussion:
### Comment 1
The subject device consists of Vial Adapters that are 13mm, 20mm for different diameter standard vials. The size of Vial Adapter with filter was demonstrated to meet specification according to internal performance standards. This difference does not raise new questions of safety and effectiveness.
### Comment 2
The material of the subject device is polycarbonate, polyethylene, acrylic/polyamide, and the predicate device material is polycarbonate. Although there is a difference in the material, the biocompatibility test of the subject device complies with the standard requirements. This difference does not raise new questions of safety and effectiveness.
### Comment 3
The connector of the Vial Adapter with filter have only one type, while the predicated device has two types. This difference was addressed through ISO 80369-7 testing and therefore, this difference does not raise new questions of safety and effectiveness.
### Comment 4
The inclusion of a Filtration Membrane in the subject device (K233284) differentiates it from the predicate device (K222718), which does not include a filter. Despite this difference, the additional filtration functionality in K233284 does not raise new questions of safety and effectiveness. It is supported by adherence to standards detailed below, which set requirements for transfer sets and infusion equipment, including the performance of fluid filters. The subject device has been tested to ensure that the filter performs adequately. Testing includes visual inspection for defects, dimensional measurements of the membrane, and filtration efficiency tests, all of which the device has passed. Therefore, the difference does not raise different questions of safety and effectiveness.
### Comment 5
Performance testing of the Vial adapter with filter inlet and outlet functions demonstrates that the difference does not raise new questions of safety and effectiveness.
### Comment 6
The shelf life of the subject device is 4 years which is longer than the predicated device; however, performance testing after accelerated aging passed all applicable and internal test methods. The difference does not raise new questions of substantially equivalence on safety
{8}------------------------------------------------
and effectiveness.
#### VIII Performance data
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility testing
Biocompatibility of the Vial Adapter was evaluated in accordance with ISO 10993- 1:2018 for the body contact category of "External communication device – Blood path indirect" with a contact duration of "Limited (< 24 hours)". The following tests were performed, as recommended:
| Cytotoxicity | ISO 10993-5: 2009 |
|---------------------------|--------------------|
| Skin sensitization | ISO 10993-10: 2010 |
| Hemolysis | ISO 10993-4: 2017 |
| Intracutaneous reactivity | ISO 10993-23: 2021 |
| Acute systemic toxicity | ISO 10993-11: 2017 |
| Pyrogenicity | ISO 10993-11: 2017 |
## Sterilization and shelf life testing
The sterilization method has been validated to ISO 11137-1 and ISO 11137-2, which has thereby determined the routine control and monitoring parameters. The sterilization process
is validated to a minimum SAL 10-6.
The shelf life of the Vial Adapter is determined based on stability study which includes ageing test. The testing is performed according to the following standards:
> ISO 11607-1: 2019 Packaging for terminally sterilized medical devices - Part 1:
Requirements for materials, sterile barrier systems and packaging systems
- > ISO 11607-2: 2019 Packaging for terminally sterilized medical devices - Part 2:
Validation requirements for forming, sealing and assembly processes
> ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
A ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
ISO 80369-7:2021 Small-bore connectors for liquids and gases in healthcare A
{9}------------------------------------------------
applications Part 7: Connectors for intravascular or hypodermic applications
» ISO 8536-4:2019 Infusion equipment for medical use Part 4: Infusion sets for single use, gravity feed
A ASTM D4169-22 Standard Practice for Performance Testing of Shipping Containers and Systems
### Performance testing
Performance testing Summary
| Items | Testing standard | Result | |
|------------------------|--------------------------------------|--------------------------------|------|
| Appearance | Internal performance standards | Pass | |
| Particulate | ISO 22413 -2021 | Pass | |
| Tensile strength | ISO 22413 -2021 | Pass | |
| Leakage | ISO 22413 -2021 | Pass | |
| Unobstructed | ISO 22413 -2021 | Pass | |
| Piercing Spike | ISO 22413 -2021 | Pass | |
| Puncture force | ISO 22413 -2021 | Pass | |
| puncture chip | ISO 22413 -2021 | Pass | |
| Dimension | Internal performance standards | Pass | |
| Housing | ISO 22413 -2021 | Pass | |
| Luer Connector | ISO 80369-7 | Pass | |
| Detachment force | Internal performance standards | Pass | |
| Spike tip ductility | Internal performance standards | Pass | |
| Filtration rate | ISO 22413 -2021 | Pass | |
| Chemical property | ISO 8536-4:2019 | Pass | |
| Chemical<br>Properties | Reducing substances<br>(easy oxides) | Internal performance standards | Pass |
| | Metal ions | | |
| | pH | | |
| | Evaporation residues | | |
| | UV absorbance | | |
| Sterile | USP46-NF41 | Pass | |
| Bacterial endotoxin | USP-NF<71> | Pass | |
#### IX Clinical Data
Clinical data was not conducted to support a substantial equivalence determination.
#### × Conclusion
The Vial Adapter with Filter are substantially equivalent to the predicate device (Vial Adapter). The non-clinical testing demonstrates that the differences between the subject and
{10}------------------------------------------------
predicate device do not raise different questions of safety and effectiveness.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.